(204 days)
TENS(1,3,4,5,6,21): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9126
TENS(3,4,5,6,11,12,16): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,8,9,10,13,14,15): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9186
TENS(3,4,7,8,11): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,5,6,9,10,12): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9196
TENS(1015): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.9): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
PMS(1
TENS & PMS is a portable and DC 3.7V battery powered multifunction device with multiple models, offering both Transcutaneous Electrical Nerve Stimulation (TENS) and Powered Muscle Stimulation (PMS) qualities in one device. The device is equipped with accessories of electrode pads, electrode cables, a battery charger, and one USB cable. The electrode cables are used to connect the pads to the device; the USB cable is used to connect the charger and the built-in lithium battery. All accessories, including USB cables, electrode pads, electrode cables, chargers can only be changed or replaced by a qualified person.
The provided text describes a 510(k) premarket notification for a TENS & PMS device. This submission aims to demonstrate substantial equivalence to an already legally marketed predicate device (K121719). The document focuses on comparing the technical characteristics and performance of the new device models (SM9079, SM9126, SM9186, SM9196) against the predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria are implicitly defined by the safety and performance parameters of the legally marketed predicate device (K121719) and compliance with various recognized standards. The reported device performance is demonstrated through the comparative tables (Table 1, 2, 3, 4) where the new device's parameters are listed alongside the predicate device's, and through the statement of compliance with specific IEC and ANSI/AAMI standards.
Since the tables in the document are already in a comparative format, I will summarize the key parameters and their reported performance for one of the new models (SM9079) against the predicate. The "Remark" column in the original tables already serves as a concise statement of whether the new device meets or is similar to the predicate.
Table: Acceptance Criteria and Reported Device Performance (Example: Model SM9079 vs. Predicate K121719)
| Parameter / Acceptance Criteria (Derived from Predicate) | Reported Device Performance (New Device SM9079) | Meets Criteria? | Notes |
|---|---|---|---|
| General Characteristics | |||
| Manufacturer | Hong Qiangxing (Shenzhen) Electronics Limited | Yes | Same |
| Intended Use (TENS) | Temporary relief of pain for specific body parts due to strain | Yes | Same (Expanded list of programs, but comparable) |
| Intended Use (PMS) | Stimulate healthy muscles to improve performance | Yes | Same (Expanded list of programs, but comparable) |
| Type of Use | OTC | Yes | Same |
| Power Source | DC 3.7V lithium battery | Yes | Same |
| Method of Line Current Isolation | Type BF | Yes | Same |
| Patient Leakage Current (Normal Condition) | < 10μA | Yes | Same |
| Patient Leakage Current (Single Fault Condition) | < 50μA | Yes | Same |
| Average DC current through electrodes | < 0.01μA | Yes | Same |
| Number of Output Modes | 24 | Similar | More treatment programs, all comparable to predicate's mode 3. No new safety/effectiveness issues. |
| Number of Output Channels | 2 | Yes | Same |
| Synchronous or Alternating Channel Output | Alternating | Similar | Different from predicate's synchronous, but same technology and treatment effect. No new safety/effectiveness issues. |
| Method of Channel Isolation | Voltage transformer Isolation | Yes | Same |
| Regulated Current or Voltage | Voltage control | Yes | Same |
| Software/Firmware/Microprocessor Control | Software | Yes | Same |
| Automatic Overload Trip? | No | Yes | Same |
| Automatic No-Load Trip? | No | Yes | Same |
| Automatic Shut Off? | Yes | Yes | Same |
| User Override Control? | Yes | Yes | Same |
| Indicator (On/Off Status) | Yes | Yes | Same |
| Indicator (Low Battery) | Yes | Yes | Same |
| Indicator (Voltage/Current Level) | Yes | Yes | Same |
| Timer Range | 10 ~ 60 minutes, 10 min/step | Yes | Same |
| Biocompatibility | All user directly contacting materials comply with ISO10993-5 and ISO10993-10 requirements. | Yes | Same |
| Accessories | Self-adhesive electrodes, electrode wires, Battery charger, USB cable | Yes | Same |
| Physical Characteristics | |||
| Weight | 102g±5g | Similar | Difference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues. |
| Dimensions | 6012312mm | Similar | Difference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues. |
| Housing Materials & Construction | ABS | Yes | Same |
| Output Specifications (Max Values from Table 5 for SM9079 vs. Predicate Mode 3/1/2) | |||
| Waveform | Pulsed, symmetric, biphasic | Yes | Same |
| Shape | Rectangular | Yes | Same |
| Max Output Voltage (@500Ω) | 42V±20% | Similar | Predicate 42V±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Max Output Current (@500Ω) | 84mA±20% | Similar | Predicate 84mA±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Pulse Width | 190µs (max) | Similar | Predicate 100µs. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Pulse Period | 40-1070ms (range) | Similar | Predicate 9.3~850ms. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Max Pulse Frequency | 0.935-25Hz±10% (range) | Similar | Predicate 110Hz. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Net Charge (per pulse) | 0µC @500Ω; Balanced waveform | Yes | Same |
| Max Phase Charge (@500Ω) | 13.30µC@500Ω | Similar | Predicate 16.8µC@500Ω. Differences in these parameters allowed as they don't exceed safety limits and passed IEC 60601-2-10 tests. |
| Max Current Density (@500Ω) | 0.065mA/cm² | Yes | Meets max current density <2mA/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126). |
| Max Average Power Density (@500Ω) | 2.64mW/cm² | Yes | Meets max average power density <0.25W/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126). |
| Compliance with Standards | |||
| ANSI AAMI ES60601-1 | Yes | Yes | Same |
| IEC 60601-2-10 | Yes | Yes | Same |
| ANSI AAMI IEC 60601-1-2 | Yes | Yes | Same |
| IEC 62133 | Yes | Yes | Same |
| ANSI AAMI HA60601-1-11 | Yes | Yes | Same |
| 21 CFR 898 (if applicable) | Yes | Yes | Same |
2. Sample size used for the test set and the data provenance
The document explicitly states that the new devices have "passed testings according to the following standards" and that the "conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device." These "testings" refer to compliance with various medical device standards (ANSI AAMI ES60601-1, IEC 60601-2-10, ANSI AAMI IEC 60601-1-2, IEC 62133, ANSI AAMI HA60601-1-11).
However, the document does not provide details about a clinical study or a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective). The submission is based on a comparison of technical specifications and performance against a predicate device and adherence to recognized standards, rather than a separate clinical trial with a defined sample size. The testing appears to be functional and electrical safety testing as per the listed standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Given that a clinical study or human-in-the-loop performance evaluation with a test set requiring expert ground truth is not described in the provided text, information on the number and qualifications of experts for establishing ground truth is not available. The evaluation is based on technical specifications and compliance with standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Since no clinical test set requiring human adjudication is mentioned, there is no information provided on any adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
The device described is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS). It is a physical therapy device and not an AI-assisted diagnostic or imaging interpretation device. Therefore, a Multi Reader Multi Case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance would be irrelevant and is not mentioned in this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Again, this is a physical therapy device, not an algorithm, so no standalone algorithm performance evaluation is applicable or mentioned.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this type of device, the "ground truth" for demonstrating substantial equivalence and safety/effectiveness relies on:
- Compliance with recognized electrical safety and performance standards: This includes specific limits for voltage, current, pulse characteristics, leakage current, and electromagnetic compatibility.
- Comparison of technical specifications with a legally marketed predicate device: The predicate device itself has established safety and effectiveness. The "ground truth" is that the predicate device is safe and effective when used as intended, and the new device demonstrates similar performance and characteristics to be considered equivalent.
- Biocompatibility testing: Compliance with ISO10993-5 and ISO10993-10 for materials directly contacting the user.
There is no mention of expert consensus, pathology, or outcomes data from a clinical trial for this specific submission. The "ground truth" is established by adherence to existing regulatory frameworks and engineering standards.
8. The sample size for the training set
As this is not an AI/machine learning device, the concept of a "training set" is not applicable and therefore, no sample size for a training set is provided.
9. How the ground truth for the training set was established
As the concept of a training set is not applicable, no information is provided on how its ground truth would be established.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 11, 2020
Hong Qiangxing (Shen Zhen) Electronics Limited % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No. 19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China
Re: K201354
Trade/Device Name: TENS & PMS Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX. NUH Dated: September 3, 2020 Received: September 14, 2020
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201354
Indications for Use (Describe)
Device Name TENS & PMS
| Model: | SM9079 |
|---|---|
| TENS | (1,3,4,5,6,21): |
| To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. | |
| PMS | (2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24): |
| It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | |
| Model: | SM9126 |
| TENS | (3,4,5,6,11,12,16): |
| To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. | |
| PMS | (1,2,7,8,9,10,13,14,15): |
| It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | |
| Model: | SM9186 |
| TENS | (3,4,7,8,11): |
| To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. | |
| PMS | (1,2,5,6,9,10,12): |
| It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | |
| Model: | SM9196 |
| TENS | (10~15): |
| To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. | |
| PMS | (1~9): |
| It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ----------------------------------------------------------------------- | --------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
510(k) Summary
[As required by 21 CFR 807.92]
1. Submission Information:
| 510(k) Number: | K201354 |
|---|---|
| Date: | December 11, 2020 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | Hong Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'anDistrict, Shenzhen City, Guangdong, China 518126 |
| Contact: | Doris Dong[Consultant, from Shanghai CV Technology Co., Ltd.]Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 ChinaE-mail: doris_d@126.comTel: 86 21-31261348 / Fax: 86 21-57712250 |
| 2. Device Description: | |
| Proprietary Name: | TENS & PMS |
| Common Name: | TENS & PMS |
| Classification Name: | Powered muscle stimulatorTranscutaneous electrical nerve stimulator for pain relief |
| Regulation Number: | 21 CFR 890.5850, 21 CFR 882.5890 |
| Product Code: | NGX, NUH |
| Device Class: | II |
| Review Panel: | Neurology & Physical Medicine |
| Device Description: | TENS & PMS is a portable and DC 3.7V battery powered multifunctiondevice with multiple models, offering both Transcutaneous Electrical NerveStimulation (TENS) and Powered Muscle Stimulation (PMS) qualities inone device.The device is equipped with accessories of electrode pads, electrode cables,a battery charger, and one USB cable. The electrode cables are used toconnect the pads to the device; the USB cable is used to connect the chargerand the built-in lithium battery. All accessories, including USB cables,electrode pads, electrode cables, chargers can only be changed or replacedby a qualified person. |
| Indications for use: | Model: SM9079TENS(1,3,4,5,6,21):To be used for temporary relief of pain associated with sore and achingmuscles in the shoulder, waist, back, neck, upper extremities (arm), andlower extremities (leg) due to strain from exercise or normal householdwork activities.PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24):It is intended to be used to stimulate healthy muscles in order to improveand facilitate muscle performance. |
{5}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Model: SM9126
TENS(3.4.5.6.11.12.16):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,8,9,10,13,14,15):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9186
TENS(3,4,7,8,11):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,5,6,9,10,12):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9196
TENS(10~15):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1~9):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
{6}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 3. Working principle
The device uses electrodes to send a specific electric current to the skin. TENS(Transcutaneous electrical nerve stimulation) mode is to relieve pain using electrical stimulation. PMS(Powered Muscle Stimulation) mode is to stimulate healthy muscles in order to improve performance by electrical stimulation.
4. Substantial Equivalence to Predicate device: Basic Unit Characteristics
Table 1-
| Parameters | New Device | Predicate Device | Remark | |
|---|---|---|---|---|
| 510(k) Number | K201354 | K121719 | -- | |
| Device Name | TENS & PMS | SM TENS & PMS | -- | |
| Model | SM9079 | No | -- | |
| Manufacturer | Hong Qiangxing (Shenzhen)Electronics Limited | Hong Qiangxing (Shenzhen)Electronics Limited | Same | |
| Intended use | TENS(1,3 | TENS(1,3,4,5,6):To be used for temporary reliefof pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(1,2,3,6):It is intended to be used tostimulate healthy musclesinorder to improve and facilitatemuscle performance. | Same | |
| Type of use | OTC | OTC | Same | |
| Power Source(s) | DC 3.7V lithium battery | DC3.7Vlithium battery | Same | |
| - Method of LineCurrent Isolation | Type BF | Type BF | Same | |
| Patient Leakage | -- | Same | ||
| Current | ||||
| Normal | < 10μΑ | < 10μΑ | ||
| Condition (μΑ) | ||||
| - Single Fault | < 50μA | < 50μA | ||
| Condition (μΑ) | ||||
| Average | DC | < 0.01μA | < 0.01μA | Same |
| current | through | |||
| electrodes | when | |||
| device is on but no | ||||
| pulses are being | ||||
| applied (µA) | ||||
| Number of Output | 24 | 6 | Similar | |
| Modes | Note 1 | |||
| Number of Output | 2 | 2 | Same | |
| channels: | ||||
| Alternating | Synchronous | Similar Note 2 | ||
| Synchronous | Or | |||
| Alternating? | ||||
| - Method of | Voltage transformer | Voltage transformer | Same | |
| Channel Isolation | Isolation | Isolation | ||
| Regulated Current | Voltage control | Voltage control | Same | |
| or | Regulated | |||
| Voltage? | ||||
| Software/Firmwar | ||||
| Software | Software | Same | ||
| e/MicroprocessorControl? | ||||
| Automatic | No | No | Same | |
| Overload Trip? | ||||
| Automatic | No | No | Same | |
| No-Load Trip? | ||||
| Automatic | Shut | Yes | Yes | Same |
| Off? | ||||
| User | Override | Yes | Yes | Same |
| Control? | ||||
| Indicato | On/Off | Yes | Yes | Same |
| r | Status | |||
| Display | Low | Yes | Yes | Same |
| Battery | ||||
| Voltage | Yes | Yes | Same | |
| /Curren | ||||
| t Level | ||||
| Timer | Range | 10 ~ 60 minutes, 10 | 10 ~ 60 minutes, 10 | Same |
| (minutes) | min/step | min./step | ||
| Compliance | with | Yes. | Yes. | Same |
| Voluntary | AAMI/ANSI ES 60601-1 | AAMI/ANSI | ||
| Hong Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||||
| Standards? | IEC 60601-1-2, IEC60601-2-10, IEC 62133,HA 60601-1-11 | ES60601-1, IEC 60601-1-2, IEC60601-2-10, IEC62133, IEC 60601-1-11 | Same | |
| Compliance with21 CFR 898? | Yes | Yes | Same | |
| Weight (grams) | 102g±5g | 140g | Similar Note 1 | |
| Dimensions (mm)[W x H x D] | 6012312mm | 83429mm | ||
| Housing Materials& Construction | ABS | ABS | Same | |
| Waveform | Pulsed, symmetric,biphasic | Pulsed, symmetric,biphasic | Same | |
| Shape | Rectangle | Rectangle | Same | |
| Maximum OutputVoltage (volts) | 42V±20% @500Ω | 42V±10% @500Ω | Similar Note 3 | |
| Maximum OutputCurrent (specifyunits) | 84mA±20% @500Ω | 84mA±10% @500Ω | ||
| Pulse width (µsec) | 190µs | 100µs | ||
| Pulse Period(msec) | 40-1070ms | 9.3~850ms | ||
| Max. pulsefrequency (Hz) [orRate (pps)] | 0.935-25Hz±10% | 110Hz | ||
| Net Charge (µCper pulse) | 0µC @500Ω; Method:Balanced waveform | 0µC @500Ω; Method: Balancedwaveform | Same | |
| Maximum PhaseCharge, (µC) | 13.30µC@500Ω | 16.8µC @500Ω | Similar | |
| MaximumAverage Current,(mA) | 0.78mA | 0.924mA | Similar Note 4 | |
| Maximum CurrentDensity, (mA/cm²,r.m.s.) | 0.065mA/cm² @500Ω | 0.462mA/cm² | ||
| MaximumAverage PowerDensity,(mW/cm²) | 2.64mW/cm² @500Ω | 9.702mW/cm² | ||
| Accessories | Self-adhesive electrodes,electrode wires, Battery charger,USB cable | Self-adhesive electrodes,electrode wires,Battery charger,USB cable | Same | |
| Biocompatibility | All user directly | All user directly | Same |
{7}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zon, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
{8}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
{9}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| contacting materials arecompliance withISO10993-5 and ISO10993-10requirements. | contacting materials arecompliance withISO10993-5 and ISO10993-10requirements. |
|---|---|
| -------------------------------------------------------------------------------------------- | -------------------------------------------------------------------------------------------- |
Comparison in details:
Note 1:
The proposed device SM9079 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9079 are a little different from predicate device K121719, but these differences will not raise any safety or effectiveness issue.
Note 2:
The output channels of the proposed device is alternating while the predicate device is synchronous. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.
Note 3:
There are some differences on the pulse width, frequency, pulse period and deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. Therefore, these differences won't raise any new safety and effectiveness issues.
Note 4:
The maximum average current of proposed device is smaller than that of the predicate device which means the better safety. The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them meet maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectivenessissues.
Final conclusion:
The subject device SM9079 is substantial Equivalent to the predicate device K121719.
Table 2-
| Parameters | New Device | Predicate Device | Remark |
|---|---|---|---|
| 510(k) Number | K201354 | K121719 | -- |
{10}------------------------------------------------
| 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||||||||
|---|---|---|---|---|---|---|---|---|
| Device Name | TENS & PMS | SM TENS & PMS | -- |
|---|---|---|---|
| Model | SM9126 | No | -- |
| Manufacturer | Hong Qiangxing (Shenzhen)Electronics Limited | Hong Qiangxing (Shenzhen)Electronics Limited | Same |
| Intended use | TENS(3 | TENS(1,3,4,5,6):To be used for temporary relief ofpain associated with sore and achingmuscles in the shoulder, waist, back,neck, upper extremities (arm), andlower extremities (leg) due to strainfrom exercise or normal householdwork activities.PMS(1,2,3,6):It is intended to be used to stimulatehealthy muscles in order to improveand facilitate muscle performance. | Same |
| Type of use | OTC | OTC | Same |
| Power Source(s) | DC 3.7V lithium battery | DC 3.7V lithium battery | Same |
| - Method of LineCurrent Isolation | Type BF | Type BF | Same |
| - Patient LeakageCurrent | -- | -- | Same |
| -- NormalCondition (μΑ) | < 10μΑ | < 10μΑ | |
| - Single FaultCondition (μΑ) | < 50μA | < 50μA | |
| Average DC currentthrough electrodeswhen device is onbut no pulses arebeing applied (µA) | < 0.01μA | < 0.01μA | Same |
| Number of OutputModes | 16 | 6 | SimilarNote 1 |
| Number of Outputchannels: | 2 | 2 | Same |
| - Synchronousor Alternating? | Synchronous | Synchronous | Same |
| - Method ofChannel Isolation | Voltage transformer Isolation | Voltage transformer Isolation | Same |
| Regulated Currentor RegulatedVoltage? | Voltage control | Voltage control | Same |
| Software/Firmware/ | Software | Software | Same |
{11}------------------------------------------------
| Hồng Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||||
|---|---|---|---|---|
| MicroprocessorControl? | ||||
| Automatic OverloadTrip? | No | No | Same | |
| Automatic No-LoadTrip? | No | No | Same | |
| Automatic Shut Off? | Yes | Yes | Same | |
| User | OverrideControl? | Yes | Yes | Same |
| Indicator | On/OffStatus? | Yes | Yes | Same |
| Display | LowBattery? | Yes | Yes | Same |
| Voltage/CurrentLevel? | Yes | Yes | Same | |
| Timer(minutes) | Range | 10 ~ 60 minutes, 10 min/step | 10 ~ 60 minutes, 10 min./step | Same |
| Compliance | withVoluntaryStandards? | Yes.AAMI/ANSI ES 60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, HA 60601-1-11 | Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, IEC 60601-1-11 | Same |
| Compliance with 21CFR 8988? | Yes | Yes | Same | |
| Weight (grams) | 55g±5g | 140g | Similar | |
| Dimensions (mm)[W x H x D] | 50.3793.410.4mm | 83429mm | Note 1 | |
| Housing | Materials& Construction | ABS | ABS | Same |
| Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Same | |
| Shape | Rectangular | Rectangular | Same | |
| Maximum OutputVoltage (volts) | 46.4V±20% @500Ω | 42V±10% @500Ω | SimilarNote 2 | |
| Maximum | OutputCurrent (specifyunits) | 92.8mA±20% @500Ω | 84mA±10% @500Ω | Same |
| Pulse width (usec) | 220μs | 100μs | Same | |
| Pulse Period (msec) | 5-680ms | 9.3~850ms | Same | |
| Max. pulsefrequency (Hz) [orRate (pps)] | 1.47-212.5Hz±10% | 110Hz | Same | |
| Net Charge (μC perpulse) | 0μC @500Ω; Method: Balancedwaveform | 0μC @500Ω; Method: Balancedwaveform | Same | |
| Maximum | PhaseCharge, (μC) | 17.26μC@500Ω | 16.8μC @500Ω | Similar |
{12}------------------------------------------------
| 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||
|---|---|---|
| Maximum AverageCurrent, (mA) | 2.056mA | 0.924mA | SimilarNote 3 |
|---|---|---|---|
| Maximum CurrentDensity, (mA/cm²,r.m.s.) | 0.119mA/cm² @500Ω | 0.462mA/cm² | |
| Maximum AveragePower Density,(mW/cm²) | 5.21mW/cm² @500Ω | 9.702mW/cm² | |
| Biocompatibility | All user directly contactingmaterials are compliance withISO10993-5 and ISO10993-10requirements. | All user directly contacting materialsare compliance with ISO10993-5 andISO10993-10 requirements. | Same |
| Accessories | Self-adhesive electrodes,electrode wires, Battery charger,USB cable | Self-adhesive electrodes, electrodewires, Battery charger, USB cable | Same |
Comparison in details:
Note 1:
The proposed device SM9126 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 1 to mode 3 of the proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9126 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.
Note 2:
There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.
Note 3:
The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them met maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.
Final conclusion:
The subject device SM9126 is substantial Equivalent to the predicate device K121719.
{13}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 Table 3-
| Parameters | New Device | Predicate Device | Remark | ||
|---|---|---|---|---|---|
| 510(k) Number | K201354 | K121719 | -- | ||
| Device Name | TENS & PMS | SM TENS & PMS | -- | ||
| Model | SM9186 | No | -- | ||
| Manufacturer | Hong Qiangxing(Shenzhen) ElectronicsLimited | Hong Qiangxing (Shenzhen)Electronics Limited | -- | ||
| Intended use | TENS(3,4,7,8,11):To be used for temporaryrelief of pain associatedwith sore and achingmuscles in the shoulder,waist, back, neck, upperextremities (arm), andlower extremities (leg) dueto strain from exercise ornormal household workactivities.PMS(1,2,5,6,9,10,12):It is intended to be used tostimulate healthy musclesin order to improve andfacilitatemuscleperformance. | TENS(1,3,4,5,6):To be used for temporaryrelief of pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(1,2,3,6):It is intended to be used tostimulate healthy muscles inorder to improve and facilitatemuscle performance. | Same | ||
| Type of use | OTC | OTC | Same | ||
| Power Source(s) | DC 3.7V lithium battery | DC 3.7V lithium battery | Same | ||
| - Method of Line Current Isolation | Type BF | Type BF | Same | ||
| - Patient Leakage Current- Normal Condition (μΑ) | < 10μA | < 10μA | Same | ||
| - Single Fault Condition (μΑ) | < 50μA | < 50μA | |||
| Average DC current throughelectrodes when device is on but nopulses are being applied (µA) | < 0.01μA | < 0.01μA | Same | ||
| Number of Output Modes | 12 | 6 | SimilarNote 1 | ||
| Number of Output channels: | 2 | 2 | Same | ||
| - Synchronous or Alternating? | Alternating | Synchronous | SimilarNote 2 | ||
| - Method of Channel Isolation | Voltage transformerIsolation | Voltage transformer Isolation | Same | ||
| RegulatedVoltage? | Current or Regulated | Voltage control | Voltage control | Same | |
| Software/Firmware/MicroprocessorControl? | Software | Software | Same | ||
| Automatic Overload Trip? | No | No | Same | ||
| Automatic No-Load Trip? | No | No | Same | ||
| Automatic Shut Off? | Yes | Yes | Same | ||
| User Override Control? | Yes | Yes | Same | ||
| IndicatorOn/Off Status | Yes | Yes | Same | ||
| Display | Low Battery | Yes | Yes | Same | |
| Voltage/CurrentLevel | Yes | Yes | Same | ||
| Timer Range (minutes) | 10 ~ 60 minutes, 10min/step | 10 ~ 60 minutes, 10 min./step | Same | ||
| ComplianceStandards? | withVoluntary | Yes.AAMI/ANSI ES 60601-1 | Yes.AAMI/ANSI ES60601-1 IEC | Same | |
| 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | |||||
| IEC 60601-1-2, IEC 60601-2-10, IEC 62133, HA 60601-1-11 | IEC 60601-1-2, IEC 60601-2-10, IEC 62133, IEC 60601-1-11 | ||||
| Compliance with 21 CFR 8988? | Yes | Yes | Same | ||
| Weight (grams) | $50\pm5$ g | 140g | Similar | ||
| Dimensions (mm) [W x H x D] | 52.99211mm | 83429mm | Note 1 | ||
| Housing Materials & Construction | ABS | ABS | Same | ||
| Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Same | ||
| Shape | Rectangular | Rectangular | Same | ||
| Maximum Output Voltage (volts) | $68V\pm20% @500\Omega$ | $42V\pm10% @500\Omega$ | Similar | ||
| Maximum Output Current (specify units) | $136mA\pm20% @500\Omega$ | $84mA\pm10% @500\Omega$ | Note 3 | ||
| Pulse width (μsec) | 220-230μs | 100μs | |||
| Pulse Period (msec) | 6.4-901ms | 9.3~850ms | |||
| Max. pulse frequency (Hz) [or Rate (pps)] | 1.11-155.6Hz | 110Hz | |||
| Net Charge (μC per pulse) | 0μC @500Ω; Method: Balanced waveform | 0μC @500Ω; Method: Balanced waveform | Same | ||
| Maximum Phase Charge, (μC) | 26μC@500Ω | 16.8μC @500Ω | SimilarNote 4 | ||
| Maximum Average Current, (mA) | 5.54mA@500Ω | 0.924mA | |||
| Maximum Current Density, (mA/cm², r.m.s.) | 0.46mA/cm² @500Ω | 0.462mA/cm² | Same | ||
| Maximum Average Power Density, (mW/cm²) | 31.40mW/cm² @500Ω | 9.702mW/cm² | SimilarNote 4 | ||
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | Same | ||
| Accessories | Self-adhesive electrodes, electrode wires, Battery charger, USB cable | Self-adhesive electrodes, electrode wires, Battery charger, USB cable | Same | ||
| Table 4- | |||||
| Parameters | New Device | Predicate Device | Remark | ||
| 510(k) Number | K201354 | K121719 | -- | ||
| Device Name | TENS & PMS | SM TENS & PMS | -- | ||
| Model | SM9196 | No | -- | ||
| Manufacturer | Hong Qiangxing (Shenzhen)Electronics Limited | Hong Qiangxing (Shenzhen)Electronics Limited | Same | ||
| Intended use | TENS(10 | TENS(1,3,4,5,6):To be used for temporary relief ofpain associated with sore and achingmuscles in the shoulder, waist, back,neck, upper extremities (arm), andlower extremities (leg) due to strainfrom exercise or normal householdwork activities.PMS(1,2,3,6):It is intended to be used to stimulatehealthy muscles in order to improveand facilitate muscle performance. | Same | ||
| Type of use | OTC | OTC | Same | ||
| Power Source(s) | DC 3.7V lithium battery | DC 3.7V lithium battery | Same | ||
| - Method of LineCurrent Isolation | Type BF | Type BF | Same | ||
| - Patient LeakageCurrent | --Normal Condition (μΑ)< 10μΑSingle Fault Condition (μΑ)< 50μΑ | --< 10μΑ< 50μΑ | Same | ||
| Average DC current through electrodeswhen device is on but no pulses arebeing applied (μΑ) | < 0.01μA | < 0.01μA | Same | ||
| Number of OutputModes | 15 | 6 | SimilarNote 1 | ||
| Number of Outputchannels:- Synchronousor Alternating? | 4Synchronous & Alternating | 2Synchronous | SimilarNote 2 | ||
| - Method ofChannel Isolation | Voltage transformer Isolation | Voltage transformer Isolation | |||
| 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | |||||
| Regulated Current orRegulated Voltage? | Voltage control | Voltage control | Same | ||
| Software/Firmware/MicroprocessorControl? | Software | Software | Same | ||
| Automatic OverloadTrip? | No | No | Same | ||
| Automatic No-LoadTrip? | No | No | Same | ||
| Automatic Shut Off? | Yes | Yes | Same | ||
| UserOverrideControl? | Yes | Yes | Same | ||
| Indicator | On/OffStatus? | Yes | Yes | Same | |
| Display | LowBattery? | Yes | Yes | Same | |
| Voltage/CurrentLevel? | Yes | Yes | Same | ||
| Timer(minutes) | Range | 60 minutes | 10 ~ 60 minutes, 10 min./step | SimilarNote 2 | |
| CompliancewithVoluntaryStandards? | Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, HA 60601-1-11 | Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, IEC 60601-1-11 | Same | ||
| Compliance with 21CFR 8988? | Yes | Yes | Same | ||
| Weight (grams) | $100g\pm5g$ | 140g | SimilarNote 1 | ||
| Dimensions(mm)[W x H x D] | 58mm180mm15.5mm | 83429mm | SimilarNote 1 | ||
| Housing Materials &Construction | ABS | ABS | Same | ||
| Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Same | ||
| Shape | Rectangle | Rectangle | Same | ||
| MaximumVoltage (volts) | Output | $54V\pm20% @500\Omega$ | $42V\pm10% @500\Omega$ | SimilarNote 3 | |
| MaximumCurrent(specifyunits) | Output | $108mA\pm20% @500\Omega$ | $84mA\pm10% @500\Omega$ | ||
| Pulse width (μsec) | 200-210μs | 100μs | |||
| Pulse Period (msec) | 20.4-620ms | 9.3~850ms | |||
| Max.frequency (Hz) [orRate (pps)] | pulse | 1.613-49.02Hz | 110Hz | ||
| Net Charge (µC perpulse) | 0µC @500Ω; Method: Balanced waveform | 0µC @500Ω; Method: Balanced waveform | Same | ||
| MaximumCharge, (µC) | Phase | $17\mu C@500\Omega$ | $16.8\mu C @500\Omega$ | Similar | |
| Maximum Average | Current, (mA) | 1.88mA | 0.924mA | SimilarNote 4 | |
| MaximumDensity,r.m.s.) | Current(mA/cm2, | $0.08mA/cm^2 @500\Omega$ | $0.462mA/cm^2$ | ||
| MaximumPower(mW/cm2) | AverageDensity, | $3.65mW/cm^2 @500\Omega$ | $9.702mW/cm^2$ | ||
| Biocompatibility | All user directly contacting materials | Same |
{14}------------------------------------------------
Comparison in details:
Note 1:
The proposed device SM9186 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 4 to mode 6 are variation waveform of the mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9186 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.
Note 2:
The output channels of the proposed device is alternating while the predicate device is synchronous. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.
Note 3:
There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum
{15}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.
Note 4:
The maximum phase charge of the proposed device SM9186 is larger than that of the predicate device. Both of them have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. The maximum average power density of proposed device is different from the predicate device due to they are calculated by different electrode area. Both of them meet maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.
Final conclusion:
The subject device SM9186 is substantial Equivalent to the predicate device K121719.
{16}------------------------------------------------
{17}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| are compliance with ISO10993-5 andISO10993-10 requirements. | are compliance with ISO10993-5 andISO10993-10 requirements. | ||
|---|---|---|---|
| Accessories | Self-adhesive electrodes, electrode wires, Battery charger, USB cable | Self-adhesive electrodes, electrode wires, Battery charger, USB cable | Same |
Comparison in details:
Note 1:
The proposed device SM9196 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9196 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.
Note 2:
The proposed device has 4 output channels are alternating & synchronous while the predicate device has 2 output channels and these channels are synchronous. The proposed device gives users more choices. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.
Note 3:
There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.
Note 4:
The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them meet maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.
Final conclusion:
The subject device SM9196 is substantial Equivalent to the predicate device K121719.
{18}------------------------------------------------
4. Substantial Equivalence to Predicate device: Output Specifications
Table 5-SM 9079
| Parameter | New Device | Predicate | New Device | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Program 1 | Program 3 | Program 4 | Program 5 | Program 6 | Program 21 | Mode 3 | Program 2 | Program 7 | Program 8 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum | @500Ω | 34.4 | 42 | 34.4 | 34.4 | 40.8 | 40.8 | 42 | 38.4 | 33.6 | 33.6 |
| Output Voltage | @2kΩ | 58.4 | 83 | 58.4 | 58.4 | 70 | 70 | 84 | 65.2 | 56.8 | 56.8 |
| (V) | @10kΩ | 107 | 144 | 107 | 107 | 123 | 123 | 130 | 117 | 104 | 104 |
| Maximum | @500Ω | 68.8 | 84 | 68.8 | 68.8 | 81.6 | 81.6 | 84 | 76.8 | 67.2 | 67.2 |
| Output Current | @2kΩ | 29.2 | 41.5 | 29.2 | 29.2 | 35 | 35 | 42 | 32.6 | 28.4 | 28.4 |
| (mA) | @10kΩ | 10.7 | 14.4 | 10.7 | 10.7 | 12.3 | 12.3 | 13 | 11.7 | 10.4 | 10.4 |
| Pulse Width(μs) | 190 | 190 | 190 | 190 | 190 | 190 | 100 | 190 | 190 | 190 | |
| Frequency (Hz) | 23.8 | 1.613 | 24.04 | 1.667~25.0 | 0.935~7.69 | 0.935~25 | 1.17 | 4.762 | 24.04 | 23.8 | |
| For interferential modes only: | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| -Beat Frequency (Hz) | |||||||||||
| For multiphasic | Symmetricalphases | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| waveforms | |||||||||||
| only: | Phaseduration | 190 | 190 | 190 | 190 | 190 | 190 | 100 | 190 | 190 | 190 |
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 11.094 | 13.87 | 11.36 | 11.36 | 13.32 | 13.22 | 16.8 | 12.12 | 11.35 | 11.08 | |
| Maximum Current Density (mA/cm²) | 0.052 | 0.004 | 0.05 | 0.054 | 0.02 | 0.065 | 0.0049 | 0.0116 | 0.051 | 0.051 | |
| Maximum Power Density, (mW/cm²) | 1.78 | 0.18 | 1.8 | 1.87 | 0.81 | 2.64 | 0.103 | 0.44 | 1.72 | 1.70 | |
| Burst Mode | (a) Pulses perburst | NA | NA | NA | NA | NA | NA | NA | 17 | 79 | 49 |
| (b) Bursts persecond | NA | NA | NA | NA | NA | NA | NA | 0.196 | 0.164 | 0.327 | |
| (c) Burstduration(seconds) | NA | NA | NA | NA | NA | NA | NA | 3.52 | 3.3 | 2.05 | |
| (d) DutyCycle: Line(b) x Line (c) | NA | NA | NA | NA | NA | NA | NA | 0.69 | 0.54 | 0.67 | |
| ON Time (seconds) | NA | NA | NA | NA | NA | NA | NA | 3.52 | 3.3 | 2.05 | |
| OFF Time (seconds) | NA | NA | NA | NA | NA | NA | NA | 1.58 | 2.79 | 1 | |
| Additional Features | N/A | N/A | N/A | N/A | NA | NA | N/A | N/A | N/A | N/A | |
| Continue | |||||||||||
| ParameterNew device | Predicate | New Device | |||||||||
| Mode or program name | Mode 2 | Program 9 | Program 10 | Program 11 | Program 12 | Program 13 | Program 15 | Program 16 | Program 17 | ||
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | ||
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum | @500Ω | 32.8 | 42 | 34.8 | 34.8 | 34.4 | 34.4 | 34.4 | 34.4 | 34.4 | 34.4 |
| Output Voltage | @2kΩ | 56 | 84 | 58.8 | 58.8 | 58.4 | 58.4 | 58.4 | 58.4 | 58.4 | 58.4 |
| (V) | @10kΩ | 103 | 130 | 107 | 107 | 107 | 107 | 107 | 107 | 107 | 107 |
| Maximum | @500Ω | 65.6 | 84 | 69.6 | 69.6 | 68.8 | 68.8 | 68.8 | 68.8 | 68.8 | 68.8 |
| Output Current | @2kΩ | 28 | 42 | 29.4 | 29.4 | 29.2 | 29.2 | 29.2 | 29.2 | 29.2 | 29.2 |
| (mA) | @10kΩ | 10.3 | 13 | 10.7 | 10.7 | 10.7 | 10.7 | 10.7 | 10.7 | 10.7 | 10.7 |
| Pulse Width(us) | 100 | 190 | 190 | 190 | 190 | 190 | 190 | 190 | 190 | ||
| Frequency (Hz) | 25 | 12.5~55.5 | 25 | 23.8 | 25 | 23.8 | 24.39 | 25 | 24.04 | 25 | |
| For interferential modes only:Beat Frequency (Hz) | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| For multiphasicwaveforms | Symmetricalphases | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| only: | PhaseDuration | 190 | 100 | 190 | 190 | 190 | 190 | 190 | 190 | 190 | 190 |
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 10.56 | 16.8 | 11.415 | 11.42 | 11.17 | 11.08 | 11.49 | 11.53 | 11.55 | 11.65 | |
| Maximum | CurrentDensity(mA/cm²) | 0.052 | 0.2331 | 0.055 | 0.053 | 0.054 | 0.052 | 0.053 | 0.054 | 0.052 | 0.054 |
| Maximum | PowerDensity,(mW/cm²) | 1.70 | 4.895 | 1.92 | 1.83 | 1.87 | 1.78 | 1.83 | 1.87 | 1.8 | 1.87 |
| Burst Mode | (a) Pulses perburst | 130 | 250~1110 | 291 | 129 | 124 | 218 | 161 | 76.5 | 149 | 291 |
| (b) Bursts persecond | 0.124 | 0.048 | 0.067 | 0.133 | 0.167 | 0.083 | 0.116 | 0.245 | 0.139 | 0.067 | |
| (c) Burstduration(seconds) | 5.18 | 20 | 11.62 | 5.4 | 4.96 | 9.14 | 6.62 | 3.06 | 6.2 | 11.64 | |
| (d) DutyCycle: Line(b) x Line (c) | 0.64 | 0.96 | 0.78 | 0.72 | 0.83 | 0.76 | 0.77 | 0.75 | 0.86 | 0.78 | |
| ON Time (seconds) | 5.18 | 20 | 11.62 | 5.4 | 4.96 | 9.14 | 6.62 | 3.06 | 6.2 | 11.64 | |
| OFF Time (seconds) | 2.94 | 1 | 3.24 | 2.12 | 1 | 2.96 | 1.97 | 1 | 1 | 3.26 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| Continue |
{19}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
{20}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| Parameter | New Device | Predicated Device | ||||||
|---|---|---|---|---|---|---|---|---|
| Mode or program name | Program 18 | Program 19 | Program 20 | Program 22 | Program 23 | Program 24 | Mode 1 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum | @500Ω | 34.4 | 34.4 | 34.4 | 38 | 34.4 | 34.4 | 42 |
| Output Voltage | @2kΩ | 58.4 | 58.4 | 58.4 | 58.4 | 58.4 | 58.4 | 84 |
{21}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| (V) | @10kΩ | 107 | 107 | 107 | 115 | 107 | 107 | 130 |
|---|---|---|---|---|---|---|---|---|
| Maximum | @500Ω | 68.8 | 68.8 | 68.8 | 76 | 68.8 | 68.8 | 84 |
| Output | @2kΩ | 29.2 | 29.2 | 29.2 | 29.2 | 29.2 | 29.2 | 42 |
| Current(mA) | @10kΩ | 10.7 | 10.7 | 10.7 | 11.5 | 10.7 | 10.7 | 13 |
| Pulse Width(us) | 190 | 190 | 190 | 190 | 190 | 190 | 100 | |
| Frequency (Hz) | 24.5 | 24.5 | 24.04 | 4.76~25.0 | 23.8 | 23.8 | 68 | |
| For interferential modes only: | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| -Beat Frequency (Hz) | ||||||||
| For multiphasic | Symmetrical phases | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| waveforms | Phase Duration | 190 | 190 | 190 | 190 | 190 | 190 | 100 |
| only: | ||||||||
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 11.32 | 11.40 | 11.55 | 12.29 | 11.06 | 11.59 | 16.8 | |
| Maximum Current Density (mA/cm2) | 0.053 | 0.053 | 0.052 | 0.06 | 0.052 | 0.052 | 0.2856 | |
| Maximum Power Density, (mW/cm2) | 1.84 | 1.84 | 1.8 | 2.29 | 1.78 | 1.78 | 5.998 | |
| Burst Mode | (a) Pulses per burst | 141 | 127 | 208 | 34-179 | 404 | 415 | 204 |
| (b) Bursts per second | 0.137 | 0.162 | 0.095 | 0.098 | 0.045 | 0.045 | 0.219 | |
| (c) Burst duration (seconds) | 5.76 | 5.17 | 8.66 | 7.14 | 16.98 | 17.44 | 3 | |
| (d) Duty Cycle: Line (b) x Line (c) | 0.79 | 0.84 | 0.82 | 0.7 | 0.76 | 0.78 | 0.66 | |
| ON Time (seconds) | 5.76 | 5.17 | 8.66 | 7.14 | 16.98 | 17.44 | 3 | |
| OFF Time (seconds) | 1.52 | 1 | 1.96 | 3.02 | 5.41 | 4.78 | 1.56 | |
| Additional Features (specify, if applicable) | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
Table 6-SM 9126
| Parameter | New Device | Predicate | New Device | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Program 1 | Program 8 | Program 9 | Program 10 | Program 13 | Program 14 | Mode 1 | Program 2 | Program 7 | Program 15 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| @500Ω | @2kΩ | @10kΩ | @500Ω | @2kΩ | @10kΩ | ||||||
| MaximumOutput Voltage(V) | 30.8 | 84 | 159 | 30.4 | 72.4 | 124 | 30.8 | 30.8 | 30.8 | 4.6 | |
| MaximumOutput Current(mA) | 61.6 | 42 | 15.9 | 60.8 | 36.2 | 12.4 | 61.6 | 61.6 | 84 | 9.2 | |
| 42 | 41.2 | 30.4 | 4.6 | ||||||||
| 84 | 113 | 72.4 | 15.6 | ||||||||
| 159 | 163 | 123 | 24 | ||||||||
| 61.6 | 82.4 | 60.8 | 9.2 | ||||||||
| 42 | 56.5 | 36.2 | 7.8 | ||||||||
| 13 | 16.3 | 12.3 | 2.4 | ||||||||
| Pulse Width(µs) | 220 | 220 | 220 | 220 | 220 | 220 | 220 | 100 | 220 | 220 | 220 |
| Frequency (Hz) | 52.08 | 40.98 | 40.98 | 40.98 | 41.67 | 52.63 | 68 | 9.26 | 38.46 | 57.47 | |
| For interferential modes only:-Beat Frequency (Hz) | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| For multiphasicwaveformsonly: | Symmetricalphases | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
| Phaseduration | 220 | 220 | 220 | 220 | 220 | 220 | 100 | 220 | 220 | 220 | |
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 13.7 | 10.5 | 13.7 | 13.7 | 13.7 | 13.7 | 16.8 | 15 | 10.5 | 1.4 | |
| Maximum Current Density(mA/cm²) | 0.117 | 0.091 | 0.093 | 0.093 | 0.094 | 0.119 | 0.2856 | 0.03 | 0.09 | 0.019 | |
| Maximum Power Density,(mW/cm²) | 3.62 | 2.78 | 2.85 | 2.85 | 2.9 | 3.66 | 5.998 | 1.15 | 5.21 | 0.09 | |
| Burst Mode | (a) Pulses perburst | 212 | 520 | 254 | 127 | 521 | 216 | 204 | 34 | 149 | 488 |
| (b) Bursts persecond | 0.17 | 0.061 | 0.12 | 0.230 | 0.0625 | 0.173 | 0.219 | 0.18 | 0.145 | 0.098 | |
| (c) Burstduration(seconds) | 4.07 | 12.7 | 6.21 | 3.09 | 12.48 | 4.1 | 3 | 3.72 | 3.87 | 8.5 | |
| Cycle: Line(b) x Line (c) | |||||||||||
| ON Time (seconds) | 4.07 | 12.7 | 6.21 | 3.09 | 12.48 | 4.1 | 3 | 3.72 | 3.87 | 8.5 | |
| OFF Time (seconds) | 1.68 | 3.52 | 2.3 | 1.28 | 3.52 | 1.66 | 1.56 | 1.84 | 2.98 | 1.7 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
{22}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
{23}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Continue
| Parameter | Predicate | New Device | Predicate | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Mode 2 | Program 3 | Program 4 | Program 5 | Program 6 | Program 11 | Program 12 | Program 16 | Mode 3 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| MaximumOutput Voltage | @500Ω | 42 | 46.4 | 30.8 | 28.8 | 37.6 | 4.6 | 11 | 4.6 | 42 |
| @2kΩ | 84 | 126 | 84 | 68 | 100 | 15.6 | 20.8 | 15.6 | 84 | |
| (V) | @10kΩ | 130 | 161 | 159 | 119 | 153 | 24 | 37.2 | 24 | 130 |
| MaximumOutput Current | @500Ω | 84 | 92.8 | 61.6 | 57.6 | 75.2 | 9.2 | 22 | 9.2 | 84 |
| @2kΩ | 42 | 63 | 42 | 34 | 50 | 7.8 | 10.4 | 7.8 | 42 | |
| (mA) | @10kΩ | 13 | 16.1 | 15.9 | 11.9 | 15.3 | 2.4 | 3.72 | 2.4 | 13 |
| Pulse Width(μs) | 100 | 220 | 220 | 220 | 220 | 220 | 220 | 220 | 100 | |
| Frequency (Hz) | 12.5~55.5 | 1.47 | 40.98 | 3.33~37.04 | 1.79~13.51 | 212.5 | 113.6 | 57.47 | 1.17 | |
| For interferential modes only: | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| Beat Frequency (Hz) | ||||||||||
| For multiphasicwaveformsonly: | SymmetricalphasesPhaseDuration | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| 100 | 220 | 220 | 220 | 220 | 220 | 220 | 220 | 100 | ||
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 16.8 | 17.26 | 13.7 | 10.39 | 14.54 | 2.02 | 4.18 | 2.02 | 16.8 | |
| Maximum Current(mA/cm²) | Density | 0.2331 | 0.005 | 0.093 | 0.078 | 0.037 | 0.072 | 0.092 | 0.019 | 0.0049 |
{24}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
| and the contraction of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution4F,Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||
|---|---|---|
| 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Maximum | Power Density(mW/cm²) | 4.895 | 0.23 | 2.85 | 2.25 | 1.4 | 0.33 | 1.01 | 0.09 | 0.103 |
| Burst Mode | (a) Pulses per burst | 250~1110 | NA | NA | NA | NA | NA | NA | NA | NA |
| (b) Bursts per second | 0.048 | NA | NA | NA | NA | NA | NA | NA | NA | |
| (c) Burst duration(seconds) | 20 | NA | NA | NA | NA | NA | NA | NA | NA | |
| (d) Duty Cycle:Line (b) x Line (c) | 0.96 | NA | NA | NA | NA | NA | NA | NA | NA | |
| ON Time (seconds) | 20 | NA | NA | NA | NA | NA | NA | NA | NA | |
| OFF Time (seconds) | 1 | NA | NA | NA | NA | NA | NA | NA | NA | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
Table 7-SM 9186
| Parameter | New Device | Predicate | New Device | Predicate | New Device | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Program 1 | Program 6 | Program 9 | Program 12 | Mode 1 | Program 2 | Program 5 | Program 10 | Mode 2 | Program 3 |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular |
| Maximum @500Ω | 43.8 | 48 | 38.4 | 42.4 | 42 | 43.8 | 36.8 | 7 | 42 | 63 |
| Output @2kΩ | 57 | 74 | 59.6 | 76 | 84 | 81 | 66 | 13.1 | 84 | 116 |
| Voltage (V) @10kΩ | 142 | 156 | 109 | 152 | 130 | 149 | 126 | 24.2 | 130 | 149 |
| Maximum @500Ω | 87.6 | 96 | 76.8 | 84.8 | 84 | 87.6 | 73.6 | 14 | 84 | 126 |
| Output @2kΩ | 28.5 | 37 | 29.8 | 38 | 42 | 40.5 | 33 | 6.55 | 42 | 58 |
| Current @10kΩ (mA) | 14.2 | 15.6 | 10.9 | 15.2 | 13 | 14.9 | 12.6 | 2.42 | 13 | 14.9 |
{25}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| Pulse Width(μs) | Frequency (Hz) | For interferential modes only:-Beat Frequency (Hz) | Formultiphasicwaveforms | SymmetricalphasesPhaseduration | Net Charge (@500Ω) [uC] | Maximum Phase Charge (uC) | Maximum Current Density (mA/cm²) | Maximum Power Density, (mW/cm²) | Burst Mode | (a) Pulses per burst | (b) Bursts per second | (c) Burst duration (seconds) | (d) Duty Cycle: Line (b) x Line (c) | ON Time (seconds) | 220 | 220 | 230 | 220μs | 100 | 220 | 220 | 230 | 100 | 224 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 31.25-58.82 | N/A | Yes | 212 | 0.186 | 3.6 | 0.67 | 3.6 | 32.26~58.82 | 34.48~58.82 | 37.03~59.52 | 68 | N/A | 8.33~52.63 | 92.59 | 155.6 | 12.5~55.5 | 1.11 | |||||||
| 0 | 16.7 | 0.19 | 8.28 | 1200 | 0.034 | 20.4 | 0.7 | 20.4 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 14.5 | 0.21 | 9.94 | 711 | 0.064 | 12.08 | 0.77 | 12.08 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 14.5 | 0.17 | 6.65 | 333 | 0.125 | 5.6 | 0.7 | 5.6 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 15.2 | 0.185 | 7.85 | 204 | 0.219 | 3 | 0.66 | 3 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 16.8 | 0.2856 | 5.998 | 2253 | 0.023 | 42.8 | 0.98 | 42.8 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 16.7 | 0.17 | 7.4 | 1639 | 0.04 | 17.7 | 0.73 | 17.7 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 15.1 | 0.25 | 9.2 | 1556 | 0.085 | 10 | 0.85 | 10 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 2.6 | 0.084 | 0.58 | 250~1110 | 0.048 | 20 | 0.96 | 20 | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 16.8 | 0.2331 | 4.895 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | ||||||||||
| 0 | 25.07 | 0.005 | 0.33 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
{26}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| OFF Time (seconds) | 1.8 | 8.6 | 3.6 | 2.4 | 1.56 | 1 | 6.5 | 1.76 | 1 | N/A | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |||||
| Continue | |||||||||||||||
| Parameter | New Device | Predicate | |||||||||||||
| Mode or program name | Program 4 | Program 7 | Program 8 | Program 11 | Mode 3 | ||||||||||
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | ||||||||||
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | ||||||||||
| Maximum Output Voltage (V) | @500Ω43.8 | @2kΩ66 | @10kΩ128 | @500Ω59 | @2kΩ94 | @10kΩ168 | @500Ω68 | @2kΩ117 | @10kΩ174 | @500Ω26.8 | @2kΩ44 | @10kΩ80 | @500Ω42 | @2kΩ84 | @10kΩ130 |
| Maximum Output Current(mA) | @500Ω76 | @2kΩ33 | @10kΩ12.8 | @500Ω118 | @2kΩ47 | @10kΩ16.8 | @500Ω136 | @2kΩ58.5 | @10kΩ17.4 | @500Ω53.6 | @2kΩ22 | @10kΩ8 | @500Ω84 | @2kΩ42 | @10kΩ13 |
| Pulse Width(μs) | 220 | 220 | 220 | 220 | 100 | ||||||||||
| Frequency (Hz) | 50~92.59 | 1.136~59.52 | 1.11~92.59 | 92.59 | 1.17 | ||||||||||
| For interferential modes only: -Beat Frequency (Hz) | N/A | N/A | N/A | N/A | N/A | ||||||||||
| For multiphasic waveforms only: | Symmetrical phasesYes | Phase Duration220 | Yes220 | Yes220 | Yes220 | Yes100 | |||||||||
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | ||||||||||
| Maximum Phase Charge (uC) | 16.72 | 21.5 | 26 | 9.8 | 16.8 | ||||||||||
| Maximum Current Density (mA/cm²) | 0.258 | 0.26 | 0.46 | 0.182 | 0.0049 | ||||||||||
| Maximum Power Density, (mW/cm²) | 9.8 | 15.19 | 31.4 | 4.88 | 0.103 | ||||||||||
| Burst Mode | (a) Pulses per burstN/A | (b) Bursts per secondN/A | (c) Burst duration (seconds)N/A | (d) Duty Cycle: Line (b) x Line (c)N/A | N/A | ||||||||||
| ON Time (seconds) | N/A | N/A | N/A | N/A | N/A | ||||||||||
| OFF Time (seconds) | N/A | N/A | N/A | N/A | N/A |
{27}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| Additional Features (specify, if applicable) | N/A | N/A | N/A | N/A | N/A |
|---|---|---|---|---|---|
| ---------------------------------------------- | ----- | ----- | ----- | ----- | ----- |
Table 8-SM 9196
| Parameter | New Device | Predicate | New Device | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Program 1 | Program 2 | Program 3 | Program 4 | Program 6 | Program 7 | Mode 1 | Program 5 | Program 8 | Program 9 | ||
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | ||
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | ||
| MaximumOutput Voltage(V) | @500Ω | 49.2 | 48 | 48 | 42.8 | 48 | 47.2 | 42 | 46 | 52.4 | 50 | |
| @2kΩ | 73.6 | 73.6 | 74 | 72.8 | 74.4 | 72.8 | 84 | 71.6 | 78.8 | 101 | ||
| @10kΩ | 118 | 124 | 123 | 120 | 120 | 120 | 130 | 119 | 129 | 133 | ||
| MaximumOutput Current(mA) | @500Ω | 98.4 | 96 | 96 | 85.6 | 96 | 94.4 | 84 | 92 | 104.8 | 100 | |
| @2kΩ | 36.8 | 36.8 | 37 | 36.4 | 37.2 | 36.4 | 42 | 35.8 | 39.4 | 50.5 | ||
| @10kΩ | 11.8 | 12.4 | 12.3 | 12 | 12 | 12 | 13 | 11.9 | 12.9 | 13.3 | ||
| Pulse Width(us) | 200 | 200 | 200 | 200 | 200 | 200 | 100 | 200 | 200 | 210 | ||
| Frequency (Hz) | 47.17 | 47.62 | 47.17 | 48.08 | 49.02 | 48.08 | 68 | 48.08 | 9.434 | 1.613 | ||
| For interferential modes only:-Beat Frequency (Hz) | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | ||
| For multiphasicwaveformsonly: | Symmetricalphases | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
| Phaseduration | 200 | 200 | 200 | 200 | 200 | 200 | 100 | 200 | 200 | 210 | ||
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| Maximum Phase Charge (uC) | 15.84 | 15.20 | 15.20 | 14.5 | 15.20 | 14.8 | 16.8 | 14.6 | 16.48 | 16.8 | ||
| MaximumCurrent(mA/cm²) | Density | 0.07 | 0.07 | 0.073 | 0.07 | 0.08 | 0.07 | 0.2856 | 0.07 | 0.02 | 0.0027 | |
| MaximumPower(mW/cm²) | Density | 3.65 | 3.51 | 3.48 | 2.82 | 3.61 | 3.43 | 5.998 | 3.26 | 0.83 | 0.14 | |
{28}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| Burst Mode | (a) Pulses per burst | 257 | 238 | 434 | 303 | 152 | 173.1 | 204 | 249.1 | 33.4 | 7.74 |
|---|---|---|---|---|---|---|---|---|---|---|---|
| (b) Bursts per second | 0.132 | 0.166 | 0.082 | 0.12 | 0.244 | 0.2 | 0.219 | 0.122 | 0.194 | 0.16 | |
| (c) Burst duration (seconds) | 5.44 | 5 | 9.2 | 6.3 | 3.1 | 3.6 | 3 | 5.18 | 3.54 | 4.8 | |
| (d) Duty Cycle: Line (b) x Line (c) | 0.72 | 0.83 | 0.75 | 0.76 | 0.76 | 0.71 | 0.66 | 0.63 | 0.69 | 0.76 | |
| ON Time (seconds) | 5.44 | 5 | 9.2 | 6.3 | 3.1 | 3.6 | 3 | 5.18 | 3.54 | 4.8 | |
| OFF Time (seconds) | 2.14 | 1.02 | 3 | 2 | 1 | 1.5 | 1.56 | 3 | 1.62 | 1.5 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
Continue
| Parameter | Predicate | New Device | Predicate | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Mode or program name | Mode 2 | Program 10 | Program 11 | Program 12 | Program 13 | Program 14 | Program 15 | Mode 3 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum | @500Ω | 42 | 45.6 | 46 | 54 | 47.2 | 46.4 | 46 | 42 |
| Output | @2kΩ | 84 | 72 | 72 | 87.6 | 73.6 | 70.4 | 73.6 | 84 |
| Voltage (V) | @10kΩ | 130 | 112 | 118 | 131 | 119 | 117 | 124 | 130 |
| Maximum | @500Ω | 84 | 91.2 | 92 | 108 | 94.4 | 92.8 | 92 | 84 |
| Output | @2kΩ | 42 | 36 | 36 | 43.8 | 36.8 | 35.2 | 36.8 | 42 |
| Current(mA) | @10kΩ | 13 | 11.2 | 11.8 | 13.1 | 11.9 | 11.7 | 12.4 | 13 |
| Pulse Width(us) | 100 | 210 | 200 | 200 | 200 | 200 | 210 | 100 | |
| Frequency (Hz) | 12.5~55.5 | 49.02 | 3.33~47.62 | 1.887~15.63 | 48.08 | 48.08 | 47.17 | 1.17 | |
| For interferential modes only: -Beat Frequency | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| (Hz) |
{29}------------------------------------------------
Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
| Formultiphasicwaveformsonly: | Symmetrical phases | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
|---|---|---|---|---|---|---|---|---|---|
| Phase Duration | 100 | 210 | 200 | 200 | 200 | 200 | 210 | 100 | |
| Net Charge (@500Ω) [uC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (uC) | 16.8 | 13.6 | 14.6 | 17.0 | 14.8 | 14.9 | 15.9 | 16.8 | |
| Maximum Current Density (mA/cm²) | 0.2331 | 0.08 | 0.07 | 0.03 | 0.07 | 0.07 | 0.07 | 0.0049 | |
| Maximum Power Density, (mW/cm²) | 4.895 | 3.42 | 3.22 | 1.46 | 3.43 | 3.31 | 3.35 | 0.103 | |
| Burst Mode | (a) Pulses per burst | 250~1110 | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| (b) Bursts per second | 0.048 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| (c) Burst duration (seconds) | 20 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| (d) Duty Cycle: Line (b) x Line (c) | 0.96 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| ON Time (seconds) | 20 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| OFF Time (seconds) | 1 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| Additional Features (specify, if applicable) | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
{30}------------------------------------------------
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
5. Test summary:
TENS & PMS are safe and effective as the predicate devices cited above. The new devices have passed testings according to the following standards:
- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012
(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);
-
IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators:
-
ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances --Requirements And Tests;
-
IEC 62133 Edition 2.0 2012-12, IEC 62133 Edition 2.0 2012-12 Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications [Including: Corrigendum 1 (2013)]:
-
ANSI AAMI HA60601-1-11 2015 Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment (IEC 60601-1-11:2015 MOD);
The conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device. Furthermore, the new device complies with the recognized standards and performs its intended tasks as well as the legally marketed predicate devices.
§ 890.5850 Powered muscle stimulator.
(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).