K Number
K192771
Manufacturer
Date Cleared
2020-10-21

(387 days)

Product Code
Regulation Number
878.4300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.

Device Description

The C4 Fiducial Marker consists of a sealed polyether ether ketone (PEEK) polymer tube containing a solution of up to 1% cobalt chloride and up to 2% N-Acetylcysteine, The ratio of cobalt chloride to N-Acetylcysteine is 1:2. Zirconium oxide is sealed within the polymer shell of the device. The length of the polymer capsule is between 5.5 mm and 10.0 mm, and the diameter is 1.0 mm (+/- 0.2 mm).

AI/ML Overview

This document is a 510(k) premarket notification for a medical device, the C4 Fiducial Marker, and does not contain information about a study that measures the device's performance against specific acceptance criteria in the manner requested (e.g., diagnostic accuracy, sensitivity, specificity, or human reader improvement with AI assistance).

Instead, this document describes the device, its intended use, and demonstrates substantial equivalence to previously cleared predicate devices through performance testing related to device integrity and visualization capabilities rather than clinical efficacy metrics.

Here's a breakdown of the information that can be extracted, and what is not available in the provided text:

Information Available:

  • Device Name: C4 Fiducial Marker
  • Intended Use: To be implanted into the body to mark the location of specific anatomy (normal or diseased) for a future medical procedure. The device can be visualized using MRI, CT, or X-ray, and provides a reference for guiding other procedures.
  • Predicate Devices:
    • Mixed media marker (K102506) from Cortex Manufacturing
    • BiomarC Fiducial Marker (K063193) from Carbon Medical Technologies
  • Reference Device: C4 Imaging MRI Marker NS (K171487)
  • Testing Performed (categorized as "Performance Testing"):
    • Structural and functional integrity testing (negative and positive pressure, axial and lateral load testing).
    • MRI, CT, and X-ray imaging demonstrating functional performance.
    • Separate MRI, CT, and X-ray imaging tests performed in tissue equivalent phantoms.
    • Biocompatibility (ISO 10993) assessment.
    • Endotoxin tests.

Missing Information (not detailed in the provided text for this 510(k) summary):

  1. Table of Acceptance Criteria and Reported Device Performance (Clinical/Diagnostic Accuracy): The document does not provide specific acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, localization accuracy in a clinical context) or reported values against such criteria. The "performance testing" described is for device integrity and basic visualization, not clinical effectiveness.
  2. Sample Size used for the test set and data provenance: Not applicable in the context of clinical accuracy or a test set as described for AI evaluation. The device integrity and visualization tests would have used a sample of the manufactured markers, but specific numbers are not provided, nor is information on the provenance of "data" in a clinical sense.
  3. Number of experts used to establish the ground truth for the test set and their qualifications: Not applicable. The "ground truth" for the tests performed would be related to physical properties, not expert consensus on clinical findings.
  4. Adjudication method for the test set: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: No, this type of study was not mentioned. The device is a physical marker, not an AI or imaging interpretation tool that would involve human readers.
  6. Stand-alone (algorithm only without human-in-the-loop performance) study: No, this is not an AI algorithm.
  7. Type of ground truth used: For the device integrity tests, the ground truth would be the physical integrity of the device (e.g., no leaks, ability to withstand loads). For visualization, the ground truth is simply the visual presence and distinctness of the marker on imaging modalities, often evaluated in phantoms. This is not clinical ground truth like pathology or outcomes data.
  8. Sample size for the training set: Not applicable, as this is not an AI algorithm requiring a training set.
  9. How the ground truth for the training set was established: Not applicable.

Summary of what the document implies about "acceptance criteria" and "study":

The "acceptance criteria" for this device (as inferred from the "Performance Testing" section) relate to:

  • Structural and Functional Integrity: The marker must maintain its integrity under various pressures and loads. While specific numerical acceptance criteria (e.g., "withstanding X psi pressure" or "withstanding Y N axial load") are not stated, the tests imply that the device must pass these challenges without failure (e.g., leaking, breaking).
  • Visualization: The marker must be visible on MRI, CT, and X-ray in tissue-equivalent phantoms. The acceptance criterion is likely successful visualization that meets a standard for medical markers (e.g., clear, distinct signal/contrast).
  • Biocompatibility: The materials of the device must be biocompatible according to ISO 10993 standards, and endotoxin levels must be acceptable. The acceptance criteria here would be passing the specified ISO tests and meeting endotoxin limits.

The "study" or "testing" proving these criteria are met consists of the aforementioned structural, functional, visualization, and biocompatibility tests. The document states these tests were performed and implies the device met the necessary standards for substantial equivalence to predicates.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

October 21, 2020

C4 Imaging LLC Andrew Bright President and CEO 196 West Ashland Street Doylestown, Pennsylvania 18901

Re: K192771

Trade/Device Name: C4 Fiducial Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: September 26, 2019 Received: September 30, 2019

Dear Mr. Bright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192771

Device Name C4 Fiducial Marker

Indications for Use (Describe)

The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[as required by section 807.92(c)]

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

General Information

Submitted by:C4 Imaging, LLC196 West Ashland StreetSuite 109Doylestown, PA 18901 USAPhone: +1 (609) 933-5895Email: abright@c4imaging.comContact Person: Andrew Bright
Date Prepared:October 20, 2019

Device Name

Trade Name:C4 Fiducial Marker
Common Name(s):Anatomical tissue marker

Classification

Regulation:21 CFR §878.4300
Class:Class II
Product Code:NEU
Classification name:Marker, Radiographic, Implantable

{4}------------------------------------------------

Predicate Devices

Mixed media markerK102506Cortex Manufacturing
BiomarC Fiducial MarkerK063193Carbon MedicalTechnologies

Device Description

The C4 Fiducial Marker consists of a sealed polyether ether ketone (PEEK) polymer tube containing a solution of up to 1% cobalt chloride and up to 2% N-Acetylcysteine, The ratio of cobalt chloride to N-Acetylcysteine is 1:2. Zirconium oxide is sealed within the polymer shell of the device. The length of the polymer capsule is between 5.5 mm and 10.0 mm, and the diameter is 1.0 mm (+/- 0.2 mm).

Indication

The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.

Comparison to Predicate Devices

The indication and technological characteristics of the C4 Fiducial Marker are substantially equivalent to the Mixed Media Marker (Cortex Manufacturing -K102506) and the BiomarC Fiducial Marker (Carbon Medical Technologies -K063193). Like the Mixed Media Marker, the C4 Fiducial Marker is comprised of a sealed polyether ether ketone (PEEK) capsule that encloses an internal marker for MRI. CT and X-ray imaging. In both cases the PEEK capsule is the only patient contacting material. Like the BiomarC Fiducial, the C4 Fiducial Marker has a sealed biocompatible shell that encloses a zirconium oxide marker.

The C4 Fiducial Marker was also compared to a reference device. the C4 Imaging MRI Marker NS (K171487), which has similar technological characteristics. The C4 Fiducial Marker and the MRI Marker NS both comprise a PEEK shell enclosing a cobalt chloride N-acetyl cysteine solution.

A risk evaluation, based on a Failure Mode Effect Analysis (FEMA), determined that technological differences were not considered to have an impact on the safety or effectiveness of the C4 Fiducial Marker. Adoption evaluation of biocompatibility, MR compatibility, sterilization, structural integrity, packaging and shelf life data during the risk assessment confirmed the C4 Fiducial Marker is substantially equivalent to the predicate devices and reference device.

{5}------------------------------------------------

Performance Testing

The C4 Fiducial Marker was subject to structural and functional integrity testing involving negative and positive pressure, as well as axial and lateral load testing. These integrity tests were followed by MRI, CT and x-ray imaging demonstrating functional performance. Separate MRI, CT and x-ray imaging tests were performed in tissue equivalent phantoms. A biocompatibility (ISO 10993) assessment and endotoxin tests were performed on the device.

Conclusion

The C4 Fiducial Marker is substantially equivalent to the Mixed Media Marker (K102506) and to the BiomarC Fiducial Marker (K063193). There are no differences that would affect the substantial equivalence of the proposed device

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.