K Number
K201634
Date Cleared
2021-03-26

(283 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RG 3ml Medication Cartridge is designed for use in hospitals and outpatient care environments with the CADD-MS®3 Ambulatory Infusion Pump for subcutaneous infusion of medication in adults.

Device Description

The RG 3ml Medication Cartridge is a sterile, single-use, non-pyrogenic, 3.0 ml piston syringe intended for use with the Smiths Medical MD, Inc. CADD-MS®3 Ambulatory Infusion Pump. The CADD-MS®3 Ambulatory Infusion Pump requires use of a 3 ml medication cartridge and an infusion set. The RG 3ml Medication Cartridge consists of a hollow barrel, movable plunger with O-rings for sealing and a male Luer lock fitting at the distal end. The male Luer lock fitting of the cartridge is connected to the female Luer fitting of an infusion set. The reservoir is placed in the CADD-MS®3 Ambulatory Infusion Pump to achieve its intended use. The RG 3ml Medication Cartridge is available with a 0.7mm (22G) x 12.5mm (1/2in.) needle and cap. The needle is attached to the end of the RG 3ml Medication Cartridge and operates as a syringe to withdraw medication from a vial.

After the medication is filled in the RG 3ml Medication Cartridge, the plunger stopper is removed to allow the RG 3ml Medication Cartridge to fit inside the Smiths CADD-MS®3 Infusion Pump. A cartridge cap from the Smiths CADD-MS®3 Infusion Pump secures the RG 3ml Medication Cartridge into place. An infusion set is then attached to the luer lock fitting of the RG 3ml Medication Cartridge.

AI/ML Overview

The provided text describes the regulatory clearance of the RG 3ml Medication Cartridge and outlines studies conducted to demonstrate its substantial equivalence to a predicate device, primarily focusing on non-clinical performance testing.

Here's an analysis based on your request, highlighting what information is available and what is not:

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Test/StandardReported Device Performance
BiocompatibilityISO 10993-1:2009 (Use of International Standard ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process) and relevant guidance- Cytotoxicity per ISO 10993-5:2009 - Dermal Sensitization per ISO 10993-10:2010 - Intracutaneous Irritation Test per ISO 10993-10:2010 - Acute Systemic Toxicity per ISO 10993-11:2017 - Subacute Toxicity per ISO 10993-11:2017 - Pyrogenicity per ISO 10993-11:2017 - Hemolysis (Direct Contact and Indirect Contact) per ISO 10993-4:2017Result: Complied with biocompatibility requirements and are considered biocompatible.
Sterility & Shelf LifeSterility Validation per FDA recognized standard ISO 11135:2014 (Ethylene oxide sterilization), Shelf-life testingResult: A sterility assurance level (SAL) was validated to 10⁻⁶. The device demonstrated a 2-year shelf life, supported by sterility, packaging, and performance testing.
Packaging IntegrityASTM F2096 (Bubble Test), ASTM F1929 (Dye Penetration Test), ASTM F1886 (Visual Inspection), ASTM D4169 (Simulated Shipping), ISO 11607-2 (General Packaging Standard), Seal strength testingResult: Packaging testing was conducted via these FDA recognized standards.
Functional Performance (with CADD-MS®3 Ambulatory Infusion Pump)ISO 7886-2:2020 (Flow rate testing), Occlusion Alarm Testing, Cartridge Loading and Detection Testing, Cartridge Volume Alarm TestingResult: The RG 3ml Medication Cartridge, when used with the CADD-MS®3 Ambulatory Infusion Pump, passed all testing required in ISO 7886-2 and flow rate testing within pump specifications, demonstrating adequate performance and compatibility.

Important Note: The document specifies that "differences between the intended use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness." This implies that the acceptance criteria for the new device were primarily demonstrated through equivalency testing against established standards and the performance of the predicate device.

2. Sample size used for the test set and the data provenance

The document describes non-clinical performance testing for a medical device (a medication cartridge). Such tests typically involve specific numbers of units (samples) of the device and may use standardized test fluids or conditions. However, the exact sample sizes used for each test (e.g., how many cartridges for flow rate testing, how many animals for biocompatibility) are NOT specified in this document.

The data provenance is from non-clinical laboratory testing and is not tied to human subjects, countries of origin, or retrospective/prospective studies in the way clinical trials would be.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the type of testing described. The "ground truth" for non-clinical performance testing is established by recognized international standards (e.g., ISO, ASTM) and the specifications of the compatible medical device (the CADD-MS®3 Ambulatory Infusion Pump). These are objective measurements against defined criteria, not subjective expert interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable to the type of testing described, as there is no subjective interpretation or consensus among experts required for the results of the performance tests. The tests yield objective measurements against predefined acceptance criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study is designed for evaluating diagnostic tools where human interpretation is involved, often with AI assistance (e.g., radiology images). This information is not applicable as the device is a medication cartridge, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable. The device is a physical medical component, not a software algorithm.

7. The type of ground truth used

The ground truth used for evaluating the RG 3ml Medication Cartridge performance includes:

  • International Standards: e.g., ISO 10993 series for biocompatibility, ISO 11135 for sterilization, ISO 7886-2 for flow rate, ASTM standards for packaging.
  • Predicate Device Specifications: The CADD-MS®3 Ambulatory Infusion Pump's specifications for compatible cartridges and its performance characteristics served as a benchmark for comparison.
  • Defined Acceptance Criteria: Each test had specific pass/fail criteria derived from these standards and specifications.

8. The sample size for the training set

This is not applicable. The device is a physical product, not a machine learning model that requires a training set. The tests performed are engineering and biological validations, not algorithm training.

9. How the ground truth for the training set was established

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2021

Chengdu Shifeng Medical Technology Co., Ltd. % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180

Re: K201634

Trade/Device Name: RG 3ml Medication Cartridge Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: MRZ Dated: February 23, 2021 Received: February 23, 2021

Dear Maureen O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201634

Device Name RG 3ml Medication Cartridge

Indications for Use (Describe)

The RG 3ml Medication Cartridge is designed for use in hospitals and outpatient care environments with the CADD-MS3 Ambulatory Infusion Pump for subcutaneous infusion of medication in adults.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201634 510(k) SUMMARY

Chengdu Shifeng Medical Technology Co., Ltd. RG 3ml Medication Cartridge

510(k) Owner

Chengdu Shifeng Medical Technology Co., Ltd. No 31, Yinyuan Road Jinhua Town Xinjin County Chengdu, Sichuan, China

Submission Correspondent

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245

Date Prepared: March 26, 2021

Subject Device

Trade Name:RG 3ml Medication Cartridge
Common or Usual Name:Infusion Pump Syringe
Classification Name:Pump, Infusion
Device Classification:21 CFR 880.5725
ClassII
Product Code:MRZ
Classification Panel:General Hospital

Predicate Device

Trade Name:Smiths Medical MD, Inc. CADD-MS®3 Ambulatory Infusion Pump
510(k) Reference:K051568
Common or Usual Name:Infusion Pump Syringe
Classification Name:Pump, Infusion
Device Classification:21 CFR 880.5725
ClassII
Product Code:FRN
Classification Panel:General Hospital

Device Description

The RG 3ml Medication Cartridge is a sterile, single-use, non-pyrogenic, 3.0 ml piston syringe intended for use with the Smiths Medical MD, Inc. CADD-MS®3 Ambulatory Infusion Pump. The CADD-MS®3 Ambulatory Infusion Pump requires use of a 3 ml medication cartridge and an infusion set. The RG 3ml Medication Cartridge consists of a hollow barrel, movable plunger with

{4}------------------------------------------------

O-rings for sealing and a male Luer lock fitting at the distal end. The male Luer lock fitting of the cartridge is connected to the female Luer fitting of an infusion set. The reservoir is placed in the CADD-MS®3 Ambulatory Infusion Pump to achieve its intended use. The RG 3ml Medication Cartridge is available with a 0.7mm (22G) x 12.5mm (1/2in.) needle and cap. The needle is attached to the end of the RG 3ml Medication Cartridge and operates as a syringe to withdraw medication from a vial.

After the medication is filled in the RG 3ml Medication Cartridge, the plunger stopper is removed to allow the RG 3ml Medication Cartridge to fit inside the Smiths CADD-MS®3 Infusion Pump. A cartridge cap from the Smiths CADD-MS®3 Infusion Pump secures the RG 3ml Medication Cartridge into place. An infusion set is then attached to the luer lock fitting of the RG 3ml Medication Cartridge.

Indications for Use

The RG 3ml Medication Cartridge is designed for use in hospitals and outpatient care environments with the CADD-MS®3 Ambulatory Infusion Pump for subcutaneous infusion of medication in adults.

Comparison of Technological Characteristics with the Predicate Device

Table 1 provides a tabular presentation of the RG 3ml Medication Cartridge compared with the predicate device. Both the subject device and the 3 ml cartridge component of the predicate device are intended for use in delivering medication utilizing the CADD-MS®3 Ambulatory Infusion Pump. The table below also assess the differences to demonstrate that there no new or different questions of safety and effectiveness between the subject and predicate device.

RG 3ml Medication Cartridge Substantial Equivalence
CharacteristicRG 3ml MedicationCartridgeSmiths Medical MD,Inc. CADD-MS®3Ambulatory InfusionPump, 3 mLCartridge ReservoirDiscussion ofDifferences
510(k) NumberK201634K051568N/A
Product CodeMRZFRNBoth are withinInfusion pumpregulation : 21 CFR880.5725, whichsupports substantialequivalence
Indications for UseThe RG 3ml MedicationCartridge is designedfor use in hospitals andoutpatient careenvironments with theCADD-MS®3Ambulatory InfusionPump for subcutaneousThe Smiths MedicalInc. 3 ml CartridgeReservoir Indicationsfor use are:The Smiths MedicalMD, Inc. 3-mlCartridge Reservoir isdesigned for use withThe predicate deviceincludes both a pumpand cartridge. Thesubject deviceindications arecompared to thecartridge indicationsfor use and support
Table 1
RG 3ml Medication Cartridge Substantial Equivalence

{5}------------------------------------------------

infusion of medicationin adults.the CADD-MS 3 fordelivering medication.substantialequivalence
Use TypePrescription UsePrescription UseSame
Cartridge UseWith CADD-MS ® 3Ambulatory InfusionPumpWith CADD-MS 3Ambulatory InfusionPumpSame
Volume3 ml3 mlSame
Sterility StatusSterile, single useSterile, single useSame
BiocompatibilityYes per ISO 10993-1Yes per ISO 10993-1Same

Non-Clinical Performance Testing

The following non-clinical/performance testing was used to support substantial equivalence:

The materials used for the RG 3ml Medication Cartridge comply with biocompatibility requirements outlined in ISO 10993-1:2009 and the Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process and are considered to be biocompatible.

The following biocompatibility testing was performed:

TestStandard
CytotoxicityISO 10993-5:2009
Dermal SensitizationISO 10993-10:2010
Intracutaneous Irritation TestISO 10993-10 :2010
Acute Systemic ToxicityISO 10993-11 :2017
Subacute ToxicityISO 10993-11 :2017
PyrogenicityISO 10993-11 :2017
Hemolysis (Direct Contactand Indirect Contact)ISO 10993-4:2017

The device has a 2-year shelf life. This was supported by sterility, packaging, and performance testing.

Ethylene Oxide is used to sterilize the device and validation was provided per FDA recognized standard ISO 11135: 2014 - Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices. A sterility assurance level (SAL) was validated to 10-6.

Packaging testing was also conducted via the following FDA recognized standards: ASTM F2096 (Bubble Test), ASTM F1929 (Dye Penetration Test), ASTM F1886 (Visual Inspection), STM D4169 (Simulated Shipping), and ISO 11607-2 (General Packaging Standard). In addition, seal strength testing was conducted on the packaging.

The following performance testing was conducted using Smiths Medical MD, Inc. CADD-MS®3 Ambulatory Infusion Pump:

{6}------------------------------------------------

  • . ISO 7886-2:2020 (Flow rate testing using the Smiths Medical MD, Inc. CADD-MS®3 Ambulatory Infusion Pump)
  • . Occlusion Alarm Testing
  • Cartridge Loading and Detection Testing ●
  • Cartridge Volume Alarm Testing ●

In order to establish compatibility with the CADD-MS 3 Ambulatory Infusion Pump and demonstrate adequate performance within the pump specifications as compared to the Smiths Medical CADD-MS 3 Cartridge Reservoir, flow rate accuracy testing was conducted using the RG 3 mL Medication Cartridge and the CADD-MS 3 Ambulatory Infusion Pump. Results showed that the RG 3ml Medication Cartridge when used with the CADD-MS®3 Ambulatory Infusion Pump passed all testing required in ISO 7886-2 and flow rate testing within pump specifications, supporting substantial equivalence to the predicate device.

Conclusion

Differences between the intended use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing and risk management activities. The RG 3ml Medication Cartridge is Substantially Equivalent (SE) to the Smiths Medical MD. Inc. CADD-MS®3 Ambulatory Infusion Pump, 3 mL Cartridge Reservoir, cleared under K051568.

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).