K Number
K192834
Device Name
MECTA Sigma
Manufacturer
Date Cleared
2020-04-26

(207 days)

Product Code
Regulation Number
882.5940
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The MECTA Sigma ECT device is for use in treating catatonia or severe major depressive episode (MDE) associated with major depressive disorder (MDD) or bipolar disorder (BPD) in patients age 13 years or older who are treatment-resistant or who require a rapid response due to the severity of their psychiatric or medical condition.
Device Description
Not Found
More Information

None

Not Found

No
The provided 510(k) summary does not contain any mention of AI, ML, deep neural networks (DNN), image processing, or any other terms typically associated with AI/ML technology in medical devices. The description focuses solely on the intended use of the device for electroconvulsive therapy (ECT).

Yes
The device is indicated for treating medical conditions (catatonia, severe major depressive episode), which aligns with the definition of a therapeutic device.

No
The device's intended use is for "treating catatonia or severe major depressive episode," indicating a therapeutic rather than a diagnostic purpose.

No

The device is described as the "MECTA Sigma ECT device," which strongly suggests a hardware component (an ECT machine). The summary does not provide a device description, but the name itself indicates it is not software-only.

Based on the provided information, the MECTA Sigma ECT device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use describes treating psychiatric conditions (catatonia, severe major depressive episode) in patients. This involves direct treatment of a patient, not the examination of specimens derived from the human body.
  • Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
    • Analyzing biological samples (blood, urine, tissue, etc.)
    • Providing information about a patient's health status based on sample analysis
    • Using reagents or calibrators

IVDs are devices used to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. The MECTA Sigma ECT device is a therapeutic device used to deliver electroconvulsive therapy directly to the patient.

N/A

Intended Use / Indications for Use

The MECTA Sigma ECT device is for use in treating catatonia or severe major depressive episode (MDE) associated with major depressive disorder (MDD) or bipolar disorder (BPD) in patients age 13 years or older who are treatment-resistant or who require a rapid response due to the severity of their psychiatric or medical condition.

Product codes

OGH

Device Description

Not Found

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

13 years or older

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Not Found

Key Metrics

Not Found

Predicate Device(s)

Not Found

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 882.5940 Electroconvulsive therapy device.

(a)
Identification. An electroconvulsive therapy device is a prescription device, including the pulse generator and its stimulation electrodes, used for treating severe psychiatric disturbances by inducing in the patient a major motor seizure by applying a brief intense electrical current to the patient's head.(b)
Classification. (1) Class II (special controls) when the device is intended to treat catatonia or a severe major depressive episode (MDE) associated with major depressive disorder (MDD) or bipolar disorder (BPD) in patients age 13 years and older who are treatment-resistant or who require a rapid response due to the severity of their psychiatric or medical condition. The special controls for this device are:(i) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge and energy, and the type of impedance monitoring system must be fully characterized to ensure that the device performance characteristics are consistent with existing clinical performance data.
(ii) Non-clinical testing data must confirm the electrical characteristics of the output waveform.
(iii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iv) Performance data must demonstrate electrical and mechanical safety and the functioning of all safety features built into the device including the static and dynamic impedance monitoring system.
(v) Appropriate analysis/testing must validate electromagnetic compatibility.
(vi) Appropriate software verification, validation, and hazard analysis must be performed.
(vii) Performance data must demonstrate electrical performance, adhesive integrity, and physical and chemical stability of the stimulation electrodes.
(viii) The labeling for the device must include the following:
(A) Information related to generic adverse events associated with electroconvulsive therapy device (ECT) treatment;
(B) Instructions must contain the following specific recommendations to the user of the device:
(
1 ) Conduct of pre-ECT medical and psychiatric assessment (including pertinent medical and psychiatric history, physical examination, anesthesia assessment, dental assessment, and other studies as clinically appropriate);(
2 ) Use of patient monitoring during the procedure;(
3 ) Use of general anesthesia and neuromuscular blocking agents;(
4 ) Use of mouth/dental protection during the procedure;(
5 ) Use of EEG monitoring until seizure termination;(
6 ) Instructions on electrode placement, including adequate skin preparation and use of conductive gel; and(
7 ) Cognitive status monitoring prior to beginning ECT and during the course of treatment via formal neuropsychological assessment for evaluating specific cognitive functions (e.g., orientation, attention, memory, executive function).(C) Clinical training needed by users of the device;
(D) Information on the patient population in which the device is intended to be used;
(E) Information on how the device operates and the typical course of treatment;
(F) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(G) A detailed summary of the device technical parameters;
(H) Where appropriate, validated methods and instructions for reprocessing of any reusable components;
(I) The following statement, prominently placed: “Warning: ECT device use may be associated with: disorientation, confusion, and memory problems”; and
(J) Absent performance data demonstrating a beneficial effect of longer term use, generally considered treatment in excess of 3 months, the following statement, prominently placed: “Warning: When used as intended this device provides short-term relief of symptoms. The long-term safety and effectiveness of ECT treatment has not been demonstrated.”
(ix) Patient labeling must be provided and include:
(A) Relevant contraindications, warnings, precautions;
(B) A summation of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(C) Information on how the device operates and the typical course of treatment;
(D) The potential benefits;
(E) Alternative treatments;
(F) The following statement, prominently placed: “Warning: ECT device use may be associated with: Disorientation, confusion, and memory problems”;
(G) Absent performance data demonstrating a beneficial effect of longer term use, generally considered treatment in excess of 3 months, the following statement, prominently placed: “Warning: When used as intended this device provides short-term relief of symptoms. The long-term safety and effectiveness of ECT treatment has not been demonstrated”; and
(H) The following statements on known risks of ECT, absent performance data demonstrating that these risks do not apply:
(
1 ) ECT treatment may be associated with disorientation, confusion and memory loss, including short-term (anterograde) and long-term (autobiographical) memory loss following treatment. Based on the majority of clinical evidence, these side effects tend to go away within a few days to a few months after the last treatment with ECT. Although the incidence of permanent cognitive memory loss was not supported by the clinical literature, some patients have reported a permanent loss of memories of personal life events (i.e., autobiographical memory);(
2 ) Patients treated with ECT may experience manic symptoms (including euphoria and/or irritability, impulsivity, racing thoughts, distractibility, grandiosity, increased activity, talkativeness, and decreased need for sleep) or a worsening of the psychiatric symptoms they are being treated for; and(
3 ) The physical risks of ECT may include the following (in order of frequency of occurrence):(
i ) Pain/somatic discomfort (including headache, muscle soreness, and nausea);(
ii ) Skin burns;(
iii ) Physical trauma (including fractures, contusions, injury from falls, dental and oral injury);(
iv ) Prolonged or delayed onset seizures;(
v ) Pulmonary complications (hypoxemia, hypoventilation, aspiration, upper-airway obstruction);(
vi ) Cardiovascular complications (cardiac arrhythmias, heart attack, high or low blood pressure, and stroke); and(
vii ) Death.(2)
Classification: Class III (premarket approval) for the following intended uses: schizophrenia, bipolar manic states, schizoaffective disorder, schizophreniform disorder, and catatonia or a severe MDE associated with MDD or BPD in:(i) Patients under 13 years or
(ii) Patients 13 years and older who are not treatment-resistant or who do not require a rapid response due to the severity of their psychiatric or medical condition.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 26, 2019, for any electroconvulsive therapy device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 26, 2019, been found to be substantially equivalent to any electroconvulsive therapy device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other electroconvulsive therapy device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

April 26, 2020

Mecta Corporation Robin Nicol President 19799 SW 95th Ave Tualatin, Oregon 97062

Re: K192834

Trade/Device Name: MECTA Sigma Regulation Number: 21 CFR 882.5940 Regulation Name: Electroconvulsive therapy device Regulatory Class: Class II Product Code: OGH Dated: March 26, 2020 Received: March 27, 2020

Dear Robin Nicol:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K192834

Device Name MECTA Sigma

Indications for Use (Describe)

The MECTA Sigma ECT device is for use in treating catatonia or severe major depressive episode (MDE) associated with major depressive disorder (MDD) or bipolar disorder (BPD) in patients age 13 years or older who are treatment-resistant or who require a rapid response due to the severity of their psychiatric or medical condition.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: