(789 days)
IAF-SYSTEM is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good health. IAF-SYSTEM simulates kneading of tissues by using an inflatable garment.
Subject device consists of an air pump, air pressure sensor, and sleeves working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; each sleeve has four (4) compression chambers. The compression massage direction is from limb end to body center. By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. Software controls the timing and pressure reflected by the sensor, cycling airflow into and out of the sleeves to compress body
The provided text describes a 510(k) premarket notification for the IAF-SYSTEM, a powered inflatable tube massager. It details the device's characteristics and compares them to a predicate device (Rapid Reboot Compression Therapy System, K182668) to establish substantial equivalence.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in the traditional sense of a clinical trial or performance study comparing a specific metric against a predefined threshold. Instead, it aims to demonstrate substantial equivalence to a predicate device by comparing various technological characteristics and safety standards.
The acceptance criteria are implicitly derived from the similarity to the predicate device and compliance with relevant safety standards. The "reported device performance" is essentially the detailed description of the subject device's features and its conformity to the standards.
While not a direct "acceptance criteria" table, the provided comparison table in the 510(k) summary (pages 5-8) serves as the basis for demonstrating equivalence. Key comparisons include:
| Characteristic | Predicate Device Performance (Rapid Reboot Compression Therapy System) | Subject Device Performance (IAF-SYSTEM) | Comparison / Acceptance |
|---|---|---|---|
| Indications for Use (IFU) | Temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas in people in good health; simulates kneading and stroking of tissues using inflatable garment. | Temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas in people in good health; simulates kneading and stroking of tissues using inflatable garment. | Same. This is a critical acceptance criterion for substantial equivalence. |
| Classification | Class II Device, IRP (21 CFR 890.5650) | Class II Device, IRP (21 CFR 890.5650) | Same. Essential for regulatory pathway. |
| OTC or Rx | OTC | OTC | Same. |
| Environment of Use | Clinics, hospital, athlete training, and home environments | Clinics, hospital, athlete training, and home environment. | Same. |
| Standards | IEC 60601-1:2014; IEC 60601-1-2:2014 | ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012; IEC 60601-1-2 Edition 4.0 2014-02 | Same. Compliance with recognized safety standards is an acceptance criterion. |
| Biocompatibility | EN ISO 10993-5:2009 & EN ISO 10993-10:2010 | Not required for subject device, as users wear clothing under the garments. No direct contact. | Different. Justification provided that direct patient contact is avoided, making biocompatibility not required and thus substantially equivalent. |
| Mode of Compression | Sequential/Peristaltic | Sequential "Seawave"/ Peristaltic | Same. Fundamental functional characteristic. |
| Max Pressure / Min Pressure | 0-200 mmHg | 0-119 mmHg | Different. Subject device provides substantially equivalent pressure despite the different maximum pressure. This is an accepted difference based on the intent of the device. |
| Number of Chambers | 4 Chambers | 4 Chambers | Same. |
| Compression Garment Material | Nylon with a Polyurethane laminate | Nylon with a Polyurethane laminate | Same. |
| Therapy Time | User determines; 10, 20, or 30 min, with option to add 10 min. | User determines; 10, 20, 25 or 30 min. | Same. Minor variation in specific time increments is considered substantially equivalent for user-determined therapy. |
| Cycle Time | 1 min 20 sec (80 sec) | Varies by program (e.g., Program 1: 50 / 38; Program 6: 80 / 68; Program 8: 260/140). Program 6 provides same 1st cycle time as predicate. All others offer substantially same cycle time, varying between 34-260 seconds (1st cycle) and 33-203 seconds (subsequent cycles). | Same and Different. The document argues differences are within the same order of magnitude and do not raise new concerns, making it substantially equivalent. |
| Modes (Inflation Sequences) | 2 modes: "A" (bottom up, one at a time), "B" (bottom up, maintains pressure in lower chambers). | 2 modes: "A" (bottom up, one at a time), "B" (bottom up, maintains pressure in selected chambers). The document also mentions "Mode seawave" with overlapping bands and pressure control, similar to Mode B's description. | Same. Both devices have sequential modes that are functionally similar. The detailed descriptions for "Mode seawave" clarify the equivalence of the underlying mechanism. |
| Safety Features | Button on display allows user to stop or pause therapy. | Button on display allows user to stop or pause therapy. | Same. Critical safety feature. |
| Technology | Compressor and valve system which sequentially inflates cells of appliance. | Compressor and valve system which sequentially inflates cells of appliance. | Same. Core technological principle. |
| Non-clinical Tests | IEC 60601-1; IEC 60601-1-2 | IEC 60601-1, IEC 60601-1-2, Earth circuit continuity, leakage current, current absorption, main switch, air hose connector, delivered pressure, display function, programs, top card function, safety pressure switch intervention, maximum setting pressure, integrity of inflatable parts. | Compliance reported. Device complies with acceptance criteria, documenting safety and effectiveness on validation and production units. This serves as the "study" for non-clinical performance. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): The document states, "No clinical tests were conducted." Therefore, there is no "test set" in the context of human subjects or clinical data for performance evaluation.
- Data Provenance: Not applicable, as no clinical tests were performed. The evidence for substantial equivalence relies on non-clinical testing and comparison of device specifications to the predicate device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable, as no clinical tests were performed and no "ground truth" for a test set was established using expert consensus. The "ground truth" for demonstrating non-clinical safety is the compliance with recognized standards.
4. Adjudication Method for the Test Set
- Not applicable, as no clinical tests were performed and no test set requiring adjudication was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is not an AI-assisted diagnostic or interpretation system that would typically involve human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical therapy device, not an algorithm or AI system subject to standalone performance evaluation in that sense. Its "standalone" performance is measured by its physical and electrical characteristics and compliance with safety standards.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- For the non-clinical tests, the "ground truth" is compliance with established engineering and electrical safety standards (e.g., IEC 60601-1, IEC 60601-1-2). The performance of the device against these standards serves as the ground truth for its safety and functional equivalence.
8. The Sample Size for the Training Set
- Not applicable. This device does not employ machine learning or AI that would require a "training set."
9. How the Ground Truth for the Training Set was Established
- Not applicable, as there is no training set for this type of device.
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the (implicit) acceptance criteria consists of non-clinical tests and comparison to a predicate device.
-
Non-Clinical Tests:
- Tests Performed: Earth circuit continuity test, leakage current, current absorption, main switch, air hose connector, delivered pressure, display function, programs functionality, top card functionality, safety pressure switch intervention, maximum setting pressure, integrity of inflatable parts.
- Findings: "device complies with the acceptance criteria of these various tests, documenting that safety and effectiveness are achieved on validation units and on production units (tests are conducted on on-going basis)." This indicates that the device's physical, electrical, and functional characteristics meet the established benchmarks for safety and performance required by the relevant standards (ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012; IEC 60601-1-2 Edition 4.0 2014-02).
-
Predicate Device Comparison:
- The core "study" for substantial equivalence is the detailed comparison of the IAF-SYSTEM to the Rapid Reboot Compression Therapy System (K182668) across all critical aspects (indications for use, technological characteristics, safety features, operational parameters, etc.).
- The conclusion drawn is that "The technological characteristics are substantially equivalent to the predicate device, meeting the same technical standards." Any differences (e.g., max pressure, cycle time, biocompatibility where not required) are justified as not raising new questions of safety or effectiveness.
In summary, the FDA clearance for the IAF-SYSTEM is based on a demonstration of substantial equivalence to an already legally marketed device through a comprehensive comparison of specifications and compliance with recognized non-clinical safety and performance standards, rather than a clinical study with a test set of human subjects.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
August 22, 2022
Foggiare Technologies LLC % Jay Mansour Principal Mansour Consulting LLC 845 Aronson Lake Court Roswell, Georgia 30075
Re: K201730
Trade/Device Name: IAF-SYSTEM Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: July 6, 2022 Received: July 7, 2022
Dear Jay Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201730
Device Name IAF-SYSTEM
Indications for Use (Describe)
IAF-SYSTEM is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good health. IAF-SYSTEM simulates kneading of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) #K201730
510k Summary:
As required by 21 CFR 807.92 (c)
1 - Date Summary Prepared: August 13, 2022
- 2 Owner/Submitter/Sponsor/Applicant information:
OWNER/SUBMITTER/SPONSOR/APPLICANT
Contact name: Edouard Sezionale- Foggiare Technologies LLC Title: owner/CEO Telephone: (561) 923-9161 Email address: thierry@foggiare.com Address: 426 E Palmetto Park Rd. Boca Raton, FL 33432 USA
APPLICATION CORRESPONDENT
Contact name: Jay Mansour, MSQA, BE, RAC- Mansour Consulting LLC Title: regulatory consultant- Principal Telephone: (678) 908-8180 Email address: jay@mansourconsulting.com Address: 845 Aronson Lake Court Roswell, GA 30075 USA
3 - Device Information:
Common/usual/classification name:
massager, powered inflatable tube
Device name:
| Device Model Name | Model Number | |
|---|---|---|
| 1 | IAF-SYSTEM | (NO MODEL NUMBER) |
| FDA 3 Letter Code | IRP |
|---|---|
| FDA regulation number: 21 CFR | 890.5650 |
| Regulation medical specialty | Physical Medicine |
| Review panel | Physical Medicine |
| Class | 2 |
{4}------------------------------------------------
4 - Substantial equivalency is claimed against the following predicate device(s):
| 510kNumber | Trade or Proprietary or Model Name | Manufacturer | Primary Predicate |
|---|---|---|---|
| K182668 | Rapid Reboot Compression Therapy System | Rapid RebootRecoveryProducts, LLC | Yes |
5 - Description of the device
Subject device consists of an air pump, air pressure sensor, and sleeves working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; each sleeve has four (4) compression chambers. The compression massage direction is from limb end to body center. By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. Software controls the timing and pressure reflected by the sensor, cycling airflow into and out of the sleeves to compress body
6 - Intended Use + indications for use
| Intended Use: | The IAF-SYSTEM is intended for the temporary relief of minor muscle achesand pains and for the temporary increase in circulation to the treated areas in |
|---|---|
| indications for use: | people who are in good health. The IAF-SYSTEM simulates kneading andstroking of tissues by using an inflatable garment. |
7 - Basis for determination of substantial equivalency:
(a) Indications for use:
The indication for use is identical to the predicate's.
(b) Technological characteristics:
The technological characteristics are substantially equivalent to the predicate device, meeting the same technical standards.
Substantial equivalency between our device and predicate device is tabulated below:
{5}------------------------------------------------
| Device | Subject DevicePrimary Predicate | Comparison | |
|---|---|---|---|
| Manufacturer | DAM SRL | Rapid Reboot RecoveryProducts, LLC | NA |
| 510(k)Number | TO BE DETERMINED | K182668 | NA |
| Model Name | IAF-SYSTEM | Rapid Reboot CompressionTherapy System | NA |
| Classification | Class II Device,IRP (21 CFR890.5650) | Class II Device, IRP (21 CFR890.5650) | Same |
| Indications forUse (IFU) | The IAF-SYSTEM is intended for thetemporary relief of minor muscle achesand pains and for the temporary increasein circulation to the treated areas inpeople who are in good health. The IAF-SYSTEM simulates kneading and strokingof tissues by using an inflatable garment. | The Rapid RebootCompression Therapy System isintended for the temporary relief ofminor muscle aches and pains and forthe temporary increase in circulationto the treated areas in people whoare in good health. The Rapid RebootCompression Therapy Systemsimulates kneading and stroking oftissues by using an inflatablegarment. | Same |
| OTC or Rx | OTC | OTC | Same |
| Environmentof Use: | Clinics, hospital, athlete training, andhome environment. | Clinics, hospital, athlete training, andhome environments | Same |
| Standards | ANSI AAMI ES60601- 1:2005/ (R)2012 AndA1:2012IEC 60601-1-2 Edition 4.02014-02 | IEC 60601-1:2014; IEC60601-1-2:2014 | Same |
| Biocompatibility not required | EN ISO 10993-5:2009 & EN ISO10993-10:2010 | Different.Biocompatibilitynot required forsubject device, asusers wearclothing underthe garments. Nodirect contact | |
| Mode ofCompression | Sequential"Seawave"/ Peristaltic | Sequential/Peristaltic | Same |
| PowerSource | 120 V, 50Hz | 120V, 60Hz | Same |
{6}------------------------------------------------
| Therapy Time | User determines therapytime. Choose from 10, 20,25 or 30minute session time. | User determines therapytime. Choose from 10, 20, or 30minute session time, with optionto add additional 10 minutes toany therapy time. | Same |
|---|---|---|---|
| Max PressureMin Pressure | 0-119 mmHg | 0-200 mmHg | Different.Subject deviceprovidessubstantiallyequivalentpressure, despitethe differentmaximumpressure |
| Number ofChambers | 4 Chambers | 4 Chambers | Same |
| CompressionApplicatorGarmentsSleeveMaterial | Nylon with a Polyurethane laminate | Nylon with a Polyurethanelaminate | Same |
| HousingMaterialsAndConstructions | Molded ABS enclosure | Molded ABS enclosure | Same |
| Patient contact | Non-conductive attachments | Non-conductive attachments | Same |
| PowerConsumption | 400 W | 30W | Different due tophysicalconfiguration |
{7}------------------------------------------------
| Cycle time | Leg (consisting of foot, calf, knee, upper | 1 min 20 sec | Same and different | |||||
|---|---|---|---|---|---|---|---|---|
| leg) | depending on the | |||||||
| The cycle time for all programs is | program. See | |||||||
| tabulated below. | below. | |||||||
| The first cycle can be different from the | ||||||||
| following ones, because it starts with | Information below | |||||||
| totally empty sectors, while the following | lists 1st cycle, | |||||||
| cycles can start with some air in the | subsequent cycle | |||||||
| cycles change over the course of the | sectors; moreover, in many programs the | and predicate cycle,respectively, all in | ||||||
| program. This is illustrated within the | seconds, comparing | |||||||
| user manual.minutes | each of the 12 | |||||||
| programs with the | ||||||||
| Prog. | Treatment | Sector | cycle | cycle | subsequent | predicate cycle | ||
| n.ਹੈ | time25 | pauseਹ | pause3 | time50 | cycle time38 | time: | ||
| 2 | 25 | 1 | 3 | 47 | ਚ ਹ | Program 1: 50 / 38 / 80 | ||
| m | 25 | ਹੈ | 3 | ರೇ | 33 | Program 2: 47 / 41 / 80 | ||
| বাঁ | 4 | ਚਰੇ | Program 3: 46 /33 / 80 | |||||
| 25 | 2 | ਤੌਰੇ | Program 4: 49 / 39 / 80 | |||||
| ട | 25 | 2 | 4 | ਦਰੇ | eo | Program 5: 69 / 60 / 80 | ||
| б | 25 | 2 | 4 | 80 | e8 | Program 6: 80 / 68 / 80 | ||
| 7 | 25 | 1 | 3 | 97 | 35 | Program 7: 97 / 35 / 80 | ||
| 8 | 30 | 1 | 3 | 260 | 140 | Program 8: 260/140/80 | ||
| Program 9: 65 / 98 / 80 | ||||||||
| ਰੇ | 25 | ਹ | 3 | દર | ಕಿ8 | Program 10: 206/203/80 | ||
| 10 | 20 | 2 | 4 | 206 | 203 | Program 11: 34 / 28 /80Program 12: 34 / 33 / 80 | ||
| 11 | 30 | 2 | 3 | 34 | 28 | |||
| 12 | 20 | 2 | 3 | 34 | 33 | Program 6 provides the | ||
| same exact (1st) cycle | ||||||||
| The cycle time is the time needed to fill air in | time as predicate device. | |||||||
| the sectors with set pressure in the set | ||||||||
| sequence + sectors pause. | All other programs offer | |||||||
| substantially the same | ||||||||
| Sector pause is the time between the end of | cycle time, varying | |||||||
| fill in of air in a sector and the start of fill in | between 34 and 260 | |||||||
| of the air in the sequel one. | seconds (for 1st cycle) and | |||||||
| The cycle pause is the time between the end | between 33 and 203 | |||||||
| of a cycle and the start of the next one. | seconds (for subsequent | |||||||
| 1st cycle time is the time needed for the 1st | 3 | cycles). | ||||||
| complete cycle (it needs a little more | The differences in cycle | |||||||
| because start with total empty sectors). | times are within the | |||||||
| Subsequent cycle time is the time need for | same order of magnitude | |||||||
| other cycles ( being little bit shorter because | as predicate device, and | |||||||
| the sectors are not totally empty). | do not raise new | |||||||
| Exception in program 9: This program | concerns. | |||||||
| includes a first section where the plantar | ||||||||
| thrust provides pressure starting from thebottom-base of the thigh, then mid-thigh, | ||||||||
| then knee and calf, and finally knee and | ||||||||
| ankle. The effect is therefore drainage from | ||||||||
| the top. After that, it starts a treatment like | ||||||||
| program 1. |
{8}------------------------------------------------
| Size | 49″ x 21″ x 14.5″ | 10″ x 6.5″ x 5″ | Same.(Tabletop orhandheld portable) |
|---|---|---|---|
| Weight | 77 pounds with base | 5.8 pounds | Different due to physicalconfiguration |
| Modes (Inflationsequences, allpreprogrammed) | 2 modes: "A" mode inflates and deflateschambers from bottom up (distal toproximal chambers), one at a time. "B"mode also inflates chambers frombottom up, but maintains pressure inselected chambers as works its way totop. Then all chambers release pressureat same time once all chambers havesequentially inflated. | 2 modes: "A" mode inflates anddeflates chambers from bottom up(distal to proximal chambers), oneat a time. "B" mode also inflateschambers from bottom up, butmaintains pressure in lowerchambers as works its way to top.Then all chambers release pressureat same time once all chambershave sequentially inflated. | Same.Both devices havesequential mode(distal toproximalchambers). |
{9}------------------------------------------------
| Modes (visualdescription) | "Sequential:" | Mode A: | Same |
|---|---|---|---|
| With "lock" on specific area | |||
| Mode seawave: | Mode B: | SameThe machinepermits applyinga set pressure ina targeted wayto each single"lock" sector todetermine theeffect andeffectiveness onthe patient.One or moresectors can bekept underpressure so thatduring programexecution, thesedo not deflateduring the cycle. | |
| "SEA WAVE" LEG-PIECEWith overlapping bands, pressure which decreases towardsthe top and programmable control of sectors,the flow is physiological. |
4
{10}------------------------------------------------
| "Leg" Attachment | Leg (consisting of foot, calf, knee, upper leg) | Leg (consisting of foot, calf, knee, upper leg) | Same |
|---|---|---|---|
| Leg AttachmentPhotos | Image: Leg attachment photo | Image: Leg attachment photo | Same |
| Leg AttachmentSizes | 51"x 26" | X-Short: 14" x 41"Short: 14" x 43"Medium: 14" x 45"Long: 14" x 48"X-Long: 14" x 52" | Different.Slight variation. |
| "Hip" Attachment | Hip (consisting of upper legs, glutes, hips, lower back) | Hip (consisting of upper legs, glutes, hips, lower back) | Same |
{11}------------------------------------------------
| "Arm"Attachment | Arm (consisting of entirearm, shoulder, upper chest and back) | Arm (consisting of entire arm,shoulder, upper chest and back) | Same |
|---|---|---|---|
| Arm AttachmentPhotos | 66 cm72 cm | Same. | |
| Arm AttachmentSizes | One size: 26" x 28 "72 cm65 cm | Regular: 18" x 38"Long: 18" x 44" | Same.Slightvariation. |
| Safety Features | Button on display allows user to stop orpause therapy session at any time. | Button on display allows user tostop or pause therapy session atany time. | Same |
| SW/Firmware/MicroprocessorControl | Microprocessor | Microprocessor | Same |
| Technology | Compressor and valve system whichsequentially inflates cells of appliance | Compressor and valve systemwhich sequentially inflates cells ofappliance | Same |
(c) Non-clinical tests- brief discussion:
Safety testing and EMC testing successfully completed. In particular, the following tests were performed:
- 1- Earth circuit continuity test, leakage current, current absorption, current absorption: Summary of findings- device complies with these IEC 60601-1 safety requirements, in the lab as well as on production units (tests are conducted on on-going basis).
- 2- Main switch, air hose connector, delivered pressure, display function, programs, top card function, safety pressure switch intervention, maximum setting pressure, integrity of the inflatable parts (multiple)
{12}------------------------------------------------
Summary of findings- device complies with the acceptance criteria of these various tests, documenting that safety and effectiveness are achieved on validation units and on production units (tests are conducted on on-going basis).
(d) Clinical tests- brief discussion
Not applicable. No clinical tests were conducted.
(e) Non-clinical and clinical tests- conclusions drawn demonstrating that the device is as safe and as effective, and performs as well as the predicate device(s): Safety of subject device is confirmed via testing. Effectiveness is ensured by similar parameter of a well-understood device mechanism.
§ 890.5650 Powered inflatable tube massager.
(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).