K Number
K192331
Manufacturer
Date Cleared
2019-11-12

(77 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LaseMD LEO Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sunspots) and ephelides (freckles).

Device Description

The LaseMD LEO Laser System is a thulium laser, producing a pulsed beam of coherent near-infrared light (1927 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LCD touch screen gives the user control over the necessary laser system parameters. The LaseMD LEO Laser System is equipped with a 658 nm aiming beam.

AI/ML Overview

The provided text is a 510(k) summary for the Lutronic Corporation LASEMD LEO Laser System. It focuses on demonstrating substantial equivalence to predicate devices rather than presenting acceptance criteria and the results of a study designed to prove the device meets specific performance criteria.

Therefore, the document does not contain most of the information requested in your prompt. Specifically:

  • Acceptance Criteria and Reported Device Performance: This information is not explicitly stated in a table format as requested. The document discusses "performance data" as an "Animal Study Report" and "Bench Testing," implying that these were used to support substantial equivalence rather than to meet predefined acceptance criteria against specific performance metrics for the device's clinical indications.
  • Sample size, data provenance, number of experts, adjudication method, MRMC study, and standalone performance: There is no mention of these details for any clinical study. The "Animal Study Report" is mentioned to demonstrate equivalence, but no specifics about its design, participants, or outcomes proving adherence to performance metrics for the indications are provided.
  • Type of ground truth, training set sample size, and how training set ground truth was established: These concepts are typically relevant for AI-powered devices or diagnostic tools. The LaseMD LEO Laser System is described as a "Laser Surgical Instrument," suggesting it's a physical device for dermatological procedures, not an AI/ML diagnostic tool. Therefore, these categories are not applicable to the information provided.

What is available:

The document states:

  • 7. Performance Data:
    • Animal Study Report: Included to demonstrate equivalence between the C1 tip of the Subject device and the Predicate devices. No other details are provided.
    • Bench Testing: The LASEMD LEO Laser System complies with all applicable standards, including ISO 13485:2016, ISO 60601-1 for electrical safety, and IEC 60601-1-2 for electromagnetic compatibility. This means the device met engineering and safety standards, not clinical performance criteria in terms of efficacy for the indications for use.

In summary, the provided document does not contain the detailed information about acceptance criteria and clinical study results that you requested, as it is a 510(k) summary focused on demonstrating substantial equivalence to existing predicate devices based on technological characteristics and a limited animal study, rather than proving performance against specific clinical efficacy criteria in human studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

November 12, 2019

Lutronic Global James Childs Clinical Science Director 19 Fortune Drive Billerica, Massachusetts 01821

Re: K192331

Trade/Device Name: LaseMD LEO Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 11, 2019 Received: September 13, 2019

Dear James Childs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

ce determination does not mean that FDA

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192331

Device Name LaseMD LEO Laser System

Indications for Use (Describe)

The LaseMD LEO Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sunspots) and ephelides (freckles).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K192331

510(k) Summary for the Lutronic Corporation LASEMD LEO Laser System

This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

1. General Information
Applicant:Lutronic Corporation219, Sowon-roHaengsin-dong, Deogyang-guGoyang-si, Gyeonggi-do, 410-722Republic of KoreaTel: (82) 31-908-3440Fax: (82) 31-907-3440
Contact Person:James ChildsClinical Science DirectorLutronic Global19 Fortune DriveBillerica, MA 01821Tel: 978-584-3464Email: jchilds@lutronic.com
Summary Preparation Date:August 23, 2019
2. Names
Trade Name:LASEMD LEO Laser System
Common Name:Dermatology Laser
Classification Name:Laser instrument, Surgical,Powered
Product Code: GEXPanel: General and Plastic

3. Predicate Devices

The LASEMD Laser System and Fraxel DUAL Laser System are surgical instruments for performing dermatological procedures in the treatment of actinic keratosis and treatment of benign pigmented lesions.

Surgery

{4}------------------------------------------------

510(K) NumberK171009K130193
Company NameLutronic CorporationSolta Medical, Inc.
Device NameLASEMD Laser SystemFraxel DUAL 1550/1927Laser System
Classification Regulation21 CFR § 878.481021 CFR § 878.4810
Classification NamePowered Laser SurgicalInstrumentPowered Laser SurgicalInstrument
Product CodeGEXGEX
Device PanelGeneral & Plastic SurgeryGeneral & Plastic Surgery

4. Device Description

The LaseMD LEO Laser System is a thulium laser, producing a pulsed beam of coherent near-infrared light (1927 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LCD touch screen gives the user control over the necessary laser system parameters. The LaseMD LEO Laser System is equipped with a 658 nm aiming beam.

5. Indications for Use

The LaseMD LEO Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sunspots) and ephelides (freckles).

6. Substantial Equivalence

The LaseMD LEO Laser System offers the user the same wavelength, same pulse energy range, same optical train, similar spot sizes, same spot fluence settings and similar per-pass treatment density range as the legally marketed Solta Medical Fraxel DUAL Laser System that is the subject of 510(k) K130193, and LASEMD Laser System (K171009) legally marketed by Lutronic Corporation. The subject and the predicate devices also all operate in pulsed-power mode. The pulsed power setting is user selectable in Expert Mode between 1 – 20W. The maximum power of 20W with the LaseMD LEO system is greater than the maximum pulsed powers available in the predicate devices and is for user convenience. The peak power densities (W/cm2) remain substantially below the ablation threshold and the mechanism of action - thermal coagulation -- is identical to that of the predicate devices. Further, an Animal Study report is included in the submission that establishes equivalence between LaseMD LEO in Energy mode (20W) with the Fraxel Dual device (12W) at similar energy settings. In addition, the LaseMD LEO system operating at 5W or

{5}------------------------------------------------

lower power settings is identical in design and performance to the LaseMD system at the same power and energy settings. The C5 tip with 350 um spot size provides fluences within the range of the predicate devices. Further, the shot density of the C5 tip includes 144 mB/cm2 and 225 mB/cm2 shot density settings. The shot density of the Predicate device is up to 426 MTZ/cm². At 20 mJ beam energy treatment, the maximum average shot area energy density (J/cm2) for C5 is 0.02 J x 144 mB/cm2 = 2.88 J/cm² and 0.02 J x 225 mB/cm² = 4.5 J/cm² for the 144 mB/cm² and 225 mB/cm2 shot densities, respectively. The average shot area energy density for the Predicate device at the same beam energy is up to 0.02 J x 426 MTZ/cm2 = 8.52 J/cm². Therefore the C5 tip average shot energy density settings are within those of the Predicate device and do not raise any new concerns about safety. No new concerns about effectiveness arise with the 144 mB/cm2 and 225 mB/cm2 shot densities because they are greater than the densities of the cleared Predicate device with 100 mB/cm2 (comparing the square tip in Stamping mode). Final treatment coverage is also controlled by the number of passes and is the same as those of the predicate devices. Therefore, the technological differences between the subject device and the predicate devices are considered to be minor and no new concerns for safety or effectiveness arise.

DeviceLaseMD LEO LaserSystemFraxel DUAL1550/1927Laser SystemLASEMD Laser System
510(k) NumberTo be assignedK130193K171009
Indications for UseThe LaseMD LEO LaserSystem is indicated foruse in dermatologicalprocedures requiring thecoagulation of softtissue, treatment ofactinic keratosis, andtreatment of benignpigmented lesions suchas, but not limited tolentigos (age spots),solar lentigos (sun spots)and ephelides (freckles).The Fraxel 1927 nmlaser is indicated for usein dermatologicalprocedures requiring thecoagulation of softtissue, treatment ofactinic keratosis, andtreatment of benignpigmented lesions suchas, but not limited tolentigos (age spots),solar lentigos (sun spots)and ephelides (freckles).The LASEMD LaserSystem is indicated foruse in dermatologicalprocedures requiring thecoagulation of softtissue, treatment ofactinic keratosis, andtreatment of benignpigmented lesions suchas, but not limited tolentigos (age spots),solar lentigos (sun spots)and ephelides (freckles).
ClassificationGEX, 21 CFR §878.4810GEX, 21 CFR §878.4810GEX, 21 CFR §878.4810
Laser TypeThulium laserThulium laserThulium laser
Laser Wavelength1927 nm1927 nm1927 nm
Aiming Beam658 nm ≤ 5 mW658 nm ≤ 5 mW658 nm ≤ 5 mW
Beam DeliveryFiber and HandpieceFiber and HandpieceFiber and Handpiece
Emission ControlFootswitchFootswitchFootswitch
Display ScreenYesYesYes
Power20 W12 W5 W
Max Pulse Energy20 mJ20 mJ20 mJ
Max Pulse Width20 ms10 ms20 ms
Pulse Repetition RateC1 Tip: 43 -323 HzC5 Tip: 46 - 632 Hz0 – 3 kHz43.5 – 307.7 Hz
Laser Beam Spot Size(Spot Fluence Range)[Treatment Method]200 µm (All C1 Tips)(3 - 60 J/cm²)[Stamping or Rolling]350 µm (C5 Tip)(1 – 20 J/cm²)[Stamping]200 µm (All Tips)(3 - 60 J/cm²)[Rolling]100 µm (Cap 100)(12 - 240 J/cm²)[Stamping or Rolling]200 µm (Cap200)(3 - 60 J/cm²)[Stamping]
Rolling Tx TipsFunction:Scan Area/ShotDensityPulses/ShotShot Energy RangeC1 with RollersRandom:2.5 mm x 10 mmDensity 40 mB/cm²(10 Pulses per shot)10 - 200 mJDynamic:2 mm x 10 mmDensity 25 mB/cm²(5 Pulses per shot):5 - 100 mJCW:4 mm x 10 mmDensity 100 mB/cm²(40 Pulses per Shot)40 - 800 mJSmall tip: 7 mm(Max at least 426MTZ/cm²)Large tip: 15 mm(Max at least 426MTZ/cm²)Caps 100, 200Random:2.5 mm x 10 mmDensity 40 mB/cm²(10 Pulses per shot)10 - 200 mJDynamic:2 mm x 10 mmDensity 25 mB/cm²(5 Pulses per shot):5 - 100 mJCW:4 mm x 10 mmDensity 100 mB/cm²(40 Pulses per Shot)40 - 800 mJ
Stamping Tx TipsFunction:Shot AreaDensityPulses/ShotShot Energy RangeC1 (Square, Comb Tip)Static:4 mm x 10 mmDensity: 100 mB/cm²(40 Pulses per Shot)40 - 800 mJC5 TipStatic:Not ApplicableC1 (Square, CombTip)Static:4 mm x 10 mmDensity: 100 mB/cm²(40 Pulses per Shot)40 - 800 mJ
6 mm x 10 mmDensity 100 mB/cm2(60 Pulses per Shot):60 – 1200 mJ
6 mm x 10 mmDensity 144 mB/cm2(84 Pulses per Shot):84 – 1680 mJ
6 mm x 10 mmDensity 225mB/cm2(135 Pulses per Shot):135 – 2700 mJ
Dimensions of LaserConsole460 mm (W) x 613 mm(L) x 1744 mm (H)489 mm (W) x 451 mm(L) x 444.5 mm (D)413 mm (W) x 324 mm(L) x 255.5 mm (D)
Weight of Laser System33kg33.1kg33.1 kg
Electrical RatingAC 100-240 V (FuseAC250V/6.3A), 50/60Hz, Power consumptionAC 100-240V, 50/60Hz, PowerConsumptionAC 100-240V, 50/60Hz, PowerConsumption

{6}------------------------------------------------

{7}------------------------------------------------

7. Performance Data

The Company's performance data for the LASEMD LEO Laser System is as follows:

An Animal Study Report is included to demonstrate the equivalence between the C1 tip of the Subject device and the Predicate devices.

Bench Testing

The LASEMD LEO Laser System complies with all applicable standards, including ISO 13485:2016, ISO 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility. See in Appendix 7.

8. Conclusion

The intended use of the LASEMD LEO Laser System is virtually identical to the intended use of the predicate devices and the technological characteristics of the LASEMD LEO Laser System. Any differences between the LASEMD LEO Laser System and the predicate devices have no significant influence on safety or effectiveness of the LASEMD LEO

{8}------------------------------------------------

Laser System. Therefore, the LASEMD LEO Laser System is substantially equivalent to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.