(113 days)
The ASEPT® Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
ASEPT® Surgical Face Mask is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond polypropylene and the middle layer is constructed of melt blown polypropylene filter. The mask is held in place over the mouth and nose by two elastic loops welded to the facemask. The elastic loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with aluminum wire and allows the user to fit the facemask around their nose. The ASEPT® Surgical Face Mask is sold non-sterile and is intended to be a single use, disposable device.
The document describes the acceptance criteria and performance of the ASEPT® Surgical Face Mask.
1. A table of acceptance criteria and the reported device performance
| Item | Standard | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Fluid Resistance Performance | ASTM F1862: Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood | At least 29 out of 32 specimens show passing results at 120 mmHg | All samples met the predetermined acceptance criteria. |
| Bacterial Filtration Efficiency | ASTM F2100-19: Standard Specification for Performance of Materials Used in Medical Face Masks | ≥ 98% | All samples met the predetermined acceptance criteria. |
| Differential Pressure (Delta P) | ASTM F2100-19: Standard Specification for Performance of Materials Used in Medical Face Masks | < 6.0 mm H2O/cm² | All samples met the predetermined acceptance criteria. |
| Particulate Filtration Efficiency | ASTM F2100-19: Standard Specification for Performance of Materials Used in Medical Face Masks | ≥ 98% | All samples met the predetermined acceptance criteria. |
| Flammability | 21 CFR 1610 | Class I, Does not Ignite | All samples met the predetermined acceptance criteria. |
| Cytotoxicity | ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Non-cytotoxic | All samples met the predetermined acceptance criteria. |
| Irritation | ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Non-irritating | All samples met the predetermined acceptance criteria. |
| Sensitization | ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Non-sensitizing | All samples met the predetermined acceptance criteria. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Fluid Resistance Performance: At least 32 specimens were used for testing (acceptance criteria states "29 out of 32 specimens").
- For Bacterial Filtration Efficiency, Differential Pressure, Particulate Filtration Efficiency, Flammability, Cytotoxicity, Irritation, and Sensitization, the document states "All samples met the predetermined acceptance criteria," implying that a sample set was tested for each, but the exact number of samples for these tests is not explicitly stated.
- Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device (surgical face mask) is evaluated based on non-clinical performance standards and biocompatibility tests, not clinical evaluations requiring expert interpretation of results. Therefore, the concept of "ground truth" established by experts in a diagnostic context does not apply here. The results are from standardized laboratory tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as the evaluation is based on objective laboratory testing against pre-defined standards, not on subjective expert consensus or adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (surgical face mask) with performance evaluated through non-clinical laboratory testing, not a diagnostic AI device requiring human reader studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device is established by the specified standards and acceptance criteria for each performance characteristic (e.g., ASTM F1862 for fluid resistance, ASTM F2100-19 for filtration efficiency, ISO 10993 for biocompatibility). The device's performance is measured directly against these objective benchmarks in a laboratory setting.
8. The sample size for the training set
Not applicable. This is a physical medical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is a physical medical device, not a machine learning model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 19, 2020
PFM Medical, Inc Jessica Jho Director of Regulatory Affairs 1916 Palomar Oaks Way, Suite 150 Carlsbad, California 92008
Re: K201137
Trade/Device Name: Asept Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 19, 2020 Received: July 23, 2020
Dear Jessica Jho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
Asept Surgical Face Mask
Indications for Use (Describe)
The ASEPT® Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
K201137
As required by 21CFR 807.92 Date of Preparation: July 19, 2020
Applicant Information
510(k) Applicant: PFM Medical, Inc. Applicant Address: 1916 Palomar Oaks Way, Suite 150, Carlsbad, CA 92008 Contact Person: Jessica Jho, RAC Email: JJho@pfmmedicalusa.com
Device Information
Trade Name: ASEPT® Surgical Face Mask Common Name: Mask, Surgical Classification Name: Surgical Apparel Product Code: FXX Regulation Number: 21 CFR §878.4040
Predicate Device Information
Trade Name: Surgical Masks Premarket Notification: K111559 Product Code: FXX, Regulation Number: 21 CFR §878.4040 Manufacturer: Shanghai Neo-Medical Import & Export Co.
Device Description
ASEPT® Surgical Face Mask is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond polypropylene and the middle layer is constructed of melt blown polypropylene filter. The mask is held in place over the mouth and nose by two elastic loops welded to the facemask. The elastic loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with aluminum wire and allows the user to fit the facemask around their nose. The ASEPT® Surgical Face Mask is sold non-sterile and is intended to be a single use, disposable device.
{4}------------------------------------------------
Indications for Use
The ASEPT® Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
Comparison of Technological Characteristics Between the Predicate and Subject Devices
The following is a summary of the technological characteristics of the ASEPT® Surgical Face Mask as compared to the predicate device. The subject and predicate devices are the identical devices manufactured by the identical supplier.
| Items | Subject Device | Predicate Device | Comparison | |
|---|---|---|---|---|
| Manufacturer | PFM Medical, Inc. | Shanghai Neo-Medical Import & Export Co. | N/A | |
| 510(k) Number | K201137 | K111559 | N/A | |
| FDA Product Code | FXX | FXX | Identical | |
| Indications for Use | The ASEPT® Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids. | The Surgical Masks are indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism and body fluid. | Identical with grammatical correction. | |
| Materials | Inner and Outer Layers | Spun-bond polypropylene | Spun-bond polypropylene | Identical |
| Middle Layer | Melt blown polypropylene filter | Melt blown polypropylene filter | Identical | |
| Ear loops | Polyester | Polyester | Identical | |
| Nose Piece | Malleable polyethylene with aluminum wire | Malleable polyethylene with aluminum wire | Identical | |
| Dimensions | 17.5cm length x 9.5cm height | 17.5cm length x 9.5cm height | Identical | |
| Mask Style | Pleated | Pleated | Identical | |
| Design Features | Malleable nosepiece, flat pleated, elastic ear loops | Malleable nosepiece, flat pleated, elastic ear loops | Identical | |
| Sterility | Non-sterile | Non-sterile | Identical | |
| Use | Single Use, Disposable | Single Use, Disposable | Identical | |
| Color | Blue and White | Blue and White | Identical | |
| ASTM F2100 Level | Level 2 | Level 2 | Identical | |
| Biocompatibility | Cytotoxicity, ISO 10993-5:2009 | Non-cytotoxic | Non-cytotoxic | same |
| Irritation, ISO 10993-10:2002 | Non-irritating | Non-irritating | same | |
| Sensitization, ISO 10993-10:2002 | Non-sensitizing | Non-sensitizing | same |
{5}------------------------------------------------
Summary of Non-Clinical Testing
| Per FDA document Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification |
|---|
| [510(k)] Submissions, the below testing has been completed on the subject device: |
| ltem | Standard | Acceptance Criteria | Results |
|---|---|---|---|
| Fluid ResistancePerformance | ASTM F1862: Standard TestMethod for Resistance of MedicalFace Masks to Penetration bySynthetic Blood | At least 29 out of 32specimens show passingresults at 120 mmHg | All samples met thepredeterminedacceptance criteria. |
| Bacterial FiltrationEfficiency | ASTM F2100-19: StandardSpecification for Performance ofMaterials Used in Medical FaceMasks | ≥ 98% | All samples met thepredetermined acceptancecriteria. |
| DifferentialPressure (Delta P) | ASTM F2100-19: StandardSpecification for Performance ofMaterials Used in Medical FaceMasks | < 6.0 mm H2O/cm² | All samples met thepredetermined acceptancecriteria. |
| ParticulateFiltration Efficiency | ASTM F2100-19: StandardSpecification for Performance ofMaterials Used in Medical FaceMasks | ≥ 98% | All samples met thepredetermined acceptancecriteria. |
| Flammability | 21 CFR 1610 | Class I, Does not Ignite | All samples met thepredeterminedacceptance criteria. |
| Cytotoxicity | ISO 10993-5: Biological evaluationof medical devices - Part 5: Testsfor in vitro cytotoxicity | Non-cytotoxic | All samples met thepredeterminedacceptance criteria. |
| Irritation | ISO 10993-10: Biologicalevaluation of medical devices -Part 10: Tests for irritation andskin sensitization | Non-irritating | All samples met thepredetermined acceptancecriteria. |
| Sensitization | ISO 10993-10: Biologicalevaluation of medical devices -Part 10: Tests for irritation andskin sensitization | Non-sensitizing | All samples met thepredeterminedacceptance criteria. |
Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) K201137, the ASEPT® Surgical Face Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K111559.
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.