K Number
K192214
Date Cleared
2019-10-11

(57 days)

Product Code
Regulation Number
888.3060
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CarboClear Vertebral Body Replacement (VBR) System is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor or trauma (e.g., fracture).

The CarboClear Vertebral Body Replacement device is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

The use of allograft or autograft with the CarboClear Vertebral Body Replacement device is optional.

Device Description

The CarboClear VBR System comprises implants (spacers) in different dimensions, and instruments. The CarboClear VBR spacers are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), with titanium alloy endplates.

AI/ML Overview

The provided document is a 510(k) summary for the CarboClear® VBR System, a medical device for spinal intervertebral body fixation. It does not describe an AI/ML powered device, nor does it detail a study proving device performance against acceptance criteria in the manner requested (e.g., with specific metrics like sensitivity, specificity, or human reader improvement with AI assistance).

The document focuses on demonstrating substantial equivalence of the CarboClear® VBR System to predicate devices through an assessment of its intended use, design, materials, technological characteristics, principles of operation, and mechanical performance characteristics.

Therefore, I cannot extract the information required for an AI/ML powered device study from this document. The sections you requested (acceptance criteria table, sample size, expert ground truth, MRMC study, standalone performance, etc.) are deeply specific to the evaluation of AI/ML models, not for a physical implant device like the CarboClear® VBR System.

The closest relevant information concerning "performance" in this document refers to mechanical testing standards for an implant:

  • Performance characteristics included static and dynamic axial compression and torsion testing per ASTM F 2077, expulsion testing, and endplate shear strength testing.
  • Performance characteristics are comparable to those of predicate devices (as applicable), thus demonstrating that the device is substantially equivalent to the predicates.

This indicates that the "acceptance criteria" were likely defined by meeting or exceeding the mechanical performance standards established in ASTM F 2077 and demonstrating comparability to predicate devices for these physical attributes. However, specific numerical acceptance criteria (e.g., a minimum axial compression strength) and reported performance values are not provided in this summary for confidentiality reasons, as this is a 510(k) summary, not the full submission. Similarly, the "study" proving this performance would be the mechanical testing described, not a clinical study involving human outcomes or an AI algorithm's performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2019

CarboFix Orthopedics, Ltd. Ms. Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim St. Herzeliya 4672411 Israel

Re: K192214

Trade/Device Name: CarboClear® VBR System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: MQP Dated: August 13, 2019 Received: August 15, 2019

Dear Ms. Rubin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Ronald P. Jean. PhD Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192214

Device Name CarboClear® VBR System

Indications for Use (Describe)

The CarboClear Vertebral Body Replacement (VBR) System is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor or trauma (e.g., fracture).

The CarboClear Vertebral Body Replacement device is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

The use of allograft or autograft with the CarboClear Vertebral Body Replacement device is optional.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

CarboFix Orthopedics Ltd.

CarboClear® VBR System

Applicant Name

CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel

Contact Person

Yael Rubin CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939

Date Prepared

October 2019

Trade/Proprietary Name

CarboClear® VBR System

Common Name Spinal Vertebral Body Replacement Device

Classification Name

Spinal intervertebral body fixation orthosis (21 CFR §888.3060; Class II; Product Code MQP).

{4}------------------------------------------------

Predicate Devices

Primary

  • Giza® Vertebral Body Replacement (Eden Spine, LLC; K112429, and more)
    Additional

  • I X-MESH™ Cage System (DePuy Spine; K173787, and more)

  • I Ocelot Stackable Cage System (DePuy Spine; K140759, and more)

  • . Bengal Cage System (DePuy Synthes Spine; K073649, and more)

  • . VERTE-STACKTM (Medtronic Sofamor Danek; K070173, and more)

Reference

  • CarboClear Pedicle Screw System (CarboFix Orthopedics Ltd.; K173487, . K182377)
  • FORZA® Spacer System (Orthofix Inc.; K152475, and more) ■

Indications for Use

The CarboClear Vertebral Body Replacement (VBR) System is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g., fracture).

The CarboClear Vertebral Body Replacement device is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

The use of allograft or autograft with the CarboClear Vertebral Body Replacement device is optional.

System Description

The CarboClear VBR System comprises implants (spacers) in different dimensions, and instruments.

{5}------------------------------------------------

The CarboClear VBR spacers are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), with titanium alloy endplates.

Substantial Equivalence

The CarboClear VBR System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable.

Performance characteristics included static and dynamic axial compression and torsion testing per ASTM F 2077, expulsion testing, and endplate shear strength testing. Performance characteristics are comparable to those of predicate devices (as applicable), thus demonstrating that the device is substantially equivalent to the predicates.

In addition, bacterial endotoxin testing was conducted for the CarboClear System.

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.