(255 days)
The indication for use of the BOUNDER Power Wheelchair is to provide mobility to persons limited to a sitting position.
The subject device is an update to the existing previously cleared BOUNDER Power Wheelchair (K901210). The BOUNDER utilizes components found on most power wheelchairs. These include a rigid or "nonfolding", rear wheel drive power wheelchair base with two drive wheels in the rear with two anti-tip wheels mounted behind and two casters in the front. It has two motors and is powered by two 12V DC Batteries (connected in series), and electronic controller. It also includes a seat, footrests, arms, positioning strap and charger. The design offers a modular approach and can be configured in multiple versions to suit different user needs.
The new models will include: BOUNDER 300, BOUNDER Plus 300, BOUNDER 450, BIG BOUNDER 600 & BIG BOUNDER 1000.
The updated subject version of the BOUNDER Power Wheelchair has the following changes:
- Increase in the weight capacity of the previously cleared device from 500 lb to 1000 lb
- Add Power Tilt, Power Anterior Tilt, Power Recline and Power Seat Elevator options
- Addition of LiNX and R-net electronics
- Add Independent Drive Wheel Suspension
- Add an Off-Road tire option for BOUNDER 300, BOUNDER Plus 300 and BOUNDER 450
- Top speed: 8.5 mph
The provided document is a 510(k) summary for the BOUNDER Power Wheelchair. It describes a medical device, specifically a power wheelchair, and the testing conducted to demonstrate its substantial equivalence to a predicate device. The document does not describe an AI/ML powered device, nor does it present acceptance criteria or performance data in a way that aligns with the requested format for such a device.
Therefore, I cannot extract the specific information requested about acceptance criteria and study design for an AI/ML device from this document. The sections you asked for, such as sample size for test and training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, and standalone performance, are not applicable or present in this regulatory submission for a physical medical device.
The document discusses "Design Verification and Validation Testing Data" where the BOUNDER Power Wheelchair was tested against various ANSI/RESNA WC standards. These standards cover aspects like static stability, dynamic stability, brake effectiveness, energy consumption, dimensions, mass, maneuvering space, maximum speed, acceleration, deceleration, climatic tests, obstacle-climbing ability, material properties, and electromagnetic compatibility. The acceptance criteria would be meeting the performance requirements outlined in each of these specific ANSI/RESNA WC standards. The study is essentially the comprehensive suite of tests conducted according to these standards.
Here's what I can provide based on the document, framed to indicate what's present and what is not applicable:
-
A table of acceptance criteria and the reported device performance
- Acceptance Criteria: Meeting the performance requirements as defined by the "Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard" (which specifically cites numerous ANSI/RESNA WC-1 and WC-2 standards, and ISO 7176-30:2018[E]). These are detailed engineering and performance specifications for power wheelchairs.
- Reported Device Performance: The document states, "Design verification and validation testing was performed on the subject device and was found to meet performance requirements. Further, it is substantially equivalent to the predicate device... The acceptance criteria for the full verification of the design and acceptance criteria for each section of the Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard have been met." Specific numerical performance values for each criterion are not provided in this summary.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable / Not Provided: This is a physical device, and testing would typically involve one or more physical units of the wheelchair model(s) undergoing various mechanical and electrical tests. The concept of "test set" or "data provenance" as applied to AI/ML data does not directly translate here. The testing was non-clinical.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable / Not Provided: This device relies on standardized performance testing, not expert interpretation or ground truth establishment in a medical imaging or diagnostic context.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable / Not Provided: This concept is for diagnostic interpretation and is not relevant to the physical performance testing of a wheelchair.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No: This is a physical medical device, not an AI-powered diagnostic tool. No MRMC study was performed.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable: This is a physical device, not an algorithm. Therefore, "standalone performance" in the context of an algorithm is not applicable. The device's performance was tested according to standardized methods.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Standardized Performance Requirements: The "ground truth" here is compliance with established engineering and safety standards (ANSI/RESNA WC, ISO 7176-30). The device must perform within the specified parameters of these standards.
-
The sample size for the training set
- Not Applicable / Not Provided: This is a physical device and does not involve AI/ML training sets.
-
How the ground truth for the training set was established
- Not Applicable / Not Provided: This device does not have a training set in the context of AI/ML.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
September 3, 2020
21st Century Scientific, Inc. RD Davidson COO 4931 N Manufacturing Way Coeur d'Alene, Idaho 83815
Re: K193599
Trade/Device Name: BOUNDER 300 Power Wheelchair, BOUNDER Plus 300 Power Wheelchair, BOUNDER 450 Power Wheelchair, BIG BOUNDER 600 Power Wheelchair, BIG BOUNDER 1000 Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 11, 2020 Received: August 12, 2020
Dear RD Davidson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather Dean, Ph.D. Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193599
Device Name
BOUNDER Series Power Wheelchair. Models: BOUNDER 300, BOUNDER 450, BIG BOUNDER 600, BIG BOUNDER 1000
Indications for Use (Describe)
The indication for use of the BOUNDER Power Wheelchair is to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
SECTION 5 510(k) SUMMARY
BOUNDER Power Wheelchair
SUBMITTER per 21 CFR 807.92(a)(1):
21st Century Scientific, Inc. 4931 N Manufacturing Way Coeur d'Alene, ID 83815 Phone: (208) 667-8800 Email: rdd(@wheelchairs.com
Contact Person
RD Davidson, COO
Manufacturer
21st Century Scientific, Inc. 4931 N Manufacturing Way Coeur d'Alene, ID 83815
FDA Registration number of Manufacturer
2027797
DATE PREPARED per 21 CFR 807.92(a)(1):
November 15, 2019
PRIOR SUBMISSIONS
The previous 510(k) for the BOUNDER series power wheelchair was K90210.
{4}------------------------------------------------
Section 5 ... : 510(k) Summary (continued)
DEVICE INFORMATION per 21 CFR 807.92 (a) (2)
Name of Device.................................: BOUNDER Series Power Wheelchair. Models: BOUNDER 300, BOUNDER Plus 300, BOUNDER 450, BIG BOUNDER 600, BIG BOUNDER 1000
| Common or Usual Name............: | Wheelchair, Powered |
|---|---|
| Classification Name................: | Powered Wheelchair 21 CFR § 890.386 |
| Product Code................: | ITI |
| Classification of New Device......: | Class II |
| Patient Population....................: | Physically Challenged Persons |
PREDICATE DEVICE per 21 CFR 807.92 (a) (3)
| Name of Device | : AMYPOWER ALLTRACK Series Power Wheelchair (K092225) |
|---|---|
| Common or Usual Name | : Wheelchair, Powered |
| Classification Name | : Powered Wheelchair 21 CFR § 890.3860 |
| Product Code | : ITI |
| Classifications of Predicate Device | : Class II |
| Patient Population | : Physically Challenged Persons |
DEVICE DESCRIPTION per 21 CFR 808.92(a)(4)
The subject device is an update to the existing previously cleared BOUNDER Power Wheelchair (K901210). The BOUNDER utilizes components found on most power wheelchairs. These include a rigid or "nonfolding", rear wheel drive power wheelchair base with two drive wheels in the rear with two anti-tip wheels mounted behind and two casters in the front. It has two motors and is powered by two 12V DC Batteries (connected in series), and electronic controller. It also includes a seat, footrests, arms, positioning strap and charger. The design offers a modular approach and can be configured in multiple versions to suit different user needs.
The new models will include: BOUNDER 300, BOUNDER Plus 300, BOUNDER 450, BIG BOUNDER 600 & BIG BOUNDER 1000.
The updated subject version of the BOUNDER Power Wheelchair has the following changes:
- Increase in the weight capacity of the previously cleared device from 500 lb to 1000 lb ●
- . Add Power Tilt, Power Anterior Tilt, Power Recline and Power Seat Elevator options
- Addition of LiNX and R-net electronics
- Add Independent Drive Wheel Suspension .
- Add an Off-Road tire option for BOUNDER 300, BOUNDER Plus 300 and BOUNDER 450 ●
- . Top speed: 8.5 mph
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|---|---|---|
| Page 2 of 4 | ||
| \SWORK2014\Drive D - Data\Data\QSM\Design History...\FDA 510K\Submission\Revised 510(k) Summary 2020-09-02.docx |
{5}------------------------------------------------
Section 5 ... : 510(k) Summary (continued)
INTENDED USE per 21 CFR 807.92(A)(5):
The intended use of the BOUNDER Power Wheelchair and the predicate device is to provide mobility to persons limited to a sitting position.
INDICATIONS FOR USE per FORM FDA 3881:
The indication for use of the BOUNDER Power Wheelchair is to provide mobility to persons limited to a sitting position.
COMPARISON OF DEVICE TECHNOLOGICAL CHARACTERISTICS TO PREDICATE DEVICE per CFR 807.92(A)(6):
The device comparison showed that the subject device is substantially equivalent in design, materials, and operational principles to the previously cleared predicate device in regard to providing mobility to persons limited to a sitting position. There is no difference in intended use and indications for use between the predicate device and the subject device.
This device has similar technological characteristics as the predicate devices utilize steel and aluminum in their frames and components, and standard foams and covers for seat cushions and backs. The user controls the chair by using a joystick or other equivalent command mode through a controller. It includes two motors and operates at 24 Volts using rechargeable batteries as a source of energy. The operating speeds, maneuverability, power modules, hand controls, seat types, and climbing ability are substantially equivalent. Both the BOUNDER Power Wheelchair and the predicate are recommended for indoor and outdoor use. The standard accessories and components are common to all power wheelchair devices.
BASIS OF SUBSTANTIAL EQUIVALENCE per 21 CFR 807.100(b)(2)(3):
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, and other device related features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A). The subject device components are as safe and effective as the predicate device and do not raise different questions of safety and effectiveness. The design verification and validation testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate device.
DESIGN VERIFICATION AND VALIDATION TESTING DATA
Design verification and validation testing was performed on the subject device and was found to meet performance requirements. Further, it is substantially equivalent to the predicate device identified throughout this submission and does not raise any new questions of safety and effectiveness. The acceptance criteria for the full verification of the design and acceptance criteria for each section of the Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard have been met.
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|---|---|---|
| Page 3 of 4 | ||
| \SWORK2014\Drive D - Data\Data\QSM\Design History\...\FDA 510K\Submission\Revised 510(k) Summary 2020-09-02.docx |
{6}------------------------------------------------
Section 5 ... : 510(k) Summary (continued)
Non-Clinical Test per 21 CFR 807.92(b)(1) Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing was performed to demonstrate that the BOUNDER Power Wheelchair meets the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise any new questions of safety and effectiveness.
The following testing was performed according to FDA recognized consensus standards:
- · ANSI/RESNA WC-1:2009 Section 1: Determination of Static Stability
- · ANSIRESNA WC-2:2009 Section 2: Determination of Dynamic Stability of electrically Powered Wheelchairs
- · ANSI/RESNA WC-2:2009 Section 3: Determination of Effectiveness of Brakes
- · ANSI/RESNA WC-2:2009 Section 4: Energy Consumption of Electrically Powered Wheelchairs and Scooters for Determination of Theoretical Distance Range
- · ANSI/RESNA WC-1:2009 Section 5: Determination of Dimensions, Mass and Maneuvering Space
- · ANSI/RESNA WC-2:2009 Section 6: Determination of Maximum Speed, Acceleration and Deceleration of Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of Seating and Wheel Dimensions
- · ANSI/RESNA WC-1:2009 Section 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
- · ANSI/RESNA WC-2:2009 Section 9: Climatic Tests for Electrically Powered Wheelchairs
- · ANSI/RESNA WC-2:2009 Section 10: Determination of Obstacle-Climbing Ability of Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 11: Test Dummies
- · ANSI/RESNA WC-1:2009 Section 13: Determination of Coefficient of Friction of Test Surfaces
- · ANSI/RESNA WC-2:2009 Section 14: Power and Control Systems for Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 15: Requirements for Information Disclosure, Documentation and Labeling
- · ANSI/RESNA WC-1:2009 Section 16: Resistance to Ignition of Upholstered Parts
- · ANSI/RESNA WC-2: 2009 Section 21: Requirements and Test Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Motorized Scooters
- · ANSI/RESNA WC-1:2009 Section 22: Set-up Procedures
- ISO 7176-30:2018[E] Part 30: Wheelchairs for Changing Occupant Posture
Verification testing demonstrated that the BOUNDER Power Wheelchair is substantially equivalent to the marketed predicate device.
Animal Study
Animal testing was not required for this submission.
Clinical Testing
Clinical testing was not required for this submission.
CONCLUSIONS per 21 CFR 807.92(b)(3)
The subject device has the same intended use and technological characteristics as the predicate device. The design verification and validation data support the safety and performance of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions. Therefore, the subject BOUNDER Power Wheelchair is comparable with the previously cleared AMYPOWER ALLTRACKS Series Power Wheelchair and demonstrates a basis of substantial equivalence.
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|---|---|---|
| Page 4 of 4 | ||
| \SWORK2014\Drive D - Data\Data\QSM\Design History...\FDA 510K\Submission\Revised 510(k) Summary 2020-09-02.docx |
N/A