(55 days)
Not Found
No
The summary explicitly states "The software does not provide any diagnostic assistance to the physician" and "Any diagnostic determination or treatment is solely determined by a physician and not the software." There is no mention of AI, ML, or any automated analysis or interpretation features. The device is described as a DICOM viewer and management tool.
No.
The device is described as a "primary diagnostic and analysis tool for diagnostic images" and a "software device that does not contact the patient, nor does it control any life sustaining devices." It explicitly states that "Any diagnostic determination or treatment is solely determined by a physician and not the software," indicating it does not provide therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states, "i-Rapha Solution is intended for use as a primary diagnostic and analysis tool for diagnostic images..."
Yes
The device description explicitly states, "The i-Rapha solution is a software device." It functions as a web-based application for viewing and managing DICOM images, relying on standard web browsers and not including any proprietary hardware.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states it's for displaying and analyzing diagnostic images from various imaging modalities (CR, DR, CT, US, MR, RF). IVDs are used to examine specimens (like blood, urine, tissue) in vitro (outside the body) to provide information about a person's health.
- Device Description: The description focuses on software for viewing and managing DICOM images. It explicitly states the software "does not contact the patient" and "does not provide any diagnostic assistance to the physician." IVDs are typically physical devices or kits used to perform tests on biological samples.
- Lack of Mention of Biological Samples: There is no mention of analyzing blood, urine, tissue, or any other biological specimen.
- Focus on Image Interpretation: The entire description revolves around the display and manipulation of medical images for interpretation by a physician.
Therefore, the i-Rapha Solution is a medical image viewing and management software, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
i-Rapha Solution is intended for use as a primary diagnostic and analysis tool for diagnostic images for hospitals, imaging centers, radiologists and any user who requires and is granted access to patient image, demographic.
i-Rapha View, a component of i-Rapha Solution, displays and manages diagnostic quality DICOM images.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display.
Product codes (comma separated list FDA assigned to the subject device)
LLZ
Device Description
The i-Rapha solution is a professional DICOM web browser application that conforms HTML5 standards to receive, store, and view DICOM images; utilizable under any web browser that supports HTML 5 protocols (e.g. Chrome Browser). This web browser application supports various types of annotation for medical images interpretation.
The i-Rapha solution is a software device that does not contact the patient, nor does it control any life sustaining devices. The software does not provide any diagnostic assistance to the physician.
The i-Rapha solution allows displaying of image studies that may not in the location as the modality. With its Web features, it is possible to review and manipulate the images of the studies located in a remote server. The i-Rapha View, a component of i-Rapha Solution, is an HTML5-based DICOM Viewer. The i-Rapha View can be used in any operating system because it can be executed by using a web browsers and especially Google Chrome that supports HTML5 regardless of the type of operating system.
Any diagnostic determination or treatment is solely determined by a physician and not the software. A physician, providing ample opportunity for competent human intervention, interprets images and information being displayed and printed.
The i-Rapha solution allows users to take full advantage of the radiographic images from various modalities in order to obtain invaluable mission critical diagnostic data and images.
The users can access their own diagnostic environment anywhere, anytime on PC.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
CR, DR, CT, US, MR, RF and other DICOM formats
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
hospitals, imaging centers, radiologists and any user who requires and is granted access to patient image, demographic.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Test Summary:
i-Rapha solution contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.
Software information is provided in accordance with FDA guidance: the content of premarket submissions for software contained in medical devices, issued on May 11, 2005.
Clinical Test Summary:
No clinical studies were considered necessary and performed.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
IRM, Inc % Kim Yun Ik Quality Manager 4th-D, Myung-Woo Bldg. 169, Yeoksam-ro Gangnam-gu, Seoul 06247 REPUBLIC OF KOREA
July 9, 2020
Re: K201296
Trade/Device Name: i-Rapha Solution Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 8, 2020 Received: May 15, 2020
Dear Kim Yun Ik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
510(k) Number (if known) K201296
Device Name i-Rapha Solution
Indications for Use (Describe)
i-Rapha Solution is intended for use as a primary diagnostic and analysis tool for diagnostic images for hospitals, imaging centers, radiologists and any user who requires and is granted access to patient image, demographic.
i-Rapha View, a component of i-Rapha Solution, displays and manages diagnostic quality DICOM images.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the logo for IRM. The logo consists of the letters "IRM" in bold, black font, followed by three colored circles. The first circle is orange, the second is dark blue, and the third is light blue with a white outline.
510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared [21 CFR 807.92(a)(a)] 1.
May 8, 2020
2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: | IRM Inc. |
---|---|
- Address: | 4th-D, Myung-Woo Bldg, 169, Yeoksam-ro, Gangnam-gu, Seoul, 06247, |
Republic of Korea | |
- Contact Name: | Kim Yun Ik / Quality Manager of IRM Inc. |
- Telephone No.: | +82-70-8230-6690 |
- Fax No.: | +82-70-8230-6693 |
- Email Address: | yikim@irm.kr |
3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
Trade Name | i-Rapha solution |
---|---|
Regulation Number | 21 CFR 892.2050 |
Regulation Name | Picture archiving and communications system |
Regulation Class | II |
Product Code | LLZ |
Product Code Name | System, Image Processing, Radiological |
4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
- 510(k) Number: | K152977 |
---|---|
- Applicant: | DICOM Grid, Inc. |
- Regulation Name: | Picture archiving and communications system |
- Product Code: | LLZ |
- Device Class: | II |
- Device Name: | DG PACS |
4
Image /page/4/Picture/0 description: The image shows the logo for IRM. The letters IRM are in bold black font. To the right of the letters are three circles, one dark blue, one orange, and one light blue with a white swirl pattern.
5. Description of the Device [21 CFR 807.92(a)(4)]
The i-Rapha solution is a professional DICOM web browser application that conforms HTML5 standards to receive, store, and view DICOM images; utilizable under any web browser that supports HTML 5 protocols (e.g. Chrome Browser). This web browser application supports various types of annotation for medical images interpretation.
The i-Rapha solution is a software device that does not contact the patient, nor does it control any life sustaining devices. The software does not provide any diagnostic assistance to the physician.
The i-Rapha solution allows displaying of image studies that may not in the location as the modality. With its Web features, it is possible to review and manipulate the images of the studies located in a remote server. The i-Rapha View, a component of i-Rapha Solution, is an HTML5-based DICOM Viewer. The i-Rapha View can be used in any operating system because it can be executed by using a web browsers and especially Google Chrome that supports HTML5 regardless of the type of operating system.
Any diagnostic determination or treatment is solely determined by a physician and not the software. A physician, providing ample opportunity for competent human intervention, interprets images and information being displayed and printed.
The i-Rapha solution allows users to take full advantage of the radiographic images from various modalities in order to obtain invaluable mission critical diagnostic data and images.
The users can access their own diagnostic environment anywhere, anytime on PC.
6. Indications for Use [21 CFR 807.92(a)(5)]
i-Rapha Solution is intended for use as a primary diagnostic and analysis tool for diagnostic images for hospitals, imaging centers, radiologists and any user who requires and is granted access to patient image, demographic.
i-Rapha View, a component of i-Rapha Solution, displays and manages diagnostic quality DICOM images.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display.
5
Image /page/5/Picture/0 description: The image shows the logo for IRM. The logo consists of the letters "IRM" in bold, black font. To the right of the letters are three circles. The first circle is orange, the second circle is dark blue, and the third circle is light blue with a white outline.
7. Technological Characteristics
There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. The below table is summarized and compared with the technological characteristics between the i-Rapha solution and the predicate device:
Proposed Device | Predicate Device #1 | SE note | ||
---|---|---|---|---|
K Number | K201296 | K152977 | - | |
Model | i-Rapha Solution | DG PACS | - | |
Manufacturer | IRM Inc. | DICOM Grid, Inc. | - | |
Classification Name | System, Image Processing, Radiological | System, Image Processing, Radiological | Same | |
Regulatory Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same | |
Product Code | LLZ | LLZ | Same | |
Classification | ll | ll | Same | |
Indications for use | i-Rapha Solution is intended for use as | |||
a primary diagnostic and analysis tool | ||||
for diagnostic images for hospitals, | ||||
imaging centers, radiologists and any | ||||
user who requires and is granted | ||||
access to patient image, demographic. | ||||
i-Rapha View, a component of i-Rapha | ||||
Solution, displays and manages | ||||
diagnostic quality DICOM images. | ||||
Lossy compressed mammographic | ||||
images and digitized film screen | ||||
images must not be reviewed for | ||||
primary image interpretations. | ||||
Mammographic images may only be | ||||
interpreted using cleared monitors | ||||
intended for mammography display. | DG PACS software is intended for use | |||
as a primary diagnostic and analysis | ||||
tool for diagnostic images for | ||||
hospitals, imaging centers, | ||||
radiologists, reading practices and any | ||||
user who requires and is granted | ||||
access to patient image, demographic | ||||
and report information. | ||||
DG Viewer, a component of DG PACS, | ||||
displays and manages diagnostic | ||||
quality DICOM images. | ||||
Lossy compressed mammographic | ||||
images and digitized film screen | ||||
images must not be reviewed for | ||||
primary image interpretations. | ||||
Mammographic images may only be | ||||
interpreted using cleared monitors | ||||
intended for mammography display. | ||||
Not intended for diagnostic use on | ||||
mobile devices. | Same | |||
Type of Use | Prescription Use | Prescription Use | Same | |
Component | Standalone software | Standalone software | Same | |
Web Standard | HTML5 | HTML5 | Same | |
Utilization of | ||||
Standards | Digital Imaging and Communications | |||
in Medicine (DICOM) | ||||
Joint Photographic Experts Group | ||||
(JPEG) | ||||
Society of Motion Picture and | ||||
Television Engineers (SMPTE) | Digital Imaging and Communications in | |||
Medicine (DICOM) | ||||
Joint Photographic Experts Group | ||||
(JPEG) | ||||
Society of Motion Picture and | ||||
Television Engineers (SMPTE) | Same | |||
Modalities | ||||
supported on | ||||
display | CR, DR, CT, US, MR, RF and other | |||
DICOM formats | CR, DR, CT, US, MR, RF and other | |||
DICOM formats | Same | |||
Proposed Device | Predicate Device #1 | SE note | ||
Operating System | ||||
(Web Browser) | Microsoft Internet Explorer 11 or | |||
Later**** | ||||
Apple Safari | ||||
Google Chrome | ||||
Firefox | Microsoft Internet Explorer 9 or Later | |||
Apple Safari | ||||
Google Chrome | ||||
Firefox | Differences | |||
Image Communication | ||||
Standard | DICOM | DICOM | Same | |
Mobile | ||||
Viewer | Mobile | |||
Phone | N/A* | Support | Differences | |
Tablet | N/A* | Support | Differences | |
Feature/ Functionality | ||||
User Authentication | Log-in | |||
Log-out | Log-in | |||
Log-out | Same | |||
Worklist (Study List) | Search and display Study List | |||
Search Studies | ||||
View Image | ||||
Series List | ||||
Save Search Options | ||||
Import DICOM | ||||
Export DICOM | ||||
Study Management | ||||
Image Preview** | Search and display Study List | |||
Search Studies | ||||
View Image | ||||
Series List | ||||
Save Search Options | ||||
Import DICOM | ||||
Export DICOM | ||||
Study Management | ||||
Report*** | ||||
Share Study*** | Similar | |||
DICOM | ||||
Query/Retrieve | DICOM Q/R | DICOM Q/R | Same | |
Patient List | Display Patient List** | |||
Search Patient** | Unknown | Differences | ||
Viewer | Image Display | |||
Viewer Layout function | ||||
Grid function | ||||
Image maximize function | ||||
Related Study List | ||||
Series Thumbnail | ||||
Study Information display function | ||||
Transformation | ||||
Image Viewing function | ||||
Window W/L | ||||
Measurement | ||||
Annotation | ||||
Image Processing | ||||
Image Reset | ||||
Export Image | Image Display | |||
Viewer Layout function | ||||
Grid function | ||||
Image maximize function | ||||
Related Study List | ||||
Series Thumbnail | ||||
Study Information display function | ||||
Transformation | ||||
Image Viewing function | ||||
Window W/L | ||||
Similar | ||||
Proposed Device | Predicate Device #1 | SE note | ||
Modify Image*** | ||||
Record*** | ||||
Report*** | ||||
DICOM Storage SCP | DICOM Storage SCP (SDHServer) | DICOM Storage SCP (Ambra Gateway) | Same |
6
Image /page/6/Picture/0 description: The image shows the logo for IRM. The logo consists of the letters "IRM" in bold, black font, followed by three circles of different colors. The first circle is orange, the second is dark blue, and the third is light blue with a white swirl inside.
7
Image /page/7/Picture/0 description: The image shows the logo for IRM. The logo consists of the letters "IRM" in a bold, sans-serif font, followed by three colored circles. The first circle is orange, the second is dark blue, and the third is light blue with a white outline.
- The subject device supports only PC Viewer. This is described in the User's Manual, so the user will be aware of this.
- ** There is additional functionality in the subject device. The added features and functions of the subject device are described in the user's manual, so the user will be aware of the functionality. The additional items do not change the Intended Use of the subject device. The differences do not raise any new potential safety risks and therefore, we believe there is no impact on safety or efficacy for the subject device.
- *** There is no functionality in the subjective device. The functionality in the predicate device is not essential for safety and efficacy and therefore, we believe there is no impact on safety or efficacy for the subject device.
- ****Microsoft Internet Explorer 9, 10 are no longer supported by Microsoft. The differences indicated, do not raise any new potential safety risks and therefore, there is no impact on safety or efficacy for the subject device.
Non-Clinical Test Summary
i-Rapha solution contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.
Software information is provided in accordance with FDA guidance: the content of premarket submissions for software contained in medical devices, issued on May 11, 2005.
Clinical Test Summary:
No clinical studies were considered necessary and performed.
8. Determination of Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between i-Rapha solution and the predicate device, K152977 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics
9. Conclusion [21 CFR 807.92(b)(3)]
The i-Rapha solution has similar intended use and technical characteristics to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness. In addition, performance testing conducted demonstrate that the subject devices are as safe and effective as the predicate. Therefore, the subject devices are substantially equivalent to the predicate.