K Number
K192969

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-07-10

(261 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, 3D- visualization, and comparison of MR medical images. The images can be viewed in a number of output formats including volume rendering.

Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly.

Ezra Plexo Software is designed to support the oncological workflow by helping the user to confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.

Device Description

Ezra Plexo is a medical image application for 2D and 3D visualization, comparison, and manipulation of medical images. Plexo can be accessed through a web browser. It provides radiologists the ability to view and manipulate volumetric data such as MRI. The product allows volumetric segmentation of regions of interest, while enabling users to edit such segmentations and to take quantitative measurements.

Plexo does not interface directly with the MR scanner or any other data collection equipment. Instead, it uploads data files previously generated by such equipment. Its functionality is independent of the acquisition equipment vendor. Its analysis results are available on screen and can be saved for review.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for Ezra Plexo Software:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with numerical performance metrics. Instead, it states that "The test results demonstrated that the Ezra Plexo Software performs to its intended use, is deemed acceptable for clinical use, and does not introduce new questions of safety or efficacy."

This suggests the acceptance criteria were likely qualitative and based on the software successfully performing its intended functions (viewing, manipulation, 3D-visualization, and comparison of MR medical images; visualization of information; and supporting oncological workflow by helping confirm absence/presence of lesions, evaluation, quantification, and documentation) without errors or safety/efficacy concerns in the validation setting.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: 150 patient exams
  • Data Provenance: Not explicitly stated, but it was reviewed by five U.S. board-certified expert radiologists. It's not specified if this data was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Five (5)
  • Qualifications of Experts: U.S. board-certified expert radiologists. Specific years of experience are not mentioned.

4. Adjudication method for the test set

  • Adjudication Method: "Consensus ground truth created by five U.S. board certified expert radiologists." This implies a form of consensus process, where the five radiologists collectively reviewed the 150 patient exams to establish the definitive ground truth for the study. The exact method of achieving consensus (e.g., majority vote, discussion until agreement) is not detailed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not conducted or reported in this document. The study was a "nonclinical software validation" focused on the software's performance against expert consensus ground truth, not on human reader improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: The description of the validation test, which involved "performing nonclinical software validation tests on 150 patient exams utilizing consensus ground truth," strongly suggests a standalone (algorithm only) performance evaluation. The device is for "viewing, manipulation, 3D-visualization, and comparison," and it "enables visualization of information that would otherwise have to be visually compared disjointedly." The focus on "performs to its intended use" suggests the algorithm's output was compared to the ground truth. However, the exact metrics (e.g., sensitivity, specificity for lesion detection/quantification) are not provided.

7. The type of ground truth used

  • Type of Ground Truth: Expert consensus. Specifically, "consensus ground truth created by five U.S. board certified expert radiologists."

8. The sample size for the training set

  • Training Set Sample Size: The document does not specify the sample size for the training set. It only describes the "nonclinical software validation tests" on a test set of 150 patient exams.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: The document does not provide information on how ground truth was established for the training set, as it only details the validation (test) set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a stylized human figure on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" in a square and the rest of the text in a smaller font.

July 10, 2020

Ezra AI Inc. % Mr. David Girard Quality Manager 79 Madison Ave., Ste 545 NEW YORK NY 10016

Re: K192969

Trade/Device Name: Ezra Plexo Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QIH Dated: June 9, 2020 Received: June 11, 2020

Dear Mr. Girard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192969

Device Name Ezra Plexo Software

Indications for Use (Describe)

Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, 3D- visualization, and comparison of MR medical images. The images can be viewed in a number of output formats including volume rendering.

Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly.

Ezra Plexo Software is designed to support the oncological workflow by helping the user to confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CER 801 Subnart D)Over-The-Counter Ise (21 C.FR 801 Subnart C

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary 1

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

Submitter Information 1.1

Submitter's NameEzra Al Inc.
Address79 Madison Avenue, Suite 545, New York, NY, 10016 USA
Telephone Number888-402-3972
Contact PersonDavid Girard
Date of Summary PreparationJuly 2, 2020

Subject Device Information 1.2

Device NameEzra Plexo Software
Common NamePlexo
ClassificationII
Review PanelRadiology
Product CodeQIH
Regulations892.2050 - Automated radiological image processing software

1.3 Predicate Device Information

Arterys Oncology DL Device Name 510(k) Number K173542

{4}------------------------------------------------

Device Description വ

Ezra Plexo is a medical image application for 2D and 3D visualization, comparison, and manipulation of medical images. Plexo can be accessed through a web browser. It provides radiologists the ability to view and manipulate volumetric data such as MRI. The product allows volumetric segmentation of regions of interest, while enabling users to edit such segmentations and to take quantitative measurements.

Plexo does not interface directly with the MR scanner or any other data collection equipment. Instead, it uploads data files previously generated by such equipment. Its functionality is independent of the acquisition equipment vendor. Its analysis results are available on screen and can be saved for review.

Indications for Use 3

Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, and comparison of medical MR images can be viewed in a number of output formats including volume rendering.

Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly.

Ezra Plexo Software is designed to support the oncological workflow by helping the user confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions.

Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.

{5}------------------------------------------------

Summary of Technological 4 Characteristics Comparison

Table 1 shows that Plexo Software and the Predicate Device (K173542) are equivalent in technological characteristics.

FeatureSubject Device:Ezra Plexo SoftwarePredicate Device:Arterys Oncology DL(K173542)
PlatformChrome DesktopClient-server, Chrome Desktop
DICOM CompliantYesYes
Type of ScansMRMR, CT
Image NavigationToolsPan, zoom, rotate, slice scroll (viewmultiple slices)Adjust window level, minimize/maxi-mizeOrientation labels, scroll sync, rulerPan, zoom, rotate, slice scroll (viewmultiple slices)Adjust window level, minimize/maxi-mizeOrientation labels, cross-reference in-dicator
Image Display ModesYes, staticYes, static and cine
Text Annotation onthe ImageYesYes
LayoutCan select a viewport layoutand add series to itCan select a viewport layoutand add series to it
2D Image ReviewYesYes
3D Image ReviewYes, 3D review of segmented regionYes, 3D review of the full image
Create MPR ImagesNoYes
Manual SegmentationYesYes
2D Semi-AutomatedSegmentationYes, available only for MR prostatestudiesYes, available only for lung CT andliver MR studies
Linear Dimension Cal-culationYesYes
Volume CalculationYesYes

Table 1: Subject and Predicate Device Comparison

{6}------------------------------------------------

Performance Testing 5

The Ezra Plexo Software has been developed in a manner consistent with accepted standards for software development and evaluated in accordance with design specifications and applicable performance standards through software verification, validation, and usability testing.

The nonclinical software validation tests were performed on 150 patient exams utilizing consensus ground truth created by five U.S. board certified expert radiologists.

The test results demonstrated that the Ezra Plexo Software performs to its intended use, is deemed acceptable for clinical use, and does not introduce new questions of safety or efficacy. The testing was conducted in accordance to the software validation/verification plans and protocols. A full description of the software functionality, device hazard analysis, software requirements, verification, validation and usability study is provided in this submission.

Conformance Standards б

There are no applicable FDA mandated performance standards for this device. However, voluntary standards have been utilized in the production of the software. The device was designed and developed in accordance to the following conformance standards:

  • · ISO 14971:2007- Medical devices Application of risk management to medical devices
  • · IEC 62304:2006/A1:2015 Medical device software Software life cycle processes
  • · IEC 62366-1:2015 Medical Devices Application of usability engineering to medical devices
  • · NEMA PS 3.1-3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) set

Substantial Equivalence Conclusion 7

The Ezra Plexo Software has the same intended use and similar technological characteristics as the predicate Arterys Oncology DL (K173542) device. As demonstrated in this submission, the subject and predicate device are identical in indications for use and intended use. The subject Ezra Plexo Software is substantially equivalent to the predicate Arterys Oncology DL (K173542) device, and the minor differences in the technological characteristics of the subject and predicate device do not raise any new or different questions of safety and effectiveness.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).