(137 days)
The Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
The Dental Zirconia are intended for use with CAD/CAM technology to produce all ceramic dental restorations as prescribed by a dentist.
The Dental Zirconia divided into two categories: Unshaded (Model: DUT) and Pre-Shaded (Model: DUTC).
Both Unshaded Dental Zirconia (Model: DUT) and Pre-Shaded Dental Zirconia (Model: DUTC) are available in discs shape, and in various specifications, which are shares same diameter (98 mm) and combinations of height (10-25 mm).
The color model for pre-shaded dental zirconia includes A1; A2; A3; A3.5; B2; C2; 1M1; 1M2; 2L1.5; 2M2; 3M2; 3M3; 4M2; 5M3 for difference color.
Both Unshaded Dental Zirconia (Model: DUT) and Pre-Shaded Dental Zirconia (Model: DUTC) have same base materials and derived from zirconia powder that has been processed via uni-aixial die pressing, followed by isostatic pressing, to achieve various shapes of uniform density and distribution. The ceramic blocks can be fabricated into various prosthetic dental devices.
The zirconia powder of Dental Zirconia is composed of ZrO2+Y2O3+HfO2+Al2O3. The performance of the Dental Zirconia conforms to ISO 6872:2015 Dentistry: Ceramic Materials.
The Dental Zirconia are disposable device, and provided as non-sterile
This document is a 510(k) premarket notification for a medical device called "Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia". It describes the device, its intended use, and demonstrates its substantial equivalence to a predicate device.
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Standard) | Reported Device Performance (Compliance) |
|---|---|
| ISO 6872:2015 Dentistry - Ceramic Materials | Complies |
| ISO 10993-3:2014 Biological evaluation of medical devices - Tests for genotoxicity, carcinogenicity and reproductive toxicity | Complies (Tested for Bacterial Reverse Mutation) |
| ISO 10993-5:2009 Biological evaluation of medical devices - Tests for In Vitro cytotoxicity | Complies (Tested for Cytotoxicity) |
| ISO 10993-10:2010 Biological evaluation of medical devices - Tests for irritation and skin sensitization | Complies (Tested for Oral Mucosa Irritation and Sensitization) |
| ISO 10993-11:2006 Biological evaluation of medical device - Tests for systemic toxicity | Complies (Tested for Acute Systemic Toxicity) |
| Safety and Efficacy for intended use with CAD/CAM technology in producing all ceramic dental restorations | Determined to be Substantially Equivalent to predicate device K172761, which is presumed to be safe and effective. |
2. Sample size used for the test set and the data provenance:
- Sample size for test set: Not explicitly stated in the provided document for each specific test. The document generally states "Non clinical tests were conducted to verify that the proposed device met all design specifications". The tests performed (Cytotoxicity, Oral Mucosa Irritation, Acute Systemic Toxicity, Sensitization, Bacterial Reverse Mutation) typically involve specific sample sizes relevant to the biological or material testing methodology, which are not detailed here.
- Data provenance: Not explicitly stated. However, the tests were conducted by the manufacturer, Qinhuangdao Audental Metal Technology Co., Ltd. in China, or by a testing facility commissioned by them. Given the context of a 510(k) submission, these would be considered prospective tests performed specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This document does not describe the use of human experts to establish "ground truth" for the performance of the device in the way one might for an AI/ML device in clinical settings. The "ground truth" for this device (dental zirconia) is established by its adherence to internationally recognized material and biocompatibility standards (e.g., ISO 6872, ISO 10993 series). The expertise involved would be in the form of certified laboratories and their technical staff, who conduct the tests and interpret results against the established criteria of these standards. The qualifications of these individuals are implied by their affiliation with accredited testing facilities competent in performing these specific ISO standard tests.
4. Adjudication method for the test set:
Not applicable in the context of this device and testing. The compliance with ISO standards is typically assessed through objective measurements and comparison against predefined thresholds specified in the standards, rather than expert adjudication in the sense of consensus on subjective interpretations.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a material (dental zirconia) used for manufacturing dental restorations, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. As mentioned, this is a physical medical device (dental zirconia), not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device's performance is established by objective material and biological test results directly measured and compared against the predefined requirements and thresholds of international standards (ISO 6872:2015 and ISO 10993 series). This differs from "expert consensus," "pathology," or "outcomes data" typically used for diagnostic or predictive algorithms.
8. The sample size for the training set:
Not applicable. This device is a material, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. No training set is involved.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 31, 2020
Qinhuangdao Audental Metal Technology Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, 102401 CHINA
Re: K192535
Trade/Device Name: Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: November 1, 2019 Received: November 4, 2019
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia
Indications for Use (Describe)
The Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Tab #8 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K192535
-
- Date of Preparation
01/30/2020
- Date of Preparation
-
- Sponsor
Qinhuangdao Audental Metal Technology Co., Ltd.
Floor 2, 16 North D, 1 Yanghe Road, Development Area, Qinhuangdao, 066004 Hebei, China Contact Person: Wang Xi Position: Product Manager Tel: +86 335 8691916 Fax: +86 335 8691917 Email: 13521721724@163.com
-
- Submission Correspondent
Mr. Ray Wang
- Submission Correspondent
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
-
- Identification of Proposed Device
Trade Name: Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia Common Name: Porcelain Powder for Clinical Use Model(s): DUT and DUTC
- Identification of Proposed Device
Regulatory Information: Classification Name: Porcelain Powder for Clinical Use Classification: II; Product Code: EIH; Regulation Number: 21 CFR 872.6660; Review Panel: Dental;
Indications for Use Statement:
The Unshaded Dental Zirconia and Pre-Shaded Dental Zirconia are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
-
- Device Description
The Dental Zirconia are intended for use with CAD/CAM technology to produce all ceramic dental restorations as prescribed by a dentist.
- Device Description
The Dental Zirconia divided into two categories: Unshaded (Model: DUT) and Pre-Shaded (Model: DUTC).
Both Unshaded Dental Zirconia (Model: DUT) and Pre-Shaded Dental Zirconia (Model: DUTC) are available in discs shape, and in various specifications, which are shares same diameter (98 mm) and combinations of height (10-25 mm).
The color model for pre-shaded dental zirconia includes A1; A2; A3; A3.5; B2; C2; 1M1; 1M2; 2L1.5; 2M2; 3M2; 3M3; 4M2; 5M3 for difference color.
Both Unshaded Dental Zirconia (Model: DUT) and Pre-Shaded Dental Zirconia (Model: DUTC) have same base materials and derived from zirconia powder that has been processed via uni-aixial die pressing, followed by isostatic pressing, to achieve various shapes of uniform density and distribution. The ceramic blocks can be fabricated into various prosthetic dental devices.
The zirconia powder of Dental Zirconia is composed of ZrO2+Y2O3+HfO2+Al2O3. The performance of the Dental Zirconia conforms to ISO 6872:2015 Dentistry: Ceramic Materials.
The Dental Zirconia are disposable device, and provided as non-sterile
{5}------------------------------------------------
-
- Identification of Primary Predicate Device
Primary Predicate Device:
- Identification of Primary Predicate Device
510(k) Number: K172761
Product Name: New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded) Manufacturer: Shanghai New Century Dental Materials Co., Ltd.
- Non-Clinical Test Conclusion 7.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ISO 6872:2015 Dentistry - Ceramic Materials
ISO 10993-3: 2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
ISO 10993-11:2006 Biological evaluation of medical device – Part 11: Tests for systemic toxicity.
The biocompatibility tests conducted on the subject device includes:
Cytotoxicity Testing (ISO 10993-5:2009), Oral Mucosa Irritation Testing (ISO 10993-10:2010), Acute Systemic Toxicity Testing (ISO 10993-11:2006), Sensitization (ISO 10993-10:2010); Bacterial Reverse Mutation (ISO 10993-3:2014)
-
- Clinical Test Conclusion
No Clinical Test conducted.
- Clinical Test Conclusion
{6}------------------------------------------------
9. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device (K172761) | Remark |
|---|---|---|---|
| Product Code | EIH | EIH | SAME |
| Regulation No. | 21 CFR 872.6660 | 21 CFR 872.6660 | SAME |
| Class | II | II | SAME |
| Indications for Use | The Unshaded Dental Zirconia andPre-Shaded Dental Zirconia are intendedfor use with CAD/CAM technology toproduce all ceramic dental restorations asprescribed by a dentist. | New Century All-Ceramic DentalZirconia Blocks (Un-Shaded &Pre-Shaded) are Intended for use withCAD/CAM technology to produce allceramic dental restorations (full contourcrowns, bridges, inlays, and Veneers) asprescribed by a dentist. | SAME |
| Basic DesignCharacteristic | Disc | Blocks, Disc, and Rod | SAME |
| Materials | ZrO2+Y2O3+HfO2+Al2O3Meet the requirements of ISO 6872 | ZrO2+Y2O3+HfO2+Al2O3Meet the requirements of ISO 6872 | SAME |
| Processing | Sintering at temperature around 1500 °C | Sintering at temperature around 1500 °C | SAME |
| Dimension | Various | Various | SAME |
| Single Use | Yes | Yes | SAME |
| Color | None and Pre-shaded | None, and Pre-shaded (for pre-shadedseries) | SAME |
| Sterile | Non-sterile | Non-sterile | SAME |
| PerformanceTest | Comply with ISO 6872 | Comply with ISO 6872 | SAME |
| Biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SAME |
| Label andLabeling | Conforms to FDA RegulatoryRequirements | Conforms to FDA RegulatoryRequirements | SAME |
Table 1 General Comparison
The proposed device is highly similar to the predicate device in the terms of indication for use, materials processing, color, sterile etc.. The only differences between the proposed and predicate device are:
- The subject device only has disc, while the predicate device has blocks, disc and rod. This a. difference does not raise any safety or efficacy concerns, because both proposed and predicate has the disc shapes, and the proposed device has no other shapes outside the scope of predicate device.
- Both subject device and predicate device has various dimension. This difference does no raise an b. safety or efficacy concerns, because no matter what size will not affect the physical/mechanical and biocompatibility properties of the proposed product, both the subject and predicate device have similar physical/mechanical and biocompatibility properties that met the requirements of ISO 6872 and ISO 10993.
{7}------------------------------------------------
10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
§ 872.6660 Porcelain powder for clinical use.
(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.