(345 days)
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures.
The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
The HipXpert system provides a patient-specific hip arthroplasty surgical plan allowing for accurate acetabular positioning using CT mapping of a patient's pelvis using a software application and a reusable, manual, mechanical navigation instrument.
The HipXpert software planning application uses patient image data to create a detailed 3D model of the pelvis as well as the instrument settings necessary for proper acetabular cup orientation.
The HipXpert mechanical instrument has three legs which are secured to the pelvis. The legs form three points which define the sextant plane. The HipXpert mechanical instrument has two protractors that are adjusted to orientate an indicator pin in the direction of the desired orientation of the acetabular component.
The subject HipXpert 3D Display and Anchoring Application utilizes the previously cleared (K093491) HipXpert planning application and HipXpert tools in addition to a mixed reality headset (Microsoft HoloLens2) and QR target. The HoloLens2 is an off-the-shelf component developed and manufactured by Microsoft which is used to view superimposed 3D images from the HipXpert planning application on the real HipXpert tool. In order to properly orient the 3D images displayed by the HoloLens2, a QR target is used to anchor these 3D images in space as they are overlaid on the real HipXpert tool.
The HipXpert System, specifically the 3D Display and Anchoring application with the HoloLens2, was evaluated to ensure accurate alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
Here's a breakdown of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text details various validation and verification activities rather than specific quantitative acceptance criteria in a table format. However, the performance reported for the HipXpert System focused on aspects like software validation, system orientation and position accuracy, image registration and tracking accuracy, and headset display performance.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Software functionality for file identification, QR recognition, and anchoring. | Successfully validated. |
| Overall system orientation and position accuracy. | Verified in a cadaveric model using screws placed into the acetabulum. |
| System accuracy, image registration accuracy, and tracking accuracy under varying conditions (light, motion, distance, angle). | Verified using methods analogous to ASTM F2554-10. (Specific quantitative results not provided in the summary, but the methods were established for accuracy evaluation). |
| Headset display performance (Field of View, resolution, luminance, distortion, contrast ratio, contrast of physical object, location of virtual image, stability of virtual objects due to motion). | Demonstrated by verifying all listed elements. (Specific quantitative results not provided in the summary, but verification confirms performance met internal standards). |
2. Sample Size Used for the Test Set and Data Provenance
The non-clinical testing included a cadaveric model for verifying overall system orientation and position accuracy. The specific number of cadaveric models used is not specified.
Regarding data provenance for the accuracy testing, it was conducted internally as part of the device's verification and validation. The text does not specify the country of origin of the data or whether it was retrospective or prospective, but given it's a device validation, it would be considered prospective for the purpose of the study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts used to establish ground truth for the test set.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method for establishing ground truth for the test set. The validation methods mentioned (cadaveric model using screws placed into the acetabulum and methods analogous to ASTM F2554-10) imply objective measurement rather than expert consensus on subjective evaluations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No MRMC comparative effectiveness study was mentioned. The HipXpert system is a surgical planning and alignment tool, and the focus of this submission is on the accuracy of the augmented reality display for guiding tool placement, not on interpreting images or improving human reader performance in a diagnostic context.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The HipXpert system, including the 3D Display and Anchoring application, is inherently a human-in-the-loop system. It provides visual guidance to a human surgeon. Therefore, a standalone (algorithm only) performance evaluation without human interaction would not be applicable or relevant to its intended use and was not described. The validation focused on the system's accuracy while being used by a human, even if a cadaver was the "patient."
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the accuracy evaluations appears to be based on objective physical measurements and established engineering standards. For instance, in the cadaveric model, the ground truth for "overall system orientation and position accuracy" would likely be the known, precise placement of screws in the acetabulum, validated by direct measurement or pre-defined landmarks. For system accuracy, image registration accuracy, and tracking accuracy, the ground truth would be precise measurements and positional data defined by the test setup, as per methods analogous to ASTM F2554-10, which are standards for evaluating the accuracy of medical image processing software and surgical navigation systems using precise physical fiducials and defined spatial relationships.
8. The Sample Size for the Training Set
The document does not provide information on the sample size used for the training set of any machine learning algorithms within the HipXpert System. The text primarily focuses on the validation of the system's functionality and accuracy, specifically the 3D display and anchoring application, rather than the development of the underlying planning algorithms (which existed in the predicate device K093491).
9. How the Ground Truth for the Training Set Was Established
Since no specific information about a training set for machine learning was provided, the method for establishing its ground truth is also not mentioned. The HipXpert software planning application itself, which creates the 3D model and instrument settings, was part of a previously cleared device (K093491). The current submission focuses on the augmented reality display portion.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 28, 2021
Surgical Planning Associates, Inc % Kellen Hills Quality and Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484
Re: K200384
Trade/Device Name: HipXpert System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OSF, OLO Dated: December 23, 2020 Received: December 29, 2020
Dear Kellen Hills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For; Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200384
Device Name HipXpert System
Indications for Use (Describe)
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures.
The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K200384
| (a)(1) | Date Prepared: | December 23, 2020 |
|---|---|---|
| Submitted By: | Surgical Planning Associates, Inc.New England Baptist Hospital,125 Parker Hill Ave Suite 545 | |
| Phone: | Boston, MA 02120617-277-4434 | |
| Contact: | Stephen B Murphy, M.D. | |
| Prepared By: | Orchid Design80 Shelton Technology Center,Shelton, CT 06484 | |
| Phone: | (203) 922 0105 | |
| (a)(2) | Proprietary Name: | HipXpert System |
| Common Name: | Patient specific manual orthopedicstereotaxic system | |
| Classification Name and Reference: | 21CFR 882.4560 - Stereotaxic instrument | |
| Product Codes: | OSF - Patient Specific Manual OrthopedicStereotaxic SystemOLO - Orthopedic Stereotaxic Instrument | |
| (a)(3) | Predicate Devices: | |
| Primary: | Surgical Planning Associates, Inc - Hip | |
| Sextant Instrument System (K093491) | ||
| Reference: | Augmedics, Ltd - xvision Spine (K190929) |
(a)(4) Device Description:
The HipXpert system provides a patient-specific hip arthroplasty surgical plan allowing for accurate acetabular positioning using CT mapping of a patient's pelvis using a software application and a reusable, manual, mechanical navigation instrument.
The HipXpert software planning application uses patient image data to create a detailed 3D model of the pelvis as well as the instrument settings necessary for proper acetabular cup orientation.
{4}------------------------------------------------
The HipXpert mechanical instrument has three legs which are secured to the pelvis. The legs form three points which define the sextant plane. The HipXpert mechanical instrument has two protractors that are adjusted to orientate an indicator pin in the direction of the desired orientation of the acetabular component.
The purpose of this submission is to introduce a variant of the device that was cleared in K093491 which uses a different software application to allow the user to see the existing 3D objects provided in the surgery plans in three-dimensions on mixed reality lenses. Additionally, the indications for use have been amended to add clarity and align with the "Physical State" as specified by product code OSF.
The subject HipXpert 3D Display and Anchoring Application utilizes the previously cleared (K093491) HipXpert planning application and HipXpert tools in addition to a mixed reality headset (Microsoft HoloLens2) and QR target. The HoloLens2 is an off-the-shelf component developed and manufactured by Microsoft which is used to view superimposed 3D images from the HipXpert planning application on the real HipXpert tool. In order to properly orient the 3D images displayed by the HoloLens2, a QR target is used to anchor these 3D images in space as they are overlaid on the real HipXpert tool.
- (a)(5) Indications for Use:
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures.
The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
(a)(6) Comparison of Technological Characteristics:
The HipXpert system is substantially equivalent to the previously cleared hip sextant system based on similarities in intended use, design, materials, packaging, sterilization and performance. The only difference is an additional software application which allows the user to view existing 3D objects provided in the surgery plans in three-dimensions using mixed reality lenses. The use of a headset for presenting stereoscopic augmented reality (AR) display of patient's anatomy is not a new feature and has been previously cleared under K190929 for the xvision Spine (reference device). The technological characteristics do not raise any new questions of safety and efficacy.
(b)(1) Non-clinical testing:
The subject device modification was successfully evaluated according to the following:
- Validation of the software used for file identification, QR target recognition and file anchoring
- . Verification of overall system orientation and position accuracy in a cadaveric model using screws placed into the acetabulum
{5}------------------------------------------------
- . Verification of system accuracy, image registration accuracy and tracking accuracy using methods analogous to ASTM F2554-10 under varying conditions of light, motion, distance, and angle
- . Performance of the Headset display was demonstrated by verifying the following elements: Field of View (FOV), resolution, luminance, distortion, contrast ratio, contrast of physical object, location of the virtual image, and stability of the virtual objects due to motion
(b)(2) Clinical testing:
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
(b)(3) Conclusions:
Based on the information provided in this premarket notification, we believe that the subject HipXpert system demonstrates substantial equivalence to the predicate Hip Sextant Instrument System.
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).