(173 days)
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
MammoScreen is a software-only device for aiding interpreting physicians in identifying focal findings suspicious for breast cancer in screening FFDM (full-field digital mammography) acquired with compatible mammography systems. The product consists of a processing server and a web interface. The software applies algorithms for recognition of suspicious calcifications and soft tissue lesions. These algorithms have been trained on large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue. MammoScreen automatically processes FFDM and the output of the device can be used by radiologists concurrently with the reading of mammograms. The user interface of MammoScreen has several functions: a) Activation of computer aided detection (CAD) marks to highlight locations, known as findings, where the device detected calcifications or soft tissue lesions suspicious for cancer. b) Association of findings with a score, known as the MammoScreen Score, which characterizes findings on a 1-10 scale, with increasing level of suspicion. Only the most suspicious findings (with a MammoScreen score equal or greater than 5) are initially marked to limit the number of findings to review. The user shall also review findings with score of 4 or lower. c) Indication, with matching markers, when findings corresponding to the same findings are detected in multiple views of the FFDM. MammoScreen is configured as a DICOM Web compliant node in a network and receives its input images from another DICOM node, called "the DICOM Web Server". The MammoScreen output will be displayed on the screen of a personal computer compliant with requirements specified in the User Manual. The image analysis unit includes machine learning components trained to detect positive findings (calcifications and soft tissue lesions).
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The provided document defines acceptance criteria primarily through comparison with a predicate device and through the results of a clinical reader study. The core acceptance criterion for the clinical study appears to be an improvement in radiologist performance when using MammoScreen assistance compared to unaided reading.
Table of Acceptance Criteria and Reported Device Performance
| Criterion Type | Specific Criterion | Reported Device Performance (MammoScreen) | Met? |
|---|---|---|---|
| Premarket Equivalence (vs. Predicate Device K181704 Transpara) | |||
| Classification Regulation | 21 CFR 892.2090 | SAME | Yes |
| Medical Device Class | Class II | SAME | Yes |
| Product Code | QDQ | SAME | Yes |
| Level of Concern | Moderate | SAME | Yes |
| Intended Use | Concurrent reading aid for physicians interpreting screening FFDM to identify findings and assess their level of suspicion. | SAME | Yes |
| Target patient population | Women undergoing FFDM screening mammography. | SAME | Yes |
| Target user population | Physicians interpreting FFDM screening mammograms. | SAME | Yes |
| Design | Software-only device. | SAME | Yes |
| Scoring System | While not identical, the principle (level of suspicion from low to high) should be substantially equivalent. | 10-point scale vs. predicate's 1-100. Manufacturer claims interpretability benefits. Exam-level score provided. Deemed "substantially equivalent." | Yes |
| Finding Discovery | Reducing the number of findings the user has to review. | Default display for scores ≥ 5, user request for scores ≤ 4. Deemed "equivalent." | Yes |
| Performance Comparison | Overall performance gains should be comparable and not raise new safety/effectiveness questions. | AUC: unaided = 0.769, assisted = 0.798 (Difference: 0.028; P = 0.035). Predicate reported unaided = 0.866, assisted = 0.887. Deemed "still comparable." | Yes |
| Fundamental Scientific Technology | Involves medical image processing and machine learning, particularly deep learning for suspicious findings. | SAME | Yes |
| Clinical Performance (Reader Study) | |||
| Radiologist Performance | Radiologist performance with MammoScreen assistance is superior to unaided performance (main objective). | Average AUC improved from 0.769 (unaided) to 0.798 (with MammoScreen) (Difference = 0.028; P = 0.035). | Yes |
| Reading Time | Should not significantly increase. | Average reading time increased by 14% for scores > 4, but decreased by 2% for scores ≤ 4 in the second session. Overall, maximum increase did not exceed 15s. | Yes |
| Standalone Performance | Non-inferior to average unaided radiologist performance. | Standalone AUC = 0.790; Non-inferior to average unaided radiologist AUC = 0.770 (absence of statistical effect (p>0.05) and lower CI of diff > -0.03). | Yes |
| Sensitivity | Sensitivity of readers tended to increase with the use of MammoScreen without decreasing specificity (conclusion statement). | Reported overall performance improvement was statistically significant at breast (AUC) and lesion (pAUC) level, confirming trend. Specific values not explicitly in acceptance criteria here. | Yes |
Study Details for Device Acceptance
-
Sample Size Used for the Test Set and Data Provenance:
- Test Set Size: 240 mammographic screening images (cases).
- Data Provenance: Acquired at a US center. The text states "US FFDM acquired on Hologic® devices, and performance comparison with FFDM acquired on GE® devices," indicating images from at least two major mammography system manufacturers in the US.
- Retrospective/Prospective: Retrospective. The study "collected" images after they were acquired, and "For each exam, the cancer status has been verified... and used as gold standard."
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., years of experience). It only states that "the cancer status has been verified by either biopsy results (for all cancer positive cases and some of the negative cases) or an adequate follow-up (for negative cases only) and used as gold standard." This implies clinical data and follow-up was the primary ground truth, not consensus of a specific number of experts.
-
Adjudication Method for the Test Set:
- The document does not explicitly describe an adjudication method for establishing ground truth from multiple expert reads. Ground truth was established via biopsy or adequate follow-up, which are objective clinical outcomes, not subjective reader interpretations.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- Yes, an MRMC study was performed.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:
- Average AUC: Increased from 0.769 (unaided) to 0.798 (with MammoScreen assistance).
- Difference: 0.028 (P = 0.035), indicating a statistically significant improvement.
- The AUC was higher with MammoScreen aid for 11 of the 14 radiologists.
- Performance improvement was also statistically significant at the breast (in terms of AUC) and lesion (in terms of pAUC) level.
-
Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
- Yes, a standalone performance study was conducted.
- Standalone Performance: MammoScreen's standalone performance (AUC = 0.790) was found to be non-inferior to the average performance of unaided radiologists (AUC = 0.770). The lower confidence interval of the difference of AUC was equal to or superior to the effect size (-0.03), and the P-value was >0.05, confirming non-inferiority.
- Detailed standalone performance metrics were also provided for mammogram, breast, and finding levels (soft tissue lesions and calcifications), including ROC AUC, sensitivity, and specificity for Hologic, GE, and combined datasets.
-
Type of Ground Truth Used:
- Clinical Outcomes Data: The primary ground truth was established by:
- Biopsy results (for all cancer-positive cases and some negative cases).
- Adequate follow-up (for negative cases only).
- Clinical Outcomes Data: The primary ground truth was established by:
-
Sample Size for the Training Set:
- The document states that the algorithms "have been trained on large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue." However, it does not specify the exact sample size of the training set.
-
How the Ground Truth for the Training Set Was Established:
- The ground truth for the training set was established using "biopsy proven examples of breast cancer, benign lesions and normal tissue." This implies a similar methodology to the test set, relying on objective clinical outcomes (histopathology from biopsy) rather than expert consensus on images.
{0}------------------------------------------------
March 25, 2020
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Therapixel % Ms. Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroihet Drive HILLSBOROUGH CA 94010
Re: K192854
Trade/Device Name: MammoScreen Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: February 10, 2020 Received: February 11, 2020
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K192854
Device Name MammoScreen
Indications for Use (Describe)
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K192854
This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR § 807.92.
Applicant Information:
510(k) Owner: Therapixel Village By CA - Le Theseus, Rue Claude Daunesse, 06560 Valbonne France Phone: +33 9 72 55 20 39
Submission Correspondent:
Cindy Domecus, R.A.C. (US & EU) Regulatory Consultant to Therapixel Phone: 650.343.4813 Fax: 650.343.7822 Email: domecusconsulting@comcast.net
Additional Contact:
Quentin de Snoeck RA/QA/CA Manager at Therapixel Phone: +33 9 72 55 20 39 Email: qdesnoeck(@therapixel.com
Date Summary Prepared: March 18th, 2020
{4}------------------------------------------------
Device Information:
| Trade Name: | MammoScreen |
|---|---|
| Common Name: | Computer-Assisted Detection Device |
| Device Classification Name: | Radiological Computer Assisted Detection/Diagnosis Software ForLesions Suspicious For Cancer |
| Regulation Number: | 892.2090 |
| Regulation Class: | Class II |
| Product Code: | QDQ |
| Submission type | Traditional 510(k) |
Predicate Device:
The predicate device is Transpara™, cleared under K181704.
Device Description:
MammoScreen is a software-only device for aiding interpreting physicians in identifying focal findings suspicious for breast cancer in screening FFDM (full-field digital mammography) acquired with compatible mammography systems. The product consists of a processing server and a web interface. The software applies algorithms for recognition of suspicious calcifications and soft tissue lesions. These algorithms have been trained on large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue.
MammoScreen automatically processes FFDM and the output of the device can be used by radiologists concurrently with the reading of mammograms. The user interface of MammoScreen has several functions:
a) Activation of computer aided detection (CAD) marks to highlight locations, known as findings, where the device detected calcifications or soft tissue lesions suspicious for cancer.
b) Association of findings with a score, known as the MammoScreen Score, which characterizes findings on a 1-10 scale, with increasing level of suspicion. Only the most suspicious findings (with a MammoScreen score equal or greater than 5) are initially marked to limit the number of findings to review. The user shall also review findings with score of 4 or lower.
c) Indication, with matching markers, when findings corresponding to the same findings are detected in multiple views of the FFDM.
MammoScreen is configured as a DICOM Web compliant node in a network and receives its input images from another DICOM node, called "the DICOM Web Server". The MammoScreen output will be displayed on the screen of a personal computer compliant with requirements specified in the User Manual.
The image analysis unit includes machine learning components trained to detect positive findings (calcifications and soft tissue lesions).
{5}------------------------------------------------
Indication for Use:
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
Intended patient population
The device is intended to be used in the population of women undergoing screening FFDM.
Warnings and precautions
Patient management decisions should not be made solely on the basis of analysis by MammoScreen.
| Subject DeviceMammoScreen | Predicate DeviceTranspara™K181704 | SubstantiallyEquivalent? | |
|---|---|---|---|
| ClassificationRegulation | 21 CFR 892.2090Radiological Computer AssistedDetection And Diagnosis Software | SAME | Yes, identical. |
| MedicalDeviceClassification | Class II | SAME | Yes, identical. |
| Product Code | QDQ | SAME | Yes, identical. |
| Level ofConcern | Moderate | SAME | Yes, identical. |
| Intended Use | A concurrent reading aid forphysicians interpreting screeningFFDM acquired with compatiblemammography systems, to identifyfindings and assess their level ofsuspicion. | SAME | Yes, identical. |
| Target patientpopulation | Women undergoing FFDM screening mammography | SAME | Yes, identical. |
Predicate and Subject Device Comparison:
{6}------------------------------------------------
| Subject Device | Predicate Device | SubstantiallyEquivalent? | |
|---|---|---|---|
| MammoScreen | Transpara™ | ||
| K181704 | |||
| Target userpopulation | Physicians interpreting FFDMscreening mammograms | SAME | Yes, identical |
| Design | Software-only device | SAME | Yes, identical |
| Indication forUse | MammoScreen™ is intended foruse as a concurrent reading aid forinterpreting physicians, to helpidentify findings on screeningFFDM acquired with compatiblemammography systems and assesstheir level of suspicion. Output ofthe device includes marks placed onfindings on the mammogram andlevel of suspicion scores. Thefindings could be soft tissue lesionsor calcifications. The level ofsuspicion score is expressed at thefinding level, for each breast andoverall for the mammogram. Patientmanagement decisions should notbe made solely on the basis ofanalysis by MammoScreen™. | The ScreenPoint Transpara™system is intended for use as aconcurrent reading aid forphysicians interpretingscreening mammograms, toidentify regions suspicious forbreast cancer and assess theirlikelihood of malignancy.Output of the device includesmarks placed on suspicioussoft tissue lesions andsuspicious calcifications;region-based scores, displayedupon the physician's query,indicating the likelihood thatcancer is present in specificregions; and an overall scoreindicating the likelihood thatcancer is present on themammogram.Patient management decisionsshould not be made solely onthe basis of analysis byTranspara™. | Yes, identical |
{7}------------------------------------------------
| Subject Device | Predicate Device | SubstantiallyEquivalent? | |
|---|---|---|---|
| MammoScreen | Transpara™K181704 | ||
| Score | Finding level:10-point scale score indicating thelevel of suspicion of malignancy(from low suspicion to highsuspicion).Breast level:The same 10-point scale score asfinding level. The score of a breastis equal to the maximum score ofthe findings detected in this breast. | Finding level:Continuous score 1-100indicating the level ofsuspicion of malignancy (fromlow suspicion to highsuspicion).Breast level:None | Both scores aresubstantiallyequivalent.Both scores increasewith the level ofsuspicion. Theminimum (resp. themaximum) of the bothscores describes thesame status.At the Exam level,both scores have a 10-point scale. |
| Exam level:Exam-level of suspicion resultingdirectly from the maximum score ofboth breasts (1-to-1 mappingbetween the score and the exam-level of suspicion). | Exam level:10-point scale score indicativeof higher frequency of cancerpositive | ||
| Findingdiscovery | Findings are by-default displayedwhen score is equal or higher to 5.Upon user request for findings ofscore equal or less to 4. | Upon user request by clickingin a position of the image alsodetected by Transpara™. | Both are thedemonstration of thesame intention:reducing the numberof findings the userhas to review. In thissense, both areequivalent. |
{8}------------------------------------------------
| Subject Device | Predicate Device | Substantially | |
|---|---|---|---|
| MammoScreen | Transpara™ | Equivalent? | |
| K181704 | |||
| Performances | Reader study: | Reader study: | |
| - 240 cases- 14 radiologists | - 240 cases- 14 radiologists | Despite a slightlyhigher performance ofMammoScreencompared to | |
| Reading time (two sessions mean): | Reading time: | Transpara™, gains | |
| - 57,67 seconds (unaidedsession)- 64, 13 seconds (withMammoScreen) | - 146 seconds (unaidedsession)- 149 seconds (withTranspara™) | are still comparableand do not raise newquestions regardingsafety andeffectiveness of thedevice. | |
| AUC: | AUC: | ||
| - radiologists AUC(unaided) = 0,769- standalone AUC = 0,786 | - radiologists AUC(unaided) = 0,866- standalone AUC =0,887 | ||
| Features | Distinguishes two types ofsuspicious findings (calcificationsand soft tissue lesions). | Distinguishes two types ofsuspicious findings(calcifications and soft tissuelesions). | Despite somedifferences betweenthe predicate deviceand MammoScreen, |
| The CAD output provided by theserver includes the location and theoutline of findings. | The CAD output provided bythe server includes the locationand the outline of findings. | features are stillcomparable and donot raise newquestions regardingsafety and | |
| MammoScreen processing server isa standalone system WITH a userinterface. | Transpara™ processing serveris a standalone systemWITHOUT a user interface. | effectiveness of thedevice. | |
| ImagingModality | FFDM | SAME | Yes, identical |
{9}------------------------------------------------
| Subject DeviceMammoScreen | Predicate DeviceTranspara™ | SubstantiallyEquivalent? | |
|---|---|---|---|
| K181704 | |||
| Fundamentalscientifictechnology | In MammoScreen, a range ofmedical image processing andmachine learning techniques areimplemented. The system includes'deep learning' modules forrecognition of suspiciouscalcifications and soft tissuelesions. These modules are trainedwith very large databases of biopsy-proven examples of breast cancerand normal tissue. | SAME | Yes, identical |
Table 7.1 Comparison between subject and predicate device
Discussion:
The indication for use of MammoScreen is very similar to that of the predicate device. Both devices are intended to be used by clinicians interpreting digital mammogram images, to help them with localizing and characterizing suspicious findings. The devices are both intended to be used concurrently with the reading of images and are not intended as a replacement for the review of a clinician or their clinical judgement. Both devices target the same findings, the same patient and user populations and use the same imaging modality. Also, both devices have a Moderate Level of Concern.
Both devices give a score. The scoring system is not the same, but is based on similar principles: providing users with a level of suspicion for malignancy from low (score 1) to high (score 10 or 100). TransparaTM exhibits some minor differences though:
- at the finding level, Transpara™ uses a 1-100 score of Transpara™ while the MammoScreen score has a 10-point scale. Indeed, Therapixel believes, based on questionnaire submitted to 15 radiologists, that interpretability of less granular scale is easier for users (how shall a difference between a score of 41 and 43 be interpreted for instance?)
- at the breast level. Transpara™ does not provide a score. Therapixel considers that this level is required for better interpretability at exam level, scores are equivalent. However, using Transpara™, the user cannot determine with precision to which of the breast (if not both), and where in the breast(s) this overall score applies. On the contrary, Therapixel's score (and the indication of the exam-level of suspicion that directly results from it) is directly connected to one of the breasts (or both) and indicated as such, and exactly where in the breast (finding) thanks to the scoring consistency (a breast inherits from the maximum score of its findings, and the exam inherits from the maximum score of both breasts). Doing so makes the algorithm decision more explicit, and easier to interpret from a user point of view.
{10}------------------------------------------------
Both scores increase with the level of suspicion. The minimum (resp. the maximum) of the both scores describes the same status. At the Exam level, both scores have a 10-point scale.
Both scores are substantially equivalent, the minor differences do not raise further or different questions of performance or safety.
In conclusion, these differences do not raise different questions of safety and effectiveness of the device when used as labeled. The overall design of MammoScreen is very similar to that of the predicate device. Both devices have the same intended use, very similar indications for use, and minor differences between MammoScreen and the predicate device do not raise different questions of safety and effectiveness which are equivalent.
Non-clinical Testing
MammoScreen is a software-only device. The level of concern for the device is determined as Moderate Level of Concern.
Tests have been performed in compliance with the following recognized consensus standards:
- IEC 62304 Edition 1.1 2015-06 Medical device software Software life-cycle processes ●
- IEC 62366-1 Edition 1.0 2015-02 Medical devices Application of usability engineering ● to medical devices.
MammoScreen has successfully completed integration and verification testing and beta validation. In addition, potential hazards have been evaluated and mitigated, and have acceptable levels.
{11}------------------------------------------------
Standalone Performance Testing
Standalone performance testing of MammoScreen assesses the performance of MammoScreenalgorithms in the absence of a clinician and includes mammograms of women acquired with devices from two manufacturers: Hologic® and GE®.
The following graph shows the ratio of positive cases over all cases processed by both GE® and Hologic® devices that fall into each score:
Image /page/11/Figure/3 description: The image is a bar graph titled "Repartition by MS score (combined)". The x-axis is labeled "MammoScreen score" and ranges from 1 to 10. The y-axis is labeled "Repartition of positive cases" and ranges from 0.0 to 1.0. The bar graph shows an increase in the repartition of positive cases as the MammoScreen score increases.
Figure 7.2.-Repartition of positive cases per MammoScreen scores for Hologic® and GE® combined
Mammogram level
Standalone performance of the MammoScreen-AI algorithm in characterizing positive and negative US FFDM acquired on Hologic® devices, and performance comparison with FFDM acquired on GE® devices, are given in the table below:
| Hologic® | GE® | Combined | FOMGE-FOMHOL | |
|---|---|---|---|---|
| ROCAUC | 0.868 (0.851, 0.885) | 0.887 (0.875, 0.898) | 0.883 (0.873, 0.892) | 0.018 (-0.002, 0.039) |
{12}------------------------------------------------
| Sensitivity | 0.844 (0.815, 0.872) | 0.849 (0.827, 0.871) | 0.847 (0.829, 0.864) | 0.005 (-0.032, 0.042)1 |
|---|---|---|---|---|
| Specificity | 0.705 (0.689, 0.722) | 0.738 (0.728, 0.747) | 0.729 (0.721, 0.737) | 0.032 (0.014, 0.051) |
Image /page/12/Figure/1 description: The image is a Receiver Operating Characteristic (ROC) curve, which is a graphical plot that illustrates the diagnostic ability of a binary classifier system as its discrimination threshold is varied. The plot shows the true positive rate (y-axis) versus the false positive rate (x-axis). There are three ROC curves plotted: AUC Hologic = 0.868, AUC GE = 0.887, and AUC Combined = 0.882, along with data points labeled MS6.
Figure 7.3: ROC curves of MammoScreen-AI algorithm at mammogram level on US Hologic® FFDM (blue), GE® FFDM (orange), and on both types combined (green). Sensitivity and specificity at the chosen standalone regime (MS6) are shown with a blue dot for each case.
Breast level
Standalone performance of the MammoScreen-AI algorithm in characterizing positive and negative breasts on US FFDM acquired on acquired on Hologic® devices, and performance comparison with FFDM acquired on GE® devices, are given in the table below:
| Hologic® | GE® | Combined | FOMGE-FOMHOL | |
|---|---|---|---|---|
| ROCAUC | 0.901 (0.886, 0.915) | 0.916 (0.906, 0.926) | 0.911 (0.902, 0.919) | 0.015 (-0.002, 0.033) |
| Sensitivity | 0.813 (0.781, 0.843) | 0.830 (0.808, 0.853) | 0.823 (0.805, 0.841) | 0.017 (-0.023, 0.056) |
| Specificity | 0.840 (0.831, 0.848) | 0.846 (0.840, 0.851) | 0.844 (0.839, 0.849) | 0.006 (-0.004, 0.016) |
4 Due to the lower bound of the 95% Cl being close to the non-inferiority margin (-0.03), 10,000 bootstrap replicates were used instead of 2,000.
{13}------------------------------------------------
Image /page/13/Figure/0 description: The image shows a receiver operating characteristic (ROC) curve. The ROC curve plots the true positive rate against the false positive rate. There are three ROC curves plotted on the graph, representing different models: AUC Hologic = 0.901, AUC GE = 0.916, and AUC Combined = 0.911. The ROC curves are all relatively close to each other, indicating that the models have similar performance.
Figure 7.4: ROC curves of ManmoScreen-AI algorithm at breast level on US Hologic® FFDM (blue), GE® FFDM (orange), and on both types combined (green). Sensitivity at the chosen standalone regime (MSG) are shown with a blue dot for each case.
Finding level
Standalone performance of the MammoScreen-AI algorithm in detecting and characterizing positive findings (soft tissue lesions and calcifications) on FFDM acquired on Hologic® devices, and comparison with FFDM acquired on GE® devices, are given in the tables below.
Two characteristics are considered: The Free-Response ROC (FROC) and the Localized ROC (LROC).
FROC curve shows the sensitivity of finding detection and characterization as a function of the average number of false marks per image. It illustrates how performant an algorithm is at detecting and characterizing positive findings while keeping the number of false marks as low as possible. The drawback of FROC curves is that the x-axis has virtually no limit and that the curve area is not bound to unit square. Therefore, no simple unique measure (such as the AUC) may be derived from FROC curves to compare between several of such curves. However, it is still valuable information as it accounts for the average number of false marks per image.
LROC curves are generally used when only one positive finding per case shall be localized. It shows the sensitivity of correctly marking a positive case (a case is considered as correctly marked
{14}------------------------------------------------
if the algorithm placed a mark at a reasonable distance from the ground truth location – here in a 15mm radius from the ground truth) as a function of the false positive rate (a negative case is considered a false positive if the algorithm has placed one or more marks on it). LROC curve areas are bound to unit square, allowing to derive single measures to describe them such as AUC.
For MammoScreen, LROC curves at the breast level are used. This allows to measure how performant the algorithm is at marking the correct finding within a breast (in either of the 2 views of the breast) as a function of the false positive breast rate (i.e., rate of placing false marks in either of the 2 views of a negative breast).
In addition to FROC and LROC analysis, the performance of four detection Operating Points (OP) is reported. Those correspond to the three OP made corresponding to visualization levels made available to users via the MammoScreen user interface, using the Filtering slider (refer to section 5.3.1.1 Filtering of findings) and corresponding to MammoScreen scores 1, 3 and 5 respectively.
| MammoScreen Score | MS1 | MS3 | MS5 |
|---|---|---|---|
| Filteringsliderposition | Left (LowestSuspicion shown) | Middle (LowestSuspicion hidden) | Right (Low Suspicionhidden) |
| Soft tissue lesions | |||||
|---|---|---|---|---|---|
| Hologic® | GE® | Combined | FOMGE-FOMHOL | ||
| LROC | AUC (primary) | 0.837 (0.811, 0.861) | 0.900 (0.884, 0.916) | 0.877 (0.862, 0.890) | 0.064 (0.034, 0.095) |
| Sensitivity @ MS1 | 0.942 (0.921, 0.963) | 0.976 (0.964, 0.987) | 0.962 (0.951, 0.973) | 0.034 (0.010, 0.060) | |
| Sensitivity @ MS3 | 0.926 (0.902, 0.949) | 0.966 (0.951, 0.978) | 0.950 (0.937, 0.963) | 0.039 (0.012, 0.067) | |
| Sensitivity @ MS5 | 0.774 (0.735, 0.811) | 0.852 (0.824, 0.879) | 0.822 (0.799, 0.843) | 0.080 (0.032, 0.128) | |
| Specificity @ MS1 | 0.109 (0.102, 0.116) | 0.166 (0.161, 0.172) | 0.150 (0.146, 0.155) | 0.058 (0.048, 0.067) | |
| Specificity @ MS3 | 0.233 (0.222, 0.242) | 0.222 (0.216, 0.228) | 0.225 (0.219, 0.230) | -0.011 (-0.022, 0.002) | |
| Specificity @ MS5 | 0.780 (0.771, 0.791) | 0.801 (0.795, 0.807) | 0.795 (0.790, 0.800) | 0.021 (0.010, 0.032) | |
| FROC | Sensitivity @ MS1 | 0.895 (0.872, 0.915) | 0.952 (0.940, 0.963) | 0.930 (0.918, 0.940) | 0.057 (0.033, 0.081) |
| Sensitivity @ MS3 | 0.878 (0.854, 0.901) | 0.943 (0.930, 0.955) | 0.918 (0.906, 0.929) | 0.064 (0.038, 0.090) | |
| Sensitivity @ MS5 | 0.750 (0.718, 0.782) | 0.835 (0.814, 0.855) | 0.802 (0.784, 0.819) | 0.084 (0.047, 0.121) | |
| Avg false marks @ MS1 | 1.819 (1.815, 1.824) | 1.268 (1.266, 1.270) | 1.424 (1.422, 1.426) | NA | |
| Avg false marks @ MS3 | 1.182 (1.171, 1.193) | 1.026 (1.021, 1.031) | 1.070 (1.066, 1.075) | NA | |
| Avg false marks @ MS5 | 0.294 (0.286, 0.302) | 0.218 (0.214, 0.222) | 0.239 (0.235, 0.243) | NA |
{15}------------------------------------------------
| Calcifications | |||||
|---|---|---|---|---|---|
| Hologic® | GE® | Combined | FOMGE-FOMHOL | ||
| LROC | AUC (primary) | 0.974 (0.959, 0.985) | 0.930 (0.912, 0.948) | 0.942 (0.928, 0.954) | -0.044 (-0.067, -0.021) |
| Sensitivity @ MS1 | 0.994 (0.979, 1.000) | 0.971 (0.953, 0.987) | 0.978 (0.964, 0.989) | -0.023 (-0.043, -0.001) | |
| Sensitivity @ MS3 | 0.994 (0.979, 1.000) | 0.971 (0.953, 0.987) | 0.978 (0.964, 0.989) | -0.023 (-0.043, -0.001) | |
| Sensitivity @ MS5 | 0.962 (0.930, 0.988) | 0.804 (0.764, 0.844) | 0.851 (0.820, 0.879) | -0.158 (-0.209, -0.108) | |
| Specificity @ MS1 | 0.549 (0.537, 0.561) | 0.645 (0.638, 0.652) | 0.619 (0.613, 0.625) | 0.096 (0.083, 0.110) | |
| Specificity @ MS3 | 0.584 (0.573, 0.597) | 0.658 (0.652, 0.665) | 0.638 (0.632, 0.644) | 0.074 (0.061, 0.088) | |
| Specificity @ MS5 | 0.869 (0.861, 0.878) | 0.909 (0.905, 0.913) | 0.898 (0.894, 0.901) | 0.040 (0.031, 0.048) | |
| FROC | Sensitivity @ MS1 | 0.994 (0.984, 1.000) | 0.941 (0.923, 0.956) | 0.957 (0.944, 0.969) | -0.052 (-0.072, -0.033) |
| Sensitivity @ MS3 | 0.994 (0.984, 1.000) | 0.938 (0.920, 0.955) | 0.955 (0.941, 0.968) | -0.055 (-0.075, -0.036) | |
| Sensitivity @ MS5 | 0.942 (0.916, 0.968) | 0.779 (0.747, 0.809) | 0.828 (0.805, 0.851) | -0.163 (-0.202, -0.124) | |
| Avg false marks @ MS1 | 0.870 (0.867, 0.873) | 0.531 (0.529, 0.532) | 0.627 (0.625, 0.628) | NA | |
| Avg false marks @ MS3 | 0.753 (0.748, 0.759) | 0.479 (0.477, 0.482) | 0.557 (0.554, 0.559) | NA | |
| Avg false marks @ MS5 | 0.238 (0.232, 0.245) | 0.122 (0.119, 0.125) | 0.155 (0.152, 0.158) | NA |
Image /page/15/Figure/1 description: The image contains two ROC curves, one titled "Breast Localized ROC (LROC)" and the other titled "Image Free-Response ROC (FROC)". Each plot shows three curves representing different methods: Hologic, GE, and Combined, with corresponding AUC values of 0.837, 0.900, and 0.877 respectively. The plots also include data points for MS1, MS3, MS5, and MS6, represented by red, green, blue, and yellow markers, respectively, on each curve.
Figure 7.5: LROC (left) and FROC (right) curves on US soft tissue lesions only.
{16}------------------------------------------------
Image /page/16/Figure/0 description: The image contains two ROC curves, one titled "Breast Localized ROC (LROC)" and the other titled "Image Free-Response ROC (FROC)". Each ROC curve plots the true positive rate against the false positive rate for different models. The LROC plot shows the AUC for Hologic is 0.974, GE is 0.930, and Combined is 0.941. Both plots also show data points for MS1, MS3, MS5, and MS6.
Figure 7.6: LROC (left) and FROC (right) curves on US calcifications only.
{17}------------------------------------------------
Per-MammoScreen score analysis
Mammogram level
Image /page/17/Figure/2 description: This image contains two bar charts comparing Hollogic and GE mammogram scores. The left chart shows the proportion of screening mammograms per score, while the right chart shows the probability of positive mammograms per score. The MammoScreen score ranges from 1 to 10 on the y-axis of both charts. The proportion of screening mammograms per score is highest for Hollogic at a score of 4, with 33.744%, while GE is highest at a score of 4, with 37.931%.
Image /page/17/Figure/3 description: The image contains text describing figure 7.8. The figure shows the repartition of screening FFDM per MammoScreen score at the mammogram level. The left side of the figure shows the repartition of screening FFDM per score for Hologic and GE manufacturers. The right side of the figure shows the probability of discovering a positive FFDM per score.
{18}------------------------------------------------
Breast level
Image /page/18/Figure/1 description: This image contains two horizontal bar charts comparing the performance of Hologic and GE mammography systems. The left chart shows the proportion of screening mammograms per score, while the right chart displays the probability of positive mammograms per score, with scores ranging from 1 to 10. The proportion of screening mammograms per score is highest for both systems at score 4, with Hologic at 30.729% and GE at 33.821%, while the probability of positive mammograms per score is 100% for both systems at score 10.
Image /page/18/Figure/2 description: The image shows the title of a figure, "Figure 7.9: Repartition of screening FFDM per MammoScreenTM score at breast level." There are two descriptions below the title. The first description is "Left: Repartition of screening FFDM per score for Hologic® and GE® manufacturers." The second description is "Right: Probability of discovering a positive FFDM per score."
{19}------------------------------------------------
Table 7.2: NPV, specificity, NPV, sensitivity at mannoscreen™ scores on a similated screening distribution (9% confidence interval in brackets).
Hologic®
| FOM | MS1 | MS2 | MS3 | MS4 | MS5 | MS6 | MS7 | MS8 | MS9 | MS10 |
|---|---|---|---|---|---|---|---|---|---|---|
| NPV | 1.0000(1.0000 - 1.0000) | 1.0000(1.0000 - 1.0000) | 0.9991(0.9987 - 0.9995) | 0.9984(0.9977 - 0.9991) | 0.9979(0.9970 - 0.9987) | 0.9956(0.9946 - 0.9964) | 0.9939(0.9927 - 0.9948) | 0.9928(0.9916 - 0.9938) | 0.9922(0.9911 - 0.9932) | 0.9922(0.9911 - 0.9932) |
| Specificity | 0.0162(0.0129 - 0.0196) | 0.0460(0.0390 - 0.0533) | 0.2175(0.2051 - 0.2309) | 0.5632(0.5424 - 0.5805) | 0.7185(0.7015 - 0.7332) | 0.9690(0.9627 - 0.9739) | 0.9942(0.9922 - 0.9968) | 0.9987(0.9968 - 0.9998) | 1.0000(1.0000 - 1.0000) | 1.0000(1.0000 - 1.0000) |
| PPV | 0.0078(0.0068 - 0.0089) | 0.0079(0.0069 - 0.0091) | 0.0082(0.0071 - 0.0094) | 0.0097(0.0084 - 0.0111) | 0.0158(0.0134 - 0.0180) | 0.0221(0.0185 - 0.0250) | 0.1049(0.0863 - 0.1267) | 0.2443(0.1804 - 0.3734) | 0.3788(0.1662 - 0.7885) | 1.0000(1.0000 - 1.0000) |
| Sensitivity | 1.0000(1.0000 - 1.0000) | 1.0000(1.0000 - 1.0000) | 1.0000(1.0000 - 1.0000) | 0.9764(0.9617 - 0.9862) | 0.8882(0.8397 - 0.9302) | 0.8073(0.7331 - 0.8737) | 0.4584(0.4040 - 0.5177) | 0.2294(0.1956 - 0.2693) | 0.0844(0.0643 - 0.1036) | 0.0000(0.0000 - 0.0000) |
GE®
| FOM | MS1 | MS2 | MS3 | MS4 | MS5 | MS6 | MS7 | MS8 | MS9 | MS10 |
|---|---|---|---|---|---|---|---|---|---|---|
| NPV | 1.0000 | 1.0000 | 0.9996 | 0.9992 | 0.9988 | 0.9973 | 0.9965 | 0.9955 | 0.9953 | 0.9952 |
| (1.0000 - 1.0000) | (1.0000 - 1.0000) | (0.9994 - 0.9997) | (0.9987 - 0.9994) | (0.9982 - 0.9991) | (0.9969 - 0.9978) | (0.9960 - 0.9969) | (0.9950 - 0.9960) | (0.9947 - 0.9958) | (0.9947 - 0.9958) | |
| Specificity | 0.0409 | 0.0610 | 0.4238 | 0.6339 | 0.7741 | 0.9808 | 0.9973 | 1.0000 | 1.0000 | 1.0000 |
| (0.0372 - 0.0448) | (0.0558 - 0.0655) | (0.4108 - 0.4338) | (0.6232 - 0.6442) | (0.7659 - 0.7822) | (0.9786 - 0.9838) | (0.9962 - 0.9982) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | |
| PPV | 0.0048 | 0.0050 | 0.0051 | 0.0079 | 0.0115 | 0.0166 | 0.1015 | 0.3154 | 1.0000 | 1.0000 |
| (0.0042 - 0.0053) | (0.0044 - 0.0055) | (0.0045 - 0.0056) | (0.0070 - 0.0088) | (0.0103 - 0.0127) | (0.0148 - 0.0184) | (0.0861 - 0.1235) | (0.2385 - 0.4166) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | |
| Sensitivity | 1.0000 | 1.0000 | 1.0000 | 0.9626 | 0.8936 | 0.7992 | 0.4513 | 0.2579 | 0.0516 | 0.0078 |
| (1.0000 - 1.0000) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (0.9529 - 0.9731) | (0.8288 - 0.9233) | (0.7325 - 0.8514) | (0.4054 - 0.5004) | (0.2236 - 0.2983) | (0.0398 - 0.0635) | (0.0035 - 0.0130) |
{20}------------------------------------------------
Clinical Testing
This investigation was a retrospective multi reader multi case study meant to compare cancer detection performance of radiologists reading with the aid of MammoScreen to the reading results of the same cohort of radiologists without any decision support.
The main objective of this investigation was to determine whether the radiologist performance when using MammoScreen is superior to unaided radiologist performance for interpretation of 2D Full Field Digital Screening Mammograms.
To do this, 240 mammographic screening images acquired at a US center have been collected. For each exam, the cancer status has been verified by either biopsy results (for all cancer positive cases and some of the negative cases) or an adequate follow-up (for negative cases only) and used as gold standard. The images have been read by 14 radiologists with and without the aid of MammoScreen, the interpretation of the standalone MammoScreen has been recorded as well. Finally, the algorithm's interpretation, the radiologist interpretation using the help of the algorithm and the unaided radiologist interpretation of every mammogram have been compared to the "gold standard". Those pairwise comparisons, have been used to numerically compute the primary and secondary endpoints (AUC under the ROC curve, Sensitivity and Specificity) to compare radiologist performance when using the system and without using it and standalone algorithm performance.
The performance characteristics of radiologists taking part to the clinical investigation was improved when using MammoScreen support, with the average AUC going from 0.77 to 0.8 (difference = 0.028; P = 0.035) (Figure 7.10.1). The AUC was higher with the aid of MammoScreen for 11 of the 14 radiologists (Figure 7.10.2).
Image /page/20/Figure/5 description: The image is a plot of the receiver operating characteristic (ROC) curve. The x-axis is the false positive rate, and the y-axis is the true positive rate. There are two ROC curves plotted on the graph, one for the unassisted case and one for the assisted case. The area under the curve (AUC) for the unassisted case is 0.769, and the AUC for the assisted case is 0.798.
Figure 7.10.1– Average ROC curves of all readers when unassisted (blue) and assisted (orange) with MammoScreen.
{21}------------------------------------------------
Image /page/21/Figure/0 description: The image contains two receiver operating characteristic (ROC) curves, one labeled "unassisted" and the other "assisted". Each ROC curve plots the true positive rate against the false positive rate for multiple readers and a multi-specialty (MS) reader. The area under the curve (AUC) values for each reader are displayed in the legend, ranging from 0.694 to 0.845 for the unassisted ROC and 0.745 to 0.847 for the assisted ROC, with the AUC for the MS reader being 0.790 in both cases.
Figure 7.10.2- Left: ROC curves of all readers when unassisted. Right: ROC curves of all readers when assisted with MammoScreen.
Performances were also measured at breast and lesion level; the overall performance improvement was found to be statistically significant at both breast (in terms of AUC) and lesion (in terms of pAUC) level confirming the trend of the analysis at mammogram level (Figure 7.10.3).
Image /page/21/Figure/3 description: The image contains two receiver operating characteristic (ROC) curves, labeled 'a' and 'b'. Each plot shows two curves, one for 'AUC unassisted' and another for 'AUC assisted'. In plot 'a', the AUC for unassisted is 0.794 and for assisted is 0.827, while in plot 'b', the AUC for unassisted is 0.592 and for assisted is 0.644. Both plots also include a dashed line representing the line of no discrimination.
Figure 7.10.3 – Average ROC curves of all readers when unassisted (blue) and assisted (orange) with MammoScreen at breast level (a) and lesion level (b).
{22}------------------------------------------------
On average, reading time per case increased when using MammoScreen in both reading sessions. During the first session the average reading time was 60.82 seconds (95% confidence interval: 59.25 seconds, 62.39 seconds) for the unaided reading condition and 68.65 seconds (95% confidence interval: 66.92 seconds, 70.39 seconds) when using MammoScreen. For the second reading session the average time was lower with respect to the first session, 54.52 seconds (95% confidence interval: 52.97 seconds, 56.07 seconds) for the unaided reading condition and 59.61 seconds (95% confidence interval: 58.05 seconds, 61.17 seconds) when using MammoScreen. Differences in reading time with and without the use of MammoScreen changed as a function of the MammoScreen score (P < 0.05): for score equal or lower than 4 during the first reading session radiologists increased their reading time by 1% when using MammoScreen, while during the second reading session the use of MammoScreen made them decrease their reading time by 2%. For score higher than 4 in both reading sessions the use of MammoScreen increased the average reading time of about 14%, the maximum increase in reading time did not exceed 15s.
The standalone analysis of MammoScreen has shown that the algorithm exhibits performances comparable (non-inferior) to those of an experienced radiologist as confirmed by the absence of statistical effect (p>0.05) and by the lower confidence interval of the difference of AUC being equal or superior to the effect size (-0.03). Indeed, the performance of the standalone MammoScreen (AUC = 0.79) was found to be non-inferior to the average performance of unaided radiologists (AUC = 0.77) (Figure 7.10.4).
Image /page/22/Figure/2 description: This image is a receiver operating characteristic (ROC) curve. The plot shows the true positive rate on the y-axis and the false positive rate on the x-axis. There are three ROC curves plotted: one for an unassisted model with an AUC of 0.768, one for an assisted model with an AUC of 0.797, and one for a standalone model with an AUC of 0.790.
Figure 7.10.4- Comparison of unassisted radiologists, assisted radiologists, and standalone MammoScreen ROC curve
{23}------------------------------------------------
Conclusions
Non-clinical and clinical performance tests demonstrate that MammoScreen is safe and effective.
Results of the primary analysis of the clinical test demonstrate that use of MammoScreen improves detection of breast cancer in mammograms. Descriptively, improvement was observed to depend neeligibly on lesion type and the total reading time was not observed to increase with the use of MammoScreen. In addition, the sensitivity of the readers tended to increase with the use of MammoScreen without decreasing specificity.
Radiologists improved their diagnostic performance in the detection of breast cancer with 2D FFDM (Full-Field Digital Mammography) by using MammoScreen. The overall conclusion of this clinical investigation is that the MammoScreen improves the diagnostic performance of radiologists in the detection of breast cancer without slowing down their average reading time. Finally, in standalone testing, MammoScreen breast cancer detection performance was observed to approach the average performance of the clinical study radiologists when reading mammograms unaided.
Based on the Intended Use. Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, MammoScreen has been shown to be substantially equivalent to the cited predicate device.
§ 892.2090 Radiological computer-assisted detection and diagnosis software.
(a)
Identification. A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.