K Number
K201626

Validate with FDA (Live)

Date Cleared
2021-03-24

(281 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.

Device Description

The Safety Subcutaneous Infusion Set is an intravascular administration set used to administer medication subcutaneously. The device consists of subcutaneous stainless-steel needle mounted to a wing stabilizer at 90° at the distal end of the infusion set. The subject device has closing type wings to allow the user to insert the needle into the subcutaneous tissue and also secure the needles on the infusion site. A female luer connector is provided at the proximal end to allow connection to external devices. Each set consists of the administration set and standard medical dressing to secure the needles on the infusion site. The subject device is provided in various needle gauges and lengths. Each needle is equipped with an open end slide clamp on the tubing to facilitate stop flow when required during priming of individual needles.

AI/ML Overview

This document is a 510(k) summary for the "Safety Subcutaneous Infusion Set" and does not contain information related to an AI/ML powered device. As such, it does not include acceptance criteria directly applicable to AI/ML performance, nor a study proving that an AI/ML device meets such criteria.

The information provided pertains to the substantial equivalence determination for a medical device that facilitates subcutaneous infusion of medication. Therefore, I cannot extract the requested information regarding AI device acceptance criteria or studies from this document.

However, I can provide the acceptance criteria and performance data for the physical device described in this document, which are based on non-clinical tests:

Acceptance Criteria and Device Performance (Non-Clinical for Physical Device):

Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Conclusion
BiocompatibilityISO 10993-1:2018Biocompatibility studies performed
Particulate MatterUSP <788>Particulate matter testing performed
SterilizationETO, SAL of 10-6Sterilization validation performed, achieved SAL of 10-6 (Same as predicate)
Ethylene Oxide ResidualsISO 10993-7:2008Ethylene oxide residuals testing performed
Shelf LifeASTM F1980-16Shelf life testing performed
Sharps ProtectionISO 23908:2011Sharps protection compliant to ISO 23908:2011 (Same as predicate)
Tensile Strength & Water TightnessISO 8536-8:2015Tensile strength and water tightness testing performed
Luer Lock ConnectionISO 80369-7Luer lock connection test performed
Comparison to Predicate (Technological Characteristics)(See detailed comparison table in document)Found to have similar design, intended use, proximal interface, needle lengths, number of needles, needle safety feature, wing type, and sterilization method. Differences in protective cap design, priming volume, and materials did not raise new safety/effectiveness questions.

The following requested information is not applicable to this document as it describes a physical medical device, not an AI/ML product:

  • Sample size used for the test set and the data provenance
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  • Adjudication method for the test set
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
  • If a standalone (i.e. algorithm only without human-in-the loop performance) was done
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc)
  • The sample size for the training set
  • How the ground truth for the training set was established

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 24, 2021

Epic Medical Pte. Ltd. % Roshana Ahmed Sr. Consultant, Regulatory Affairs Medical Devices Quaras, LLC 25 Independence Blvd. Warren, New Jersey 07059

Re: K201626

Trade/Device Name: Safety Subcutaneous Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA Dated: February 17, 2021 Received: February 19, 2021

Dear Roshana Ahmed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201626

Device Name Safety Subcutaneous Infusion Set

Indications for Use (Describe)

The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201626 510(k) Summary

I. Submitter

Epic Medical Pte. Ltd. 105 Cecil Street #20-04 The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292

Contact Person: Mr. Freddie Lee, Chief Executive Officer Date Prepared: February 16, 2021

II. Device

Device Proprietary Name:Safety Subcutaneous Infusion Set
Common or Usual Name:Subcutaneous Administration Set
Classification Name:Intravascular Administration Set
Regulation Number:21 CFR 880.5440
Product Code:FPA
Device ClassificationII

III. Predicate Device

Substantial equivalence is claimed to the following device:

  • SUB-Q Subcutaneous Tissue Infusion Set, K140131, EMED Technologies Corporation ●

Device Description IV.

The Safety Subcutaneous Infusion Set is an intravascular administration set used to administer medication subcutaneously. The device consists of subcutaneous stainless-steel needle mounted to a wing stabilizer at 90° at the distal end of the infusion set. The subject device has closing type wings to allow the user to insert the needle into the subcutaneous tissue and also secure the needles on the infusion site. A female luer connector is provided at the proximal end to allow connection to external devices.

Each set consists of the administration set and standard medical dressing to secure the needles on the infusion site. The subject device is provided in various needle gauges and lengths. Each needle is equipped with an open end slide clamp on the tubing to facilitate stop flow when required during priming of individual needles.

{4}------------------------------------------------

V. Indications for Use

The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.

VI. Comparison of Technological Characteristics

The Safety Subcutaneous Infusion Set has the same intended use as the predicate device (K140131). The Safety Subcutaneous Infusion Set has the same general design construction. proximal interface, number of lumens, range of needle gauge and length, tubing length, number of needles, needle safety feature, clamp type, wing type, and sterilization method/level as the predicate device (K140131).

The subject device differs from the predicate device with respect to needle protective cap design, materials of construction and packaging materials.

The table below compares the subject and predicate devices.

Safety Subcutaneous InfusionSUB-Q Subcutaneous TissueComparison
SetInfusion Set
(K201626)(K140131)
Indication forUseThe Safety Subcutaneous InfusionSet is indicated for subcutaneousinfusion of medicationadministered by an externalinfusion pump or syringe.SUB-Q Subcutaneous InfusionSet is intended to providesubcutaneous infusion ofmedicine from an externalinfusion pump or syringe.Same as predicate
ProximalinterfaceFemale Luer LockFemale Luer LockSame as predicate
Tubing length40 cm5 cm ~ 91.4 cmSimilar to predicate
Needle gauge24, 26, and 2724 and 27Similar to predicate
Needle lengths6mm, 9mm, 12mm6mm, 9mm, 12mm, 14mm and16mmSame as predicate
Number ofneedles1 - 41 - 4 (up to 8 with Y connector)Same as predicate
Needle lubricantPolydimethylsiloxane (PDMS)Polydimethylsiloxane (PDMS)Same as predicate
Wing typeClosed, Needle protectingStandard and ClosedSame as predicate
Needleprotective capSlidingTubular sheathDifferent

{5}------------------------------------------------

Safety Subcutaneous InfusionSet(K201626)SUB-Q Subcutaneous TissueInfusion Set(K140131)Comparison
Needle safetyfeatureNeedle inaccessible to user beforeengaging.Sharp protection featurecompliant to ISO 23908:2011.Needle inaccessible to user beforeengaging.Sharp protection featurecompliant to ISO 23908:2011.Same as predicate
SterilizationETO, SAL of 10-6ETO, SAL of 10-6Same as predicate
Priming volume24-gauge needle: 0.4 ~ 1.427-gauge needle: 0.1 ~ 0.424-gauge needle: 0.16 ~ 0.827-gauge needle: 0.12 ~ 0.8Similar to predicate
TransparentdressingOff-the-shelf dressing purchasedfrom:3M HealthcareOff-the-shelf dressing purchasedfrom:3M HealthcareSame as predicate
ComponentmaterialsLuer lock: ABSLuer cap: LDPETubing: PVCWings: Polypropylene andThermoplastic elastomer(locking)Protective cap: PolycarbonateNeedles: Stainless steelSlide clamp: PolyethyleneBonding agent: CyanoacrylateLuer lock: PVCLuer cap: PolypropyleneTubing: PVCWings: PVC (std) andPolypropylene (locking)Protective cap: PolypropyleneNeedles: Stainless steelSlide clamp: ABSBonding agent: CyanoacrylateDifferent

Discussion

The Safety Subcutaneous Infusion Set has the same intended use as the predicate device and both the subject and predicate device share the same design, proximal interface, needle lengths, number of needles and needle safety feature, and wing type. The subject device tubing length and needle gauges and fall within the range of tubing length and needle gauges for the predicate device.

The subject device differs from the predicate device with respect to protective cap design (sliding cap design vs. tubular sheath design), priming volume (slight difference in priming volume), and materials of construction for a subset of components. These technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness and are addressed by the performance and biological safety testing identified below.

{6}------------------------------------------------

VII. Performance Data

The following non-clinical data were provided in support of the substantial equivalence determination:

  • biocompatibility studies per ISO 10993-1:2018
  • particulate matter per USP <788>
  • sterilization validation
  • ethylene oxide residuals per ISO 10993-7:2008 ●
  • shelf life per ASTM F1980-16
  • sharps protection per ISO 23908:2011
  • . tensile strength and water tightness per ISO 8536-8:2015
  • luer lock connection test per ISO 80369-7

VIII. Conclusion

The information provided above supports that the Safety Subcutaneous Infusion Set is as safe and effective as the predicate device. Although there are minor technological differences between the subject and predicate devices, these difference do not raise different questions of safety and effectiveness. Therefore, it is concluded that the Safety Subcutaneous Infusion Set is substantially equivalent to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.