K Number
K201211

Validate with FDA (Live)

Date Cleared
2020-09-03

(121 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3MP Color LCD Monitors C32S+, C32SP+ are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The displays are not intended for mammography.

The 3MP Monochrome LCD Monitors G32S+, G32SP+ are intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The devices are not specified for digital mammography systems.

Device Description

3MP Color LCD monitors C32S+, C32SP+; 3MP Monochrome LCD monitors G32S+, G32SP+ are specifically designed to provide the high definition image outputs for general Radiography.

The products have been strictly calibrated so they meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panels, supporting resolution 2048 x 1536. The built-in brightness stabilization control circuits make sure the brightness of these monitors are stable in their life and the calibrations are continuous, so the products meet the demands of high precision medical imaging. For the C32SP+ and G32SP+ surface protection panels with anti-reflection coating, there are characteristics such as anti-reflection, easy cleaning and anti-scratch screen.

Model variations are distinguished by characters. C means the Color monitor, G means the Monochrome monitor, and P means the monitor with an additional glass screen. For example, C32S+ is a color LCD monitor; G32SP+ is a monochrome LCD monitor with the additional glass screen.

AI/ML Overview

The provided document is a 510(k) Pre-Market Notification for medical display monitors (3MP Color LCD Monitors C32S+, C32SP+ and 3MP Monochrome LCD Monitors G32S+, G32SP+). It focuses on demonstrating substantial equivalence to predicate devices rather than providing a study proving performance against specific clinical acceptance criteria for an AI-powered device.

Therefore, many of the requested items (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth types and training set details) are not applicable or not present in this type of submission. This document describes the technical specifications and bench test results of the monitors themselves, not the performance of an AI algorithm based on image interpretation.

Here's an analysis based on the information available:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in a pass/fail table format for an AI algorithm's performance. Instead, it lists technical specifications of the proposed monitors and compares them to predicate devices, along with bench test results to verify compliance with display standards. The performance reported relates to display characteristics, not diagnostic accuracy of an AI.

However, we can infer some "acceptance criteria" based on the comparison tables and testing performed, which are primarily related to meeting technical standards for medical displays.

Acceptance Criteria (Inferred from testing and predicate comparison)Reported Device Performance (Proposed Devices)
• Conformance to DICOM GSDF (AAPM TG18 guideline)Verified conformance
• Luminance non-uniformity characteristics (AAPM TG18 guideline)Measured results
• Luminance stability with temperature (0℃, 25℃, 40℃) (AAPM TG18)Measured results
• Luminance at various viewing angles (30°, 45° diagonal, horizontal, vertical; center, 4 corners) (AAPM TG18)Measured results
• Temporal responseMeasured using typical panel manufacturer data
• Absence of miscellaneous artifacts (TG18 guideline)Visually checked
• Spatial noise (by noise power spectrum)Measured
• Spatial resolution (MTF)Measured
• Maximum number of pixel defects/faultsMeasured/Specified
• Maximum, minimum, achievable, and recommended luminanceMeasured
• Color tracking (primary colors and color gamut)Measured
• Display technology: Color (IPS) for C-series, TFT Grayscale LCD (IPS) for G-seriesColor (IPS) for C-series, TFT Grayscale LCD (IPS) for G-series
• Screen size: 54.1cm/21.3"54.1cm/21.3"
• Native resolution: 1536 x 20481536 x 2048
• Viewable image size (H x V): ~324.86 x 433.15 mm~324.86 x 433.15 mm
• Pixel pitch: 0.2115 x 0.2115 mm0.2115 x 0.2115 mm
• Recommended brightness for: 500 cd/m²500 cd/m²
• Backlight type: LEDLED
• Brightness stabilization: YesYes
• Digital uniformity equalizer: YesYes

Note: Specific numerical results for each measurement (e.g., exact luminance non-uniformity values) are not provided in this summary, but the document states they were measured and found to be equivalent to predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable as the document describes tests on the physical display monitors, not an AI algorithm processing medical images. The "test set" here refers to the monitors themselves.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. The "ground truth" for these display monitors is defined by technical specifications and adherence to standards like DICOM and AAPM TG18 guidelines, measured by equipment and engineering assessments, not by expert medical image interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are relevant for clinical studies involving human interpretation or AI output, not for technical verification of display monitors.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This document is for medical display monitors and does not involve AI assistance or MRMC studies for diagnostic performance improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. No algorithm is being evaluated in this submission.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the performance evaluation of these medical displays is established by adherence to technical and performance standards for display devices, such as DICOM Part 3.14 and AAPM Task Group 18 (TG18 guideline) for display performance. It's measured through physical bench testing and comparison to established engineering specifications and predicate devices.

8. The sample size for the training set

This information is not applicable. There is no AI algorithm with a training set being described in this document.

9. How the ground truth for the training set was established

This information is not applicable. There is no AI algorithm with a training set being described in this document.

In summary, the provided text is a 510(k) summary for medical display monitors, focusing on their technical specifications and substantial equivalence to predicate devices, as demonstrated through bench testing against established display standards. It does not contain information related to AI algorithm performance or clinical studies of diagnostic accuracy.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

Shenzhen Beacon Display Technology Co., Ltd. % Fu Ailing Document Engineer 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road Shenzhen, Guangdong 518055 CHINA

Re: K201211

Trade/Device Name: 3MP Color LCD Monitors C32S+, C32SP+, 3MP Monochrome LCD Monitors G32S+, G32SP+ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: July 28, 2020 Received: August 6, 2020

September 3, 2020

Dear Fu Ailing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K201211

Device Name

3MP Color LCD Monitors C32S+, C32SP+; 3MP Monochrome LCD Monitors G32S+, G32SP+

Indications for Use (Describe)

The 3MP Color LCD Monitors C32S+, C32SP+ are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The displays are not intended for mammography.

The 3MP Monochrome LCD Monitors G32S+, G32SP+ are intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The devices are not specified for digital mammography systems.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201211

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

April 15, 2020

2. Submitter's Information [21 CFR807.92 (a) (1)]

Name of Sponsor:Shenzhen Beacon Display Technology Co., Ltd.
Address:12F, Block B1, Nanshan Zhiyuan, No. 1001 Xueyuan Road, NanshanDistrict, Shenzhen, 518055 China
Contact Name:Fu Ailing
Telephone No.:+86-0755-36868363-5150
Fax No.:+86-0755-26995755
Email Address:fuailing@beacon-display.cn

3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]

Trade Name/Model: 3MP Color LCD Monitors C32S+, C32SP+;

3MP Monochrome LCD Monitors G32S+, G32SP+

  • Common Name: 3MP LCD Monitors C32S+, C32SP+, G32S+, G32SP+
    Classification Name: Picture archiving and communications system

Requlation Number: 21 CFR 892.2050

Product code: PGY

Classification Panel: Radiology

Device Class: Il

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate within this submission is as follows:

{4}------------------------------------------------

EIZO Corporation, RadiForce RX360, RX360-AR has been cleared by FDA through 510(k) No. K182591 (Decision Date - October 18, 2018).

EIZO NANAO Corporation, 3MP Monochrome LCD Monitor, RadiForce GX340 has been cleared by FDA through 510(k) No. K113784 (Decision Date - May 4, 2012).

5. Description of the Device [21 CFR 807.92(a) (4)]

3MP Color LCD monitors C32S+, C32SP+; 3MP Monochrome LCD monitors G32S+, G32SP+ are specifically designed to provide the high definition image outputs for general Radiography.

The products have been strictly calibrated so they meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panels, supporting resolution 2048 x 1536. The built-in brightness stabilization control circuits make sure the brightness of these monitors are stable in their life and the calibrations are continuous, so the products meet the demands of high precision medical imaging. For the C32SP+ and G32SP+ surface protection panels with anti-reflection coating, there are characteristics such as anti-reflection, easy cleaning and anti-scratch screen.

Model variations are distinguished by characters. C means the Color monitor, G means the Monochrome monitor, and P means the monitor with an additional glass screen. For example, C32S+ is a color LCD monitor; G32SP+ is a monochrome LCD monitor with the additional glass screen.

6. Intended Use [21 CFR 807.92(a)(5)]

The 3MP Color LCD Monitors C32S+, C32SP+ are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The displays are not intended for mammography.

The 3MP Monochrome LCD Monitors G32S+, G32SP+ are intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The devices are not specified for digital mammography systems.

{5}------------------------------------------------

7. Technological Characteristics [21 CFR 807.92(a)(6)]

Monitor characteristics:
Screen technology21.3", TFT, color, LCD screen, anti-glare, hard coating
Active area(H x V)324.86 x 433.15 mm
Pixel pitch0.2115 mm (H) x 0.2115 mm (V)
Resolution1536 x 2048
Contrast ratio1500:1 (typical), 1200:1 (minimal)
Viewing angle (CR > 10 )Horizontal: 178° (typical)
Vertical: 178° (typical)
Screen brightness1000 cd/m² (typical)
Refresh rate60 Hz
BacklightingLEDs
Lifetime of backlight50000 hours
Response time (Ton + Toff)25 ms (typical)
Power supply:
Line voltage12 V
Current consumption6.67 A
Power consumption80 W (maximum)
Standby< 5 W
Adaptor1:GSM90A12 Input: 100240V, 1.3-0.6A, 50/60Hz,Output: +12VDC, 6.67A2:GSM120A12 Input: 100240V, 1.4-0.7A, 50/60Hz,Output: +12VDC, 8.5A
Control and connection:
Front1 operation LED, 6 functional keys
BackGround1 Power switch1 DVI-D1 DP1 USB Upstream Port1 USB Downstream Port2
Mechanical characteristics :
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Protection levelIP 20
IP 65 (Applicable to the front of C32SP+)
Climatic conditions :
Working temperature0°C - 40°C
Working humidity15% - 85% Relative humidity, no condensation
Working atmospheric700 hPa - 1060 hPa
Transport and storage-20°C - 60°C
Transport and storage10% - 90% Relative humidity, no condensation
Transport and storage700 hPa - 1060 hPa
Safety regulations :
Safety standardsIEC60601-1, EN60601-1
ANSI/AAMI ES60601-1: 2005 + A2(R2012) + A1
CAN/CSA-C22.2 NO. 60601-1
ConformityCCC, CE, TUV, FCC
Dimension:
Dimensions (W x H x D) in369 x 511.5 ~ 596.15 x 220 mm
With packing (W x H x D)608 x 561 x 327 mm
Weight :
Net weight9.5 ± 0.5 kg
Gross weight11.0 ± 0.5 kg

C32S+, C32SP+ Color LCD Monitors

{6}------------------------------------------------

G32S+, G32SP+ Monochrome LCD monitors

Monitor characteristics:
Screen technology21.3", TFT, grayscale, LCD screen, anti-glare, hard
Active area( H x V)324.86 x 433.15 mm
Pixel pitch0.2115 mm (H) x 0.2115 mm (V)
Resolution1536 x 2048
Contrast ratio1500:1 (typical), 1200:1 (minimal)
Viewing angle ( CR > 10 )Horizontal: 170° (typical)
Vertical: 170° (typical)
Screen brightness2000 cd/m² (typical)
Refresh rate60 Hz
BacklightingLEDs
Lifetime of backlight50000 hours
Response time ( Ton + Toff )28 ms (typical)

{7}------------------------------------------------

Power supply:
Line voltage12 V
Current consumption6.67 A
Power consumption50 W (maximum)
Standby< 5 W
Adaptor1:GSM90A12 Input: 100240V, 1.3-0.6A, 50/60HzOutput: +12VDC, 6.67A2:GSM120A12 Input: 100240V, 1.4-0.7A, 50/60HzOutput: +12VDC, 8.5A
Control and connection:
Front1 operation LED, 6 functional keys
BackGround1 Power switch1 DVI-D1 DP1 USB Upstream Port1 USB Downstream Port2
Mechanical characteristics :
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Protection levelIP 20
Climatic conditions :
Working temperature0℃ - 40℃
Working humidity15% - 85% Relative humidity, no condensation
Working atmospheric700 hPa - 1060 hPa
Transport and storage-20℃ - 60℃
Transport and storage10% - 90% Relative humidity, no condensation
Transport and storage700 hPa - 1060 hPa
Safety regulations :
Safety standardsIEC60601-1, EN60601-1
ConformityCCC, CE, TUV, FCC
Dimension:
Dimensions (W x H x D) in369 x 511.15 ~ 596.15 x 220 mm
With packing (W x H x D)608 x 561 x 327 mm
Weight :
Net weight9.0 ± 0.5 kg
Gross weight11.0 ± 0.5 kg

{8}------------------------------------------------

8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

8.1 Intended uses:

IDComparisonItemProposed Device3MP Color LCD MonitorsC32S+, C32SP+Predicate Device3MP Color LCD MonitorRadiForce RX360, RX360-AR
1Intended UseThe 3MP Color LCD MonitorsC32S+, C32SP+ are indicated foruse in displaying radiologicalimages for review, analysis, anddiagnosis by trained medicalpractitioners. The displays arenot intended for mammography.The Product is indicated for usein displaying radiological imagesfor review, analysis, anddiagnosis by trained medicalpractitioners. The display is notintended for mammography.

Table 1 Intended Use Comparison of C32S+, C32SP+

Table 2 Intended Use Comparison of G32S+, G32SP+

IDComparisonltemProposed Device3MP Monochrome LCDMonitorsG32S+, G32SP+Predicate Device3MP Monochrome LCDMonitorRadiForce GX340
2Intended UseThe 3MP Monochrome LCDMonitors G32S+, G32SP+ areintended to be used in displayingand viewing digital images fordiagnosis of X-ray or MRI, etc. bytrained medical practitioners. Thedevices are not specified fordigital mammography systems.The RadiForce GX340 isintended to be used indisplaying and viewing digitalimages for diagnosis of X-ray orMRI, etc. by trained medicalpractitioners. The device is notspecified for digitalmammography system.

{9}------------------------------------------------

8.2 Comparison table

Table 3 General Comparison of C32S+, C32SP+

IDComparisonItemProposed Device3MP Color LCD MonitorsC32S+, C32SP+Predicate Device3MP Color LCD MonitorRadiForce RX360,RX360-ARExplanation of Differences
Display Performance/Specifications
3.1Display technologyColor (IPS)Color (IPS)-
3.2Screen size54.1cm/21.3"Aspect ratio:3:454.1cm/21.3"Aspect ratio:3:4-
3.3Viewing angle (H/V)178°/178°178°/178°-
3.4Native resolution1536 x 20481536 x 2048-
3.5Viewable image size(H x V)324.86 x 433.15 mm324.9 x 433.2 mm-
3.6Pixel pitch0.2115 x 0.2115 mm0.2115 x 0.2115 mm-
3.7Response time(typical)25 ms (on / off)12 ms (on / off)Different screen
3.8Brightness (typical)1000 cd/m²1100 cd/m²Different screen
3.9Recommendedbrightness for500 cd/m²500 cd/m²-
3.10Contrast ratio (typical)1500:11500:1-
3.11Backlight typeLEDLED-
3.12Display colors10-bit (DisplayPort): 1.073billion1024 from a palette of 16,384tones8-bit(DVI): 16.77 million 256from a palette of 16,384 tones10-bit colors (DisplayPort) :1.07 billion (maximum) colors8-bit colors: 16.77 million froma palette of 543 billion colorsTone between the predicatedevice and our proposeddevices are different. But theypass the exams in AAPM-TG184.3 "Luminance Response".Therefore, they are equivalentto the predicate device.
Video Signals
4.1Input video signalsDVI-D (dual link) x 1,DisplayPort x 1DVI-D (dual link) x 1,DisplayPort x 2Different design scheme
4.2Digital ScanningFrequency (H, V)31 - 97 kHz/59 - 61 Hz31 - 127 kHz/29 - 61.5 Hz(VGA Text: 69 - 71 Hz)Frame synchronous mode:29.5 - 30.5 Hz, 59 - 61 HzDifferent design scheme
4.3Video bandwidthDVI: 216 MHzDisplayPort: 216 MHzDVI: 215MHzDisplayPort: 215MHzDifferent design scheme
Power Related Specifications
5.1Power RequirementsDC 12 V / 6.67 AAC 100 - 240 V:50 / 60 HzDifference between built-inpower supply and built-outpower supply
5.2Maximum powerconsumption80 W74 WDifferent design scheme
5.3Power save modeLess than 5 W1 W or lessDifferent design scheme
5.4Power ManagementDVI DMPM, DisplayPort 1.2DVI DMPM, DisplayPort 1.1aDifferent design scheme
Miscellaneous Features/Specifications
6.1Quality-controlSoftwareBeacon Monitor ManageRadiCSDifferent design scheme
6.2SensorsBacklight sensorIntegrated front sensorAmbient light sensorBacklight sensorIntegrated front sensorPresence sensorAmbient light sensorDifferent design scheme
6.3Luminance calibrationtoolsIntegrated optical sensorExternal optical sensorCalibration software: BeaconMonitor ManageIntegrated optical sensorExternal optical sensorCalibration software:RadiCSDifferent design scheme
6.4USB PortsUpstreamUSB 2.0: Type-B x 1UpstreamUSB 2.0: Type-B x 2Different design scheme
6.5BrightnessstabilizationYesYes-
6.6Digital uniformityequalizerYesYes-
6.7Net weight9.58Different weight due to differentcomponents and parts
6.8Hole Spacing (VESAStandard)100 x 100mm100 x 100mm-
6.9Dimensions w/o stand(W x H x D)369 x 220 x 511.5 ~ 596.15341.3 x 200 x 481.5 - 571.5Different housing design due tothe different panel size
IDComparisonItemProposed Device3MP Monochrome LCDMonitorsG32S+, G32SP+Predicate Device3MP Monochrome LCDMonitorRadiForce GX340Explanation of Differences
Display Performance/Specifications
3.1Display technologyTFT Monochrome LCD Panel(IPS)TFT Monochrome LCD Panel(IPS)-
3.2Screen size54 cm / 21.3" (541 mmdiagonal)54 cm / 21.3" (541 mmdiagonal)-
3.3Viewing angle (H/V)170°/170°176°/176°Different screen
3.4Native resolution1536 x 2048 (3:4 aspect ratio)1536 x 2048 (3:4 aspect ratio)-
3.5Viewable image size(H x V)324.86 x 433.15 mm324.8 x 433.1 mm-
3.6Pixel pitch0.2115 mm (H) x 0.2115 mm0.2115 x 0.2115 mm-
3.7Response time(typical)28 ms (On/Off)40 ms (On/Off)Different screen
3.8Brightness (typical)2000 cd/m²1200 cd/m²Different screen
3.9Recommendedbrightness for500 cd/m²500 cd/m²-
3.10Contrast ratio (typical)1500:11400:1Different screen
3.11Backlight typeLEDLED-
3.12Grayscale Tones10-bit (DisplayPort): 1,024from a palette of 16,384 tones8-bit (DVI): 256 from a paletteof 16,384 tones10-bit (DisplayPort): 1,024from a palette of 16,369 tones8-bit: 256 from a palette of16,369 tonesTone between the predicatedevice and our proposeddevices are different. But theyAAPM-TG18 4.3 "LuminanceResponse". Therefore, theyare equivalent to the predicatedevice.
Video Signals
4.1Input video signalsDVI-D (dual link) x 1DisplayPort x 1DVI-D (dual link) x 1DisplayPort x 1-
4.2Digital ScanningFrequency (H, V)31 - 97 kHz, 60 - 70 kHz31 - 127 kHz, 29.5 - 61 Hz(VGA Text: 69 - 71 Hz)Frame synchronous mode:29.5 - 30.5 Hz, 59 - 61 HzDifferent design scheme
4.3Dot Clock216 MHz215 MHzDifferent design scheme
Power Related Specifications
5.1Power RequirementsDC 12 V / 6.67 AAC 100 - 120 V, 200 - 240 V:50 / 60 HzDifference between built-inpower supply and built-outpower supply
5.2Maximum powerconsumption50 W90 WDifferent design scheme
5.3Power save modeLess than 5 WLess than 1.6 WDifferent design scheme
5.4Power ManagementDVI DMPM, DisplayPort 1.2DVI DMPM, DisplayPort 1.1aDifferent design scheme
Miscellaneous Features/Specifications
6.1Quality-controlproceduresBeacon Monitor ManageRadiCSDifferent design scheme
6.2SensorsBacklight SensorIntegrated Front SensorAmbient Light SensorBacklight SensorIntegrated Front SensorPresence SensorAmbient Light SensorDifferent design scheme
6.3Luminance calibrationtoolsIntegrated optical sensorExternal optical sensorCalibration software: BeaconMonitor ManageIntegrated optical sensorExternal optical sensorCalibration software:RadiCSDifferent design scheme
6.4USB Ports1 upstream2 downstream / Rev. 2.01 upstream2 downstream / Rev. 2.0-
6.5BrightnessstabilizationYesYes-
6.6Digital uniformityequalizerYesYes-
6.7Net weight9.0 kg10.2 kgDifferent weight due to differentcomponents and parts
6.8Hole Spacing (VESAStandard)VESA standard 100 x 100 mmVESA standard 100 x 100 mm-
6.9Dimensions w/o stand(W x H x D)369 x 511.15 ~ 596.15 x 220376 x 520~599 x 245.5Different housing design due tothe different panel size

{10}------------------------------------------------

510(k) Submission

{11}------------------------------------------------

{12}------------------------------------------------

Table 4 General Comparison of G32S+/G32SP+

{13}------------------------------------------------

510(k) Submission

{14}------------------------------------------------

{15}------------------------------------------------

It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the C32S+, C32SP+, G32S+, G32SP+.

8.3 Performance Testing Comparation

The bench tests were performed on the proposed devices C32S+, C32SP+, G32S+, G32SP+ as below.

    1. Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline).
    1. Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 quideline.
    1. Measure the luminance stability with the temperature 0 ℃ , 25 ℃ and 40 ℃ on Luminance response by AAPM-TG18.
    1. Measure the luminance at the angles of 30° and 45° in diagonal, horizontal and vertical directions at center and four corners by AAPM-TG18.
    1. Measure the temporal response using the typical data provided by the panel manufacturer.
    1. Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 quideline.
    1. Measure the spatial noise by noise power spectrum.
    1. Measure the spatial resolution expressed as modulation transfer function (MTF)
    1. Maximum number allowed for each type of pixel defects/faults
    1. Meausure the maximum, minimum, achievable, and recommended luminance.
    1. Measure the color tracking (primary colors and color gamut).

The test results showed that C32S+, C32SP+, G32S+, G32SP+ are with display characteristics equivalent to those of the predicate devices, RadiForce RX360, RX360-AR and 3MP Monochrome LCD Monitor, RadiForce GX340 except some items, each of which was determined that it would not affect observer's performance. No animal or clinical testing is needed for C32S+, C32SP+, G32S+, G32SP+.

{16}------------------------------------------------

9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Shenzhen Beacon Display Technology Co., Ltd. concludes that:

  • The intended uses of C32S+, C32SP+, G32S+, G32SP+ are totally same as those of the predicate devices.
  • The technological characteristics differences between C32S+, C32SP+, G32S+, G32SP+ and the predicate devices do not affect the safety and effectiveness, so no new risk is raised.
  • Demonstrated by the bench tests, the display characteristics of C32S+, C32SP+, G32S+, G32SP+ are equivalent to those of the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).