(28 days)
The CarboClear® Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The CarboClear® II Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The CarboClear® and CarboClear® II Pedicle Screw Systems are composed of implants in various dimensions, used to build a spinal construct; and of a set of instruments, intended to assist in the insertion and placement of the implants.
The implants include pedicle screws, rods, locking elements and transverse connectors. The implants are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK). The threaded portion of the pedicle screws is encased within a thin titanium shell, and includes a small tantalum marker.
The implants are supplied sterile, and are intended for single use.
This document is a 510(k) Premarket Notification and does not describe acceptance criteria for a device's performance based on clinical study data, nor does it detail a study proving such performance. Instead, it focuses on demonstrating substantial equivalence to predicate devices, primarily through mechanical performance testing rather than clinical outcomes or diagnostic accuracy.
Therefore, many of the requested elements (e.g., sample size for test sets, expert adjudication, MRMC studies, ground truth for training/test sets) are not applicable or findable within this document, as they relate to clinical or AI/diagnostic performance studies, which were not the basis for this specific device clearance.
The "device" in this context is a CarboClear® Pedicle Screw System and CarboClear® II Pedicle Screw System, which are medical implants. The "performance" being evaluated here is primarily mechanical performance, not diagnostic accuracy or human-in-the-loop improvement for complex medical tasks often associated with AI/diagnostic devices.
Here's an attempt to answer the questions based on the provided document, highlighting what is (and isn't) present:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not specify quantitative "acceptance criteria" in the way one might for a diagnostic device (e.g., minimum sensitivity/specificity). Instead, it states that the performance characteristics (mechanical) were evaluated against standards, and the results were comparable to predicate devices. This implies the acceptance criterion was "comparable mechanical performance" as demonstrated by meeting specified ASTM standards.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Mechanical performance comparable to predicate devices and meeting relevant ASTM standards. | "The results of the tests are comparable to those of the predicate devices, as applicable, demonstrating substantially equivalent mechanical performance of the subject device." Specific quantitative results (e.g., fatigue strength, pull-out force) are not provided in this summary, only that they met the standard of comparability. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify mechanical test sample sizes. Mechanical testing typically involves a set number of samples per test condition (e.g., n=6 for fatigue testing). These are in vitro tests, not patient data.
- Data Provenance: Not applicable for a mechanical performance study. The tests were performed in vitro to ASTM standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
Not applicable. Ground truth and expert review are relevant for diagnostic or AI performance studies involving interpretation of medical data. This document describes mechanical testing of an orthopedic implant.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human readers/experts in diagnostic studies. This is a mechanical performance study.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its Effect Size
No. An MRMC study is relevant for diagnostic performance, evaluating how human readers perform with or without AI assistance on medical cases. This document covers mechanical testing of a pedicle screw system.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/algorithm-based diagnostic device. The performance refers to the physical mechanical properties of the implant itself.
7. The Type of Ground Truth Used
Not applicable in the context of diagnostic "ground truth." For mechanical testing, the "ground truth" is adherence to established engineering standards (ASTM F1717, F1798, F2193, F543) and demonstrable equivalence to predicate devices under these test conditions.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
Summary of why requested information is largely absent:
The provided text is a 510(k) summary for a pedicle screw system, which is an implantable medical device. The clearance is based on substantial equivalence to existing predicate devices, primarily demonstrated through mechanical performance testing (in vitro engineering tests) rather than clinical studies or diagnostic accuracy evaluations. Many of the questions posed directly relate to the methodology of AI/diagnostic device studies (e.g., human reader performance, training/test sets, expert ground truth, MRMC studies), which are not performed for this type of device clearance.
{0}------------------------------------------------
August 7, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
CarboFix Orthopedics Ltd. Ms. Hila Wachsler-Avrahami Regultory Affairs 11 Ha'hoshlim St. Herzeliya, 4672411 Israel
Re: K201926
Trade/Device Name: CarboClear® Pedicle Screw System, CarboClear® II Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: July 8, 2020 Received: July 10, 2020
Dear Hila Wachsler-Avrahami:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K201926
Device Name
CarboClear® Pedicle Screw System
Indications for Use (Describe)
The CarboClear® Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Indications for Use
Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
Device Name
CarboClear® II Pedicle Screw System
Indications for Use (Describe)
The CarboClear® II Pedicle Screw System is integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
CarboFix Orthopedics Ltd. CarboClear® Pedicle Screw System & CarboClear® II Pedicle Screw System
Applicant Name
CarboFix Orthopedics Ltd.
11 Ha'hoshlim St., Herzeliya 4672411, Israel
Contact Person
Hila Wachsler-Avrahami CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
Date Prepared
July 8, 2020
Trade/Proprietary Name
-
- CarboClear® Pedicle Screw System
-
- CarboClear® II Pedicle Screw System
Common Name
Pedicle Screw System
Regulation Number and Device Class
21 CFR §888.3070; Class II
Product Code, Regulatory Description and Review Panel
NKB; Thoracolumbosacral pedicle screw system; Orthopedic
{5}------------------------------------------------
CARBOFIX ORTHOPEDICS LTD. CARBOCLEAR® & CARBOCLEAR® II PEDICLE SCREW SYSTEMS
Predicate Devices
Primary Predicate Device:
-
- CarboClear® Pedicle Screw System (CarboFix Orthopedics Ltd.: K173487)
Additional Predicate Devices:
- CarboClear® II Pedicle Screw System (CarboFix Orthopedics Ltd.; K200519)
- Mesa Spinal System (K2M, Inc.; K052398 and more)
- . CD HORIZON® Spinal System (Medtronic Sofamor Danek; K981676 and more)
Intended Use/Indications for Use
-
- CarboClear® Pedicle Screw System
The CarboClear® Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
- CarboClear® Pedicle Screw System
2. CarboClear® II Pedicle Screw System
The CarboClear® II Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
System Description
The CarboClear® and CarboClear® II Pedicle Screw Systems are composed of implants in various dimensions, used to build a spinal construct; and of a set of instruments, intended to assist in the insertion and placement of the implants.
The implants include pedicle screws, rods, locking elements and transverse connectors. The implants are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK). The threaded portion of the pedicle screws is encased within a thin titanium shell, and includes a small tantalum marker.
The implants are supplied sterile, and are intended for single use.
{6}------------------------------------------------
Purpose of Submission
The purpose of this Special 510(k) Premarket Notification is to add pedicle screw sizes to the CarboClear® and CarboClear® II Pedicle Screw Systems.
Performance Data and Substantial Equivalence
The intended use, design, material, dimensions, technological characteristics, and principles of operation of the subject CarboClear® and CarboClear® II Pedicle Screw Systems are substantially equivalent to those of the predicate devices.
Performance characteristics included tests according to ASTM F1717, ASTM F1798, ASTM F2193 and ASTM F543. The results of the tests are comparable to those of the predicate devices, as applicable, demonstrating substantially equivalent mechanical performance of the subject device.
Conclusion
Based on the information provided in this Premarket Notification, the subject CarboClear® Pedicle Screw System and CarboClear® II Pedicle Screw System are substantially equivalent to their predicate devices.
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.