K Number
K200666
Device Name
Stryker Xia 3 Power Adaptor
Manufacturer
Date Cleared
2020-04-23

(41 days)

Product Code
Regulation Number
888.3070
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/ anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis); tumor; pseudoarthrosis; and failed previous fusion. The XIA® 4.5 Spinal System is intended for anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; failed previous fusion. The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenttal scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The intended use for the Xia 3 Power Adaptor is to facilitate the placement of pedicle screws using the power technique (corded and cordless), the use of the Xia 3 Power Adaptor is intended for exclusive use with the Stryker Cordless and Corded Power Drivers. When the power adaptors are attached, the Stryker Power Drivers (corded and cordless) provide power to rotate screwdrivers for the insertion of pedicle screws. Pedicle screws from select Stryker Spine implant systems may be implanted in the non-cervical spine using powered (corded and cordless) instrumentation. The systems included are the family of XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System.
Device Description
This special 510(k) premarket notification is to expand the intended use of the Xia 3 Power Adaptor. The Xia 3 Power Adaptor instrument accessory is intended to facilitate the insertion of pedicle screws using powered instrumentation. To facilitate the insertion of pedicle screws using the power technique, the use of the Xia 3 Power Adaptor is intended for use with Stryker power system drivers (corded and cordless). The adaptor serves as a mechanical interface between the power driver and screwdriver instrument. When the adaptor is attached, the driver (corded and cordless) provides appropriate power to rotate screw drivers for the insertion of pedicle screws. No changes have been made to the indications for use of the associated thoracolumbar spinal implant systems XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System. The indications for use of each spinal system remain consistent with their most recent 510(k) clearance.
More Information

Not Found

No
The document describes a mechanical adaptor for powered surgical tools used in spinal surgery. There is no mention of any software, algorithms, or data processing that would indicate the presence of AI or ML.

Yes
The device is described as a "Spinal System" intended for various spinal conditions, indicating it is used in the treatment or diagnosis of disease.

No

The device is an instrument accessory designed to facilitate the surgical insertion of pedicle screws, not to diagnose medical conditions.

No

The device described is a mechanical adaptor intended to connect a power driver to a screwdriver for inserting pedicle screws, which are physical implants. The description focuses on hardware components and their mechanical function, with no mention of software as the primary or sole component.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening.
  • Device Description and Intended Use: The text clearly describes the device as a spinal system and an instrument accessory used for spinal fixation and fusion. It is implanted or used during surgical procedures on the spine.
  • Lack of Specimen Analysis: There is no mention of the device analyzing any biological specimens from the patient. Its function is mechanical support and facilitation of surgical procedures.

Therefore, the device described is a surgical implant/instrumentation system, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/ anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 4.5 Spinal System is intended for anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; failed previous fusion.

The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.

When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenttal scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The intended use for the Xia 3 Power Adaptor is to facilitate the placement of pedicle screws using the power technique (corded and cordless), the use of the Xia 3 Power Adaptor is intended for exclusive use with the Stryker Cordless and Corded Power Drivers. When the power adaptors are attached, the Stryker Power Drivers (corded and cordless) provide power to rotate screwdrivers for the insertion of pedicle screws.

Pedicle screws from select Stryker Spine implant systems may be implanted in the non-cervical spine using powered (corded and cordless) instrumentation. The systems included are the family of XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System.

Product codes

NKB

Device Description

This special 510(k) premarket notification is to expand the intended use of the Xia 3 Power Adaptor.

The Xia 3 Power Adaptor instrument accessory is intended to facilitate the insertion of pedicle screws using powered instrumentation. To facilitate the insertion of pedicle screws using the power technique, the use of the Xia 3 Power Adaptor is intended for use with Stryker power system drivers (corded and cordless). The adaptor serves as a mechanical interface between the power driver and screwdriver instrument. When the adaptor is attached, the driver (corded and cordless) provides appropriate power to rotate screw drivers for the insertion of pedicle screws. No changes have been made to the indications for use of the associated thoracolumbar spinal implant systems XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System. The indications for use of each spinal system remain consistent with their most recent 510(k) clearance.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Spinal, non-cervical

Indicated Patient Age Range

Skeletally mature patients, pediatric patients

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Performance Evaluation: The Xia 3 Power Adaptor Instrument has demonstrated substantial equivalence to the predicate device. Engineering analysis demonstrated connection of the Xia 3 Power Adaptor to various Stryker Power Drivers does not represent a new worse case or affect the performance of the Xia 3 Power Adaptor.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K172724, K170496, K152632, K120434, K122845, K111478

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 23, 2020

Stryker Spine Megan Callanan Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401

Re: K200666

Trade/Device Name: Stryker Xia 3 Power Adaptor Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: March 11, 2020 Received: March 13, 2020

Dear Megan Callanan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, MBE Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K200666

Device Name Xia 3 Power Adaptor

Indications for Use (Describe)

The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/ anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 4.5 Spinal System is intended for anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; failed previous fusion.

The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.

When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenttal scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

3

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

510(k) Summary: Xia 3 Power Adaptor
Submitter:Stryker Spine
Contact Person :Name: Megan Callanan
Phone: (551)262-2429
Email: megan.callanan1@stryker.com
Date Prepared:3/11/2020
Trade Name:Xia 3 Power Adaptor
Common Name:Spinal Fixation Appliances, Instrument Accessory
Proposed Class:Class II
Classification Name:Thoracolumbosacral Pedicle Screw System
Regulation Number:888.3070
Product Code:NKB
Predicate Device:Primary Predicate is the most recently cleared system K172724.
Additional Predicate: K170496, K152632, K120434, K122845, K111478
Device Description:This special 510(k) premarket notification is to expand the intended use of the Xia 3 Power Adaptor.

The Xia 3 Power Adaptor instrument accessory is intended to facilitate the insertion of pedicle screws using powered instrumentation. To facilitate the insertion of pedicle screws using the power technique, the use of the Xia 3 Power Adaptor is intended for use with Stryker power system drivers (corded and cordless). The adaptor serves as a mechanical interface between the power driver and screwdriver instrument. When the adaptor is attached, the driver (corded and cordless) provides appropriate power to rotate screw drivers for the insertion of pedicle screws. No changes have been made to the indications for use of the associated thoracolumbar spinal implant systems XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System. The indications for use of each spinal system remain consistent with their most recent 510(k) clearance. |
| Intended Use | The intended use for the Xia 3 Power Adaptor is to facilitate the placement of pedicle screws using the power technique (corded and cordless), the use of the Xia 3 Power Adaptor is intended for exclusive use with the Stryker Cordless and Corded Power Drivers. When the power adaptors are attached, the Stryker Power Drivers (corded and cordless) provide power to rotate screwdrivers for the insertion of pedicle screws.

Pedicle screws from select Stryker Spine implant systems may be implanted in the non-cervical spine using powered (corded and cordless) instrumentation. The |
| Indications for Use: | systems included are the family of XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System. |
| | The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. |
| | The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. |
| | The XIA® 4.5 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; failed previous fusion. |
| | The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. |
| | When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, |
| | pseudarthrosis, and/or failed previous fusion. This system is intended to be used with
autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior
approach.
Except for the staples, when used for posterior non-cervical pedicle screw fixation in
pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to
fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis)
including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis.
Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients
diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or
trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be
used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a
posterior approach.
The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal
System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle |
| | fixation of the spine to provide immobilization and stabilization of spinal segments in
skeletally mature patients as an adjunct to fusion for the following indications:
degenerative disc disease (defined as back pain of discogenic origin with
degeneration of the disc confirmed by history and radiographic studies);
spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e.
scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous
fusion. |
| Summary of the
Technological
Characteristics | The Xia 3 Power Adaptor shares the same materials, fundamental scientific
technologies and design as the predicate device.
The screws are also intended to be inserted manually or with Stryker's corded and
cordless power drivers. The power adaptor accessory instrument assembly aids in
the rotation of the bone screw to facilitate insertion. |
| Non-Clinical
Performance
Evaluation | The Xia 3 Power Adaptor Instrument has demonstrated substantial equivalence to the
predicate device. Engineering analysis demonstrated connection of the Xia 3 Power
Adaptor to various Stryker Power Drivers does not represent a new worse case or
affect the performance of the Xia 3 Power Adaptor. |
| Conclusion | Based on the design features, the use of established well known materials, feature
comparisons, and indications for use the subject devices have demonstrated
substantial equivalence to the identified predicate devices. |

5