K Number
K200068
Manufacturer
Date Cleared
2020-06-29

(168 days)

Product Code
Regulation Number
888.3040
Panel
OR
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex DynaNite K-Wire is indicated for fixation of osteotomies and reconstruction of the lesser toes following correction procedure for hammertoe.

Device Description

The Arthrex DynaNite K-Wire is a double tipped, unthreaded implantable Nickel Titanium (Nitinol) K-Wire consisting of two regions: a longer, superelastic end and a shorter, malleable end to allow fixation of bone fragments while being able to deform the protruding end. The Arthrex DynaNite K-Wire family ranges from 0.86 mm to 2.5 mm in diameter and is 5.91 inches in length. The Arthrex DynaNite K-Wire is sold sterile and is single-use.

AI/ML Overview

This document, K200068, is a 510(k) premarket notification for the Arthrex DynaNite K-Wire. It's important to understand that a 510(k) submission generally focuses on demonstrating substantial equivalence to a predicate device, rather than providing the detailed clinical trial data and acceptance criteria typically associated with de novo classifications or PMA applications for novel devices or high-risk devices.

Therefore, the document describes engineering and biocompatibility testing to demonstrate performance and safety, not an AI model that requires a "study that proves the device meets acceptance criteria" in terms of performance metrics like sensitivity, specificity, or human reader improvement.

Based on the provided text, here's what can be extracted and how it differs from a typical AI/software as a medical device (SaMD) study:

1. Table of Acceptance Criteria and Reported Device Performance:

The "acceptance criteria" here are based on meeting performance requirements comparable to the predicate device and established standards for metallic bone fixation fasteners, and ensuring biocompatibility.

Acceptance Criteria (Type of Test)Reported Device Performance (Summary)
Mechanical Performance
Tensile (ASTM F2516)Proposed Arthrex DynaNite K-Wire performs statistically equivalent to the predicate device (K052736).
Reverse Bend (ISO 7801)Proposed Arthrex DynaNite K-Wire performs statistically equivalent to the predicate device (K052736).
Cantilever Static Fatigue BendProposed Arthrex DynaNite K-Wire performs statistically equivalent to the predicate device (K052736).
Dynamic Fatigue BendProposed Arthrex DynaNite K-Wire performs statistically equivalent to the predicate device (K052736).
Single Bend and RecoveryTesting was conducted. (Specific performance not detailed, but implied to be acceptable based on overall conclusion.)
Transformation Temperature (ASTM F2082)Testing was conducted. (Specific performance not detailed, but implied to be acceptable based on overall conclusion.)
Biocompatibility
Cyclic Potentiodynamic Polarization Corrosion (ASTM F2129)Testing was conducted. (Specific performance not detailed, but implied to be acceptable based on overall conclusion.)
Bacterial Endotoxins Test (BET) (ANSI/AAMI ST72:2016, USP , USP , EP 2.6.14)Meets pyrogen limit.
Cytotoxicity (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Sensitization (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Irritation (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Genotoxicity (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Systemic Toxicity (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Subchronic/Subacute Toxicity (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Implantation (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
Material Characterization (ISO 10993-1:2018)Testing was conducted, and the overall assessment states the device does not introduce additional risks.
MRI Safety
MRI force (ASTM F2052)Testing conducted in accordance with FDA guidance. (Implied safe, specific values not given.)
MRI torque (ASTM F2213)Testing conducted in accordance with FDA guidance. (Implied safe, specific values not given.)
Image artifact (ASTM F2119)Testing conducted in accordance with FDA guidance. (Implied acceptable artifact profile, specific values not given.)
Radio Frequency Induced Heating (ASTM F2182)Testing conducted in accordance with FDA guidance. (Implied acceptable heating, specific values not given.)

2. Sample Size Used for the Test Set and Data Provenance:

  • For the mechanical tests related to performance (tensile, bend, fatigue), the sample sizes are not explicitly stated but would be the number of test articles (K-wires) subjected to each specific test. These are in-vitro (lab-based) tests, not human data.
  • For biocompatibility tests, sample sizes are implied to be sufficient for the specific biological tests (e.g., cell cultures for cytotoxicity, animal models for implantation/systemic toxicity, or specific chemical analyses). These are also in-vitro or in-vivo (animal model) tests, not human data.
  • Data Provenance: The data comes from the manufacturer's internal testing as part of their design verification and validation process, conducted according to recognized ASTM and ISO standards. It is prospective in the sense that the tests were specifically performed for this submission, using newly manufactured devices. It is not patient data from a specific country.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This section is not applicable to this type of device and submission. "Ground truth" in the context of an AI/SaMD diagnostic device typically refers to clinical diagnosis or pathology verified by multiple expert readers. For a medical device like a K-wire, "ground truth" is established by adherence to engineering standards, material science principles, and biocompatibility guidelines. The "experts" are the engineers, material scientists, and toxicologists who conducted and interpreted these tests.

4. Adjudication Method for the Test Set:

This is not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical trial data involving image interpretation or clinical outcomes where there might be inter-reader variability. For engineering and biocompatibility testing, success or failure is determined by whether the device meets predefined pass/fail criteria for each test according to the relevant standard.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This is not applicable. MRMC studies are used to evaluate the impact of an AI algorithm on human reader performance for diagnostic tasks. The Arthrex DynaNite K-Wire is a physical implantable device, not a diagnostic AI.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

This is not applicable. This concept applies to AI algorithms. The K-wire's "performance" is its physical and biological properties.

7. Type of Ground Truth Used:

As explained above, the "ground truth" is defined by:

  • Engineering standards: ASTM and ISO specifications for mechanical properties (e.g., required tensile strength, fatigue life).
  • Biocompatibility standards: ISO 10993 series and USP for biological safety (e.g., acceptable cytotoxicity limits, pyrogen limits).
  • Predicate device performance: The performance of the predicate device (K132895: WMT Implantable K-Wires) serves as a benchmark for demonstrating substantial equivalence for mechanical properties.

8. Sample Size for the Training Set:

This is not applicable. "Training set" refers to data used to train machine learning models. This device does not involve machine learning.

9. How the Ground Truth for the Training Set Was Established:

This is not applicable.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.