K Number
K192487
Device Name
EZ-TRAX Containment Device
Manufacturer
Date Cleared
2019-12-23

(103 days)

Product Code
Regulation Number
880.6850
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The EZ-TRAX™ Containment Device is intended for use as an accessory in healthcare facilities to organize, enclose, reprocess, transport, and store medical devices between surgical and other medical uses. The EZ-TRAX™ Containment Device is not intended on its own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterile barrier system.
Device Description
The EZ-TRAXTM Containment Device includes bases with dividers, lids, and brackets fabricated from a variety of materials commonly used to enclose, protect, and organize non-sterile medical devices. The subject device is not intended to maintain sterility, but it is intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization wrap. The subject device protects the interior components during transportation, and storage. The EZ-TRAX TM Containment Device is composed of intrinsically stable metals and thermoplastic polymers. The trays and lids are composed of anodized aluminum with stainless steel handles. The dividers are composed of Aluminum and the posts are composed of medical grade thermoplastic polymers. The lids and bottom of bases are fully perforated with an evenly distributed hole pattern. The sides of the bases are partially perforated. The trays are used with locking lids. The bases / lid assembly was designed to be used for sterilization via steam sterilization and used in standard autoclaves found in hospitals and healthcare facilities. The trays were designed in such a way to withstand repeated steam sterilization cycles.
More Information

Not Found

No
The device description and performance studies focus on the physical properties, sterilization compatibility, and containment function of the device, with no mention of AI or ML.

No
The device is described as an accessory for organizing, enclosing, reprocessing, transporting, and storing medical devices, and is not intended to maintain sterility on its own or to provide medical treatment.

No

Explanation: The device is described as a "Containment Device" used to organize, enclose, reprocess, transport, and store medical devices. It does not perform any diagnostic function.

No

The device description explicitly details physical components made of metals and thermoplastic polymers, and the performance studies focus on cleaning, sterilization, life cycle, and biocompatibility of these physical materials. There is no mention of software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for organizing, enclosing, reprocessing, transporting, and storing medical devices. It is an accessory used in healthcare facilities for handling other medical devices.
  • Device Description: The description focuses on the physical construction and materials of the containment device, designed for sterilization and protection of other medical devices.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (blood, urine, tissue, etc.) to provide information for diagnosis, monitoring, or treatment. IVDs are typically reagents, instruments, or systems used for these purposes.
  • Performance Studies: The performance studies described relate to cleaning, sterilization, life cycle testing, and biocompatibility – all relevant to a device used for handling and sterilizing other medical devices, but not characteristic of IVD performance studies (which would involve analytical and clinical performance related to specimen testing).
  • Predicate Device: The predicate device is a "Transportation / Sterilization Cassette," which aligns with the function of a device for handling and sterilizing other medical devices, not an IVD.

In summary, the EZ-TRAX™ Containment Device is a medical device accessory used in the process of handling and sterilizing other medical devices, not a device used for in vitro diagnostic testing of human specimens.

N/A

Intended Use / Indications for Use

The EZ-TRAX™ Containment Device is intended for use as an accessory in healthcare facilities to organize, enclose, reprocess, transport, and store medical devices between surgical and other medical uses. The EZ-TRAX™ Containment Device is not intended on its own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterile barrier system.

Product codes (comma separated list FDA assigned to the subject device)

KCT

Device Description

The EZ-TRAXTM Containment Device includes bases with dividers, lids, and brackets fabricated from a variety of materials commonly used to enclose, protect, and organize non-sterile medical devices.

The subject device is not intended to maintain sterility, but it is intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization wrap.

The subject device protects the interior components during transportation, and storage.

The EZ-TRAX TM Containment Device is composed of intrinsically stable metals and thermoplastic polymers. The trays and lids are composed of anodized aluminum with stainless steel handles. The dividers are composed of Aluminum and the posts are composed of medical grade thermoplastic polymers.

The lids and bottom of bases are fully perforated with an evenly distributed hole pattern. The sides of the bases are partially perforated. The trays are used with locking lids.

The bases / lid assembly was designed to be used for sterilization via steam sterilization and used in standard autoclaves found in hospitals and healthcare facilities. The trays were designed in such a way to withstand repeated steam sterilization cycles.

The following EZ-TRAXTM models / sizes are offered:

Brand NameModelDimensions
EZ TRAXTMBASE.ASSY.AL.24.12.2(L) 22.97" (W) 11.18" (H) 2.04"
EZ TRAXTMBASE.ASSY.AL.24.12.2.4(L) 22.97" (W) 11.18" (H) 2.44"

The maximum load is 25 lbs and can be used with different types of medical devices including:

  • Devices with lumens up to 500mm in length -
  • Conjoined/mated surfaces: forceps, clamps, bending pliers, ratchet handles, retractors
  • Cannulated: drill bits, taps, guides, screwdrivers, cannulated screws

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

healthcare facilities

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Performance Data

Cleaning instructions were validated using microbiological, protein, and hemoglobin assays. Sterilization, including sterilant penetration and drying time, was performed according to ANSV/AAMI/ISO 17665-2. Life cycle (simulated usage) testing was performed, which inspection, component dimensional fit verification, and functional closure (lid- bottom latch) verification. Biocompatibility testing was performed using methods described in ANSI/AAMI/ISO 10993-5 (cytotoxicity). No clinical data were included in this submission.

As the EZ-TRAXTM Containment Device has been subjected to performance testing including:

Non-Clinical Performance Testing Table

TestStandardCommentsResult
Material CompatibilityAAMI TIR12:2010 AAMI ST79:2017 AAMI ST81:2004/R2016 ISO 17665-1:2006/R2013Material compatibility pre-vacuum 132C for 4 minutes – Mechanical Washing and Steam Sterilization. The testing subjected the EZ-TRAXTM to repetitive cleaning and sterilization processing for 25 reprocessing cycles at parameters that represented the worst case conditions. Chemical indicators were utilized to demonstrate steam penetration. The study found no degradation or lack of functionality after 25 cycles.Passed
Mechanical Cleaning Validation – HemoglobinAAMI TIR30:2011 ASTM F32018-18 ASTM F3293-18The mechanical cleaning validation of the EZ-TRAXTM concluded that the manufacturer's cleaning instructions are efficacious for removing gross amounts of soil from the EZ-TRAXTM containment system to a hemoglobin level less than 2.2 μg/cm2 per device.Passed
Mechanical Cleaning Validation – Protein AnalysisAAMI TIR30:2011 ASTM F32018-18 ASTM F3293-18The mechanical cleaning validation of the EZ-TRAXTM protein concluded that the manufacturer's cleaning instructions are efficacious for removing gross amounts of soil from the EZ- TRAXTM to a protein level less than 6.4 μg/cm2 per device.Passed
MEM Elution CytotoxicityISO 10993-5:2009 / R2014The cytotoxicity testing was conducted per ISO 10993- 5:2009/(R)2014 and concluded that EZ-TRAXTM requirements of the test and are NOT considered to have a cytotoxic potential.Passed
Sterilization ValidationAAMI ST77:2013 ISO 14937:2009 AAMI ST8:2013The sterilization validation of the EZ- TRAXTM included pre-vacuum steam 132C for 4 minutes. The conclusion could achieve a Sterility Assurance Level (SAL) of 10-6 after processing in the following pre-vacuum steam sterilization cycle 132C for 4 minutes.Passed
Thermal Profile StudyAAMI ST77:2013 AAMI ST79:2017The thermal profile study of the EZ- TRAXTM included pre-vacuum steam 132C for 4 minutes. The study demonstrated that adequate sterilant penetration can be achieved. The EZ- TRAXTM can reach and maintain a steady state thermal conditions throughout the exposure phase when processed in the following pre-vacuum steam sterilization cycle 132C for 4 minutes.Passed
Drying Time TestAAMI ST77:2013The study concluded that the EZ- TRAXTM meets or exceeds the minimum acceptance criteria for dry time of 10 minutes.Passed
Handle 100 lbs force testAAMI ST77:2013None of the tray handles broke loose, showed evidence of permanent distortion, cracking or other evidence of failure when tested with a force of 50 lbs on each handle. The study concluded that the EZ- TRAXTM can carry a minimum of four times maximum recommended weight of 25 lbs without deforming, cracking, or exhibiting other evidence of damage.Passed

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Sterility Assurance Level (SAL) of 10-6

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Medtronic Transportation / Sterilization Cassettes (K152241)

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 23, 2019

K1 Medical LLC % Joseph Azary Regulatory Consultant K1 Medical Technologies 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K192487

Trade/Device Name: EZ-TRAX™ Containment Device Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 24, 2019 Received: November 27, 2019

Dear Joseph Azary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT. Elizabeth Claverie Williams, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

510(k) Number (if known) K192487

Device Name

EZ-TRAX™ Containment Device

Indications for Use (Describe)

The EZ-TRAXIM Containment Device is intended for use as an accessory in healthcare facilities to organize, enclose, reprocess, transport, and store medical devices between surgical and other medical uses. The EZ-TRAX™ Containment Device is not intended on its own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterile barrier system.

Validated Cycle Times for Dynamic Air Removal Steam Sterilization Cycles

CycleTemperatureExposure TimeDrying Time
Dynamic Air
Removal270° F/ 132° C4 minutes10 minutes

Sterilization validations included a worst case EZ-TRAX™ Containment Device and a medical device challenge set comprising of:

  • Lumen dimensions (3) 1mm x 500mm -
  • -Conjoined/mated surfaces: forceps, clamps, bending pliers, ratchet handles, retractors
  • -Cannulated: drill bits, taps, guides, screwdrivers, cannulated screws
  • -A total weight of 42 lbs comprising of (EZ-TRAX" " Containment Device + sterile barrier system + medical device load) as a worst case challenge to the system. Healthcare facilities should not exceed 25 pounds (EZ-TRAXIM Containment Device + sterile barrier system + medical device load).

The EZ-TRAX™ is offered in the following sizes:

Brand NameModelDimensions
EZ-TRAXTMBASE.ASSY.AL.24.12.2(L) 22.97" (W) 11.18" (H) 2.04"
EZ-TRAXTMBASE.ASSY.AL.24.12.2.4(L) 22.97" (W) 11.18" (H) 2.44"

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

ZOver-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and

FORM FDA 3881 (8/14)

3

complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

4

510(K) SUMMARY

K1 Medical LLC's EZ-TRAX'M Containment Device K192487

Submitter

K1 Medical LLC 56 Newton Road Woodbridge, CT 06525 Contact Person: Joseph Azary, Regulatory Consultant

Phone: 203-242-6670

Date Prepared: December 21, 2019

Name of Device: EZ-TRAXTM Containment Device

Common or Usual Name: Sterilization Cassette

Classification Name: Sterilization Wrap

Regulatory Class: Class II, 21 CFR 880.6850

Product Code: KCT

Predicate Device: Medtronic Transportation / Sterilization Cassettes (K152241)

Intended Use / Indications for Use

The EZ-TRAX™ Containment Device is intended for use as an accessory in healthcare facilities to organize, enclose, reprocess, transport, and store medical devices between surgical and other medical uses. The EZ-TRAX™ Containment Device is not intended on its own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterile barrier system.

Validated Cycle Times for Dynamic Air Removal Steam Sterilization Cycles

CycleTemperatureExposure TimeDrying Time
Dynamic Air
Removal270° F/ 132° C4 minutes10 minutes

Sterilization validations included a worst case EZ-TRAX™ Containment Device and a medical device challenge set comprising of:

  • Lumen dimensions (3) 1mm x 500mm -
  • Conjoined/mated surfaces: forceps, clamps, bending pliers, ratchet handles, retractors -
  • Cannulated: drill bits, taps, guides, screwdrivers, cannulated screws -
  • A total weight of 42 lbs comprising of (EZ-TRAXTM Containment Device + sterile barrier system + medical device load) as a worst case challenge to the system. Healthcare facilities should not exceed 25 pounds (EZ-TRAXTM Containment Device + sterile barrier system + medical device load).

5

Device Description

The EZ-TRAXTM Containment Device includes bases with dividers, lids, and brackets fabricated from a variety of materials commonly used to enclose, protect, and organize non-sterile medical devices.

The subject device is not intended to maintain sterility, but it is intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization wrap.

The subject device protects the interior components during transportation, and storage.

The EZ-TRAX TM Containment Device is composed of intrinsically stable metals and thermoplastic polymers. The trays and lids are composed of anodized aluminum with stainless steel handles. The dividers are composed of Aluminum and the posts are composed of medical grade thermoplastic polymers.

The lids and bottom of bases are fully perforated with an evenly distributed hole pattern. The sides of the bases are partially perforated. The trays are used with locking lids.

The bases / lid assembly was designed to be used for sterilization via steam sterilization and used in standard autoclaves found in hospitals and healthcare facilities. The trays were designed in such a way to withstand repeated steam sterilization cycles.

The following EZ-TRAXTM models / sizes are offered:

Brand NameModelDimensions
EZ TRAXTMBASE.ASSY.AL.24.12.2(L) 22.97" (W) 11.18" (H) 2.04"
EZ TRAXTMBASE.ASSY.AL.24.12.2.4(L) 22.97" (W) 11.18" (H) 2.44"

The maximum load is 25 lbs and can be used with different types of medical devices including:

  • Devices with lumens up to 500mm in length -
  • -Conjoined/mated surfaces: forceps, clamps, bending pliers, ratchet handles, retractors
  • -Cannulated: drill bits, taps, guides, screwdrivers, cannulated screws

6

Subject DevicePredicate DeviceComparison
K1 Medical LLC
EZ-TRAXTM Containment
Device K192487Medtronic Transportation /
Sterilization Cassettes K152241
Indications
for UseThe EZ-TRAXTM
Containment Device is
intended for use as an
accessory in healthcare
facilities to organize, enclose,
reprocess, transport, and store
medical devices between
surgical and other medical
uses. The EZ-TRAXTM
Containment Device is not
intended on its own to
maintain sterility; it is
intended to be used in
conjunction with a legally
marketed, validated, FDA-
cleared sterile barrier system.
The following parameters
and cycles have been
validated to attain sterility
assurance level (SAL) of
10-6.
Cycle Times for Dynamic Air
Removal Steam Sterilization
Cycles
Cycle: Dynamic Air
Removal Temperature:
270F / 132C Exposure
Time: 4 minutes Drying
Time: 10 minutes
Sterilization validations
included a worst case EZ-
TRAXTM Containment
Device and a medical device
challenge set comprising of:
  • Lumen dimensions
    (3) 1mm x 500mm
  • Conjoined/mated
    surfaces: forceps,
    clamps, bending
    pliers, ratchet
    handles, retractors
  • Cannulated: drill
    bits, taps, guides,
    screwdrivers,
    cannulated screws
  • A total weight of 42
    lbs comprising of
    (EZ-TRAXTM
    Containment Device
  • sterile barrier | Intended for use in healthcare
    facilities to organize, enclose,
    sterilize, transport, and store
    medical devices and other
    instrumentation between surgical
    and other medical uses. The system
    is not intended on its own to
    maintain sterility; it is intended to
    be used in conjunction with a
    legally marketed, validated, FDA-
    cleared sterilization wrap.

Sterilization validations included
implants and common surgical
instruments such as rasps, drivers,
trials, handles, inserters, probes,
drills, etc. The validated total
weight of 28.4 lbs. The validated
worst case loading configuration
included the following worst case
lumen dimensions 363 x 1.575mm
and 247.5 x 4.1mm.

Cycle: Gravity Displacement
Temperature: 250F, 270F, and
275F Exposure Time: 30, 15, and
10 minutes Min Dry Time: 30
minutes

Cycle: Dynamic Air Removal
Temperature: 270F and 275F
Exposure Time: 4 and 3 minutes | Similar |
| | | | |
| | system + medical | | |
| | device load) as a | | |
| | worst case challenge | | |
| | to the system. | | |
| | Healthcare | | |
| | facilities should | | |
| | | | |
| | not exceed 25 | | |
| | pounds (EZ- | | |
| | TRAXTM | | |
| | Containment | | |
| | Device + sterile | | |
| | barrier system + | | |
| | medical device | | |
| | load). | | |
| Fundamental | Sterilization Cassette | Sterilization Cassette | Same |
| Scientific | | | |
| Technology | | | |
| Product Code | KCT | KCT | Same |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| Material
Composition | Thermoplastic polymers,
Aluminum,
and stainless steel | Thermoplastic polymers, Aluminum,
and stainless steel | Same |
| Design | Base, lid with locking latch and
individual inserts | Base, lid with locking latch and
individual inserts | Same |
| Dimensions | Worst case 24 x 12 x 2.4" | Worst case 22.75 x 11.26 x 5.51" | Similar |
| Configuration | Perforated bases, lids and inserts | Perforated bases, lids and inserts | Same |
| Air Permeance | Yes | Yes | Same |
| Percent
Perforation | Evenly distributed pattern | Evenly distributed hole pattern | Same |
| Sterilization
Method | Dynamic Air Removal | Dynamic Air Removal
Gravity Displacement | Similar |
| Sterilization
Parameters | Dynamic Air
Removal
Temperature: 270F
/ 132C Exposure
Time: 4 minutes
Drying Time: 10
minutes | Gravity Displacement
250F, 30 minutes, and 30 minutes
dry
270F, 15 minutes, and 30 minutes
dry
275F, 10 minutes, and 30 minutes
dry
Dynamic Air Removal
270F, 4 minutes, and 30 minutes
dry
275F, 3 minutes, and 30 minutes
dry | Similar |
| Reusable | Yes | Yes | Same |
| Microbial
Barrier
Properties | Used with FDA approved sterile
barrier system | Used with FDA approved
sterilization wrap | Same |
| Material
Compatibility | Materials compatible with
sterilization method | Materials compatible with
sterilization method | Same |
| Toxicological | Biocompatible | Biocompatible | Same |

TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE

7

8

The following technological differences exist between the subject and predicate devices:

  • The predicate device has additional processing parameters

Summary of Technological Characteristics

The subject device is composed of anodized aluminum, stainless steel and thermoplastic polymers. The base and lid is perforated with an evenly distributed pattern and designed to be used for sterilization via steam sterilization. The device is designed in such a way to withstand reported sterilization cycles.

Non-Clinical Performance Data

Cleaning instructions were validated using microbiological, protein, and hemoglobin assays. Sterilization, including sterilant penetration and drying time, was performed according to ANSV/AAMI/ISO 17665-2. Life cycle (simulated usage) testing was performed, which inspection, component dimensional fit verification, and functional closure (lid- bottom latch) verification. Biocompatibility testing was performed using methods described in ANSI/AAMI/ISO 10993-5 (cytotoxicity). No clinical data were included in this submission.

As the EZ-TRAXTM Containment Device has been subjected to performance testing including:

9

Non-Clinical Performance Testing Table

TestStandardCommentsResult
Material
CompatibilityAAMI TIR12:2010
AAMI ST79:2017
AAMI ST81:2004/R2016
ISO 17665-1:2006/R2013Material compatibility pre-vacuum 132C for 4 minutes –
Mechanical Washing and
Steam Sterilization.
The testing subjected the EZ-TRAXTM to repetitive cleaning and
sterilization processing for 25
reprocessing cycles at parameters
that represented the worst case
conditions. Chemical indicators
were utilized to demonstrate steam
penetration. The study found no
degradation or lack of functionality
after 25 cycles.Passed
Mechanical Cleaning
Validation
– HemoglobinAAMI TIR30:2011
ASTM F32018-18
ASTM F3293-18The mechanical cleaning validation of
the EZ-TRAXTM concluded that the
manufacturer's cleaning instructions are
efficacious for removing gross amounts
of soil from the EZ-TRAXTM
containment system to a hemoglobin
level less than 2.2 μg/cm2 per device.Passed
Mechanical Cleaning
Validation
– Protein AnalysisAAMI TIR30:2011
ASTM F32018-18
ASTM F3293-18The mechanical cleaning validation of
the EZ-TRAXTM protein concluded
that the manufacturer's cleaning
instructions are efficacious for removing
gross amounts of soil from the EZ-
TRAXTM to a protein level less than
6.4 μg/cm2 per device.Passed
MEM Elution
CytotoxicityISO 10993-5:2009 /
R2014The cytotoxicity testing was
conducted per ISO 10993-
5:2009/(R)2014 and concluded that
EZ-TRAXTM requirements of the
test and are NOT considered to have
a cytotoxic potential.Passed
TestStandardCommentsResult
Sterilization
ValidationAAMI ST77:2013
ISO 14937:2009
AAMI ST8:2013The sterilization validation of the EZ-
TRAXTM included pre-vacuum steam
132C for 4 minutes. The conclusion could
achieve a Sterility Assurance Level (SAL)
of 10-6 after processing in the following
pre-vacuum steam sterilization cycle 132C
for 4 minutes.Passed
Thermal Profile
StudyAAMI ST77:2013
AAMI ST79:2017The thermal profile study of the EZ-
TRAXTM included pre-vacuum steam
132C for 4 minutes. The study
demonstrated that adequate sterilant
penetration can be achieved. The EZ-
TRAXTM can reach and maintain a steady
state thermal conditions throughout the
exposure phase when processed in the
following pre-vacuum steam sterilization
cycle 132C for 4 minutes.Passed
Drying Time TestAAMI ST77:2013The study concluded that the EZ-
TRAXTM meets or exceeds the minimum
acceptance criteria for dry time of 10
minutes.Passed
Handle 100 lbs
force testAAMI ST77:2013None of the tray handles broke loose,
showed evidence of permanent distortion,
cracking or other evidence of failure when
tested with a force of 50 lbs on each handle.
The study concluded that the EZ-
TRAXTM can carry a minimum of four
times maximum recommended weight of 25
lbs without deforming, cracking, or
exhibiting other evidence of damage.Passed

10

Conclusions

The EZ-TRAXTM Containment Device has the same intended use and principles of operation, and very similar indications for use and technological characteristics, as the predicate device. The minor differences between the EZ-TRAXTM Containment Device and the predicate device does not alter its therapeutic purpose or raise new issues of safety or effectiveness, and performance data demonstrate that the EZ-TRAX™ Containment Device is as safe and effective as the predicate device. Thus, the EZ-TRAX™ Containment Device is substantially equivalent to the Medtronic Transportation / Sterilization Cassettes (K152241).

K1 Medical concludes that based on the nonclinical tests performed that the subject device, EZ- TRAXTM Containment Device, is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device, Medtronic Transportation / Sterilization Cassettes (K152241), Class II (21 CFR 880.6850), product code KCT.