(81 days)
The Cellvizio® I.V.E. with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells and vessels and their organization or architecture.
The GastroFlex™ N and ColoFlex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts. i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories.
The AlveoFlex™N Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., respiratory systems, accessed by an endoscope or endoscopic accessories.
The CholangioFlex™ N Confocal Miniprobe™ is intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscope or endoscopic accessories.
The AQ-Flex™ 19 N Confocal Miniprobe™ is intended to allow imaging of anatomical tracts. i.e.; gastrointestinal and respiratory tracts, accessed by an endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA).
The CystoFlex™ F N, CystoFlex™ R N and Uroflex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts. i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.
The CelioFlex™ 5 N Confocal Miniprobe™ is intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures.
The subject device Cellvizio® I.V.E. system with Confocal Miniprobes™ is a confocal laser imaging system that can be used with a variety of Confocal Miniprobe™ (fiber optic probes) to allow real-time imaging of the internal microstructure of tissues by direct contact with the tip of the Confocal Miniprobe™ during existing medical procedures, such as endoscopy or endoscopic or open surgical procedures. The technology provides the physician with additional information in the form of real-time images during these open or minimally invasive procedures.
The Cellvizio® I.V.E. system with Confocal Miniprobes™ represents a refinement of the currently cleared and marketed predicate device, the Cellvizio® 100 Series System with Confocal Miniprobes™ (K172844). Design modifications and refinements include:
-
- Improving the integration of the system in endoscopy, interventional, or surgical operating suites by reducing the global footprint of the system.
-
- lmproving the user interface and ease-of-use of the system by replacing the keyboard and the trackball of the Cellvizio® 100 series with a touchscreen for review and interaction with the software and the addition of a separate remote (external) display screen for live imaging. This configuration allows better positioning of the physician's line of site during procedures.
-
- Simplifying the connection of the Miniprobes to the system. This improvement has no impact on other parts of the Confocal Miniprobe design, in particular the sheathed fiber and that are the patientcontacting parts are unchanged.
-
- The addition of "autofocus," allowing automatic positioning of the laser focal point for optimal optical injection from the Optical Scanning Unit into the fibers.
The Cellvizio® I.V.E. system with Confocal Miniprobes™ is a refinement of the existing Cellvizio 100 Series System with Confocal Miniprobes™ (K172844). The provided text describes the device, its intended use, and various tests performed to demonstrate its safety and effectiveness and its substantial equivalence to the predicate device.
Here's an analysis of the acceptance criteria and study information:
1. Acceptance Criteria and Reported Device Performance
The document states that "The results from these performance evaluations demonstrated that the Cellvizio® I.V.E system with Confocal Miniprobes™ met the acceptance criteria defined in the product specification and the performance is equivalent to the predicate device." However, specific numerical acceptance criteria (e.g., minimum resolution in microns, specific electrical safety thresholds, or maximum cytotoxicity levels) are not explicitly provided in the given text. Instead, the compliance is generally stated in relation to recognized standards.
Therefore, the table below will list the tests performed and the general performance statement.
| Test Category | Acceptance Criteria (as stated or implied) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993 series (Parts 1, 5, 10, 11) for cytotoxicity, sensitization, irritation, systemic toxicity. | Tests performed according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010, and ISO 10993-11:2017. Device met acceptance criteria defined in product specification. |
| Reprocessing Effectiveness | Compliance with AAMI TIR 12:2010 and AAMI TIR 30:2011. | Tests performed according to AAMI TIR 12:2010 and AAMI TIR 30:2011. Device met acceptance criteria defined in product specification. |
| Laser Safety | Compliance with IEC 60825-1:2007, 2014, and 21 CFR 1040.10, 1040.11 with Laser Notice No. 50 and No. 56. | Tests performed compliant with IEC 60825-1:2007 and 2014, and 21 CFR 1040.10 and 1040.11 with Laser Notice No. 50 and No. 56. Device met acceptance criteria defined in product specification. |
| Imaging Quality | Compliance with ISO 8600-1:2015 for general requirements; resolution measured according to internal standards using the 1951 USAF resolution test chart. | Tests performed in compliance with ISO 8600-1:2015. Resolution measured according to internal standards using 1951 USAF resolution test chart. Device met acceptance criteria defined in product specification. |
| Software Verification & Validation | Compliance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (minor level of concern) and IEC 62304:2006 (amendment 2015). | Software verification and validation testing conducted, documentation provided as recommended by FDA guidance. Software considered "minor" level of concern. Developed in accordance with IEC 62304:2006 (amendment 2015). Device met acceptance criteria defined in product specification. |
| Electrical Safety & EMC | Compliance with IEC 60601-1:2005/Amendment 2012, IEC 60601-2-18:2009, IEC 60601-1-6:2010/A1:2013 (Ed 3.1), and IEC 60601-1-2:2014. | Tests performed. System complies with IEC 60601-1:2005/Amendment 2012, IEC 60601-2-18:2009, 60601-1-6:2010/A1:2013 (Ed 3.1) for safety and IEC 60601-1-2:2014 for EMC. Device met acceptance criteria defined in product specification. |
| Usability | Compliance with IEC 62366-1:2015, relying on design control and risk management processes, validated by an external laboratory. | Usability engineering process compliant with IEC 62366-1:2015 standards implemented. Compliance validated with an external laboratory (LNE). Device met acceptance criteria defined in product specification. |
| Overall Performance | Substantial equivalence to predicate device (K172844) in terms of safety and effectiveness, based on technological characteristics and performance. | The device is deemed substantially equivalent, with performance matching the predicate and meeting specified acceptance criteria across all tested aspects. Differences in technological characteristics do not raise new questions of safety and effectiveness. |
2. Sample Size for Test Set and Data Provenance
The document does not specify a sample size for a "test set" in the context of clinical or image-based performance as would be relevant for an AI/ML device for diagnostic interpretation. The listed tests are engineering and regulatory compliance validations rather than a clinical performance study involving patient data. For example:
- Biocompatibility tests involve material samples.
- Reprocessing tests involve device components.
- Laser safety, imaging quality (resolution using a test chart), software V&V, electrical safety, and usability tests are performed on the device system itself or its software.
Therefore, the concept of a "test set" related to patient data or clinical images, as commonly understood for AI/ML device studies, is not applicable or detailed in this submission. Data provenance (country of origin, retrospective/prospective) is also not relevant or mentioned for these types of engineering and regulatory tests.
3. Number of Experts and Qualifications for Ground Truth
Again, since this submission focuses on engineering and regulatory compliance rather than clinical diagnostic performance using patient data, the concept of "ground truth" established by experts for a test set (e.g., radiologists interpreting images) is not applicable or mentioned.
4. Adjudication Method for Test Set
No adjudication method is described because there are no human interpretations of patient data in a test set to adjudicate.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study, which typically evaluates how human readers' performance improves with AI assistance, was not performed or mentioned in this submission. This device appears to be an imaging system, not an AI-powered diagnostic interpretation tool for image analysis.
6. Standalone Performance Study
A standalone performance study (algorithm only, without human-in-the-loop performance) was not performed or mentioned. The device is an imaging system designed to provide real-time images for a physician during existing medical procedures.
7. Type of Ground Truth Used
For the engineering and regulatory tests performed:
- Biocompatibility: Ground truth is established by adherence to ISO standards and laboratory testing results against specified limits.
- Reprocessing Effectiveness: Ground truth is adherence to AAMI TIR standards and demonstration of cleanliness/sterility.
- Laser Safety, Electrical Safety & EMC: Ground truth is compliance with IEC and CFR standards, measured against defined technical specifications.
- Imaging Quality: Ground truth for resolution is based on objective measurements using a calibrated test chart (1951 USAF resolution test chart) against internal standards and ISO 8600-1.
- Software V&V: Ground truth is compliance with FDA guidance and IEC 62304.
- Usability: Ground truth is compliance with IEC 62366-1, validated by an external laboratory.
There is no mention of ground truth based on expert consensus, pathology, or outcomes data as would be used for clinical performance studies.
8. Sample Size for the Training Set
This submission describes a medical imaging device (hardware and embedded software) and its regulatory compliance. It is not an AI/ML diagnostic algorithm that would typically have a "training set" of data in the common sense. Therefore, a sample size for a training set is not applicable or mentioned.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" for an AI/ML algorithm in this context, the method for establishing its ground truth is not applicable or mentioned.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Mauna Kea Technologies % Michael Daniel President Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460
February 28, 2020
Re: K193416
Trade/Device Name: Cellvizio® I.V.E. system with Confocal Miniprobes™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: December 7, 2019 Received: December 9, 2019
Dear Michael Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193416
Device Name
Cellvizio® I.V.E. system with Confocal MiniprobesTM
Indications for Use (Describe)
The Cellvizio® I.V.E. with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells and vessels and their organization or architecture.
The GastroFlex™ N and ColoFlex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts. i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories.
The AlveoFlex™N Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., respiratory systems, accessed by an endoscope or endoscopic accessories.
The CholangioFlex™ N Confocal Miniprobe™ is intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscope or endoscopic accessories.
The AQ-Flex™ 19 N Confocal Miniprobe™ is intended to allow imaging of anatomical tracts. i.e.; gastrointestinal and respiratory tracts, accessed by an endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA).
The CystoFlex™ F N, CystoFlex™ R N and Uroflex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts. i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.
The CelioFlex™ 5 N Confocal Miniprobe™ is intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
510(k) Number: K193416
Applicant Information:
Date Prepared: February 26, 2020
| Name: | Mauna Kea TechnologiesManufacturer Contact Person: Aline Criton |
|---|---|
| Address: | 9 rue d'EnghienF-75010 Paris, France |
| Phone: | +33 1 48 24 03 45 |
| Fax: | +33 1 48 24 12 18 |
| Contact Person: | Michael A Daniel, Consultant |
| e-mail: | madaniel@clinregconsult.com |
| Phone Number: | (415) 407-0223 |
| Facsimile Number: | (610) 545-0799 |
Device Information:
| Device Trade Name: | Cellvizio® I.V.E. system with Confocal Miniprobes™ |
|---|---|
| Common Name: | Endoscope and Accessories |
| Classification Name(s): | Confocal Optical Imaging |
| Product Code/Regulation: | OWN / GCJ 21 CFR 876.1500 |
| Classification: | Class II |
Primary Predicate Device:
Cellvizio 100 Series System with Confocal Miniprobes (K172844).
Reference Predicate Device:
Cellvizio 100 Series Confocal Laser Imaging systems and their Confocal Miniprobes (K183640)
Device Description:
The subject device Cellvizio® I.V.E. system with Confocal Miniprobes™ is a confocal laser imaging system that can be used with a variety of Confocal Miniprobe™ (fiber optic probes) to allow real-time imaging of the internal microstructure of tissues by direct contact with the tip of the Confocal Miniprobe™ during existing medical procedures, such as endoscopy or endoscopic or open surgical procedures. The technology provides the physician with additional information in the form of real-time images during these open or minimally invasive procedures.
The Cellvizio® I.V.E. system with Confocal Miniprobes™ represents a refinement of the currently cleared and marketed predicate device, the Cellvizio® 100 Series System with Confocal Miniprobes™ (K172844). Design modifications and refinements include:
-
- Improving the integration of the system in endoscopy, interventional, or surgical operating suites by reducing the global footprint of the system.
-
- lmproving the user interface and ease-of-use of the system by replacing the keyboard and the trackball of the Cellvizio® 100 series with a touchscreen for review and interaction with the software and the addition of a
{4}------------------------------------------------
separate remote (external) display screen for live imaging. This configuration allows better positioning of the physician's line of site during procedures.
-
- Simplifying the connection of the Miniprobes to the system. This improvement has no impact on other parts of the Confocal Miniprobe design, in particular the sheathed fiber and that are the patientcontacting parts are unchanged.
-
- The addition of "autofocus," allowing automatic positioning of the laser focal point for optimal optical injection from the Optical Scanning Unit into the fibers.
The intended use, indications for use, and mechanism of the subject device are unchanged as compared to the 510(k) cleared Cellvizio® 100 Series System with Confocal Miniprobes™ (K172844).
Indications for Use:
The Cellvizio® I.V.E. system with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells and vessels and their organization or architecture.
The GastroFlex™ N and ColoFlex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., gastrointestinal systems, accessed by an endoscope or endoscopic accessories.
The AlveoFlex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., respiratory systems, accessed by an endoscope or endoscopic accessories.
The CholangioFlex™ N Confocal Miniprobes™ are intended to allow imaging of the upper gastrointestinal tract including biliary and pancreatic ducts, accessed by an endoscopic accessories.
The AQ-Flex™ 19 N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA).
The CystoFlex™ F N, CystoFlex™ R N and UroFlex™ N Confocal Miniprobes™ are intended to allow imaging of anatomical tracts, i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.
The CelioFlex™ 5 N Confocal Miniprobes™ are intended to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, including robot-assisted procedures.
Comparison to Predicate Device:
The Cellvizio® I.V.E. and the Cellvizio® 100 Series systems are based on the same technology and have virtually identical Confocal Miniprobes™. The two devices are substantially equivalent as:
-
- Both devices have the same Intended use and indications for use. The Cellvizio® 100 Series System with Confocal Miniprobes™ (K172844) (predicate device) and the Cellvizio® I.V.E. system with Confocal Miniprobes™ (subject device) are based on the same technology of confocal laser endomicroscopy.
-
- Both devices rely on an optical scanning unit and a fiber probe to propagate light between the observation area and the system in order to provide real-time imaging. The Cellvizio® I.V.E. system with confocal Miniprobes™ is basically a miniaturized version of the predicate Cellvizio® 100 Series System with Confocal
{5}------------------------------------------------
Miniprobes™ (K172844). Both systems use Confocal Miniprobes™ with identical optical design and properties (only the connector differs). Both systems are intended to be used by the same users and on the same intended patient populations.
- ဒေ There are no significant technological characteristic differences between the predicate devices. The design of the Cellvizio® I.V.E. system is based on the Cellvizio® 100 Series system and has evolved in order to reduce the system footprint and to improve the usability of the device. These design changes are not significant changes and raise no new or different questions of safety and effectiveness.
Therefore, technologically the Cellvizio® I.V.E. system with Confocal Miniprobes™ is substantially equivalent to the predicate Cellvizio® 100 series system with Confocal Miniprobes™.
Testing Completed:
.
The following verification and validation tests were performed on the subject device:
- Biocompatibility (Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity & Systemic toxicity) according to:
- ISO 10993-1:2009 "Biological evaluation of medical devices Part 1: Evaluation and testing within a O risk management process".
- ISO 10993-5:2009 "Biological evaluation of medical devices Part 5: Tests for in vitro cytoxicity", O
- o ISO 10993-10:2010 "Biological evaluation of medical devices - Part 10: Tests for irritation and delayedtype hypersensitivity",
- ISO 10993-11:2017 "Biological evaluation of medical devices Part 11: Tests for systemic toxicity" O
- . Effectiveness of reprocessing methods has been tested according to:
- AAMI TIR 12:2010 "Designing, testing and labeling reusable medical devices for reprocessing in health o care facilities: A quide for medical device manufacturers" and
- AAMI TIR 30:2011 "A compendium of processes, materials, test methods, and acceptance criteria for O cleaning reusable medical devices"
- . Laser safety has been tested compliant with:
- IEC 60825-1:2007 and 2014 and o
- 21 CFR 1040.10 and 21 CFR 1040.11 with Laser Notice No. 50 and No. 56. O
- . Imaging quality has been tested in compliance with the following standards:
- o ISO 8600-1:2015 "Optics and photonics -- Medical endoscopes and endotherapy devices -- Part 1: General requirements",
- o The resolution is measured according to our internal standards using 1951 USAF resolution test chart as a resolution test pattern.
- . Software Verification and Validation:
- Software verification and validation testing were conducted and documentation was provided as o recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The Cellvizio® I.V.E. software is considered as a "minor" level of concern.
- The Cellvizio® I.V.E. software was developed in accordance with the IEC 62304:2006 standard o (amendment 2015).
- Electrical safety and electromagnetic compatibility (EMC):
- Electrical safety and EMC testing were conducted on the Cellvizio® I.V.E system with Confocal o Miniprobes™. The system complies with the IEC 60601-1:2005/Amendment 2012, IEC 60601-2-18:2009 and 60601-1-6:2010/A1:2013 (Ed 3.1) standards for safety and the IEC 60601-1-2:2014 standard for EMC.
- Usability
{6}------------------------------------------------
- Mauna Kea Technologies has implemented a usability engineering process compliant with the IEC 62366 о 1:2015 standards. Compliance to this standard relies on the design control process and on the risk management process and is validated with an external laboratory (LNE).:
The results from these performance evaluations demonstrated that the Cellvizio® I.V.E system with Confocal Miniprobes™ met the acceptance criteria defined in the product specification and the performance is equivalent to the predicate device.
Conclusions:
The subject device, the Cellvizio® I.V.E system with Confocal Miniprobes™, is substantially equivalent to the predicate device Cellvizio® 100 Series System with Confocal Miniprobes™ (K172844) in terms of technological characteristics, with minor design modifications implemented. The subject device has the same principle of operation, critical performance requirements, biocompatibility, reprocessing, intended use, indications for use, and product technical information as the cleared predicate device, the Cellvizio® 100 series system with Confocal Miniprobes™.
Based upon performance testing, reprocessing parameters, biocompatibility, electromagnetic compatibility, and usability testing provided in this subject device Cellvizio® I.V.E system with Confocal Miniprobes™ can be used as safely and effectively to image the internal microstructure of tissues including, and not limited to, the identification of cells and their organization or architecture as the predicate. The differences in technological characteristics do not raise any different questions of safety and effectiveness. The Cellvizio® I.V.E. system with Confocal Miniprobes™ was determined to be substantially equivalent to the Cellvizio® 100 series system with Confocal Miniprobes™.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.