K Number
K200762

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-23

(30 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The In2Bones USA LLC, CoLink® Cfx System is indicated for stabilization of fractures, revision procedures, osteotomies and reconstruction of the small bones in the foot and ankle including the calcaneus in both pediatric and adult patients.

Device Description

The In2Bones CoLink® Cfx Plating System is a system of plates and screws and surgical instruments intended for stabilization of calcaneal fractures and osteotomies. These subject devices are part of the overarching CoLink® Plating System and will be commonly referred to as the CoLink® Cfx Plating System.

AI/ML Overview

The provided FDA document, K200762, is a 510(k) premarket notification for a medical device (CoLink® Cfx Plating System) that aims to demonstrate substantial equivalence to previously cleared predicate devices. This type of submission relies on comparisons to existing devices rather than new clinical effectiveness or performance studies against specific acceptance criteria.

Therefore, the document does not contain the information required to populate the fields in your request, as it does not describe a study proving the device meets specific acceptance criteria in the way you've outlined.

Here's why and what information is available (and what is not):

  • Acceptance Criteria & Reported Performance: Not applicable. The document focuses on demonstrating similarity to already approved devices.
  • Sample Size, Ground Truth, Adjudication, MRMC, Standalone Studies: Not applicable. These relate to studies designed to evaluate a device's performance against a ground truth, which is not the purpose of this 510(k) submission.
  • Training Set Sample Size & Ground Truth: Not applicable. This refers to AI/ML model development, which is not relevant to this mechanical implant device.

However, the document does provide the following relevant information about comparisons to predicate devices:

1. A table of acceptance criteria and the reported device performance

Criteria CategoryDescription from Document (Implicit "Acceptance" is Substantial Equivalence)Performance (as described in the document)
Indications for Use"stabilization of fractures, revision procedures, osteotomies and reconstruction of the small bones in the foot and ankle including the calcaneus in both pediatric and adult patients.""The CoLink® Cfx Plating System has similar indications to the CoLink® Plating System (K163293) and Silverback Gorilla Plating System (K182148)."
Material(Implicitly, the material should be similar to predicates)"made of Titanium Alloy" (Similar to K163293)
Design/Geometry(Implicitly, the geometry should be similar or justified as equivalent)"This submission is adding additional plates for the calcaneus that will be used with previously cleared screws." "similar in geometry."
Screws Compatibility(Implicitly, compatibility with existing screws)"The 3.5mm and 4.0mm screws initially cleared in the CoLink® Afx Plating System (K181113) can be used with the CoLink® Cfx Plating System."
Mechanical Performance(Implicitly, performance demonstrating equivalence to predicates in applicable mechanical tests.) The document refers to ASTM F382 (Standard Specification for Metallic Bone Plates for External Skeletal Fixation) for four-point bend testing."No additional mechanical testing was required for the CoLink® Cfx Plating System. The 3.5mm and 4.0mm screws are identical to previously cleared screws and no new worst-case plates were added. Engineering analysis was conducted related to the CoLink® Plating four-point bend testing performed per ASTM F382 to show the subject plates are substantially equivalent to the predicate plates."
Sterilization(Implicitly, the sterilization method should be validated or demonstrated to be covered by previous validations.)"not a worst case for sterilization...was adopted into the previous validations for the CoLink® Afx System (K181113)."
Biocompatibility(Implicitly, the device should be biocompatible or demonstrated to be covered by previous validations.)"not a worst case for...biocompatibility...was adopted into the previous validations for the CoLink® Afx System (K181113)."
Shelf Life(Implicitly, the shelf life should be validated or demonstrated to be covered by previous validations.)"not a worst case for...shelf life...was adopted into the previous validations for the CoLink® Afx System (K181113)."
Pyrogenicity(Implicitly, the device should be non-pyrogenic or demonstrated to be covered by previous validations.)"not a worst case for...pyrogenicity...was adopted into the previous validations for the CoLink® Afx System (K181113). Bacterial endotoxin testing (LAL) is performed on each lot."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable. This document describes a 510(k) submission for a mechanical implant, not an AI/ML product or a product requiring a clinical "test set" in the context of effectiveness studies. The evaluations were based on engineering analysis and comparison to predicate devices, not patient data sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No ground truth established by experts for a test set is mentioned or required for this type of submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No test set or adjudication process is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is a mechanical bone fixation system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This is a mechanical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not applicable. The "ground truth" in this context is the FDA's determination of substantial equivalence based on the provided engineering analysis and comparison to predicate devices.

8. The sample size for the training set

  • Not applicable. This is a mechanical device, not an AI/ML product that uses a training set.

9. How the ground truth for the training set was established

  • Not applicable. No training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date April 23, 2020. The text is written in a simple, sans-serif font. The date is written in the standard US format, with the month first, followed by the day and year. The text is black and the background is white.

In2Bones USA, LLC Christine Scifert VP of Quality and Regulatory 6000 Poplar Ave. Suite 115 Memphis, Tennessee 38119

Re: K200762

Trade/Device Name: CoLink Cfx Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 23, 2020 Received: March 24, 2020

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K200762

Device Name CoLink® Cfx Plating System

Indications for Use (Describe)

The In2Bones USA LLC, CoLink® Cfx System is indicated for stabilization of fractures, revision procedures, osteotomies and reconstruction of the small bones in the foot and ankle including the calcaneus in both pediatric and adult patients.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

CoLink® Cfx Plating System April 10, 2020

Company:In2Bones USA, LLC6000 Poplar Ave, Suite 115Memphis, TN 38119901-260-7931
Primary Contact:Christine Scifert
Company Contact:Rebecca Wahl
Trade Name:CoLink® Cfx Plating System
Common Name:Plate, Fixation, BoneScrew, Fixation, Bone
Classification:II
Regulation Number:888.3030 - Single/multiple component metallic bone fixation appliancesand accessories
Panel:87-Orthopedic
Product Code(s):HRS, HWC

The In2Bones CoLink® Cfx Plating System is a system of plates and screws and Device Description: surgical instruments intended for stabilization of calcaneal fractures and osteotomies. These subject devices are part of the overarching CoLink® Plating System and will be commonly referred to as the CoLink® Cfx Plating System.

Indications for Use: The In2Bones USA LLC, CoLink® Cfx System is indicated for stabilization and fixation of fractures, revision procedures, osteotomies and reconstruction of the small bones in the foot and ankle including the calcaneus in both pediatric and adult patients.

Substantial Equivalence: The subject components were demonstrated to be substantially equivalent to the following systems previously cleared by the FDA:

{4}------------------------------------------------

Primary Predicate

  • K163293 In2Bones CoLink® Plating System .
    Additional Predicates

  • K182148 – Paragon 28 Silverback Gorilla Plating System

  • K181113 In2Bones CoLink® Afx Plating System ●

Similar to the primary predicate device (K163293), the subject CoLink® Cfx Plating System is made of Titanium Alloy and provided sterile. The CoLink® Cfx Plating System has similar indications to the CoLink® Plating System (K163293) and Silverback Gorilla Plating System (K182148). This submission is adding additional plates for the calcaneus that will be used with previously cleared screws. The 3.5mm and 4.0mm screws initially cleared in the CoLink® Afx Plating System (K181113) can be used with the CoLink® Cfx Plating System. The previously cleared CoLag® screws (K170518 and K180377) that are intended to be used as stand-alone screws can also be used during the procedure with the CoLink® Cfx Plating System, but would not be used in conjunction with the plates or screws. The subject plates and screws have been demonstrated to be substantially equivalent to the previously cleared devices identified above as the products are similar in indications, materials and geometry.

Performance Testing: No additional mechanical testing was required for the CoLink® Cfx Plating System. The 3.5mm and 4.0mm screws are identical to previously cleared screws and no new worst-case plates were added. Engineering analysis was conducted related to the CoLink® Plating four-point bend testing performed per ASTM F382 to show the subject plates are substantially equivalent to the predicate plates. The CoLink® Cfx Plating System is not a worst case for sterilization, biocompatibility, shelf life and pyrogenicity and was adopted into the previous validations for the CoLink® Afx System (K181113). Bacterial endotoxin testing (LAL) is performed on each lot.

Conclusion

Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.