K Number
K193310
Date Cleared
2020-03-17

(109 days)

Product Code
Regulation Number
870.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

Device Description

The Abbott Confirm RxTM (ICM) System consists of the following key features and components:

  • Confirm Rx™ ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias.
  • Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool to implant the device subcutaneously.
  • Model 3111 Magnet: SJM donut magnet, facilitates faster startup of Bluetooth connection and provides user authentication (required for programmer sessions).
  • Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information.
  • myMerlin™ Patient App (Model APP1000 (Android) and APP1001 (iOS)): The mobile application provides the means for the patient to activate EGM recording in the Confirm RxTM device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network.
  • Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net MN5000 system allows physicians to remotely monitor and diagnose patients' cardiac events.
AI/ML Overview

The provided text describes a 510(k) premarket notification for the "myMerlin Mobile Application (Android)" and "myMerlin Mobile Application (iOS)" by Abbott (St. Jude Medical). This application is part of the Confirm Rx™ Insertable Cardiac Monitor System. The submission claims substantial equivalence to a previously cleared predicate device (K192593, Confirm Rx™ Insertable Cardiac Monitor System).

The submission focuses on updates to the myMerlin™ mobile application, primarily related to scheduled background workflows and telemetry connectivity improvements, as well as the ability to transfer a patient's remote monitoring support profile on Merlin.net.

The document states that the fundamental technological characteristics, intended use, indications for use, operating rules, cybersecurity, and usability testing remain the same as the predicate device. Therefore, the testing performed was to demonstrate that the updated mobile application maintains its safety and effectiveness for its intended use, rather than proving performance against new acceptance criteria for a novel functionality.

Given this context, the acceptance criteria and study details are presented as follows:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySource of Criteria / Performance StandardReported Device Performance
Software Functionality & PerformancePredetermined design and performance specifications (consistent with predicate)"Completion of all verification and validation activities demonstrated that the updated mobile application meets its predetermined design and performance specifications."
Telemetry ConnectivityPredetermined design and performance specifications for device communicationUpdate to improve myMerlin™ mobile application scheduled background workflows and update of the telemetry connectivity of the Confirm Rx™ ICM System, including the myMerlin™ mobile application and Confirm Rx™ ICM device.
Data Transfer (Remote Monitoring Profile)Predetermined design and performance specifications"The candidate myMerlin™ mobile application… is being updated to allow the patient's remote monitoring support profile on Merlin.net to be transferred from the implant clinic to the remote monitoring clinic."
Safety and EffectivenessEquivalent to predicate device (K192593)"The results of the testing show that the candidate Confirm Rx™ ICM System with updated myMerlin™ mobile application performs as intended and is safe for its intended use."
Risk MitigationRisk analysis"The results of the verification and validation tests and the risk analysis have demonstrated the myMerlin™ mobile application within the Confirm Rx™ ICM System functions in accordance with the product specifications."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size of a "test set" in terms of patient data or scenarios used. The testing described is primarily focused on software verification, system verification, design validation, and Bluetooth Low Energy proxy software verification. This implies testing involved the software application itself and its interaction with the ICM device, rather than a clinical study with a patient cohort.

  • Data Provenance: Not applicable in the context of a clinical test set. The provenance of the software updates is from the manufacturer, Abbott. The nature of the changes (software updates) suggests internal testing and technical verification.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not provided. Given the nature of the submission (updates to a mobile application interacting with an already cleared device), the "ground truth" would likely be established by engineering and software performance standards, rather than expert clinical consensus on patient data.

4. Adjudication Method for the Test Set

This information is not provided. As noted in point 3, the testing appears to be technical in nature rather than clinical, so a traditional clinical adjudication method would not be expected.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was mentioned. The myMerlin™ mobile application is primarily a patient interface for activating EGM recording and data pass-through. It is not an AI-assisted diagnostic tool for "human readers," so this type of study would not be relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

The submission describes the myMerlin™ mobile application (APP1000/APP1001) as a component of the Confirm Rx™ ICM System. The application enables patients to activate EGM recording and facilitates data pass-through to the Merlin.net Patient Care Network for physician review. The focus is on the functional performance of the application and its connectivity, not on a standalone algorithm for diagnosis without human input. Therefore, a standalone algorithm performance study as typically understood for diagnostic AI was not done or described. The "standalone" performance here refers to the application's ability to fulfill its intended technical functions.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the specific updates to the myMerlin™ mobile application, the "ground truth" for verification and validation would be defined by predetermined design and performance specifications or product specifications. This means the software was tested against its expected behavior and technical requirements, rather than against clinical outcomes, pathology, or expert consensus on a clinical dataset. The objective was to confirm that the updated software performed as intended and did not negatively impact the already cleared system's functionality.

8. The Sample Size for the Training Set

This information is not applicable. The myMerlin™ mobile application is not described as involving a machine learning or AI algorithm that would require a "training set" in the conventional sense (i.e., for learning to identify patterns in data). The changes are described as software updates to workflows and connectivity.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as no training set for a machine learning algorithm is mentioned or implied.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

March 17, 2020

Abbott (St. Jude Medical) Vanessa Powell Regulatory Affairs Specialist 15900 Valley View Court Sylmar, California 91342

Re: K193310

Trade/Device Name: myMerlin Mobile Application (Android), myMerlin Mobile Application (iOS) Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC, DSI Dated: March 11, 2020 Received: March 12, 2020

Dear Vanessa Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name

Confirm RxTM Insertable Cardiac Monitor System

Indications for Use (Describe)

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

The Confirm RxTM ICM has not been specifically tested for pediatric use.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Date Prepared:November 27, 2019
Submitter:Abbott (formerly St. Jude Medical), Cardiac Rhythm Management Division
Address:15900 Valley View Ct.Sylmar, CA 91342USA
Phone:(818) 362-6822
Establishment Registration:2017865
Contact Person:Vanessa PowellRegulatory Affairs Specialist(818) 493-3290vanessa.powell@abbott.comJennifer DunhamManager, Regulatory Affairs(818) 383-1630jennifer.dunham@abbott.com
Trade Name/ProprietaryConfirm Rx™ Insertable Cardiac Monitor System
Common Name:Insertable Cardiac Monitor
Model Number:DM3500 (including APP1000 (Android) and APP1001 (iOS))
Classification Name:Telephone electrocardiograph transmitter and receiver (21 CFR 870.2920)
Product Code:MXC, DSI
Classification:Class II
Pediatric Use:The Confirm Rx™ Insertable Cardiac Monitor has not been specificallytested in pediatric patients below the age of 18 years.

LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED

510(k) K192593 Confirm Rx™ Insertable Cardiac Monitor System

INDICATIONS FOR USE

There are no changes to the Indications for Use as a result of this submission. The Indications for Use for the Confirm RxTM Insertable Cardiac Monitor (ICM) system are as follows:

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

The Confirm RxTM ICM has not been specifically tested for pediatric use.

{4}------------------------------------------------

PRODUCT DESCRIPTION

The Abbott Confirm RxTM (ICM) System consists of the following key features and components:

  • 트 Confirm Rx™ ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias. Specific features include:
    • Patient-initiated triggering of EGM storage using the myMerlin™ mobile application. o This includes capability for the patient to identify symptoms, which are stored with the EGM for physician review.
    • Automated triggering of EGM storage when tachycardia, bradycardia, or pauses are o detected; with physician-programmable values for pause duration, bradycardia rate, tachycardia rate, and number of tachycardia intervals.
    • Automated triggering of EGM storage when atrial fibrillation (AF) is detected, with O physician programmable values for AF duration.
    • The ability to identify EGM anomalies as a consequence of noise or vigorous activity O and inhibit EGM storage as applicable.
    • The addition of remote care capabilities O
  • 트 Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool to implant the device subcutaneously. The implantable device is pre-loaded into the insertion tool and packaged together with the incision tool.
    • The DM3520 incision tool is used to make an angled cut, which is the sole incision O required to implant the ICM. The introducer end of the DM3510 insertion tool is inserted, creating the initial pocket, Figure 1.
    • o With the insertion tool in place, the plunger is withdrawn to drop the pre-loaded device into the insertion channel. The plunger is pushed forward to insert the device into the pocket (Figure 1). This completes insertion (implantation) of the ICM, and the incision is closed per standard of care.

Image /page/4/Figure/14 description: The image contains two figures labeled (a) and (b). Both figures show a hand holding a medical instrument, possibly a scalpel or lancet, and making an incision on a surface. In figure (b), there is an arrow indicating the direction of movement of the instrument.

Figure 1: Insertion Process

{5}------------------------------------------------

  • 트 Model 3111 Magnet: SJM donut magnet, facilitates faster startup of Bluetooth connection and provides user authentication (required for programmer sessions).
  • Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information. The Merlin PCS programmer, using the Model BLU1000 Bluetooth dongle, an off the shelf component, communicates with the Confirm RxTM device with Bluetooth telemetry (also referred to as Bluetooth Low Energy or BLE). Programmer software Model 3330 v23.0.1 and later will contain support for the Confirm Rx™ device, adding support for the Model BLU1000 Bluetooth dongle, and new tabs of Implant View and Reason for Monitoring features.
    • Implant View is designed to streamline programming at the time of implant. Upon o initial interrogation at implant, the programmer automatically displays the Implant View in which the user can immediately input device and patient information to be stored onto the device, as well as set the Reason for Monitoring.
    • Reason for Monitoring allows the user to select from a list of possible conditions for o which the patient is receiving the device (such as Syncope, Ventricular Tachycardia, Palpations, etc). The programmer then sets the AF duration parameter and EGM storage priority based on the reason selected. These parameters can be manually adjusted by the user later if customization is preferred.
  • myMerlin™ Patient App (Model APP1000 (Android) and APP1001 (iOS)): The mobile ■ application provides the means for the patient to activate EGM recording in the Confirm RxTM device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network. Patients who do not supply their own mobile device may be provided with an Abbott issued off the shelf Android mobile device, which is not a part of the medical device.
  • Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net MN5000 system allows physicians to remotely monitor and diagnose patients' cardiac events. The Merlin.net MN5000 v7.7 contains updates that are specific to Confirm RxTM.

{6}------------------------------------------------

Image /page/6/Figure/3 description: The image shows a diagram of a cardiac monitoring system. The system includes several components, such as a clinician, programmer, external magnet, ICM insertable cardiac monitor, heart, patient application, clinician portal, and patient. The ICM insertable cardiac monitor is connected to the heart via sensing, and it communicates with the programmer and patient application via telemetry communications. The clinician portal is connected to the patient application via the internet.

Figure 2: Block Diagram of Confirm RxTM ICM System

The myMerlin™M mobile application, APP1000 and APP1001, are the subject of this premarket notification for the Confirm Rx™ ICM System. The Confirm Rx™ ICM device Model DM3500 was cleared on October 18, 2019 per 510(k) K192593. The Confirm Rx™ ICM, DM3500, Implant Tools, Model DM3520 and DM3510 are not the subject of this premarket notification. The Merlin PCS programmer, Model 3330 v24.5.1 software for the Model 3650 Merlin PCS programmer and Merlin.net MN5000 v7.7 are not subject of this premarket notification. The magnet, Model 3111, is Class I exempt MDDS and not subject of this premarket notification.

TECHNOLOGICAL CHARACTERISTICS

The fundamental technological characteristics of the myMerlin™ mobile application Model APP1000 and APP1001 are the same as the predicate myMerlin™M mobile application Model APP1000 (Android) and APP1001 (iOS). Model APP1000 runs on the Android operating system and Model APP1001 runs on the iOS operating system. The myMerlin™ for Confirm RxTM mobile application provides the means for the patient to activate EGM recording in the Confirm Rx™ ICM device, with data pass-through functionality to enable physician follow-up via theMerlin.net Patient Care Network (clinician portal). The Confirm Rx™ ICM (DM3500) will continue to use the same technology. See Figure 2 for a block diagram of the Confirm Rx™ ICM system.

The myMerlin™ mobile application provides the means for the patient to activate EGM recording in the Confirm RxTM ICM device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network (clinician portal). The Confirm Rx™ ICM (DM3500) will continue to use the same technology.

The fundamental technological characteristics of the myMerlin™ mobile application are not changing. In comparison to the predicate device the myMerlin™ mobile application and the

{7}------------------------------------------------

candidate Confirm Rx™ ICM system with the updated myMerlin™ mobile application with updated mobile application have the same:

  • Intended Use and Indications for Use .
  • o Operating rules
  • . Cybersecurity
  • . Usability Testing

SUBSTANTIAL EQUIVALENCE

The Confirm RxTM ICM System, including the myMerlinTM mobile application with the proposed updates and enhancements, is substantially equivalent to the predicate Confirm Rx™ ICM System (K192593, cleared on October 18, 2019). The indications for use are not impacted by the updated mobile application operations. Both the candidate and the predicate device (APP1000 Android, and APP1001 iOS) have the same function and fundamental scientific technology. The minor differences do not raise new issues of safety and effectiveness. Thus, the candidate Confirm Rx™ ICM System, including the updated myMerlin™ mobile application Model APP1000 and APP1001, is substantially equivalent to the predicate Confirm Rx™ ICM System (K192593).

In comparison to the predicate myMerlin™ mobile application, the candidate myMerlin™ mobile application has the following differences:

  • The candidate myMerlin™ mobile application with updated telemonitoring . communication contains the following changes:
    • o Update to improve myMerlin™ mobile application scheduled background workflows
    • O Update of the telemetry connectivity of the Confirm Rx™ ICM System including the myMerlin™ mobile application and Confirm RxTM ICM device.
  • The candidate myMerlin™ mobile application, model APP1000, is being updated to ● allow the patient's remote monitoring support profile on Merlin.net to be transferred from the implant clinic to the remote monitoring clinic.

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

All necessary testing was conducted on the candidate myMerlin™ mobile application to support a determination of substantial equivalence to the predicate device, including:

  • Software Verification and System Verification ●
  • Design Validation ●
  • Bluetooth Low Energy Proxy Software Verification .

Completion of all verification and validation activities demonstrated that the updated mobile application meets its predetermined design and performance specifications and that the product is substantially equivalent to the predicate myMerlin™ mobile application, model APP1000 (Android) and APP1001(iOS). The results of the testing show that the candidate Confirm RxTM ICM System with updated myMerlin™ mobile application performs as intended and is safe for its intended use.

{8}------------------------------------------------

CONCLUSION

The myMerlin™ mobile application is used in conjunction with the Confirm Rx Insertable Cardiac Monitor (ICM) System; therefore, the indications for use are not impacted by the updates and enhancements to the myMerlin mobile application. The results of the verification and validation tests and the risk analysis have demonstrated the myMerlin™ mobile application within the Confirm Rx™ ICM System functions in accordance with the product specifications.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).