K Number
K200173
Date Cleared
2020-05-01

(99 days)

Product Code
Regulation Number
876.5010
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for endoscopic removal of biliary stones. The device is supplied sterile and intended for single use only.

Device Description

The subject Stone Retrieval Balloon is comprised of a latex balloon mounted at the distal end of a Pebax catheter with three internal lumens. For some models, the balloon can be inflated to 8.5mm, 12mm, 15mm and 18mm only. But for some models, the balloon can be inflated to three sizes, 8.5 mm, 12 mm and 15 mm diameters, or 13mm, 15mm and 18mm. Radiopaque bands placed at the distal and proximal ends of the balloon provide fluoroscopic visualization of the balloon location. The catheter length is 2000mm with a outer diameter of 2.4mm. The three lumens correspond to a balloon inflation port, a wire guide port and an injection port. A stopcock is included at the proximal end of the balloon inflation port to control air movement into or out of the balloon. For rapid exchange models, there is a small hole on the catheter which is about 200mm far away from the distal end, it is used to insert the guide wire rapidly. EO Sterilization and use for single use only.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a Stone Retrieval Balloon. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a novel device or a PMA.

Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria in the context of AI/ML device performance. The performance data provided is related to bench testing to demonstrate substantial equivalence to a predicate device, not, for example, the performance of an AI algorithm in image interpretation.

Here's a breakdown of why the requested information cannot be extracted from this document:

  • No AI/ML Component: The device described is a physical medical instrument (a stone retrieval balloon) used for endoscopic removal of biliary stones. There is no mention of any artificial intelligence, machine learning, or software component that would involve image analysis, diagnostic capabilities, or human-AI interaction.
  • Focus on Substantial Equivalence: The entire 510(k) submission is built around demonstrating that the "Stone Retrieval Balloon" is substantially equivalent to existing predicate devices (Tri-Ex Extraction Balloon and 3-Lumen Extraction Balloon). This involves comparing technical characteristics, materials, and bench performance, not the diagnostic accuracy or efficacy of an AI system.
  • Performance Data: The "Performance Data" section (5.8) explicitly states: "The Stone Retrieval Balloon meets all design specifications and medical for biocompatibility (ISO 1099) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device." This refers to physical and material properties, not AI performance metrics.
  • No Clinical Test: Section 5.9 explicitly states: "No Clinical test is included in this submission." This further reinforces that there isn't a study proving a specific performance metric against a clinical outcome or ground truth in the way an AI/ML device would be evaluated.

In summary, the provided document does not describe the acceptance criteria or a study that proves an AI/ML device meets those criteria, as the device in question is a physical medical instrument without an AI/ML component.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 1, 2020

Hangzhou AGS MedTech Co., Ltd. Chunqi Han R&D Director Building 5, Building 6, No.597 Kangxin Road Yuhang District Hangzhou, 311106 CHINA

K200173 Re: Trade/Device Name: Stone Retrieval Balloon Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: GCA Dated: January 21, 2020 Received: January 23, 2020

Dear Chunqi Han:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling

{1}------------------------------------------------

(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel G. Walter, Jr. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200173

Device Name Stone Retrieval Balloon

Indications for Use (Describe)

The device is intended for endoscopic removal of biliary stones. The device is supplied sterile and intended for single use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.

5.1 Submitter

Submitted by/Owner:Hangzhou AGS MedTech Co., Ltd.Building 5, Building 6, NO.597 Kangxin Road YuhangDistrict, Hangzhou, Zhejiang 311106 China
EstablishmentRegistration Number:3010288205
Registration Status:Active
Contact Person:Yanping FuPhone: 0086-15958493282Fax: 0086-0571-87671225Email: fuyp@bioags.com
Date Prepared:July 18, 2019

5.2 Proposed Device

Trade Name:/
Device Name:Stone Retrieval Balloon
Common Name:Stone Retrieval Balloon
Regulation class:Class II
Regulation Number:876.5010
Regulation Description:Biliary Catheter and Accessoriess
Review Panel:Gastroenterology/Urology
Product Code:GCA
Product Code Name:Biliary Catheter For Stone Removal That May AlsoAllow For Irrigation And Contrast Injection

5.3 Predicate Device

Trade Name:Tri-Ex Extraction Balloon With Multiple Sizing
Device Name:Tri-Ex Extraction Balloon With Multiple Sizing
Common Name:Extraction Balloon With Multiple Sizing
510(k) Number:K170292
Regulation class:Class II
Regulation Number:876.5010
Regulation Description:Biliary Catheter and Accessoriess
Review Panel:Gastroenterology/Urology
Product Code:GCA
Product Code Name:Biliary Catheter For Stone Removal That May Also AllowFor Irrigation And Contrast Injection

{4}------------------------------------------------

Trade Name:Single Use 3-Lumen Extraction Balloon V
Device Name:Single Use 3-Lumen Extraction Balloon V
Common Name:3-Lumen Extraction Balloon
510(k) Number:K091495
Regulation class:Class II
Regulation Number:876.5010
Regulation Description:Biliary Catheter and Accessoriess
Review Panel:Gastroenterology/Urology
Product Code:FGE/LQR
Product Code Name:Stents, Drains And Dilators For The Biliary Ducts/Dislodger, Stone, Biliary

5.4 Device Description

The subject Stone Retrieval Balloon is comprised of a latex balloon mounted at the distal end of a Pebax catheter with three internal lumens. For some models, the balloon can be inflated to 8.5mm, 12mm, 15mm and 18mm only. But for some models, the balloon can be inflated to three sizes, 8.5 mm, 12 mm and 15 mm diameters, or 13mm, 15mm and 18mm. Radiopaque bands placed at the distal and proximal ends of the balloon provide fluoroscopic visualization of the balloon location. The catheter length is 2000mm with a outer diameter of 2.4mm. The three lumens correspond to a balloon inflation port, a wire guide port and an injection port. A stopcock is included at the proximal end of the balloon inflation port to control air movement into or out of the balloon. For rapid exchange models, there is a small hole on the catheter which is about 200mm far away from the distal end, it is used to insert the guide wire rapidly. EO Sterilization and use for single use only.

5.5 Indication for use statement

The device is intended for endoscopic removal of biliary stones. The device is supplied sterile and intended for single use only.

5.6 Comparison of Technology Characteristics

Our proposed device Stone Retrieval Balloon is substantially equivalent to the predicate devices. The differences between the Stone Retrieval Balloon and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below

{5}------------------------------------------------

Difference ItemProposed devicePredicate deviceComparison
Common nameStone Retrieval BalloonTri-Ex Extraction Balloon withMultiple Sizing/
Trade name/Tri-Ex Extraction Balloon withMultiple Sizing/
Model number5051 series, 5053 series, 5451series, 5453 seriesTXR-8.5-12-15-A/
510(k) submitterHangzhou AGS MedTech Co., Ltd.Wilson-Cook Medical, Inc./
510(k) number/K170292/
TechnicalBalloonMax outerdiameter8.5/12/15/18/8.5-12-15/13-15-18mm8.5 mm, 12 mm and 15mmDifferent.Our balloon outer diameter includes a range from 8.5mm to18mm, the predicate device includes a range from 8.5mm to15mm. For the 18mm balloon, we provide the followingsupport evidence to demonstrate that it will not raise differentquestions regarding its safety and effectiveness.First, we searched in the FDA 510k Premarket Notificationdatabase for other similar retrievalballoon catheter, and we found Olympus's 3-LumenExtraction Balloon (K091495), maximum diameter of theballoon(mm) is 20mm, it is shown in the 510(k) summary:Second, there is a literature that retrieval balloon catheter isused to remove stones in the bile duct which the meancommon bile duct (CBD) dilation was 19.2mm±3.9 and themean size of stones 15.8±2.9.
Difference ItemProposed devicePredicate deviceComparison
RecommendedWireGuideDiameterinch0.035 /0.025ich0.035ichDifferent. Our proposed device has another option for wireguide, the difference don't raise question about safety andeffectiveness.
PerformanceBalloon Diamater, Balloon Deflation, BalloonStrength, etc.Balloon Diameter, Balloon Deflation,Balloon Strength, etc.Different.We conduct bench performance for our proposed device andthe predicate device, please refer to Section 18 of thissubmission, the test results show that our proposed device issubstantial equivalence with the predicate device.
BiologicalMaterialsorsubstancesin contactwith thesamehumantissue orbodyfluidsSheath: Pebax;Radiopaque band: Tal;Bind wire: PET;Latex balloon: Natural latex;Binder: UV-curing adhesive;Marker: Polyurethane ink.Catheter: nylon;Balloon: Latex.Different.Biocompatibility tests have been done for the difference.Biological risks are acceptable.
BiocompatibilityIn Vitro Cytotoxicity Test : ISO 10993-5: 2009;Skin sensitization Test : ISO 10993-10: 2010;Intracutaneous Reactivity Test: ISO 10993-10: 2010;Unknown

{6}------------------------------------------------

{7}------------------------------------------------

5.7 Applicable Guidance Document NA

5.8 Performance Data

The Stone Retrieval Balloon meets all design specifications and medical for biocompatibility (ISO 1099) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.

5.9 Clinical Test

No Clinical test is included in this submission.

5.10 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS Medtech Co., Ltd has demonstrated that proposed device Electrosurgical System is substantially equivalent to the predicate devices.

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.