K Number
K200075

Validate with FDA (Live)

Date Cleared
2020-03-13

(59 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3DMetal Femoral Cones are indicated for use with Revision and GMK Hinge knee systems, as well as the GMK femoral extension stems and offsets. Specific indications are as follows:

-Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis,
rheumatoid arthritis or polyarthritis.

-Post traumatic loss of joint configuration.

-Considerable loss of function of the knee joint.

-High-grade joint destruction requiring additional stabilization and reconstruction of bone defects.

-Primary implantation failure.

-Former revision arthroplasty.

Device Description

The 3DMetal Diaphyseal Femoral Cones are sterile implantable devices intended to be used in the diaphyseal side of the femoral component in order to fill and reconstruct large bone deficiencies and cavitary defects in the diaphysis of the distal femur. The subject devices are available in four different sizes and they are to be cemented to the extension stem used in GMK Revision and GMK Hinge systems in Total Knee Arthroplasty procedures. Analogously to the reference devices, the 3DMetal Diaphyseal Femoral Cones are manufactured using Electron Beam Melting (EBM) process with titanium alloy powder.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "3DMetal Diaphyseal Femoral Cones." This is a regulatory submission demonstrating substantial equivalence to a predicate device, not a study evaluating AI performance or a comparative effectiveness study. Therefore, most of the requested information regarding AI-specific criteria, human readers, ground truth establishment for AI, and training set details are not applicable to this document.

However, I can extract information related to the device's acceptance criteria (in terms of performance testing) and the studies conducted to show it meets those criteria.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / TestDevice Performance / Result
Non-Clinical Studies
Design Validation
Design Rationale Shape, Dimension and Range of ProductTest Report B2 + Addendum was provided in support of compliance. The text states: "The slight difference in sizes between the subject and the predicate devices does not raise new question of safety or effectiveness as demonstrated by design validation testing."
Characterization Testing
Dynamic Fatigue Test (according to IL 07.09.5173 rev.0)Report 00812-009932-1 was provided. The text implies the device passed this test to demonstrate safety and performance.
Stereological evaluation (according to ASTM F1854-15)Test Report 00812-009932-2 (Test report B3) was provided. This likely validates the porous structure (e.g., porosity of trabecular layer) to be comparable to the predicate.
BiocompatibilityThe product material (Ti6Al4V) is "commonly used for implantable medical devices," implying established biocompatibility. (No specific report number given but implied through material selection). The submission states biocompatibility is shared with the predicate.
SterilityThe product is "sterile implantable devices." The submission states sterility is shared with the predicate. (No specific report number or method for this test is explicitly given in the provided text, but it's a standard requirement for sterile devices).
Shelf LifeThe submission states shelf life is shared with the predicate. (No specific report number or method for this test is explicitly given in the provided text, but it's a standard requirement).
PackagingThe submission states packaging is shared with the predicate. (No specific report number or method for this test is explicitly given in the provided text, but it's a standard requirement).
Pyrogenicity: Bacterial endotoxin test (LAL test)According to European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>). The device is not labeled as non-pyrogenic or pyrogen free, implying it meets the general pyrogenicity requirements for its intended use, but not necessarily a "pyrogen-free" claim.
Pyrogenicity: Pyrogen test (according to USP chapter <151>)For pyrogenicity determination. The device is not labeled as non-pyrogenic or pyrogen free, implying it meets the general pyrogenicity requirements for its intended use.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated for each test. The studies referenced (e.g., dynamic fatigue, stereological evaluation) would typically involve a specific number of device samples for testing, but the exact quantity is not detailed in this summary.
  • Data Provenance: The studies are "Non-Clinical Studies" conducted to validate the device design and characterize its performance. This indicates laboratory testing of the device itself, not data derived from human subjects or clinical experience. The provenance is internal to the manufacturer (Medacta International SA), as these are design validation and characterization tests. These are prospective tests on newly manufactured samples of the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not Applicable. This is a submission for a physical medical implant, not an AI or diagnostic device that requires expert-established ground truth from images or other data. The "ground truth" here is established by engineering standards, material science, and biomechanical principles validated through laboratory testing.

4. Adjudication Method for the Test Set

  • Not Applicable. As this is not a study involving human interpretation of data, there is no adjudication method in the sense of expert consensus. The results are determined by objective measurements from laboratory tests against predefined engineering specifications or recognized standards (e.g., ASTM, European Pharmacopoeia, USP).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No. No MRMC comparative effectiveness study was done, as this is a physical medical implant, not an AI diagnostic or assistance device. The document explicitly states: "No clinical studies were conducted."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not Applicable. This is not an algorithm or AI device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • For the non-clinical tests, the "ground truth" is derived from established engineering standards, material specifications, and biomechanical principles. For example:
    • Dynamic Fatigue Test: Ground truth is the endurance limit or fatigue life expected for the material and design under specified loads, often benchmarked against predicate devices or industry standards.
    • Stereological Evaluation: Ground truth relates to the measured porosity and pore size distribution meeting predefined specifications for the 3DMetal structure, often against specific ASTM standards (e.g., F1854-15 mentioned).
    • Biocompatibility: Ground truth is that the material (Ti6Al4V) is known and accepted as biocompatible for implantable devices, and this is confirmed by testing for specific biological responses (e.g., pyrogenicity).

8. The Sample Size for the Training Set

  • Not Applicable. This is not an AI or machine learning device; therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. There is no training set for this device.

{0}------------------------------------------------

March 13, 2020

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K200075

Trade/Device Name: 3DMetal Diaphyseal Femoral Cones Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: January 13, 2020 Received: January 14, 2020

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200075

Device Name 3DMetal Diaphyseal Femoral Cones

Indications for Use (Describe)

The 3DMetal Femoral Cones are indicated for use with Revision and GMK Hinge knee systems, as well as the GMK femoral extension stems and offsets. Specific indications are as follows:

-Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis,

rheumatoid arthritis or polyarthritis.

-Post traumatic loss of joint configuration.

-Considerable loss of function of the knee joint.

-High-grade joint destruction requiring additional stabilization and reconstruction of bone defects.

-Primary implantation failure.

-Former revision arthroplasty.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

2.0 510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Director of Quality and Regulatory, Medacta USA Date Prepared: January 13, 2020

II. Device

Device Proprietary Name:3DMetal Diaphyseal Femoral Cones
Common or Usual Name:Total Knee Prosthesis
Classification Name:Knee joint patellofemorotibial polymer/metal/polymer semi-
constrained cemented prosthesis
Primary Product Code:JWH
Regulation Number:21 CFR 888.3560
Device ClassificationII

III. Predicate Device

Substantial equivalence is claimed to the following device:

  • Trabecular Metal™ Femoral Cone Augments, K103517, Zimmer Trabecular Metal A Technology, Inc
    In addition, the following Reference device is cited within the submission:

  • 3DMetal Tibial Cones (K170149), Medacta International SA A

IV. Device Description

The 3DMetal Diaphyseal Femoral Cones are sterile implantable devices intended to be used in the diaphyseal side of the femoral component in order to fill and reconstruct large bone deficiencies and cavitary defects in the diaphysis of the distal femur.

{4}------------------------------------------------

The subject devices are available in four different sizes and they are to be cemented to the extension stem used in GMK Revision and GMK Hinge systems in Total Knee Arthroplasty procedures.

Analogously to the reference devices, the 3DMetal Diaphyseal Femoral Cones are manufactured using Electron Beam Melting (EBM) process with titanium alloy powder.

V. Indications for Use

The 3DMetal Femoral Cones are indicated for use with the GMK Revision and GMK Hinge knee systems, as well as the GMK femoral extension stems and offsets. Specific indications are as follows:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, ● rheumatoid arthritis or polyarthritis.
  • Post traumatic loss of joint configuration. ●
  • Considerable loss of function of the knee joint. ●
  • High-grade joint destruction requiring additional stabilization and reconstruction of ● bone defects.
  • Primary implantation failure. ●
  • Former revision arthroplasty.

VI. Comparison of Technological Characteristics

The 3DMetal Diaphyseal Femoral Cones and the predicate Trabecular Metal Femoral Cone Augments share the following characteristics:

  • shape:
  • porosity of trabecular layer;
  • biocompatibility; ●
  • device usage:
  • sterility;
  • shelf life; and ●
  • packaging. ●

The 3DMetal Diaphyseal Femoral Cones differ from the predicate devices as follow:

  • sizes
  • material

Discussion

Medacta International SA has not made any change to the intended use, shape, device usage, biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the predicate devices.

The slight difference in sizes between the subject and the predicate devices does not raise new question of safety or effectiveness as demonstrated by design validation testing.

The different material of the 3DMetal Diaphyseal Femoral Cones with respect to the predicate devices, do not compromise device safety and performance since the material of the subject

{5}------------------------------------------------

device, Ti6Al4V is commonly used for implantable medical devices. Additionally, the subject device material and manufacturing process are shared with the reference devices, 3DMetal Tibial Cones (K170149).

The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject devices design, and supports the substantial equivalence of the 3DMetal Femoral Cones implants to the identified predicate devices.

Performance Data VII.

Based on the risk analysis, design validation and characterization testing were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:

Non-Clinical Studies

  • DESIGN VALIDATION .
    • o 3DMetal Diaphyseal Femoral Cones Design Rationale Shape, Dimension and Range of Product. Test Report B2 + Addendum.
  • CHARACTERIZATION TESTING o
    • o 3DMetal Diaphyseal Femoral Cones Dynamic Fatigue Test according to IL 07.09.5173 rev.0 and Report 00812-009932-1. Test report B1
    • o 3DMetal Diaphyseal Femoral Cones Stereological evaluation according to ASTM F1854-15 and Test Report 00812-009932-2. Test report B3
  • o PYROGENICITY:
    • o Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
    • 0 Pyrogen test according to USP chapter <151> for pyrogenicity determination
    • o The subject devices are not labeled as non-pyrogenic or pyrogen free.

Clinical Studies:

  • No clinical studies were conducted. ●

VIII. Conclusion

The information provided above supports that the 3DMetal Diaphyseal Femoral Cones are as safe and effective as the predicate devices. Therefore, it is concluded that the 3DMetal Diaphyseal Femoral Cones are substantially equivalent to the predicate device.

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.