K Number
K192394

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-12-02

(90 days)

Product Code
Regulation Number
888.3027
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Hi-Fatigue Bone Cement is intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Device Description

Hi-Fatique Bone Cement is a PMMA, radiopaque bone cement, designed for the fixation of prothesis to the living bone. Hi-Fatigue Bone Cement is a traditional bone cement product. The bone cement is made of two separate sterile components. When both components are mixed together, they become a self-hardening, radiopaque bone cement which fixes the implant. The liquid is contained in a vial and the powder in a pouch; these components are packed in blister with Tyvek lid or an aluminium pouch. The devices are sold disposable, single-use and sterile.

AI/ML Overview

This appears to be a 510(k) summary for a bone cement device, not an AI/ML medical device. Therefore, the requested information about acceptance criteria, study details, and ground truth establishment for AI/ML performance is not found in this document.

The document discusses performance data related to the physical and chemical properties of bone cement, referring to a "FDA Class II Special Controls Guidance Document: Polymethylacrylate (PMMA) Bone Cement; Guidance for Industry and FDA" dated July 17, 2002.

Here's an attempt to extract relevant information, acknowledging the mismatch with the prompt's AI/ML focus:

1. A table of acceptance criteria and the reported device performance

The document states that "The performance data demonstrate that the new devices Hi-Fatigue Bone Cement is substantially equivalent to the predicate device Palacos® R... and meet the requirements of the Special Controls Guidance document."

Unfortunately, the specific acceptance criteria values (e.g., minimum compressive strength, maximum polymerization temperature) and the reported device performance values for each of the mechanical, physical, and chemical properties are not explicitly listed in a comparative table within this 510(k) summary. The summary only lists the types of tests performed.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the 510(k) summary. The performance data section vaguely states "non-clinical performance testing was performed."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a bone cement, not an AI/ML imaging device requiring expert ground truth for interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as the device is a bone cement.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable as the device is a bone cement, not an AI/ML imaging device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a bone cement.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For a bone cement, "ground truth" would relate to the established physical, chemical, and biological properties defined by industry standards or regulatory guidance. The document states that testing was "in accordance with the 'FDA Class II Special Controls Guidance Document: Polymethylacrylate (PMMA) Bone Cement; Guidance for Industry and FDA' dated July 17, 2002." This guidance document would define the expected performance characteristics that the device must meet or be substantially equivalent to.

8. The sample size for the training set

This is not applicable as the device is a bone cement and does not involve AI/ML training.

9. How the ground truth for the training set was established

This is not applicable as the device is a bone cement and does not involve AI/ML training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

OSARTIS GmbH Volker Stirnal Director, Ouality Assurance and Regulatory Affairs Lagerstrasse 11-15 Dieburg, 64807 Germany

Re: K192394

Trade/Device Name: Hi-Fatigue Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: LOD Dated: August 30, 2019 Received: September 3, 2019

Dear Mr. Stirnal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne, Ph.D. Acting Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192394

Device Name Hi-Fatigue Bone Cement

Indications for Use (Describe)

Hi-Fatigue Bone Cement is intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Hi-Fatigue Bone Cement 510(k) Summary K192394

1. General Information

1.1 Submitter and Owner of the 510(k)

OSARTIS GmbH Lagerstraße 11-15 64807 Dieburg Germany Phone: +49 6071 / 929-0 Fax: +49 6071 / 929-100

1.2 Contact Person

Volker Stirnal

1.3 Device Subject of this 510(k)

Hi-Fatigue Bone Cement

1.4 Date of Preparation

28.10.2019

2. Name of the Device and Classification Information

This traditional 510(k) has been submitted for the following device.

Hi-Fatigue Bone Cement

2.1 Trade/Proprietary Name

Hi-Fatique Bone Cement

2.2 Common/Usual Name

PMMA bone cement for orthopedics

2.3 Classification Information

Classification Name: Polymethylmethacrylate (PMMA) Bone Cement

Classification Regulation: 21 CFR § 888.3027

Regulatory Class: Class II

Product Code: LOD - Bone Cement

Orthopedic Panel:

{4}------------------------------------------------

Hi-Fatigue Bone Cement

510(k) Summary

3. Predicate Device

The predicate devices are as follows:

  • Palacos® R (510(k) application K030902) -
  • -Cobalt™ MV Bone Cement (510(k) application K091608)

4. Device Description

Hi-Fatique Bone Cement is a PMMA, radiopaque bone cement, designed for the fixation of prothesis to the living bone. Hi-Fatigue Bone Cement is a traditional bone cement product. The bone cement is made of two separate sterile components. When both components are mixed together, they become a self-hardening, radiopaque bone cement which fixes the implant. The liquid is contained in a vial and the powder in a pouch; these components are packed in blister with Tyvek lid or an aluminium pouch. The devices are sold disposable, single-use and sterile.

5. Indication for Use

Below is the indication for use:

Hi-Fatigue Bone Cement is intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

6. Comparison of the technological Characteristics with the Predicate Device

Hi-Fatigue Bone Cement shares many of the same technological characteristics compared to the predicates Palacos® R and Cobalt™ MV Bone Cement, including important considerations such as most materials.

CharacteristicsHi-Fatigue GBone CementPalacos® RK030902Cobalt™ MVBone CementK091608
MaterialpowderPolymerPoly (methylacrylate/methylmethacrylate)(PMMA)Poly (methylmethacrylate/styrene)Poly (methylacrylate/methylmethacrylate)(PMMA)Poly (methylmethacrylate)Methylmethacrylate-Styrene
InitiatorBenzoyl peroxideBenzoyl peroxideBenzoyl peroxide
Radio-pacifierZirconium dioxideZirconium dioxideZirconium dioxide
ColourAdditivesNoneChlorophyll VIIIFD&C Blue No. 2Aluminium Lake
MaterialLiquidMonomerMethylmethacrylate (MMA) stabilizedwith HydroquinoneMethylmethacrylate (MMA) stabilizedwith HydroquinoneMethylmethacrylate stabilized withHydroquinone

Table 1: Comparison of the Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

Hi-Fatique Bone Cement

510(k) Summary

CharacteristicsHi-Fatigue G Bone CementPalacos® R K030902Cobalt™ MV Bone Cement K091608
ActivatorN,N-dimethyl-p-toluidineN,N-dimethyl-p-toluidineN,N-dimethyl-p-toluidine
Colour AdditivesNoneChlorophyll VIII in oily solutionNone

7. Performance Data

This 510(k) submission provided performance data to establish the substantial equivalence of the new bone cements to the predicate bone cement. Performance testing was conducted in accordance with the "FDA Class II Special Controls Guidance Document: Polymethylacrylate (PMMA) Bone Cement; Guidance for Industry and FDA" dated July 17, 2002.

Non-clinical performance testing was performed to characterize the bone cements in accordance with special controls guidance document. This testing included the following:

  • Mixing and Application characteristics (e.g. dough time, setting time, viscosity, intrusion )
  • Chemical Composition (e.q. trace elements, residual low MW molecules, leachables)
  • Molecular weight and Polymer structure (e.g. molecular weight, glass transition temperature)
  • Physical Properties (e.g. porosity, shrinkage)
  • Stability of Components (e.g. change in monomer viscosity due to artificial aging)
  • Thermal Properties (e.g. maximum polymerization temperature)
  • Mechanical properties (e.g. cyclic fatique properties, bending properties, compressive properties, tensile properties, fracture toughness)

The performance data demonstrate that the new devices Hi-Fatigue Bone Cement is substantially equivalent to the predicate device Palacos® R 510(k) application K030902 and meet the requirements of the Special Controls Guidance document.

Bacterial endotoxins of Hi-Fatique Bone Cement have been evaluated using Recombinant Factor C Assay (EndoZyme) following Ph. Eur. 5.01.10 and 2.06.14, based on USP <161> and USP<85>. Test results meet the endotoxin limits of 20 endotoxin units (EU)/device as defined in USP and as recommended by the FDA guidance "Pyrogen and Endotoxins Testing: Question and Answers" (2012).

§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.

(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”