AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SciPharm's DFV Desensitizing Varnish is a fluoride-containing resin preparation for the treatment of dentinal hypersensitivity and for the reduction of post-operative sensitivity.

Device Description

SciPharm's DFV Desensitizing Varnish is a device that consists of a synthetic compound from a natural rosin that is intended to coat a prepared tooth cavity prior to the insertion of a restorative material by forming a film that facilitates occlusion of compromised surfaces, including open dentinal tubules. DFV Desensitizing Varnish and the predicate device consist of a viscous liquid that can be applied to the teeth using a brush or similar applicator. The purpose of the 5% sodium fluoride as the source of fluoride ions for the formation of calcium fluoride, and along with the rosin, occludes the dentin tubules. Both use denatured alcohol (ethanol) as a significant component, providing fluidity for application and to promote quick drying on the tooth surface.

SciPharm's DFV Desensitizing Varnish is a topically applied, flavored (Strawberry, Bubblegum, Spearmint, Chocolate, and Marshmallow) topical varnish containing sodium fluoride on a synthetic rosin based preparation from a natural rosin. Only a professional clinician may apply DFV Desensitizing Varnish. The varnish and applicator are packaged in a thermoformed PET/PVC opaque unit dose tray, and sealed with foil.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called DFV Desensitizing Varnish. It details the device's properties, intended use, and comparison to predicate devices, including performance data.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria & Reported Device Performance

The document does not explicitly state "acceptance criteria" in a pass/fail sense with numerical thresholds for performance. Instead, it presents a comparison of the proposed device (DFV Desensitizing Varnish) with an "Equivalent Varnish" (presumably a predicate device or a similar product used for internal comparison) across several physical and chemical tests. The implicit acceptance criteria are that the DFV Desensitizing Varnish should perform comparably or better than the equivalent device.

TestProposed DFV Desensitizing VarnishEquivalent VarnishReported Device Performance (DFV)Comments
ContaminationNo Foreign ParticlesNo Foreign ParticlesNo Foreign ParticlesMeets the comparative standard.
ColorMatch StandardMatch StandardMatch StandardMeets the comparative standard. Further context on what "standard" refers to is not provided, but it implies internal consistency.
OdorMatch StandardMatch StandardMatch StandardMeets the comparative standard. Similar to Color, "standard" is internal.
Viscosity4500 – 6500 cps7000 – 8500 cps4500 – 6500 cpsThe proposed device has a different, lower viscosity. The document does not provide context on why this difference is acceptable, but it implies it's within an acceptable range for the intended application.
Non-Pyrolizable %7.5% - 8.7%8.4% - 9.5%7.5% - 8.7%The proposed device has a slightly lower range for non-pyrolizable percentage. Similar to viscosity, acceptability is implicit.
Fluoride Content4.5% - 5.5% NaF (2.0% - 2.5% F)4.5% - 5.5% NaF (2.0% - 2.5% F)4.5% - 5.5% NaF (2.0% - 2.5% F)Exactly matches the equivalent varnish, which is a key performance metric for a fluoride varnish.
Cytotoxicity11% cell viability (Toxicon Report 17-04113-G1)11% cell viability (Toxicon Report 18-00602-G1)11% cell viabilityBoth devices showed 11% cell viability, which is below the 70% threshold. The document explains this is expected due to ingredients like ethanol and sodium fluoride, and is considered acceptable as it's consistent with existing approved varnishes.
Dentin Tubule Occlusion (SEM)Smooth with minimal surface defectsSubsurface porosity and heterogeneitySmooth with minimal surface defectsThe proposed device demonstrated a "thick uniform film with fewer defects and protrusions" than the competitor's varnish, suggesting better performance in this aspect.
Shelf LifeStable for at least 24 monthsNot specifiedStable for at least 24 monthsDemonstrated stability through internal testing (Report 2019-R-012).

2. Sample Size Used for the Test Set and Data Provenance

  • Cytotoxicity Test: Not explicitly stated for either the DFV Desensitizing Varnish or the equivalent device. It refers to "L929 cells." The provenance is not specified beyond being generated by "Toxicon Report 17-04113-G1" and "Toxicon Report 18-00602-G1."
  • Dentin Tubule Occlusion (SEM): Refers to "Disk-shaped human dentine specimens." The exact number of specimens/samples is not specified. The provenance (e.g., country) is not mentioned.
  • Physical Tests (Contamination, Color, Odor, Viscosity, Non-Pyrolizable %, Fluoride Content): The sample size for these tests is not specified.
  • Shelf Life: Not specified, but implied to be multiple units tested over time intervals (Time 0, 6 months, 12 months, 18 months, etc.) for up to 4 years.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • No expert adjudication is mentioned or implied for these performance tests. The tests are laboratory-based and objective (e.g., cell viability percentage, SEM imaging, chemical concentration measurements).

4. Adjudication Method for the Test Set

  • None. As the tests are objective laboratory measurements, there is no need for expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • Not applicable. This device is a cavity varnish, not an AI-powered diagnostic or assistive technology. It does not involve human readers interpreting cases or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. As above, this is not an AI-powered device. Performance is measured directly from the physical and chemical properties of the varnish.

7. The Type of Ground Truth Used

  • Cytotoxicity: Measured cell viability against a control and compared to a 70% threshold (ISO 10993-5 and ISO 7405 standards).
  • Dentin Tubule Occlusion (SEM): Direct microscopic observation and characterization of the film formed on dentine specimens.
  • Physical/Chemical Tests: Laboratory measurements against internal "standards" or specifications (e.g., specific viscosity range, fluoride content percentage).
  • Shelf Life: Stability over time judged by changes in contamination, color, viscosity, fluoride content, nonpyrolizable percentage, and odor against acceptable ranges.

8. The Sample Size for the Training Set

  • Not applicable. This device is a pharmaceutical product (dental varnish), not a machine learning model, so there is no concept of a "training set" for model development.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for a machine learning model, this question does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 18, 2020

Scientific Pharmaceuticals Thomas Kelly Regulatory Affairs/Quality Assurance Manager 3221 Producer Way Pomona, California 91768

Re: K192381

Trade/Device Name: DFV Desensitizing Varnish-Strawberry Flavor, DFV Desensitizing Varnish-Bubblegum (Tutti-Fruity) Flavor, DFV Desensitizing Varnish-Spearmint Flavor, DFV Desensitizing Varnish-Chocolate Flavor, DFV Desensitizing Varnish-Marshmallow Flavor

Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: November 20, 2019 Received: November 20, 2019

Dear Thomas Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192381

Device Name DFV Desensitizing Varnish

Indications for Use (Describe)

SciPharm's DFV Desensitizing Varnish is a fluoride-containing resin preparation for the treatment of dentinal hypersensitivity and for the reduction of post-operative sensitivity.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue Erlenmeyer flask on the left, followed by the text "SCI-PHARM" in blue. To the right of the text is a mortar and pestle, also in blue, with the registered trademark symbol above it.

510(k) Summary — K192381

510(k) Submission:Traditional
Device Type:Varnish, Cavity
Submitter:Scientific Pharmaceuticals, Inc.3221 Producer WayPomona, CA 91768
Contact Person:Thomas KellyRegulatory Affairs/Quality Assurance ManagerRegulatory@scipharm.comTelephone: 909-595-9922 ext. 237
Alternate Contact Person:David ButlerPresidentDave@scipharm.comTelephone: 909-595-9922 ext. 225
Date:February 18, 2020
Confidentiality:According to 21 CFR 807.95
Trade Name:Sci-Pharm DFV Desensitizing Varnish
Common Name:Cavity Varnish or Topical Varnish
Classification Nameand References:Cavity Varnish21CFR 872.3260
Product Code:LBH
Classification Panel:Dental
Classification:Medical Device, Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the address of Scientific Pharmaceuticals. The address is 3221 Producer Way, Pomona, CA 91768. The text is black and the background is white.

Image /page/4/Picture/1 description: The image shows the logo for SCI-PHARM. The logo is blue and features a graphic of two beakers on the left, the text "SCI-PHARM" in the center, and a mortar and pestle on the right. The logo also includes a registered trademark symbol.

Purpose and Description:

SciPharm's DFV Desensitizing Varnish is a device that consists of a synthetic compound from a natural rosin that is intended to coat a prepared tooth cavity prior to the insertion of a restorative material by forming a film that facilitates occlusion of compromised surfaces, including open dentinal tubules. DFV Desensitizing Varnish and the predicate device consist of a viscous liquid that can be applied to the teeth using a brush or similar applicator. The purpose of the 5% sodium fluoride as the source of fluoride ions for the formation of calcium fluoride, and along with the rosin, occludes the dentin tubules. Both use denatured alcohol (ethanol) as a significant component, providing fluidity for application and to promote quick drying on the tooth surface.

SciPharm's DFV Desensitizing Varnish is a topically applied, flavored (Strawberry, Bubblegum, Spearmint, Chocolate, and Marshmallow) topical varnish containing sodium fluoride on a synthetic rosin based preparation from a natural rosin. Only a professional clinician may apply DFV Desensitizing Varnish. The varnish and applicator are packaged in a thermoformed PET/PVC opaque unit dose tray, and sealed with foil.

The technological difference between DFV Desensitizing Varnish and the other six (6) varnishes that Scientific Pharmaceutical's manufactures is the formulation of components. The components are commonly used in the dental materials industry. The DFV Desensitizing Varnish uses a synthetic resin as the base material instead of the natural rosin used in other formulations. This synthetic rosin is the same used in the predicate device. DFV Desensitizing Varnish also includes additional base material, sweeteners, bitterness blockers, and flavoring agents.

The device is intended to prevent the penetration of restorative materials, such as amalgam, into dental tissue.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue Erlenmeyer flask on the left, followed by the text "SCI-PHARM" in blue. On the right, there is a blue mortar and pestle with a registered trademark symbol above it. The logo is simple and professional, and it conveys the company's focus on science and pharmaceuticals.

Chemical Composition:

Raw Material(Ingredient)Purpose of IngredientConcentration ofIngredient
1.Purified RosinFilm Former45.0% - 75.0%
2.PolyamideFilm Former10.0% - 25.0%
3.Ethanol,200 ProofSolvent,Surface Wettability10.0% - 25.0%
4.Sodium FluorideReduce Tooth Decay4.50% - 5.50%
5.CarbamidePeroxideWhitening Agent forcolor stability0.25% - 5.0%
6.XylitolSweetener andstimulates saliva flow0.5% - 4.0%
7.SucraloseSweetener0.25% - 5.0%
8.Bitter BlockerFlavoring Agent0.10% - 3.00 %
9.Silica FillerThickener1.0% - 5.0%
10.Flavorants(Strawberry,Bubblegum,Spearmint,Chocolate, &Marshmallow)Flavoring Agent0.25% - 10.0%

Contraindications: Patients allergic, or hypersensitive, to natural resin or colophony (kolophonium or resembling agents) should avoid the use of this product. In case of an adverse reaction, the varnish may be removed by brushing with a soft toothbrush and flossing, followed by rinsing with warm water. Direct ingestion should be avoided. Dyspnea, (difficult or labored breathing) associated with asthmatic children, has been reported in rare cases. Patients with sensitive digestive systems may experience nausea, usually after a prolonged treatment. Patients with stomatitis or ulcerative gingivitis should avoid this product. Primary Predicate Device: Young Dental Manufacturing: Oral-B 5% Sodium Fluoride Varnish (K153334). Reference Device: 3M: Vanish 5% Sodium Fluoride White Varnish (K090519) Unit dose container, Nylon Brushes for application, mixing Accessories(s):

well.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of two blue Erlenmeyer flasks, the text "SCI-PHARM" in blue, and a blue mortar and pestle. The registered trademark symbol is located to the right of the text.

Comparison to Predicate Device

DFV DesensitizingVarnishPrimary PredicateDeviceOral-B 5% SodiumFluoride VarnishReference DeviceVanish Varnish
Indications for Use(abbreviated)Treatment of dentalhypersensitivity andreduction of post-operative sensitivityK153334Treatment of dentalhypersensitivity andreduction of post-operative sensitivityK090519Treatment of dentalhypersensitivity andreduction of post-operative sensitivity
ContraindicationsHypersensitivity tocolophony should avoidthis product, as well aspatients with ulcerativegingivitis and stomatitisUlcerative gingivitis andstomatitisHypersensitivity tocolophony should avoidthis product, as well aspatients with ulcerativegingivitis and stomatitis
Mode of ActionDentin Tubule OcclusionDentin Tubule OcclusionDentin Tubule Occlusion
Method ofApplicationApplication of a thin coatof varnish of varnish ontooth surfaces only by aprofessional. Not forhome use.Application of a thin coatof varnish of varnish ontooth surfaces.Application of a thin coatof varnish of varnish ontooth surfaces only by aprofessional. Not forhome use.
Percentage ofSodium Fluoride5%5%5%
PackagingVarnish and applicatorbrush placed in athermoformed PET/PVCopaque unit dose trayand closed with apeelable foil seal.Varnish and applicatorbrush placed in a moldedopaque acclar tray andclosed with a foil seal.Varnish and applicatorbrush placed in a moldedopaque tray and closedwith a foil seal.
Unit Package0.4 mL0.4 mL0.5 mL
Basic CompositionSynthetic RosinSynthetic RosinSynthetic Rosin

Technological Characteristics:

The basic composition, fundamental technology, and intended use of DFV Desensitizing Varnish is substantially equivalent to the predicate device and comparable to devices that have been on the market for many decades. These products use components used in the dental industry such as resins, rosins, modified rosins hydroxypropyl cellulose, polyurethane, methacrylates, polyethylene glycol, amides, dimethacrylates, and other film forming ingredients to cover dentinal tubules, which provide relief from dentinal

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue graphic of two Erlenmeyer flasks on the left, the text "SCI-PHARM" in blue in the center, and a blue mortar and pestle on the right. There is a registered trademark symbol next to the mortar and pestle.

hypersensitivity. DFV Desensitizing Varnish is a varnish that contains sodium fluoride, like all of the others, and contains rosin that provides immediate and sustained relief of dental pain or sensitivity. The immediate relief of hypersensitivity is achieved by the hardened resin films that forms immediately after application and occludes the dentin tubules.

Biocompatibility Tests: The chemical components in Sci-Pharm's DFV Desensitizing Varnish have been used in a variety of predicate devices. Rosin and other rosin derivatives are used as adhesive agents in approved dental varnishes.

Biocompatibility testing is an essential element of developing a medical device. Each medical device used in dentistry shall be subjected to a structured biological evaluation program within a risk management process. In order to comply with the requirements, DFV Desensitizing Varnish was tested for cytotoxicity and to provide evidence that the formula blocks dentinal tubules.

Cytotoxicity followed the procedure outlined in ISO 10993-5 and ISO 7405, where a smear was extracted in Minimal Essential Medium (MEM) with 10% Fetal Bovine Serum for 24 hours at 37°C. Then L929 cells were exposed to the extracts and the viability of the cells were measured by their ability to uptake a vital dye. The viability percentage of the cells exposed to the extracts from DFV Desensitizing Varnish was 11%, which is below the 70% (Toxicon Report 17-04113-G1). This value is expected as it contains significant ethanol, so an additional study was conducted using an approved dental varnish that is similar, and currently being used on the market. The predicate product is another dental varnish already on the market and compared with the same methodology to determine cytotoxicity from DFV Desensitizing varnish. The viability percentage of the cells exposed to the extracts from an equivalent device was also 11%, which is also below the 70% (Toxicon Report 18-00602-G1). The cytotoxicity for both varnishes are the same, indicating that there is no difference. Based on the chemical content, the residual ethanol will disrupt cellular integrity, the sodium fluoride is a bacterial growth inhibitor, which can also affect the viability of the cells, as can the carbamide peroxide. All of these chemicals had an effect that contributed towards the low cellular viability result, but since these are the very same chemicals used for all approved dental varnishes

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for SCI-PHARM. The logo is blue and features a beaker on the left, the text "SCI-PHARM" in the middle, and a mortar and pestle on the right. The logo is simple and professional.

currently on the market, then they are found acceptable. Additional testing can be requested, but the results will be the same due to the ethanol, not the flavor.

The second test is where dentin is exposed to the varnish and viewed under SEM to verify that the varnish blocks dentinal tubulae. There were two varnishes evaluated, Sci-Pharm's DFV Desensitizing Varnish and an equivalent predicate varnish. Disk-shaped human dentine specimens had the varnishes applied separately using a plastic brush and a single stroke. The surfaces were imaged using reflection optical microscopy (ROM), allowing the identification morphology of the specimen surfaces without the dehydration effects. Then the top surface of the sectioned specimens were examined using a scanning electron microscope (SEM). The ROM images demonstrated a typical granular appearance with dentinal tubule orifices closed by smear plugs. Both varnishes tested showed highly smooth material layer covering all of the dentine and when evaluated by SEM, Sci-Pharm's DFV Desensitizing Varnish remained smooth with minimal surface defects while the other varnish showed subsurface porosity and heterogeneity. The conclusion was that Sci-Pharm's DFV Desensitizing Varnish demonstrated a thick uniform film with fewer defects and protrusions than the competitors varnish. Additional testing is not necessary as the benefits are from sodium fluoride and the rosin mixture, not the flavors.

Proposed DFVDesensitizingVarnishEquivalent Varnish
Test
ContaminationNo ForeignParticlesNo ForeignParticles
ColorMatch StandardMatch Standard
OdorMatch StandardMatch Standard
Viscosity4500 – 6500 cps7000 – 8500 cps
Non-Pyrolizable %7.5% - 8.7%8.4% - 9.5%
FluorideContent4.5% - 5.5% NaF(2.0% - 2.5% F)4.5% - 5.5% NaF(2.0% - 2.5% F)

Physical Tests:

Shelf Life

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue graphic of two Erlenmeyer flasks on the left, followed by the text "SCI-PHARM" in blue. To the right of the text is a blue mortar and pestle with a registered trademark symbol.

Shelf Life was established following Sci-Pharm's internal procedure, EX-06, where the product being tested is analyzed during intervals within the shelf life, such as Time 0, Time 6 months, Time 12 months, Time 18 months, etc., and continued for up to 4 years. The product is evaluated for any changes that may occur over time. Testing is performed on contamination, color, viscosity, fluoride content, nonpyrolizable percentage, and odor. Color changes may occur over time, but since a small thin layer that is washed off within hours, the color change in the product is not substantial. Report 2019-R-012 provides evidence that the product is stable for at least 24 months.

Substantial Equivalence: In summary, this submission demonstrates that Sci-Pharm's DFV Desensitizing Varnish is substantially equivalent in safety and effectiveness, and performs equivalently or better than the identified primary predicate device and the reference device varnishes for their intended use. Device is biocompatible when used as directed by dental professionals per ISO 10993-1. DFV Desensitizing Varnish does not introduce any new indications for use or potential hazards.

§ 872.3260 Cavity varnish.

(a)
Identification. Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.(b)
Classification. Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.