K Number
K193573
Date Cleared
2020-04-10

(109 days)

Product Code
Regulation Number
878.4460
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Skinsense polyisoprene underglove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The Biogel PI Micro Surgical Glove is a disposable device made of polyisoprene material that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

A powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

A powder-free, sterile, surgeon's glove is a disposable device made of non-latex that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The Biogel® surgical gloves that are the subject of this submission are sterile, single-use, powder-free gloves that are constructed of either synthetic polyisoprene or synthetic polychloroprene.

AI/ML Overview

The provided documents describe the acceptance criteria and the results of a study conducted to demonstrate that several Biogel® gloves meet these criteria for chemotherapy drug permeation.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criterion for each chemotherapy drug tested is implied by the testing method as a "Breakthrough detection time" in minutes (0.01 µg/cm²/mins). The general expectation for protective gloves against hazardous substances is a breakthrough time as long as possible, ideally exceeding the typical operational exposure duration. The standard testing time of 240 minutes (4 hours) appears to be a benchmark for "safe" permeation resistance.

The tables below synthesize the acceptance criteria (testing was performed according to ASTM D6978-05 (2019)) and the reported performance for each specific glove type and chemotherapy drug.


Biogel® Skinsense® Indicator® Underglove tested for use with chemotherapy agents

Drug and ConcentrationAcceptance Criteria (Breakthrough detection time in minutes (0.01µg/cm²/mins))Reported Device Performance (Breakthrough detection time in minutes)
Bleomycin 15 mg/ml>240>240
Busulfan 6 mg/ml>240>240
Carmustine 3.3 mg/mlAs long as possible, specific value reported6.6
Cisplatin 1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240>240
Cytarabine 100 mg/ml>240>240
Dacarbazine (DTIC) 10 mg/ml>240>240
Doxorubicin Hydrochloride 2 mg/ml>240>240
Ellence 2 mg/ml>240>240
Etoposide (Toposar) 20 mg/ml>240>240
Fludarabine 25 mg/ml>240>240
Fluorouracil 50 mg/ml>240>240
Idarubicin 1 mg/ml>240>240
Ifosfamide 50 mg/ml>240>240
Mechlorethamine HCl 1 mg/ml>240>240
Melphalan 5 mg/ml>240>240
Methotrexate 25 mg/ml>240>240
Mitomycin C 0.5 mg/ml>240>240
Mitoxantrone 2 mg/ml>240>240
Paclitaxel (Taxol) 6 mg/ml>240>240
Paraplatin 10 mg/ml>240>240
Rituximab 10 mg/ml>240>240
Thiotepa 10 mg/mlAs long as possible, specific value reported16.9
Vincristine Sulfate 1 mg/ml>240>240

Biogel® PI Micro tested for use with chemotherapy agents

Drug and ConcentrationAcceptance Criteria (Breakthrough detection time in minutes (0.01µg/cm²/mins))Reported Device Performance (Breakthrough detection time in minutes)
Bleomycin 15 mg/ml>240>240
Busulfan 6 mg/ml>240>240
Carmustine 3.3 mg/mlAs long as possible, specific value reported10.0
Cisplatin 1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240>240
Cytarabine 100 mg/ml>240>240
Dacarbazine (DTIC) 10 mg/ml>240>240
Doxorubicin Hydrochloride 2 mg/ml>240>240
Ellence 2 mg/ml>240>240
Etoposide (Toposar) 20 mg/ml>240>240
Fludarabine 25 mg/ml>240>240
Fluorouracil 50 mg/ml>240>240
Idarubicin 1 mg/ml>240>240
Ifosfamide 50 mg/ml>240>240
Mechlorethamine HCl 1 mg/ml>240>240
Melphalan 5 mg/ml>240>240
Methotrexate 25 mg/ml>240>240
Mitomycin C 0.5 mg/ml>240>240
Mitoxantrone 2 mg/ml>240>240
Paclitaxel (Taxol) 6 mg/ml>240>240
Paraplatin 10 mg/ml>240>240
Rituximab 10 mg/ml>240>240
Thiotepa 10 mg/mlAs long as possible, specific value reported20.3
Vincristine Sulfate 1 mg/ml>240>240

Biogel® PI UltraTouch® tested for use with chemotherapy agents

Drug and ConcentrationAcceptance Criteria (Breakthrough detection time in minutes (0.01µg/cm²/mins))Reported Device Performance (Breakthrough detection time in minutes)
Bleomycin 15 mg/ml>240>240
Busulfan 6 mg/ml>240>240
Carmustine 3.3 mg/mlAs long as possible, specific value reported24.2
Cisplatin 1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240>240
Cytarabine 100 mg/ml>240>240
Dacarbazine (DTIC) 10 mg/ml>240>240
Doxorubicin Hydrochloride 2 mg/ml>240>240
Ellence 2 mg/ml>240>240
Etoposide (Toposar) 20 mg/ml>240>240
Fludarabine 25 mg/ml>240>240
Fluorouracil 50 mg/ml>240>240
Idarubicin 1 mg/ml>240>240
Ifosfamide 50 mg/ml>240>240
Mechlorethamine HCl 1 mg/ml>240>240
Melphalan 5 mg/ml>240>240
Methotrexate 25 mg/ml>240>240
Mitomycin C 0.5 mg/ml>240>240
Mitoxantrone 2 mg/ml>240>240
Paclitaxel (Taxol) 6 mg/ml>240>240
Paraplatin 10 mg/ml>240>240
Rituximab 10 mg/ml>240>240
Thiotepa 10 mg/mlAs long as possible, specific value reported17.9
Vincristine Sulfate 1 mg/ml>240>240

Biogel® PI tested for use with chemotherapy agents

Drug and ConcentrationAcceptance Criteria (Breakthrough detection time in minutes (0.01µg/cm²/mins))Reported Device Performance (Breakthrough detection time in minutes)
Bleomycin 15 mg/ml>240>240
Busulfan 6 mg/ml>240>240
Carmustine 3.3 mg/mlAs long as possible, specific value reported26.7
Cisplatin 1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240>240
Cytarabine 100 mg/ml>240>240
Dacarbazine (DTIC) 10 mg/ml>240>240
Doxorubicin Hydrochloride 2 mg/ml>240>240
Ellence 2 mg/ml>240>240
Etoposide (Toposar) 20 mg/ml>240>240
Fludarabine 25 mg/ml>240>240
Fluorouracil 50 mg/ml>240>240
Idarubicin 1 mg/ml>240>240
Ifosfamide 50 mg/ml>240>240
Mechlorethamine HCl 1 mg/ml>240>240
Melphalan 5 mg/ml>240>240
Methotrexate 25 mg/ml>240>240
Mitomycin C 0.5 mg/ml>240>240
Mitoxantrone 2 mg/ml>240>240
Paclitaxel (Taxol) 6 mg/ml>240>240
Paraplatin 10 mg/ml>240>240
Rituximab 10 mg/ml>240>240
Thiotepa 10 mg/mlAs long as possible, specific value reported28.7
Vincristine Sulfate 1 mg/ml>240>240

Biogel® PI Indicator® Underglove tested for use with chemotherapy agents

Drug and ConcentrationAcceptance Criteria (Breakthrough detection time in minutes (0.01µg/cm²/mins))Reported Device Performance (Breakthrough detection time in minutes)
Bleomycin 15 mg/ml>240>240
Busulfan 6 mg/ml>240>240
Carmustine 3.3 mg/mlAs long as possible, specific value reported17.3
Cisplatin 1.0 mg/ml>240>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240>240
Cytarabine 100 mg/ml>240>240
Dacarbazine (DTIC) 10 mg/ml>240>240
Doxorubicin Hydrochloride 2 mg/ml>240>240
Ellence 2 mg/ml>240>240
Etoposide (Toposar) 20 mg/ml>240>240
Fludarabine 25 mg/ml>240>240
Fluorouracil 50 mg/ml>240>240
Idarubicin 1 mg/ml>240>240
Ifosfamide 50 mg/ml>240>240
Mechlorethamine HCl 1 mg/ml>240>240
Melphalan 5 mg/ml>240>240
Methotrexate 25 mg/ml>240>240
Mitomycin C 0.5 mg/ml>240>240
Mitoxantrone 2 mg/ml>240>240
Paclitaxel (Taxol) 6 mg/ml>240>240
Paraplatin 10 mg/ml>240>240
Rituximab 10 mg/ml>240>240
Thiotepa 10 mg/mlAs long as possible, specific value reported24.1
Vincristine Sulfate 1 mg/ml>240>240

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the number of individual gloves or test replicates used for each chemotherapy drug. However, per ASTM D6978, testing generally involves multiple specimens (e.g., 3 specimens per drug as per standard recommendations for breakthrough time).
  • Data Provenance: The study was a non-clinical test conducted according to ASTM D6978-05 (2019): Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. The country of origin for the data is not specified, but the use of an international standard like ASTM suggests a rigorous and replicable testing methodology. The data is prospective in the sense that the testing was performed specifically to support this FDA submission for the "tested for use with chemotherapy agents" claim.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the study did not involve human experts establishing ground truth in a medical context. The ground truth (breakthrough time) was established through objective measurement in a laboratory setting following a standardized test method (ASTM D6978-05 (2019)).

4. Adjudication method for the test set:

This information is not applicable as the study did not involve human adjudication. The results were obtained through standardized laboratory measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The submission describes a non-clinical in vitro permeation test for medical gloves, not an AI-assisted diagnostic or clinical device. There are no human readers or AI involved in this type of testing.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The submission describes a non-clinical in vitro permeation test for medical gloves. It is not an algorithm or software device.

7. The type of ground truth used:

The ground truth used was objective laboratory measurement of the "Breakthrough detection time" of chemotherapy drugs through the glove material, as defined by the ASTM D6978-05 (2019) standard. This is a direct physical measurement, not based on expert consensus, pathology, or outcomes data in a clinical sense.

8. The sample size for the training set:

This information is not applicable. There is no training set for this type of non-clinical device testing, as it does not involve machine learning or AI.

9. How the ground truth for the training set was established:

This information is not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 10, 2020

Molnlycke Health Care US LLC. Calen Souther Regulatory Affairs Specialist 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092

Re: K193573

Trade/Device Name: Biogel® Skinsense® Indicator® Underglove tested for use with chemotherapy agents, Biogel® PI UltraTouch® tested for use with chemotherapy agents, Biogel® PI Indicator® Underglove tested for use with chemotherapy agents, Biogel® PI tested for use with chemotherapy agents, Biogel® PI Micro tested for use with chemotherapy agents

Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, LZC Dated: January 16, 2020 Received: January 21, 2020

Dear Calen Souther:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193573

Device Name

Biogel PI Indicator Underglove tested for use with chemotherapy agents

Indications for Use (Describe)

The Skinsense polyisoprene underglove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

[continued on next page]

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

[K193573 Indications for Use, continued]

Biogel® PI Indicator® UndergloveTested for use with chemotherapy agents
Drug and ConcentrationBreakthrough detection time inminutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml17.3
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml24.1
Vincristine Sulfate 1 mg/ml>240
  • י Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 17.3 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 24.1 minutes

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K193573

Device Name

Biogel PI Micro tested for use with chemotherapy agents

Indications for Use (Describe)

The Biogel PI Micro Surgical Glove is a disposable device made of polyisoprene material that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

[continued on next page]

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

[K193573 Indications for Use, continued]

Biogel® PI MicroTested for use with chemotherapy agents
Drug and ConcentrationBreakthrough detection time inminutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml10.0
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml20.3
Vincristine Sulfate 1 mg/ml>240
  • י Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 10.0 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 20.3 minutes

{6}------------------------------------------------

Indications for Use

510(k) Number (if known) K193573

Device Name

Biogel PI UltraTouch tested for use with chemotherapy agents

Indications for Use (Describe)

A powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

[continued on next page]

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{7}------------------------------------------------

[K193573 Indications for Use, continued]

Biogel® PI UltraTouch
Tested for use with chemotherapy agents
Drug and ConcentrationBreakthrough detection time in minutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml24.2
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride 2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml17.9
Vincristine Sulfate 1 mg/ml>240
  • י Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 24.2 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 17.9 minutes

{8}------------------------------------------------

Indications for Use

510(k) Number (if known) K193573

Device Name

Biogel PI tested for use with chemotherapy agents

Indications for Use (Describe)

A powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

[continued on next page]

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{9}------------------------------------------------

[K193573 Indications for Use, continued]

Drug and ConcentrationBreakthrough detection time inminutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml26.7
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml28.7
Vincristine Sulfate 1 mg/ml>240
  • י Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 26.7 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 28.7 minutes

{10}------------------------------------------------

Indications for Use

510(k) Number (if known) K193573

Device Name

Biogel Skinsense Indicator Underglove tested for use with chemotherapy agents

Indications for Use (Describe)

A powder-free, sterile, surgeon's glove is a disposable device made of non-latex that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

[continued on next page]

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{11}------------------------------------------------

[K193573 Indications for Use, continued]

Drug and ConcentrationBreakthrough detection time in
minutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml6.6
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride 2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml16.9
Vincristine Sulfate 1 mg/ml>240
  • י Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 6.6 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 16.9 minutes

{12}------------------------------------------------

510(k) Summary

The information contained herein is being provided in accordance with the requirements of 21 CFR 807.92(c).

Date Prepared:March 23, 2020
Applicant:Mölnlycke Health Care US, LLC.5550 Peachtree Parkway, Suite 500Norcross, GA 30092Registration Number: 3004763499Owner/Operator Number: 8030877
Official Correspondent:Calen SoutherRegulatory Affairs SpecialistPhone: 770-595-4222Fax: 678-245-7746Email: calen.souther@molnlycke.com
Trade/Proprietary Names:Biogel® Skinsense® Indicator® Underglove tested for use withchemotherapy agentsBiogel® PI UltraTouch® tested for use with chemotherapyagentsBiogel® PI Indicator® Underglove tested for use withchemotherapy agentsBiogel® PI tested for use with chemotherapy agentsBiogel® PI Micro tested for use with chemotherapy agents
Common Name:Surgeon's Gloves
Regulation Name:Non-powdered surgeon's gloves
Device Class:Class I
Regulation Number:21 CFR 878.4460
Product Code:KGO, LZC
Predicate Device Information:K140477
Biogel® PI UltraTouch® G Surgical Glove tested for use withchemotherapy agents
Biogel® Skinsense® Surgical Glove tested for use withchemotherapy agents

{13}------------------------------------------------

Reason for 510(k) submission

The purpose of this Traditional 510(k) submission is to add a "Tested for use with chemotherapy agents" claim to the following existing, 510(k) cleared surgical gloves: Biogel® Skinsense® Underglove, Biogel® Pl UltraTouch®, Biogel® PI Indicator® Underglove, Biogel® PI, Biogel® PI Micro.

Description of Devices

The Biogel® surgical gloves that are the subject of this submission are sterile, single-use, powder-free gloves that are constructed of either synthetic polyisoprene or synthetic polychloroprene. Refer to Tables 6-1 to 6-5 for a detailed description of the technological characteristics and comparison to the applicable predicate device.

Indications for Use

Biogel® Skinsense® Indicator® Underglove tested for use with chemotherapy agents

A powder-free, sterile, surgeon's glove is a disposable device made of non-latex that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

Drug and ConcentrationBreakthroughdetection time inminutes(0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml6.6
Cisplatin 1.0 mg/ml>240
Cyclophosphamide(Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10mg/ml>240
DoxorubicinHydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240

{14}------------------------------------------------

Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml16.9
Vincristine Sulfate 1 mg/ml>240
  • Carmustine (3.3 mg/ml) has a minimum breakthrough time of 6.6 minutes.
  • . Thiotepa (10 mg/ml) has a minimum breakthrough time of 16.9 minutes

Biogel® PI Micro tested for use with chemotherapy agents

The Biogel Pl Micro Surgical Glove is a disposable device made of polyisoprene material that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

Drug and ConcentrationBreakthroughdetection time inminutes(0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml10.0
Cisplatin 1.0 mg/ml>240
Cyclophosphamide(Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10mg/ml>240
DoxorubicinHydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240

{15}------------------------------------------------

Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml20.3
Vincristine Sulfate 1 mg/ml>240
  • Carmustine (3.3 mg/ml) has a minimum breakthrough time of 10.0 minutes.
  • . Thiotepa (10 mg/ml) has a minimum breakthrough time of 20.3 minutes

Biogel® PI UltraTouch® tested for use with chemotherapy agents

A powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

Drug and ConcentrationBreakthrough detection time in minutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml24.2
Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride 2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240

{16}------------------------------------------------

Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml17.9
Vincristine Sulfate 1 mg/ml>240
  • Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 24.2 minutes.
  • . Thiotepa (10 mg/ml) has a minimum breakthrough time of 17.9 minutes

Biogel® PI tested for use with chemotherapy agents

A powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

Drug and ConcentrationBreakthroughdetection time inminutes(0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml26.7
Cisplatin 1.0 mg/ml>240
Cyclophosphamide(Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10mg/ml>240
DoxorubicinHydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240

{17}------------------------------------------------

Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml28.7
Vincristine Sulfate 1mg/ml>240
  • Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 26.7 minutes.
  • . Thiotepa (10 mg/ml) has a minimum breakthrough time of 28.7 minutes

Biogel® PI Indicator® Underglove tested for use with chemotherapy agents

The Skinsense polyisoprene underglove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.

Drug and ConcentrationBreakthroughdetection time inminutes(0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml17.3
Cisplatin 1.0 mg/ml>240
Cyclophosphamide(Cytoxan)20 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10mg/ml>240
DoxorubicinHydrochloride2 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20mg/ml>240
Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240

{18}------------------------------------------------

Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml24.1
Vincristine Sulfate 1 mg/ml>240
  • . Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 17.3 minutes.
  • Thiotepa (10 mg/ml) has a minimum breakthrough time of 24.1 minutes ■

{19}------------------------------------------------

Table 6-1: Technological Characteristics Comparison
-----------------------------------------------------
Subject DevicePredicate Device
FeatureBiogel® Skinsense® Indicator® Underglove tested for use with chemotherapy agentsBiogel® Skinsense® Surgical Glove tested for use with chemotherapy agents
510(k) clearanceOriginally cleared under K053102; subject of this premarket notification to add "tested for use with chemotherapy agents" claimK140477
ManufacturerMölnlycke Health CareMölnlycke Health Care
Regulation21 CFR 878.4460,21 CFR 878.4460,
Class NameSurgeon's GlovesSurgeon's Gloves
ClassificationClass IClass I
Product CodeKGO, LZCKGO, LZC
Indication for UseA powder-free, sterile, surgeon's glove is a disposable device made of non-latex that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:Biogel® Skinsense® Surgical Gloves are intended to be worn on the hands, usually in surgical settings, to provide barrier against potentially infectious material and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Drug and ConcentrationBreakthrough detection time in minutes (0.01µg/cm²/mins)Bleomycin 15 mg/ml >240Busulfan 6 mg/ml >240Carmustine 3.3 mg/ml 6.6Cisplatin 1.0 mg/ml >240Cyclophosphamide (Cytoxan) 20 mg/ml >240Cytarabine 100 mg/ml >240Drug and ConcentrationBreakthrough detection time in minutes (0.01µg/cm²/mins)Bleomycin 15 mg/ml >240Busulfan 6 mg/ml >240Carmustine 3.3 mg/ml 60.2Cisplatin 1.0 mg/ml >240Cyclophosphamide (Cytoxan) 20 mgl/ml >240Cytarabine 100 mg/ml >240Dacarbazine (DTIC) 10 mg/ml >240
Dacarbazine (DTIC) 10 mg/ml>240Doxorubicin Hydrochloride 2 mg/ml>240
Doxorubicin Hydrochloride 2 mg/ml>240Ellence 2 mg/ml>240
Ellence 2 mg/ml>240Etoposide (Toposar) 20 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240Fludarabine 25 mg/ml>240
Fludarabine 25 mg/ml>240Fluorouracil 50 mg/ml>240
Fluorouracil 50 mg/ml>240Idarubicin 1 mg/ml>240
Idarubicin 1 mg/ml>240Ifosfamide 50 mg/ml>240
Ifosfamide 50 mg/ml>240Mechlorethamine HCl 1 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240Melphalan 5 mg/ml>240
Melphalan 5 mg/ml>240Methotrexate 25 mg/ml>240
Methotrexate 25 mg/ml>240Mitomycin C 0.5 mg/ml>240
Mitomycin C 0.5 mg/ml>240Mitoxantrone 2 mg/ml>240
Mitoxantrone 2 mg/ml>240Paclitaxel (Taxol) 6 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240Paraplatin 10 mg/ml>240
Paraplatin 10 mg/ml>240Rituximab 10 mg/ml>240
Rituximab 10 mg/ml>240Thiotepa 10 mg/ml75.8
Thiotepa 10 mg/ml16.9Trisenox 0.1 mg/ml>240
Vincristine Sulfate 1 mg/ml>240Vincristine Sulfate 1 mg/ml>240
Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 6.6 minutes. Thiotepa (10 mg/ml) has a minimum breakthrough time of 16.9 minutesPlease note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 60.2 minutes. Thiotepa (10 mg/ml) has a minimum breakthrough time of 75.8 minutes
DesignSingle-useSterilePowder-freeHand specificBeaded cuffSingle-useSterilePowder-freeHand specificBeaded cuff
MaterialSynthetic PolychloropreneSynthetic Polychloroprene
SterilizationMethodIrradiationIrradiation
Sterility (SAL)10-610-6
ColorBlueStraw (Natural)
Shelf Life3-years3-years
Resistance toPermeation byChemotherapyDrugsMeets ASTM D6978-05 (2019)Meets ASTM D6978-05 (2019)
Freedom fromHolesAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirementsAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015) requirements
Powder ResidueMeets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)Meets requirements of ≤ 2.0 mg/glove for Powder-free designationper ASTM D3577-09 (2015)
Dimensions andPhysicalPropertiesMeets ASTM D3577-09 (2015)Meets ASTM D3577-09 (2015)

{20}------------------------------------------------

{21}------------------------------------------------

{22}------------------------------------------------

Table 6-2: Technological Characteristics Comparison

FeatureSubject DeviceBiogel® PI Micro tested for use with chemotherapy agentsPredicate DeviceBiogel® PI UltraTouch® G Surgical Glove tested for use with chemotherapy agents
510(k) clearanceOriginally cleared under K141719; subject of this premarket notification to add "tested for use with chemotherapy agents" claimK140477
ManufacturerMölnlycke Health CareMölnlycke Health Care
Regulation21 CFR 878.4460, 21 CFR 878.625021 CFR 878.4460, 21 CFR 878.6250
Class NameSurgeon's GlovesSurgeon's Gloves
ClassificationClass IClass I
Product CodeKGO, LZCKGO, LZC
Indication for UseThe Biogel PI Micro Surgical Glove is a disposable device made of polyisoprene material that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:Biogel® PI UltraTouch® G Surgical Gloves are intended to be worn on the hands, usually in surgical settings, to provide barrier against potentially infectious material and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Drug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 10.0 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mg/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240 Doxorubicin Hydrochloride 2 mg/ml >240Drug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 12.1 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mgl/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240 Doxorubicin Hydrochloride 2 mg/ml >240

{23}------------------------------------------------

Ellence 2 mg/ml>240Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml20.3Thiotepa 10 mg/ml15.5
Vincristine Sulfate 1 mg/ml>240Trisenox 0.1 mg/ml>240
Vincristine Sulfate 1 mg/ml>240
Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:
• Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 10.0 minutes.• Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 12.1 minutes.
• Thiotepa (10 mg/ml) has a minimum breakthrough time of 20.3 minutes• Thiotepa (10 mg/ml) has a minimum breakthrough time of 15.5 minutes
DesignSingle-useSterilePowder-freeHand specificBeaded cuffDesignSingle-useSterilePowder-freeHand specificBeaded cuff
MaterialSynthetic PolyisopreneMaterialSynthetic Polyisoprene
Sterilization MethodIrradiationSterilization MethodIrradiation
Sterility (SAL)10-6Sterility (SAL)10-6

{24}------------------------------------------------

ColorStraw (Natural)Straw (Natural)
Shelf Life3-years3-years
Resistance toPermeation byChemotherapyDrugsMeets ASTM D6978-05 (2019)Meets ASTM D6978-05
Freedom fromHolesAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirementsAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirements
Powder ResidueMeets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)Meets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)
Dimensions andPhysicalPropertiesMeets ASTM D3577-09 (2015)Meets ASTM D3577-09 (2015)

{25}------------------------------------------------

Table 6-3: Technological Characteristics Comparison

FeatureSubject DevicePredicate
510(k) clearanceBiogel® PI UltraTouch® tested for use with chemotherapy agentsPreviously cleared under K050184; subject of this premarket notification to add "tested for use with chemotherapy agents" claimBiogel® PI UltraTouch® G Surgical Glove tested for use with chemotherapy agentsK140477
ManufacturerMölnlycke Health CareMölnlycke Health Care
Regulation21 CFR 878.4460, 21 CFR 878.625021 CFR 878.4460, 21 CFR 878.6250
Class NameSurgeon's GlovesSurgeon's Gloves
ClassificationClass IClass I
Product CodeKGO, LZCKGO, LZC
Indication for UseA powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:Biogel® PI UltraTouch® G Surgical Gloves are intended to be worn on the hands, usually in surgical settings, to provide barrier against potentially infectious material and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Drug and ConcentrationDrug and Concentration
Breakthrough detection time in minutes (0.01µg/cm²/mins)Breakthrough detection time in minutes (0.01µg/cm²/mins)
Bleomycin 15 mg/ml>240Bleomycin 15 mg/ml>240
Busulfan 6 mg/ml>240Busulfan 6 mg/ml>240
Carmustine 3.3 mg/ml24.2Carmustine 3.3 mg/ml12.1
Cisplatin 1.0 mg/ml>240Cisplatin 1.0 mg/ml>240
Cyclophosphamide (Cytoxan) 20 mg/ml>240Cyclophosphamide (Cytoxan) 20 mgl/ml>240
Cytarabine 100 mg/ml>240Cytarabine 100 mg/ml>240
Dacarbazine (DTIC) 10 mg/ml>240Dacarbazine (DTIC) 10 mg/ml>240
Doxorubicin Hydrochloride 2 mg/ml>240Doxorubicin Hydrochloride 2 mg/ml>240
Ellence 2 mg/ml>240Ellence 2 mg/ml>240
Etoposide (Toposar) 20 mg/ml>240Etoposide (Toposar) 20 mg/ml>240
Fludarabine 25 mg/ml>240Fludarabine 25 mg/ml>240
Fluorouracil 50 mg/ml>240Fluorouracil 50 mg/ml>240
Idarubicin 1 mg/ml>240Idarubicin 1 mg/ml>240
Ifosfamide 50 mg/ml>240Ifosfamide 50 mg/ml>240
Mechlorethamine HCl 1 mg/ml>240Mechlorethamine HCl 1 mg/ml>240
Melphalan 5 mg/ml>240Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240Methotrexate 25 mg/ml>240
Mitomycin C 0.5 mg/ml>240Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2 mg/ml>240Mitoxantrone 2 mg/ml>240
Paclitaxel (Taxol) 6 mg/ml>240Paclitaxel (Taxol) 6 mg/ml>240
Paraplatin 10 mg/ml>240Paraplatin 10 mg/ml>240
Rituximab 10 mg/ml>240Rituximab 10 mg/ml>240
Thiotepa 10 mg/ml17.9Thiotepa 10 mg/ml15.5
Vincristine Sulfate 1 mg/ml>240Trisenox 0.1 mg/ml>240
Vincristine Sulfate 1 mg/ml>240
Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:- Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 24.2 minutes.- Thiotepa (10 mg/ml) has a minimum breakthrough time of 17.9 minutesPlease note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:- Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 12.1 minutes.- Thiotepa (10 mg/ml) has a minimum breakthrough time of 15.5 minutes
DesignSingle-useSterilePowder-freeHand specificBeaded cuffSingle-useSterilePowder-freeHand specificBeaded cuff
MaterialSynthetic PolyisopreneSynthetic Polyisoprene
Sterilization MethodIrradiationIrradiation
Sterility (SAL)10-610-6
ColorStraw (Natural)Straw (Natural)
Shelf Life3-years3-years
Resistance toPermeation byChemotherapyDrugsMeets ASTM D6978-05 (2019)Meets ASTM D6978-05
Freedom fromHolesAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirementsAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirements
Powder ResidueMeets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)Meets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)
Dimensions andPhysicalPropertiesMeets ASTM D3577-09 (2015)Meets ASTM D3577-09 (2015)

{26}------------------------------------------------

{27}------------------------------------------------

{28}------------------------------------------------

Table 6-4: Technological Characteristics Comparison

FeatureSubject DevicePredicate Device
Biogel® Pl tested for use with chemotherapy agentsBiogel® PI UltraTouch® G Surgical Glove tested for use with chemotherapy agents
510(k) clearancePreviously cleared under K050184/K053442; subject of this premarket notification to add "tested for use with chemotherapy agents" claimK140477
ManufacturerMölnlycke Health CareMölnlycke Health Care
Regulation21 CFR 878.4460, 21 CFR 878.625021 CFR 878.4460, 21 CFR 878.6250
Class NameSurgeon's GlovesSurgeon's Gloves
ClassificationClass IClass I
Product CodeKGO, LZCKGO, LZC
Indication for UseA powder-free sterile surgeon's glove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:Biogel® PI UltraTouch® G Surgical Gloves are intended to be worn on the hands, usually in surgical settings, to provide barrier against potentially infectious material and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Drug and ConcentrationDrug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 26.7 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mg/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240Drug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 12.1 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mgl/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240 Doxorubicin Hydrochloride 2 mg/ml >240

{29}------------------------------------------------

DoxorubicinHydrochloride2 mg/ml>240Ellence 2 mg/mlEtoposide (Toposar) 20 mg/ml>240
Ellence 2 mg/ml>240Fludarabine 25 mg/ml>240
Etoposide (Toposar) 20mg/ml>240Fluorouracil 50 mg/ml>240
Fludarabine 25 mg/ml>240Idarubicin 1 mg/ml>240
Fluorouracil 50 mg/ml>240Ifosfamide 50 mg/ml>240
Idarubicin 1 mg/ml>240Mechlorethamine HCl 1mg/ml>240
Ifosfamide 50 mg/ml>240Melphalan 5 mg/ml>240
Mechlorethamine HCl 1mg/ml>240Methotrexate 25 mg/ml>240
Melphalan 5 mg/ml>240Mitomycin C 0.5 mg/ml>240
Methotrexate 25 mg/ml>240Mitoxantrone 2 mg/ml>240
Mitomycin C 0.5 mg/ml>240Paclitaxel (Taxol) 6mg/ml>240
Mitoxantrone 2 mg/ml>240Paraplatin 10 mg/ml>240
Paclitaxel (Taxol) 6mg/ml>240Rituximab 10 mg/ml>240
Paraplatin 10 mg/ml>240Thiotepa 10 mg/ml15.5
Rituximab 10 mg/ml>240Trisenox 0.1 mg/ml>240
Thiotepa 10 mg/ml28.7Vincristine Sulfate 1mg/ml>240
Vincristine Sulfate 1mg/ml>240
Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:
Please note that Carmustine (3.3 mg/ml) and Thiotepa (10 mg/ml) have much lower permeation times compared to other chemotherapy drugs:
Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 26.7 minutes.Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 12.1 minutes.
Thiotepa (10 mg/ml) has a minimum breakthrough time of 28.7 minutesThiotepa (10 mg/ml) has a minimum breakthrough time of 15.5 minutes
DesignSingle-useSterilePowder-freeHand specificBeaded cuffSingle-useSterilePowder-freeHand specificBeaded cuff
MaterialSynthetic PolyisopreneSynthetic Polyisoprene
SterilizationMethodIrradiationIrradiation

{30}------------------------------------------------

Sterility (SAL)$10^{-6}$$10^{-6}$
ColorStraw (Natural)Straw (Natural)
Shelf Life3-years3-years
Resistance toPermeation byChemotherapyDrugsMeets ASTM D6978-05 (2019)Meets ASTM D6978-05
Freedom fromHolesAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirementsAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015)requirements
Powder ResidueMeets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)Meets requirements of ≤ 2.0 mg/glove for Powder-freedesignation per ASTM D3577-09 (2015)
Dimensions andPhysicalPropertiesMeets ASTM D3577-09 (2015)Meets ASTM D3577-09 (2015)

{31}------------------------------------------------

Table 6-5: Technological Characteristics Comparison

Subject DevicePredicate Device
FeatureBiogel® PI Indicator® Underglove tested for use with chemotherapy agentsBiogel® PI UltraTouch® G Surgical Glove tested for use with chemotherapy agents
510(k) clearancePreviously cleared under K081180; subject of this premarket notification to add "tested for use with chemotherapy agents" claimK140477
ManufacturerMölnlycke Health CareMölnlycke Health Care
Regulation21 CFR 878.4460, 21 CFR 878.625021 CFR 878.4460, 21 CFR 878.6250
Class NameSurgeon's GlovesSurgeon's Gloves
ClassificationClass IClass I
Product CodeKGO, LZCKGO, LZC
Indication for UseThe Skinsense polyisoprene underglove is a disposable device made of polyisoprene that is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:Biogel® PI UltraTouch® G Surgical Gloves are intended to be worn on the hands, usually in surgical settings, to provide barrier against potentially infectious material and other contaminants.In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Drug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 17.3 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mg/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240Drug and Concentration Breakthrough detection time in minutes (0.01µg/cm²/mins) Bleomycin 15 mg/ml >240 Busulfan 6 mg/ml >240 Carmustine 3.3 mg/ml 12.1 Cisplatin 1.0 mg/ml >240 Cyclophosphamide (Cytoxan) 20 mgl/ml >240 Cytarabine 100 mg/ml >240 Dacarbazine (DTIC) 10 mg/ml >240 Doxorubicin Hydrochloride 2 mg/ml >240

{32}------------------------------------------------

DoxorubicinHydrochloride2 mg/ml>240Ellence 2 mg/ml>240
Ellence 2 mg/ml>240Etoposide (Toposar) 20 mg/ml>240
Etoposide (Toposar) 20mg/ml>240Fludarabine 25 mg/ml>240
Fludarabine 25 mg/ml>240Fluorouracil 50 mg/ml>240
Fluorouracil 50 mg/ml>240Idarubicin 1 mg/ml>240
Idarubicin 1 mg/ml>240Ifosfamide 50 mg/ml>240
Ifosfamide 50 mg/ml>240Mechlorethamine HCl 1mg/ml>240
Mechlorethamine HCl 1mg/ml>240Melphalan 5 mg/ml>240
Melphalan 5 mg/ml>240Methotrexate 25 mg/ml>240
Methotrexate 25 mg/ml>240Mitomycin C 0.5 mg/ml>240
Mitomycin C 0.5 mg/ml>240Mitoxantrone 2 mg/ml>240
Mitoxantrone 2 mg/ml>240Paclitaxel (Taxol) 6mg/ml>240
Paclitaxel (Taxol) 6mg/ml>240Paraplatin 10 mg/ml>240
Paraplatin 10 mg/ml>240Rituximab 10 mg/ml>240
Rituximab 10 mg/ml>240Thiotepa 10 mg/ml15.5
Thiotepa 10 mg/ml24.1Trisenox 0.1 mg/ml>240
Vincristine Sulfate 1mg/ml>240Vincristine Sulfate 1mg/ml>240

{33}------------------------------------------------

Sterilization MethodIrradiationIrradiation
Sterility (SAL)10-610-6
ColorBlueStraw (Natural)
Shelf Life3-year3-year
Resistance to Permeation by Chemotherapy DrugsMeets ASTM D6978-05 (2019)Meets ASTM D6978-05
Freedom from HolesAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015) requirementsAQL meets 21 CFR 800.20 and ASTM D3577-09 (2015) requirements
Powder ResidueMeets requirements of ≤ 2.0 mg/glove for Powder-free designation per ASTM D3577-09 (2015)Meets requirements of ≤ 2.0 mg/glove for Powder-free designation per ASTM D3577-09 (2015)
Dimensions and Physical PropertiesMeets ASTM D3577-09 (2015)Meets ASTM D3577-09 (2015)

{34}------------------------------------------------

Non-clinical Testing

In support of the "tested for use with chemotherapy agents" claim, the permeation of the subject devices by specific chemotherapy drugs was evaluated according to ASTM D6978-05 (2019): Standard Proctice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Clinical Data

Clinical testing was deemed not necessary to support this submission, and therefore, was not performed.

Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject devices are as safe, as effective, and perform as well as the legally marketed devices.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).