K Number
K192715
Device Name
CRF Radiofrequency Ablation System
Date Cleared
2020-01-07

(102 days)

Product Code
Regulation Number
878.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The CRF Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue.
Device Description
The CRF Radiofrequency Ablation System consists of an RF generator, active electrode, grounding pad, and peristallic punp for electrode cooling. This device is designed to cal tissue heating at the tip of the electrode causing coagulation and ablation of tissue. The CRF Generator supplies radiofrequency energy at 480 kHz while monitoring both the tissue imperature at the CRF electrode tip. The CRF Generator monitors the power, resistance, current and temperature during energy delivery to the paient. The CRF Electrodes are sterile, single-use, hand-held for use in radiofrequency ablation procedures. The electrodes are surgically invasive, transient devices in patient contact typically for less than one hour. Cooling of the electrodes is provided by chilled sterile saline or water that is pumped through the electrode and out via outflow tubing. This is an enclosed system within the electrode and the saline does not contact. The CRF Generator automatically monitors impedance and temperature and adjusts RF output accordingly. The CRF Radiofrequency Ablation System is used in a healthcare facility or hospital setting.
More Information

No
The description focuses on standard feedback control mechanisms (monitoring impedance and temperature to adjust RF output) and does not mention AI or ML.

Yes
The device is used for 'coagulation and ablation of tissue', which is a therapeutic intervention aimed at treating medical conditions by modifying or destroying tissue.

No

The device is described as an ablation system used for tissue coagulation and ablation, not for diagnosing a condition or disease. It monitors parameters internally for energy delivery, not for diagnostic purposes.

No

The device description explicitly lists hardware components such as an RF generator, active electrode, grounding pad, and peristaltic pump, indicating it is not a software-only device.

Based on the provided information, the CRF Radiofrequency Ablation System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is for "percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue." This describes a procedure performed on the patient's body, not on a sample taken from the patient's body.
  • Device Description: The description details a system that delivers radiofrequency energy directly to tissue within the patient. It involves an electrode, grounding pad, and cooling system, all designed for direct interaction with the patient's internal tissues.
  • Mechanism of Action: The device works by causing "tissue heating at the tip of the electrode causing coagulation and ablation of tissue." This is a physical process applied directly to the patient's body.
  • Lack of IVD Characteristics: There is no mention of analyzing samples (blood, urine, tissue samples, etc.) or providing diagnostic information based on such analysis. The device is therapeutic, not diagnostic.

IVD devices are used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. The CRF Radiofrequency Ablation System does not fit this definition.

N/A

Intended Use / Indications for Use

The CRF Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue.

Product codes

GEI

Device Description

The CRF Radiofrequency Ablation System consists of an RF generator, active electrode, grounding pad, and peristaltic pump for electrode cooling. This device is designed to cause tissue heating at the tip of the electrode causing coagulation and ablation of tissue. The CRF Generator supplies radiofrequency energy at 480 kHz while monitoring both the tissue temperature at the CRF electrode tip. The CRF Generator monitors the power, resistance, current and temperature during energy delivery to the patient.

The CRF Electrodes are sterile, single-use, hand-held for use in radiofrequency ablation procedures. The electrodes are surgically invasive, transient devices in patient contact typically for less than one hour. Cooling of the electrodes is provided by chilled sterile saline or water that is pumped through the electrode and out via outflow tubing. This is an enclosed system within the electrode and the saline does not contact. The CRF Generator automatically monitors impedance and temperature and adjusts RF output accordingly. The CRF Radiofrequency Ablation System is used in a healthcare facility or hospital setting.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

healthcare facility or hospital setting

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing of the device was completed to demonstrate substantial equivalence of the CRF Radiofrequency Ablation System to the identified predicate device. Non-clinical animal testing of the device was not required to establish substantial equivalence. The device components were subjected to the following verification and validation testing, as required:

  • Mechanical testing: Mechanical verification testing was performed on subject devices to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 standards and the Company's internal test requirements.
  • Electrical testing: Electrical verification testing was conducted for the subject device to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 electrical requirements.
  • Electromagnetic compatibility: Electromagnetic compatibility testing was completed for the subject device to ensure compliance with the current IEC 60601-1-2 standard.
  • Benchtop ablation testing. Comparative ablation verification testing was performed using an ex vivo tissue model to demonstrate the substantially equivalent ablation performance of the subject and predicate devices.
  • Usability testing: Testing was performed to verify and validate the usability requirements of the subject devices as a system in compliance with IEC 62366 and IEC 60601-1-6. Elements captured included normal use errors, and testable requirements for primary operating functions and frequent use scenarios.
  • Biocompatibility testing: Biocompatibility testing of patient contacting CRF devices was performed in accordance with the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" including cytotoxicity, systemic toxicity and irritation testing. All CRF devices demonstrated compliance with listed requirements.
  • Sterilization validation testing: ETO sterilization validation for CRF devices was carried out using the overkill method and halfcycles in accordance with the current ISO 11135 standard to provide a SAL of 10-6.
  • Packaging and shelf life testing: Sterile packaging performance testing was conducted for the subject devices in their final packaging. Shelf life testing for sterile subject devices and packaging was conducted based on an accelerated aging protocol in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. Real-time age testing will be performed to add to and support the results of the accelerated age testing.
  • Software validation testing: The CRF radiofrequency generator includes software developed according to the Company's software development process in compliance with the current IEC 62304 standard. Software verification testing was conducted in accordance with the FDA Guidance for Content of Premarket Submissions for Software Contained in Medical Devices (2005) and IEC 62304.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K070446

Reference Device(s)

K052796

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG" and "ADMINISTRATION".

January 7, 2020

Cambridge Interventional LLC Michael Arnold President 78 Cambridge Street Burlington, Massachusetts 01803

Re: K192715

Trade/Device Name: CRF Radiofrequency Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 12, 2019 Received: November 19, 2019

Dear Michael Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192715

Device Name

CRF Radiofrequency Ablation System

Indications for Use (Describe)

The CRF Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

CRF Radiofrequency Ablation System

Submitter:

Cambridge Interventional LLC 78 Cambridge Street Burlington MA 01803

Telephone: 617-365-6899 Contact Name: Michael Arnold Email: marnold@cambint.com Date prepared: November 12, 2019

Device Name and Classification

Model Name: CRF Radiofrequency Ablation System Common Name: Electrosurgical Cutting and Coagulation Device and Accessories Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Regulation Number: 21 CFR 878.4400 Product Code: GEI Device Class: II

Name of Predicate Device

Valleylab Cool-tip™ Switching Controller (K070446) - Product Code GEI, Class II

Name of Reference Device

Valleylab Cool-tip™ Radiofrequency Ablation System (K052796) - Product Code GEI, Class II

Description of Device

The CRF Radiofrequency Ablation System consists of an RF generator, active electrode, grounding pad, and peristallic punp for electrode cooling. This device is designed to cal tissue heating at the tip of the electrode causing coagulation and ablation of tissue. The CRF Generator supplies radiofrequency energy at 480 kHz while monitoring both the tissue imperature at the CRF electrode tip. The CRF Generator monitors the power, resistance, current and temperature during energy delivery to the paient.

The CRF Electrodes are sterile, single-use, hand-held for use in radiofrequency ablation procedures. The electrodes are surgically invasive, transient devices in patient contact typically for less than one hour. Cooling of the electrodes is provided by chilled sterile saline or water that is pumped through the electrode and out via outflow tubing. This is an enclosed system within the electrode and the saline does not contact. The CRF Generator automatically monitors impedance and temperature and adjusts RF output accordingly. The CRF Radiofrequency Ablation System is used in a healthcare facility or hospital setting. The table below lists components for the CRF Radiofrequency Ablation System.

Catalog No.Description of ComponentProvided Sterile?
CRFCRF Radiofrequency GeneratorNo
CRF-C1020RF Cannula, 10 cm, 2 cm TipYes
CRF-C1030RF Cannula, 10 cm, 3 cm TipYes
CRF-C1510RF Cannula, 15 cm, 1 cm TipYes
CRF-C1520RF Cannula, 15 cm, 2 cm TipYes
CRF-C1530RF Cannula, 15 cm, 3 cm TipYes
CRF-C2020RF Cannula, 20 cm, 2 cm TipYes
CRF-C2030RF Cannula, 20 cm, 3 cm TipYes
CRF-C2040RF Cannula, 20 cm, 4 cm TipYes
CRF-C2050RF Cannula, 20 cm, 5 cm TipYes

4

Catalog No.Description of ComponentProvided Sterile?
CRF-C2520RF Cannula, 25 cm, 2 cm TipYes
CRF-C2530RF Cannula, 25 cm, 3 cm TipYes
CRF-E10RF Electrode, Cooled, Blunt Tip, 10 cmYes
CRF-E15RF Electrode, Cooled, Blunt Tip, 15 cmYes
CRF-E20RF Electrode, Cooled, Blunt Tip, 20 cmYes
CRF-E25RF Electrode, Cooled, Blunt Tip, 25 cmYes
CRF-N1005RF Electrode, Non-Cooled, Sharp Tip, 10cm, 0.5cm TipYes
CRF-N1510RF Electrode, Non-Cooled, Sharp Tip, 15cm, 1cm TipYes
CRF-N2020RF Electrode, Non-Cooled, Sharp Tip, 20cm, 2cm TipYes
CRF-N2520RF Electrode, Non-Cooled, Sharp Tip, 25cm, 2cm TipYes
CRF-TNTemperature Probe, CannulaYes
CRF-TPTemperature Probe, Blunt ProbeYes
CRF-U1020RF Electrode, Cooled, Sharp Tip, 10cm, 2cm TipYes
CRF-U1030RF Electrode, Cooled, Sharp Tip, 10cm, 3cm TipYes
CRF-U1507RF Electrode, Cooled, Sharp Tip, 15cm, 0.7cm TipYes
CRF-U1510RF Electrode, Cooled, Sharp Tip, 15cm, 1cm TipYes
CRF-U1520RF Electrode, Cooled, Sharp Tip, 15cm, 2cm TipYes
CRF-U1530RF Electrode, Cooled, Sharp Tip, 15cm, 3cm TipYes
CRF-U2020RF Electrode, Cooled, Sharp Tip, 20cm, 2cm TipYes
CRF-U2030RF Electrode, Cooled, Sharp Tip, 20cm, 3cm TipYes
CRF-U2050RF Electrode, Cooled, Sharp Tip, 20cm, 5cm TipYes
CRF-U2525RF Electrode, Cooled, Sharp Tip, 25cm, 2.5cm TipYes
CRF-U2535RF Electrode, Cooled, Sharp Tip, 25cm, 3.5cm TipYes
CRF-FSFootswitchNo
CRF-PUMPPumpNo
CRF-A3ECable, CRF electrodes to CRF generatorNo
CRF-AMECable, CRF electrode to Cool-tip generatorNo
CRF-AMGCable, CRF-GP to Cool-tip generatorNo
CRF-SCGStorage Case for CRF GeneratorNo
CRF-SCPStorage Case for PumpNo
CRF-TUBETubing Set for PumpYes
CRF-GBCIntroducer for CRF-CYes
CRF-GBUIntroducer for CRF-UYes
CRF-GPGround Pad, DisposableNo

Indications for Use

The CRF Radiofrequency Ablation System is intended for use in percutaneous, intraoperative coagulation and ablation of tissue.

The Indications for Use statement is consistent with the predicate Cool-tip™ device, but without specific indications.

Comparison to Predicate Device

Compared to the predicate Cool-tip™ device, the CRF Radiofrequency Ablation System has similar technological characteristics and operating principle. There are no significant design or performance specifications that affect safety of the subject device. There is substantial equivalence between the subject device and the predicate with respect to indications and intended use. The CRF Radiofrequency Ablation System is substantially equivalent to the legally marketed predicate device with respect to indications for use and technology characteristics. The table below presents side-by-side comparisons for each major component for the subject and predicate device.

5

ParameterSubject DevicePredicate DeviceComparison
DeviceCRF Radiofrequency Ablation SystemValleylab Cool-tip™ Switching
Controller (K070446)N/A
IndicationThe CRF Radiofrequency Ablation
System is indicated for use in
percutaneous, laparoscopic and
intraoperative coagulation and ablation
of tissue.The Cool-tip™ RF Switching Controller
is to be used with the Cool-tip™ RF
System and is intended for use in
percutaneous, laparoscopic and
intraoperative coagulation and ablation
of tissue, such as partial or complete
ablation of nonresectable liver tumors,
and osteoid osteoma tumors within
bone.Different
Principle of OperationProvide RF energy to patient-contacting
TC Electrode for tissue ablation.Provide RF energy to patient-contacting
TC Electrode for tissue ablation.Identical
Classification878.4400878.4400Identical
Device ClassGEIGEIIdentical
Generator Electrical
ClassificationClass 1, IPX0, Type BFClass 1, IPX0, Type BFIdentical
Generator Output Frequency480 kHz480 kHzIdentical
Generator waveformSine waveSine waveIdentical
Generator waveform crest factor1.41.4Identical
Generator Rated PowerUp to 370wUp to 200wDifferent
Generator ImpedanceUp to 1000 ohmsUp to 1000 ohmsIdentical
Generator TimerUp to 15:00 minUp to 30:00 minDifferent
Generator Temperature MonitorThermocoupleThermocoupleIdentical
Generator Footswitch controlAvailableAvailableIdentical
TC Electrode length10, 15, 20, 25cm10, 14.4, 15, 20, 25cmDifferent
TC Electrode tip exposure0.5, 0.7, 1, 2, 2.5, 3, 3.5, 4, 5cm0.7, 1, 2, 2.5, 3, 4cmDifferent
TC Electrode tip diameter0.9, 1.5, 1.8mm1.5mmDifferent
TC Electrode Cooling
MechanismElectrode cooling is provided by sterile
water, which is pumped through the
inflow tubing into the Electrode shaft,
and out through the outflow tubing.Electrode cooling is provided by sterile
water, which is pumped through the
inflow tubing into the Electrode shaft,
and out through the outflow tubing.Identical
Maximum number of Electrodes
connected simultaneouslyUp to 3Up to 3Identical
Temp probe sensorType T ThermocoupleType T ThermocoupleIdentical
Temp Probe & Introducer length20cm20cmIdentical
Temp Probe & Introducer
diameter0.9, 1.3mm0.9, 1.3mmIdentical
Active electrode materialStainless SteelStainless SteelIdentical
Electrode insulation materialPolyesterPolyesterIdentical
Electrode sterilizationETO SterilizationETO SterilizationIdentical
Electrode shelf life4 years4 yearsIdentical
Electrode usageSingle UseSingle UseIdentical
Pump typePeristaltic pumpPeristaltic pumpIdentical
Pump specification80-120 ml/min80-140 ml/minDifferent
Tubing Set ConfigurationTwo-piece set (inflow and outflow
components)Two-piece set (inflow and outflow
components)Identical
Tubing Set MaterialPVC tubingPVC tubingIdentical
Tubing Set sterilizationETO SterilizationETO SterilizationIdentical
Tubing Set shelf life4 years4 yearsIdentical
Tubing Set usageSingle UseSingle UseIdentical
Ground Pad materialConductive hydrogelConductive hydrogelIdentical
Ground Pad usageSingle UseSingle UseIdentical
Ground Pad shelf life2 years2 yearsIdentical
Ground Pad regulatory clearance510(k) clearance (K000079)510(k) clearance (K030697)Identical

6

K192715

Summary of Non-Clinical Data

Performance testing of the device was completed to demonstrate substantial equivalence of the CRF Radiofrequency Ablation System to the identified predicate device. Non-clinical animal testing of the device was not required to establish substantial equivalence. The device components were subjected to the following verification and validation testing, as required:

• Mechanical testing: Mechanical verification testing was performed on subject devices to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 standards and the Company's internal test requirements.

· Electrical testing: Electrical verification testing was conducted for the subject device to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 electrical requirements.

· Electromagnetic compatibility: Electromagnetic compatibility testing was completed for the subject device to ensure compliance with the current IEC 60601-1-2 standard.

· Benchtop ablation testing. Comparative ablation verification testing was performed using an ex vivo tissue model to demonstrate the substantially equivalent ablation performance of the subject and predicate devices.

• Usability testing: Testing was performed to verify and validate the usability requirements of the subject devices as a system in compliance with IEC 62366 and IEC 60601-1-6. Elements captured included normal use errors, and testable requirements for primary operating functions and frequent use scenarios.

· Biocompatibility testing: Biocompatibility testing of patient contacting CRF devices was performed in accordance with the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" including cytotoxicity, systemic toxicity and irritation testing. All CRF devices demonstrated compliance with listed requirements.

· Sterilization validation testing: ETO sterilization validation for CRF devices was carred out using the overkill method and halfcycles in accordance with the current ISO 11135 standard to provide a SAL of 10-6.

• Packaging and shelf life testing: Sterile packaging performance testing was conducted for the subject devices in their final packaging. Shelf life testing for sterile subject devices and packaging was conducted based on an accelerated aging protocol in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. Real-time age testing will be performed to add to and support the results of the accelerated age testing.

· Software validation testing:

The CRF radiofrequency generator includes software developed according to the Company's software development process in compliance with the current IEC 62304 standard. Software verification testing was conducted in accordance with the FDA Gudance for Content of Premarket Submissions for Software Contained in Medical Devices (2005) and IEC 62304.

Summary of Clinical Data

Clinical testing of the device was not required to establish substantial equivalence.

Conclusion

The CRF Radiofrequency Ablation System is substantially equivalent to the predicate device cleared in K070446. The subject and predicate devices share the same fundamental scientific technology, including principles of operation and mechanism of action. Differences in design and technological characteristics between the subject device do not raise any new questions of safety and effectiveness. The intended use of the subject device has not changed. The comparison of technological characteristics, non-clinical performance data, electrical safety and electromagnetic compatibility testing demonstrates that the CRP Radiofrequency Ablation System is substantially equivalent to the previously cleared predicate device.