AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface with evidence of adequate healthy bone sufficient for support of the implanted components. Candidates for unicondylar knee replacement include those with: · Joint impairment due to osteoarthritis or traumatic arthritis of the knee · Previous femoral condyle or tibial plateau fracture, creating loss of function · Valgus or varus deformity of the knee · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.

The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis. The indications for use include restoring joint function and relief of pain due to: · Painful joint disease due to osteoarthritis · Traumatic arthritis of the knee Post traumatic loss of joint function · Failed osteotomies, hemiarthroplasties and unicondylar implants The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions. The iDuo implant is intended for cemented use only.

The iTotal Posterior Stabilized {PS) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: · Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee. · Post traumatic loss of joint function. · Moderate varus, valgus or flexion deformity. · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.

The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: · Painful joint disease due to osteoarthritis, traumatic arthritis, theumatoid arthritis or osteonecrosis of the knee. · Post traumatic loss of joint function. · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.

Device Description

ConforMIS knee replacement systems are patient-specific semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis ortrauma. Using patient imaging (either CT or MR scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral components of the devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from CoCrMo and polyethylene inserts manufactured from UHMWPE or an all polyethylene tibial component. The patellar components are manufactured from UHMWPE.

AI/ML Overview

This document is a 510(k) premarket notification for knee replacement systems. It does not present a study or acceptance criteria for a device that relies on an algorithm or AI. Instead, it describes a non-clinical performance evaluation for sterilization validation of existing knee replacement systems.

Therefore, most of the requested information regarding acceptance criteria and studies for an AI/algorithm-based device simply do not apply to this document.

However, I can extract the information relevant to the non-clinical performance evaluation described:

  1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Sterilization Validation to establish a SAL of 1x10^-6 for Vaporized Hydrogen Peroxide (VHP) sterilization.Device components are effectively sterilized using Vaporized Hydrogen Peroxide, achieving a SAL of 1x10^-6.
    VHP residual testing(Implicitly met, as effective sterilization was confirmed)
    Product and packaging compatibility with VHP sterilization(Implicitly met, as effective sterilization was confirmed)
  2. Sample size used for the test set and the data provenance: Not applicable. This is a non-clinical sterilization validation, not a study involving patient data or a test set in the context of an algorithm.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is a non-clinical sterilization validation, not an analysis requiring expert ground truth for an algorithm.

  4. Adjudication method for the test set: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an AI device or an MRMC study.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document does not describe an algorithm or AI device.

  7. The type of ground truth used: For the sterilization validation, the "ground truth" would be the scientifically established standard for sterility (SAL of 1x10^-6) as defined by regulatory bodies and testing protocols.

  8. The sample size for the training set: Not applicable. This document does not describe a machine learning model.

  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

March 13, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Conformis, Inc. Nancy Giezen Manager Regulatory Affairs 600 Technology Park Drive Billerica. Massachusetts 01821

Re: K193105

Trade/Device Name: iUni Unicondylar Knee Replacement System, iDuo Bicompartmental Knee Repair System, iTotal CR Knee Replacement System, iTotal PS Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY, OOG, NPJ, HSX Dated: January 8, 2020 Received: January 14, 2020

Dear Nancy Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Dr.Ting Song, PhD, RAC Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193105

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

Device Name

iUni Unicondylar Knee Replacement System

Indications for Use (Describe)

The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface with evidence of adequate healthy bone sufficient for support of the implanted components.

Candidates for unicondylar knee replacement include those with:

  • · Joint impairment due to osteoarthritis or traumatic
  • arthritis of the knee
  • · Previous femoral condyle or tibial plateau fracture,
  • creating loss of function
  • · Valgus or varus deformity of the knee

· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name

iDuo Bicompartmental Knee Repair System

Indications for Use (Describe)

The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis.

The indications for use include restoring joint function and relief of pain due to:

  • · Painful joint disease due to osteoarthritis
  • · Traumatic arthritis of the knee
  • Post traumatic loss of joint function
  • · Failed osteotomies, hemiarthroplasties and unicondylar implants

The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions.

The iDuo implant is intended for cemented use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

iTotal Posterior Stabilized {PS) Knee Replacement System

Indications for Use (Describe)

The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee

replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

· Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee.

· Post traumatic loss of joint function.

· Moderate varus, valgus or flexion deformity.

· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral implants.

· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

iTotal Cruciate Retaining (CR) Knee Replacement System

Indications for Use (Describe)

The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

  • · Painful joint disease due to osteoarthritis, traumatic arthritis, theumatoid arthritis or osteonecrosis of the knee.
  • · Post traumatic loss of joint function.
  • · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
  • · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
  • · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{6}------------------------------------------------

510(k) SUMMARY

Submitter's Name and Address:Conformis, Inc.600 Technology Park DriveBillerica, MA 01821
Establishment Registration Number(s):3009844603 and 3004153240
Date Summary was Prepared:November 5, 2019
Contact Person:Nancy GiezenManager Regulatory AffairsTelephone: 781-345-9058

Trade/Device Name(s):

iUni Unicondylar Knee Replacement System iDuo Bicompartmental Knee Repair System iTotal Cruciate Retaining (CR) Knee Replacement System iTotal Posterior Stabilized (PS) Knee Replacement System

Common Name:

Knee Replacement System

Device Class:

Class II

Regulation Numbers:

888.3560 888.3520

Classification Names and Product Codes:

iUni

Prosthesis, knee, femorotibial, nonconstrained, cemented, metal/polymer Knee Arthroplasty Implantation System; OOG, HSX

iDuo

Prosthesis, Knee, Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Knee Arthroplasty Implantation System; NPJ

{7}------------------------------------------------

iTotal CR and PS Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerKnee Arthroplasty Implantation System; JWH, OIY, OOG

Legally Marketed Predicate Device (Primary Predicate):

iTotal Cruciate Retaining (CR) Knee Replacement System (K180906)

(Ssecondary Predicates):

iUni Unicondylar Knee Replacement System (K132640, K133256) iDuo Bicompartmental Knee Replacement System (K093513, K133256) iTotal Posterior Stabilized (PS) Knee Replacement System (K161668)

Device Description:

ConforMIS knee replacement systems are patient-specific semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis ortrauma.

Using patient imaging (either CT or MR scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral components of the devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from CoCrMo and polyethylene inserts manufactured from UHMWPE or an all polyethylene tibial component. The patellar components are manufactured from UHMWPE.

Indications for Use:

iUni

The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface in patients with evidence of adequate healthy bone sufficient for support of the implanted components.

Candidates for unicondylar knee replacement include those with:

  • · Joint impairment due to osteoarthritis or traumatic
  • arthritis of the knee
  • · Previous femoral condyle or tibial plateau fracture,
  • creating loss of function
  • · Valgus or varus deformity of the knee
  • · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

iDuo

The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with

{8}------------------------------------------------

severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis. The indications for use include restoring joint function and relief of pain due to: • Painful joint disease due to osteoarthritis

· Traumatic arthritis of the knee

  • Post traumatic loss of joint function
  • · Failed osteotomies, hemiarthroplasties and unicondylar implants

The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions.

The iDuo implant is intended for cemented use only.

iTotal CR

The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:

• Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee.

· Post traumatic loss of joint function.

· Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.

• Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.

iTotal PS

The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:

· Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis or osteonecrosis of the knee.

  • Post traumatic loss of joint function.
  • Moderate varus, valgus or flexion deformity.

{9}------------------------------------------------

· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.

• Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

Technological Characteristics:

The subject devices were compared to the predicates and the design features, materials, sizes and methods of sterilization (EO, VHP, steam) are identical. This 510(k) adds the option to sterilize device components using the Steris VHP LTS-V sterilizer.

Non-Clinical Performance Evaluation:

The following non-clinical laboratory testing was performed, confirming that the device components are effectively sterilized using Vaporized Hydrogen Peroxide.

  • · Sterilization Validation testing to establish a SAL of 1x10-6
  • · VHP residual testing
  • Product and packaging compatibility with VHP sterilization

Conclusion:

Based on the testing conducted it is concluded that the subject devices are substantially equivalent to the predicate devices and can be sterilized to a SAL of 1x 10-6 ustilizing VHP sterilization.

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.