The Mega Plus MIS Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Mega Plus MIS Spine System can be used in an open approach and percutaneous approach.
The Mega Plus MIS Spine System is a top-loading multiple component, posterior spinal fixation system which consists of pedicle screws, rods, locking bolts, and a transverse (cross) link. Various sizes of these implants are available. Mega Plus MIS Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. Mega Plus MIS Spine System can either be used in an open approach and/or a percutaneous with minimally invasive (MIS) surgical approach. All products are made of titanium alloy (ASTM F136) and CoCrMo alloy (ASTM F1537) approved for medical use. The implants will be provided non-sterile. The locking bolts utilized as part of the proposed system are identical to those of the Mega Plus Spine System (K173180, K183080). The crosslinks and additional universal instrument components that may be utilized as part of the subject system were cleared previously for the Mega Plus Spine System (K173180, K183080).
The provided text describes a 510(k) premarket notification for the "Mega Plus MIS Spine System," a medical device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing detailed clinical study data or acceptance criteria for AI/ML performance.
Therefore, the requested information regarding acceptance criteria for device performance, sample sizes, expert qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance, as you've outlined, is not present in this document.
The document primarily focuses on:
- Device Description: What the device is and its components.
- Intended Use/Indications for Use: For what medical conditions and patient populations the device is intended.
- Technological Characteristics: How it compares to predicate devices.
- Performance Data: This section mentions non-clinical testing (mechanical tests) per ASTM F1717, but it does not provide acceptance criteria or specific results in the form of a table. It simply states that "The results of this non-clinical testing show that the strength of the Mega Plus MIS Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."
In summary, this document does not contain the information needed to answer your questions because it pertains to a traditional medical device (spinal fixation system) and its mechanical performance testing for substantial equivalence, not an AI/ML-driven device requiring clinical performance criteria, expert adjudication, or separate training/test sets as described in your request.
{0}------------------------------------------------
August 11, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters. Below that is the word "ADMINISTRATION" in smaller, blue letters.
BK Meditech, Co., Ltd % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K200981
Trade/Device Name: Mega Plus MIS Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: July 10, 2020 Received: July 13, 2020
Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
510(k) Number (if known) K200981 Device Name Mega Plus MIS Spine System
Indications for Use (Describe)
The Mega Plus MIS Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Mega Plus MIS Spine System can be used in an open approach and percutaneous approach.
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information
unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
5.510(K) SUMMARY
| Submitter's Name: | BK Meditech Co., Ltd. |
|---|---|
| Submitter's Address: | 58, Eunhaengnamu-ro, Yanggam-myeon,Hwaseong-si, Gyeonggi-do 18633, Republic of Korea |
| Submitter's Telephone: | (82) 313529135 |
| Contact Person: | Nathan Wright MSEmpirical Testing Corp.719-351-0248nwright@empiricaltech.com |
| Date Summary was Prepared: | April 13, 2020 |
| Trade or Proprietary Name: | Mega Plus MIS Spine System |
| Common or Usual Name: | Thoracolumbosacral Pedicle Screw System |
| Classification: | Class II per 21 CFR §888.3070 |
| Product Code: | NKB |
| Classification Panel: | Division of Orthopedic Devices |
DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Mega Plus MIS Spine System is a top-loading multiple component, posterior spinal fixation system which consists of pedicle screws, rods, locking bolts, and a transverse (cross) link. Various sizes of these implants are available. Mega Plus MIS Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. Mega Plus MIS Spine System can either be used in an open approach and/or a percutaneous with minimally invasive (MIS) surgical approach. All products are made of titanium alloy (ASTM F136) and CoCrMo alloy (ASTM F1537) approved for medical use. The implants will be provided non-sterile. The locking bolts utilized as part of the proposed system are identical to those of the Mega Plus Spine System (K173180, K183080). The crosslinks and additional universal instrument components that may be utilized as part of the subject system were cleared previously for the Mega Plus Spine System (K173180, K183080).
The safety and effectiveness of this device has not been established when used in conjunction with bone cement or for use in patients with poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye.
INDICATIONS FOR USE
The Mega Plus MIS Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Mega Plus MIS Spine System can be used in an open approach and percutaneous approach.
{4}------------------------------------------------
TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:
- . Indications for use
- Principles of operation
- Materials of manufacture ●
- Implant sizes
- Surgical technique ●
| 510k Number | Trade or Proprietary orModel Name | Manufacturer | PredicateType |
|---|---|---|---|
| K183080 | Mega Plus Spine System | BK Meditech Co., Ltd. | Primary |
| K151648 | Valencia MIS Pedicle ScrewSystem | Altus Partners, LLC | Additional |
| K171421 | VERTICALE® PosteriorSpinal Fixation System | Silony medical GmbH | Additional |
| K110280 | REVLOK™ FenestratedScrew System | Globus Medical Inc. | Additional |
Table 5-1 Predicate Devices
Performance Data
The Mega Plus MIS Spine System has been tested in the following test modes:
- Static compression bending per ASTM F1717
- Static torsion per ASTM F1717 ●
- Static tension bending per ASTM F1717 .
- . Dynamic compression bending per ASTM F1717
The results of this non-clinical testing show that the strength of the Mega Plus MIS Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the Mega Plus MIS Spine System is substantially equivalent to the predicate device.