K Number
K201968
Date Cleared
2021-02-26

(226 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Color Doppler Palm Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), B/M, CFM, Combined (B+CFM), Pulsed Wave and Fusion Harmonic Imaging modes. It is indicated for Abdominal, Neonatal Cephalic, Adult Cephalic, Cardiac Adult, Cardiac Pediatric.

The Digital Color Doppler Palm Ultrasound System is intended for use in environments where is provided by healthcare professionals.

Device Description

The SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System is a mobile, general purpose, softwarecontrolled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data in various modes of operation. The SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System is a portable system facilitating point of care ultrasound applications. The SonoEye P3 Series Digital Color Doppler Palm Ultrasound System includes:

  • · A commercial off-the-shelf (COTS) Android mobile device
  • CHISON Ultrasound software is running as an app (Android ) on the COTS device
  • The SonoEye P3, SonoEye V3, SonoEye G3 Phase array USB transducer
AI/ML Overview

The provided text does not include information about AI/ML models or clinical studies involving human readers and AI assistance. Instead, it describes a 510(k) Premarket Notification for a medical device (Digital Color Doppler Palm Ultrasound System) and demonstrates its substantial equivalence to predicate devices, primarily through non-clinical testing and comparison of technical specifications.

Therefore, many of the requested categories for AI/ML model acceptance criteria and study design cannot be filled from the provided document. The document explicitly states: "No clinical testing was required." This indicates that the FDA clearance for this device was based on demonstrating similar technical and safety characteristics to already approved devices, rather than a clinical effectiveness study involving human users or AI.

However, I can extract the information that is present according to the prompt's structure:

Device: SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System

1. A table of Acceptance Criteria and the Reported Device Performance

Based on the provided document, the acceptance criteria are adherence to recognized medical device safety and performance standards and demonstration of substantial equivalence to predicate devices. The "reported device performance" is primarily the statement of compliance with these standards and the comparison of specifications.

CategoryAcceptance Criteria (Stated Requirements/Standards Met)Reported Device Performance (as stated in the 510(k) summary)
General ComplianceSubstantial equivalence to legally marketed predicate devices.Concluded as substantially equivalent to predicate devices (Lumify Diagnostic Ultrasound System K192226, Clarius Ultrasound System K192107, TE7 Diagnostic Ultrasound System K160381) with regard to safety and effectiveness.
Electrical SafetyCompliance with IEC 60601-1 (Medical Electrical Equipment - Part 1: General Requirements for Safety) and ANSI/AAMI ES60601-1:2005. Transducers to meet Type BF isolated applied part requirements. When used with IEC 60950-1 compliant COTS devices, the system meets IEC 60601-1 Class II requirements.Transducer and software, with representative device, verified compliant with IEC 60601-1. Transducers meet Type BF isolated applied part requirements. When used with IEC 60950-1 compliant COTS devices, the system meets IEC 60601-1 requirements for Class II equipment. The subject device passed hardware test for voltage and complied with IEC60601-1 & IEC60601-2-37 requirements. Passed temperature test and complied with IEC60601-1 & IEC60601-2-37 requirements. Complied with humidity requirements (IEC60601-1 & IEC60601-2-37) and within normal range.
EMCCompliance with IEC 60601-1-2: 2014 (Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests). Classified as Group 1, Class A equipment per CISPR 11 for radiated and conducted electromagnetic disturbances. When used with IEC 55032 compliant COTS devices, the system meets IEC 60601-1-2 requirements for Group 1, Class A.Transducer and representative Android device classified as Group 1, Class A equipment in accordance with CISPR 11. When used with IEC 55032 compliant COTS devices, the system meets IEC 60601-1-2 requirements for Group 1, Class A equipment.
BiocompatibilityCompliance with ISO 10993-1:2009 (Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process) and FDA guidance "Use of International Standard ISO 10993-1..."Device found to conform to applicable medical device safety standards in regards to biocompatibility, and conforms to ISO 10993-1.
Ultrasound SpecificCompliance with IEC 60601-2-37: 2007 (Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment) and FDA guidance "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers". Acoustic Output: Track 3; Ispta.3 ≤ 720 mW/cm², MI ≤ 1.9, TI ≤ 6.0.Device found to conform to applicable medical device safety standards in regards to thermal and mechanical safety, and fulfills the requirement of IEC 60601-2-37. Acoustic output meets criteria (Track 3; Ispta.3 ≤ 720 mW/cm², MI ≤ 1.9, TI ≤ 6.0).
SoftwareCompliance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate Level of Concern) and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". COTS Android devices: output current ≥ 1.5A, output voltage 5V±5%, full compliance with USB 2.0 standard. Representative Android devices (Samsung Galaxy A70s, Huawei MatePad Pro) configured as specified (CPU, RAM, ROM, resolution, WLAN, OS, Bluetooth).Software Documentation for a Moderate Level of Concern included. Representative Android devices meet specified current, voltage, and USB 2.0 compliance. Samsung Galaxy A70s and Huawei MatePad Pro configurations are detailed.

Note: The performance metrics provided relate to the operational specifications of the device (e.g., depth range, shades of gray, gain, focus, measurement types, transducer types, acoustic output limits, power requirements) and were deemed "Same" or had "SE (Substantial Equivalence) Analysis" where differences were addressed and determined not to affect safety or effectiveness.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not applicable in the context of this 510(k) summary, as it explicitly states "No clinical testing was required." The "testing" referred to is non-clinical verification of compliance with standards and comparison to predicate devices, not a study on a human patient population or a dataset for AI.
  • Data Provenance: Not applicable, as there was no clinical test set data. The device itself is manufactured in China (CHISON Medical Technologies Co., Ltd., Wuxi, Jiangsu, China).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable as "No clinical testing was required" and therefore no human expert-established ground truth was part of the submission for clinical performance. The ground truth for engineering/safety testing would be derived from standard test equipment and protocols.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable, as no clinical test set requiring adjudication was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. The document explicitly states "No clinical testing was required." This premarket notification is for a general diagnostic ultrasound system, not an AI-powered diagnostic aid.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No standalone algorithm performance study was done. This device is a hardware ultrasound system with integrated software, not an AI algorithm intended for standalone diagnostic performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests conducted (electrical safety, EMC, biocompatibility, acoustic output), the "ground truth" would be the specifications and requirements defined by the international standards (IEC, ISO, ANSI/AAMI, CISPR) and FDA guidance documents. These are objective engineering and physical measurements. Clinical ground truth for diagnostic accuracy was not established or required for this 510(k) clearance.

8. The sample size for the training set

  • Not applicable, as there is no mention of an AI/ML model for which a training set would be used.

9. How the ground truth for the training set was established

  • Not applicable, as there is no mention of an AI/ML model requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

CHISON Medical Technologies Co., Ltd. % Qifei Liu Regulatory Affairs Manager No.228, Changjiang East Road, Block 51 and 53, Phase 5 Shuofang Industrial Park, Xinwu District Wuxi. Jiangsu 214142 CHINA

Re: K201968

Trade/Device Name: SonoEye P3/ SonoEye V3/ Sono Eye G3 Digital Color Doppler Palm Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 13, 2021 Received: January 27, 2021

Dear Qifei Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

February 26, 2021

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201968

Device Name

SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System

Indications for Use (Describe)

The Digital Color Doppler Palm Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), B/M, CFM, Combined (B+CFM), Pulsed Wave and Fusion Harmonic Imaging modes. It is indicated for Abdominal, Neonatal Cephalic, Adult Cephalic, Cardiac Adult, Cardiac Pediatric.

The Digital Color Doppler Palm Ultrasound System is intended for use in environments where is provided by healthcare professionals.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201968

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

1. Submitter:

Submitter:CHISON Medical Technologies Co., Ltd.
Address:No.228, Changjiang East Road, Block 51 and 53, Phase 5, ShuofangIndustrial Park, Xinwu District, Wuxi, Jiangsu, China 214142
Contact:Mr. Liu Qifei
Tel:+86-510-85310019
Fax:+86-510-85310021
Date Prepared:July 3, 2020

2. Device :

SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Trade Name: Ultrasound System

Common Name: Diagnostic Ultrasound System with Transducers

Classification: Regulatory Class: II Review Category: Tier II

Classfication Name21 CFR SectionProduct Code
Ultrasonic pulsed doppler imaging system892.155090-IYN
Ultrasonic pulsed echo imaging system892.156090-IYO
Diagnostic ultrasonic transducer892.157090-ITX

3. Predicate Device(s):

DeviceModelProduct Code510(k)Number
1.Main predicate deviceLumifyDiagnosticUltrasound SystemIYN,IYO,ITXK192226
2.Reference deviceClarius Ultrasound SystemIYN,ITX,IYOK192107
3.Reference deviceTE7 Diagnostic UltrasoundSystemIYN,ITX,IYOK 160381

4. Device Description:

The SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System is a mobile, general purpose, softwarecontrolled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data in various modes of operation. The SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm

{4}------------------------------------------------

Ultrasound System is a portable system facilitating point of care ultrasound applications. The SonoEye P3 Series Digital Color Doppler Palm Ultrasound System includes:

  • · A commercial off-the-shelf (COTS) Android mobile device
  • CHISON Ultrasound software is running as an app (Android ) on the COTS device
  • The SonoEye P3, SonoEye V3, SonoEye G3 Phase array USB transducer

5. The requirement of commercial off-the-shelf (COTS) devices All models need to comply with IEC55032, IEC60950-1, among which we have selected typical models to do verification testing with SonoEye on ANSI/AAMI ES60601-1 and IEC60601-1-2, such as Samsung Galaxy A70s or Huawei MatePad Pro.

  • Electrical Safety
    The transducer and software, along with a representative device, have been verified as compliant with IEC 60601-1. The transducers meet Type BF isolated applied part requirements. When the transducer and software are used in conjunction with a device(COTS devices) compliant with IEC 60950- 1, the system meets IEC 60601-1 requirements for Class II equipment.

  • Electromagnetic Compatibility The transducer and representative Android device are classified as Group 1, Class A equipment in accordance with international standard CISPR 11 for radiated and conducted electromagnetic disturbances. When the transducer and software are used in conjunction with a device(COTS devices) compliant with IEC 55032, the system meets IEC 60601- 1-2 requirements for Group1,Class A equipment.

  • Android devices' output current and voltage The representative Android devices' output current is 1.5A and voltage is 5V. A COTS device connected to our transducer and system should meet the current and voltage range, the output current is greater than or equal to 1.5A, and the output voltage is 5V±5%.

  • Full compliance with USB 2.0 standard The representative Android devices are full compliance with USB 2.0 standard. A COTS device connected to our transducer and system should meet the standard.

  • · Representative Android devices Representative Android devices are Samsung Galaxy A70s and Huawei MatePad Pro.The configuration is as follows.

Samsung Galaxy A70s:

•CPU frequency 2.0GHz (big quad core), 1.7GHz (small quad core), Octa-core

·RAM capacity 8GB

·ROM capacity 128GB

·Main screen resolution 2400x1080 pixels

•WLAN function Dual-band WIFI, IEEE 802.11 alb/g/n/ac (support 2.4G and 5GHz)

· Operating System: Android 9

  • ·Support Bluetooth 5.0

{5}------------------------------------------------

•Huawei MatePad Pro:

•CPU frequency 2 x Cortex-A76 Based 2.86 GHz+ 2 x Cortex-A76 Based 2.09 GHz+ 4 x Cortex-A55 1.86 GHz, Octa-core

•RAM capacity 8GB

·ROM capacity 128GB

•Main screen resolution 2560x1600 pixels

•WiFi function supports dual frequency (2.4GHz+5GHz)

·Bluetooth function supports Bluetooth 5.1 module

•Operating System: Android 9

6. Indications for Use:

The Digital Color Doppler Palm Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), B/M,CFM,Combined (B+CFM), Pulsed Wave and Fusion Harmonic Imaging modes. It is indicated for Abdominal, Neonatal Cephalic, Adult Cephalic, Cardiac Adult, Cardiac Pediatric.

The Digital Color Doppler Palm Ultrasound System is intended for use in environments where healthcare is provided by healthcare professionals.

7. Summary of Non-Clinical Tests:

The Digital Color Doppler Palm Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.

The device has been found to conform to applicable medical device safety standards in reqards to thermal, mechanical and electrical safety as well as biocompatibility.

The product fulfils the requirement of: ANSI/AAMI ES60601-1:2005Medical Electrical Equipment - Part 1: General

Requirements for Safety.

IEC 60601-1-2: 2014 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.

IEC 60601-2-37: 2007 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.

ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process

The device has been found to conform to applicable FDA medical device guidance documents titled as followings:

  • Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (Document issued on: June 27, 2019)
  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005)
  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (Document Issued on: October 2, 2014)
  • Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (Document

{6}------------------------------------------------

issued on: June 16, 2016)

8. Clinical Test:

No clinical testing was required. Software Documentation for a Moderate Level of Concern, per the FDA guidance document,"Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on May 11, 2005", is also included as part of this submission.

9. Comparison to Predicate Device:

ItemsMain predicatedeviceReferencedeviceReference deviceSubmissionDeviceRemark
LumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigitalColorDopplerPalmUltrasoundSystem
Indications for UseFetal/Obstetric,Abdominal,Pediatric,Cephalic,Urology,Gynecological,Cardiac FetalEcho,Small Organ,Musculoskeletal,PeripheralVessel,Carotid,Cardiac.ophthalmic,fetal,abdominal,intraoperative(non-neurological), pediatric,small organ,cephalic(adult),trans-rectal,transvaginal,musculo-skeletal(conventional,superficial),urology,gynecology,cardiac (adult,pediatric),peripheralvessel,carotid, andproceduralguidance ofneedles intothe bodyfetal,abdominal,intra-operative(abdominal, thoracic,and vascular),pediatrics,smallorgan(breast,thyroid, testes),neonatal and adultcephalic,trans-esoph.(Cardiac), trans-vaginal,musculo-skeletal(conventional, superficial),urology,peripheral vessel,adult and pediatriccardiac,ophthalmic examsAbdominal,NeonatalCephalic,Adult Cephalic,Cardiac Adult,Cardiac,PediatricSame
ItemsMain predicatedeviceReferencedeviceReference deviceSubmissionDeviceRemark
LumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigital ColorDoppler PalmUltrasoundSystem
DesignAutocorrelationfor colorprocessing andFFT for pulseDopplerprocessing.SupportingLinear, Curve,Phase arrayand Volumeprobes.Cine play backcapabilityImage filearchiveAutocorrelationfor colorprocessing andFFT for pulseDopplerprocessing.SupportingLinear, PhaseMicroconvex array,Endocavityprobe andVolumeprobes.Cine play backcapabilityImage filearchiveAutocorrelation forcolor processingand FFT for pulseDopplerprocessing.Supporting Linear,convex andphased array.Cine play backcapabilityImage file archiveAutocorrelationfor colorprocessingand FFT forpulseDopplerprocessing.SupportingPhased probe.Cine play backcapabilityImage filearchiveSame
OperatingControlsTGCTGCTGCSTCSame
Depth Range:0.003 to >30 cmDepth Range:5.0 to 10.0cm(Depth dependon probe type)Depth depend onprobe typeDepth Range:1.0 to 22cmSEAnalysis1
256 shades ofgray256 in B-Mode30-240 in B mode256 shades ofgrayGain:0-255,1/stepSame
GainGainGainFocusSame
FocusFocusFocusFocus: adjustableSame
OperatingControlsColor boxsize/positioncan be adjustROI adjustROI adjustROI size/position:adjustableSame
N/ABaselineBaselineBaselineSame
Cine control:drag scroll bar,press orTap the cinecapture buttonto capture thecine as part ofthe exam.Cine controlCine control:step, playbackward, playcontinuouslySame
Freeze control:Toggling freezekeyFreeze control:Toggling freezekeyFreeze control:Toggling freezekeyFreeze control:Toggling freezekeySame
SafetyComplianceANSI/AAMIES60601-1IEC60601-1IEC60601-2-37IEC60601-1-2ISO 10993-1ANSI/AAMIES60601-1IEC60601-1IEC 60601-2-37IEC 60601-1-2ISO 10993-1ANSI/AAMIES60601-1IEC60601-1IEC60601-2-37IEC60601-1-2ISO 10993-1ANSI/AAMIES60601-1IEC60601-1IEC60601-2-37IEC60601-1-2ISO 10993-1Same
Main predicatedeviceReferencedeviceReference deviceSubmissionDevice
ItemsLumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigital ColorDoppler PalmUltrasoundSystemRemark
B modeB modeB modeB modeSame
N/APW modePulsed waveDoppler modePW modeSame
M modeM modeM modeB/M modeSame
Color modeColor DopplerModeColor modeCFM modeSame
N/AN/AIQFHISame
GainGainGainGainSame
DepthDepthDepthDepthSame
TGCTGCTGCSTCSame
N/ASpatialCompoundSpatial CompoundCompoundSame
FrequencyFrequencyFrequencyFrequencyScalingSame
FreezeFreezeFreezeFreeze/UnfreezeSame
ZoomZoomZoomZoomSame
2D distance2D distancedistanceB-distanceSame
CircumferenceCircumferenceCircumferenceB-CircumferenceSame
AreaAreaAreaB-AreaSame
N/AVolumeVolumeB-VolumeSame
N/ADistanceDistanceB/M-DistanceSame
OperationN/AM-TimeTimeB/M-TimeSame
ModeN/AM-HRHRB/M-HRSame
N/AVelocityVelocityPW-VelocitySame
N/ATimeTimePW-TimeSame
N/AVascularVascularVessel measurepackageSame
N/ASmall OrganAbdomenAbdomenmeasurepackageSame
N/AN/ACardiologyCardiac measurepackageSame
N/AReportReportGeneral reportSame
N/AN/AVascular ExamReportVessel reportSame
N/AN/AAbdomen ExamReportAbdomen reportSame
N/AN/ACardiac ExamReportCardiac reportSame
Languagefollow systemN/AMulti-languageInterfaceMulti-languageInterfaceSame
ThumbnailN/AThumbnail windowClipboardSame
N/AN/AiTouchInstant AIOSame
BiopsyGuideBiopsy GuideBiopsy GuideBiopsy GuideSame
ItemsMain predicatedeviceReferencedeviceReference deviceSubmissionDeviceRemark
LumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigital ColorDoppler PalmUltrasoundSystem
Line
Reacts SessionViewsN/AN/ASonoRemoteSame
N/AN/AiNeedleSonoNeedleSame
N/ANeedleEnhanceiNeedleSuperNeedleSame
Save LoopSave CineSave ClipSave CineSame
Save ImageSave ImageSave ImageSave ImageSame
Playing LoopsCine LoopCineReviewCine LoopSame
AnnotationAnnotationAnnotationAnnotationSame
N/ABodymarkBodymarkBodymarkSame
N/AN/AArrowArrow MarkSame
PatientDatabasePatientinformationPatient informationPatientmanagementSame
N/AAngle/Baselineon PWAngle/Baseline onPWVoice/Angle/Baseline on PWSame
N/AN/APhysical keyPhysical keySame
SettingSettingSetupSettingSame
PatientDatabasePatientinformationPatient datamanagementArchivesSame
WalkthroughQuick StartGuide andVideo TutorialsOperation NoteTutorialsSame
ReviewReviewReviewEasyviewSame
N/ADemoiVisionDemoSame
AboutN/ASysteminformationAboutSame
N/AAdjust sampleGate locationDrag the SV gateto place the SV onthe targetAdjust sampleGate locationSame
N/AAdjust size ofsample GatehorizontalDrag the SV gateto place the SV onthe targetAdjust size ofsample GatehorizontalSame
N/AAdjust size ofsample GateverticalDrag the SV gateto place the SV onthe targetAdjust size ofsample GateverticalSame
N/AAdjust PWsample gateDrag the SV gateto place the SV onthe targetAdjust PWsample gateSame
ItemsMain predicate deviceReference deviceReference deviceSubmission DeviceRemark
LumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigital ColorDoppler PalmUltrasoundSystem
DisplayAnnotationsLogo; HospitalName;Examdate;Examtime;MechanicalI index;Thermalindes;Probemodel;TGCCorve;Focusposition;Imagingparameters;Systemstatus;Gray/Color barLogoHospitalName(NA)Exam date;Exam time;End exam;MI;TI; TGC; Probemodel;Imagingparameters;System status;Gray/Color barProbe model,acoustic outputvalue, MI,TI,iNeedle, iTouch,frequency, TGC,System status,Depth, GainLogo; HospitalName; Examdate; Exam time;Mechanicalindex; Thermalindes; Probemodel; STC;Focus position;Imagingparameters;System status;Gray/Color barSame
Measurements2D mode:DepthDistance ,Area,CircumferenceB mode:Depth,Distance, Area,CircumferenceDoppler mode:Volume,Velocity, TimeM mode:Distance,Time, HR2D mode: Depth,Distance, Area,CircumferenceB/C mode:VolumeDoppler mode:Velocity, Time,M mode:Distance,Time, HR2D mode:Depth, Distance,Area, VolumeDoppler mode:D Velocity, Time,B/M mode:Distance, Time,HRSame
Transducer Types &ConnectorsConvex Array,Phased Array,Linear Array,USB interfaceMicro convexarray,Endocavityprobe andVolume probesLinear, convex andphased array.Phased ArrayUSB interfaceSame
UsersSitesHospitals,clinics usageHospitals,clinics usageHospitals, clinicsusageHospitals, clinicsusageSame
AcousticOutputTrack 3;Ispta.3 ≤ 720mW/cm²MI ≤ 1.9TI ≤ 6.0Track 3(ISPTA) of 720mW/ cm²MI: 0.0 to 1.9,TI: This iscontinuouslydisplayed overthe range of0.0 tomaximumoutput, basedon thescanner andapplication, inTrack 3Ispta.3 ≤ 720mW/cm²MI ≤ 1.9Track 3;Ispta.3 ≤ 720mW/cm²MI ≤ 1.9TI ≤ 6.0Same
ItemsMain predicatedeviceReferencedeviceReference deviceSubmissionDeviceRemark
LumifyDiagnosticUltrasoundSystemClariusUltrasoundSystemTE7 DiagnosticUltrasoundSystemDigitalColorDopplerPalmUltrasoundSystem
0.1
PowerRequirementsPowerrequirements:AC: 100V-240V,Frequency: 50-60HzOperatingtemperature: 5-40 ℃ ; relativehumidity15-95%;Barometricpressure: 700 to1060 hPaInput: 12 VDC,11.5 ATemperature:0°C (50°F) to40°C (113°F)Humidity: 15%to 95%sVoltage:100V-240VTemperature: 0°C(50°F) to 40°C(113°F)Humidity: 30% to85%sAtmospherepressure: 700 to1060 hPaPowerrequirements:DC 5V,Operatingtemperature: 10-38 ℃ ; relativehumidity30-75%;Barometricpressure: 700 to1060 hPaSEAnalysis2

Table 1 Substantial Equivalence Comparison

{7}------------------------------------------------

510(k) Summary

Page 5 of 10

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Comparison Analysis

SE Analysis 1:

Operation Controls, compared with the predicate device, the subject device employs the same operation controls design and has some differences in value range. But both of them comply with the requirements of IEC60601-1 & IEC60601-2-37 and meet clinical requirements. Therefore, they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

SE Analysis 2:

Operation Controls, compared with the predicate device, the subject device employs the same operation controls design and has some differences in presenting voltage, humidity and temperature.

For voltage, the voltage of subject device is different with predicate device since hardware configuration is different. The subject device has passed the hardware test and complied with requirement of IEC60601-1 & IEC60601-2-37 and also meets clinical requirements

For temperature, compared with predicate device, the temperature of subject device is lower than it, but still in a normal range. The subject device has passed temperature test and complied with requirement of IEC60601-1 & IEC60601-2-37 and meet clinical requirements For humidity, compared with predicate device, the humidity of subject device is in a normal range and complied with requirement of IEC60601-1 & IEC60601-2-37 and meet clinical requirements

{12}------------------------------------------------

Therefore, they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

10.Substantially Equivalent Conclusion:

In accordance with the Act. 21 CFR Part 807 and based on the information provided in this premarket notification, CHISON Medical Technologies Co., Ltd. concludes that the SonoEye P3/ SonoEye V3/ SonoEye G3 Digital Color Doppler Palm Ultrasound System is substantially equivalent to the predicate devices with regard to safety and effectiveness.

N/A