K Number
K192177
Device Name
SysMed S/T
Date Cleared
2020-11-25

(471 days)

Product Code
Regulation Number
868.5905
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The device provides positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome in self breathing patients weighing over 30kg (66lbs). This product can be used in the home as well as in clinical/hospital environments.
Device Description
The proposed device is a non-invasive, continuous positive airway pressure ventilator, incorporating a heated respiratory humidifier. The device is intended to treat obstructive sleep apnea by delivering a flow of positive airway pressure at a level prescribed by a physician to splint open the airway and prevent airway collapse. The proposed device is available in three types, which are Aurora, Resware and Zizmer. Each type is available in different models designed with different therapy modes and pressure ranges. The therapy modes are available in four types, which are CPAP, Spontaneous (S), Spontaneous/Timed (S/T) and Timed (T). The pressure range for the proposed devices is available in three types, which are 4-20cm H2O, 4-25cm H2O and 4-30cm H2O. Alarm module is incorporated in the device. The device will generate audio and visual alarm for airway leakage, motor malfunction and high motor temperature. The proposed device is provided non-sterile, and not to be sterilized by the user prior to use. The proposed device can be reused by single patient.
More Information

No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on standard CPAP functionalities and therapy modes.

Yes
The device is specifically designed to provide "positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome." This directly indicates it is used for medical treatment, classifying it as a therapeutic device.

No
The device provides positive pressure therapy to treat obstructive sleep apnea, it does not diagnose the condition.

No

The device description clearly outlines hardware components such as a heated respiratory humidifier, motor, and alarm module, indicating it is not software-only.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, tissue) to provide information about a person's health. This testing is done outside of the body (in vitro).
  • Device Function: The description clearly states the device provides "positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome in self breathing patients". It works by delivering air pressure to the airway within the patient's body.
  • No Specimen Analysis: There is no mention of the device collecting or analyzing any biological specimens.

The device is a therapeutic device used to treat a medical condition by directly interacting with the patient's respiratory system.

N/A

Intended Use / Indications for Use

The device provides positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome in self breathing patients weighing over 30kg (66lbs). This product can be used in the home as well as in clinical/hospital environments.

Product codes (comma separated list FDA assigned to the subject device)

BZD

Device Description

The proposed device is a non-invasive, continuous positive airway pressure ventilator, incorporating a heated respiratory humidifier. The device is intended to treat obstructive sleep apnea by delivering a flow of positive airway pressure at a level prescribed by a physician to splint open the airway and prevent airway collapse.

The proposed device is available in three types, which are Aurora, Resware and Zizmer. Each type is available in different models designed with different therapy modes and pressure ranges. The therapy modes are available in four types, which are CPAP, Spontaneous (S), Spontaneous/Timed (S/T) and Timed (T). The pressure range for the proposed devices is available in three types, which are 4-20cm H2O, 4-25cm H2O and 4-30cm H2O. The designated pressure range and therapy mode for each proposed model was provided in following table

TypeModelPressure RangeOperating Mode
Aurora-seriesAurora Bi-Level S4-25 cm H2OCPAP, S
Aurora Bi-Level S/T4-25 cm H2OCPAP, S, S/T
ZiZmer-seriesZiZ CPAP4-20 cm H2OCPAP
Resware-seriesBI 20 S4-20 cm H2OCPAP, S
BI 20 S/T4-20 cm H2OCPAP, S, S/T, T
BI 25 S4-25 cm H2OCPAP, S
BI 25 S/T4-25 cm H2OCPAP, S, S/T, T
BI 30 S/T4-30 cm H2OCPAP, S, S/T, T

Alarm module is incorporated in the device. The device will generate audio and visual alarm for airway leakage, motor malfunction and high motor temperature.

The proposed device is provided non-sterile, and not to be sterilized by the user prior to use. The proposed device can be reused by single patient.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

airway

Indicated Patient Age Range

adult

Intended User / Care Setting

Home as well as in clinical/hospital environments.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • AAMI/ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012 Medical Electrical Equipment-Part 1: General Requirements for Basic Safety and Essential Performance;
  • IEC 60601-1-8:2012 Medical Electrical Equipment-Part 1-8: General requirements for basic safety and essential performance-Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • IEC 60601-1-11:2015 Medical Electrical Equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
  • ISO 80601-2-70:2015 Medical Electrical Equipment-Part 2-70: Particular Requirements for Basic Safety and Essential Performance of Sleep Apnea Breathing Therapy Equipment;
  • ISO 80601-2-74:2017 Medical Electrical Equipment-Part 2-74: Particular Requirements for Basic Safety and Essential Performance of Respiratory Humidifying Equipment;
  • ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity.
  • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization;
  • ISO 10993-12: 2012 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
  • ISO 18562-2: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
  • ISO 18562-3: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds
  • ISO 18562-4: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 4: Tests for leachables in condensate

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K140159

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K153061

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 25, 2020

SysMed (China) Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O Box 120-119 Shanghai, 200120 China

Re: K192177

Trade/Device Name: SysMed S/T Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: October 29, 2020 Received: October 30, 2020

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K192177

Device Name SysMed S/T

Indications for Use (Describe)

The device provides positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome in seff breathing patients weighing over 30kg (66lbs). This product can be used in the home as well as in clinical/hospital environments.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K192177

    1. Date of Preparation: 11/24/2020
    1. Sponsor Identification

SysMed (China) Co., Ltd

11-2-3 No. 17 Wensu Str., Hunnan New Dis., Shenyang, 110171, China

Establishment Registration Number: 3010440667

Contact Person: Yue Jian Position: Quality Manager Tel: +86-24-23970169 Fax: +86-24-23970519 Email: office02@sysmed.cn

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net

4

4. Identification of Proposed Device

Trade Name: SysMed S/T Common Name: Bi-Level S/T

Regulatory Information

Classification Name: ventilator, non-continuous (respirator) Classification: II; Product Code: BZD; Regulation Number: 21CFR 868.5905 Review Panel: Anesthesiology;

Indication for Use Statement:

The device provides positive pressure therapy for the treatment of adult obstructive sleep apnea syndrome in self breathing patients weighing over 30kg (66lbs). This product can be used in the home as well as in clinical/hospital environments.

Device Description

The proposed device is a non-invasive, continuous positive airway pressure ventilator, incorporating a heated respiratory humidifier. The device is intended to treat obstructive sleep apnea by delivering a flow of positive airway pressure at a level prescribed by a physician to splint open the airway and prevent airway collapse.

The proposed device is available in three types, which are Aurora, Resware and Zizmer. Each type is available in different models designed with different therapy modes and pressure ranges. The therapy modes are available in four types, which are CPAP, Spontaneous (S), Spontaneous/Timed (S/T) and Timed (T). The pressure range for the proposed devices is available in three types, which are 4-20cm H2O, 4-25cm H2O and 4-30cm H2O. The designated pressure range and therapy mode for each proposed model was provided in following table

TypeModelPressure RangeOperating Mode
Aurora-seriesAurora Bi-Level S4-25 cm H2OCPAP, S
Aurora Bi-Level S/T4-25 cm H2OCPAP, S, S/T
ZiZmer-seriesZiZ CPAP4-20 cm H2OCPAP
Resware-seriesBI 20 S4-20 cm H2OCPAP, S
BI 20 S/T4-20 cm H2OCPAP, S, S/T, T
BI 25 S4-25 cm H2OCPAP, S

5

BI 25 S/T4-25 cm H2OCPAP, S, S/T, T
BI 30 S/T4-30 cm H2OCPAP, S, S/T, T

Alarm module is incorporated in the device. The device will generate audio and visual alarm for airway leakage, motor malfunction and high motor temperature.

The proposed device is provided non-sterile, and not to be sterilized by the user prior to use. The proposed device can be reused by single patient.

  • న. Identification of Predicate Device
    510(k) Number: K140159 Product Name: S9 WANDA VPAP ST

  • б. Identification of Reference Device
    510(k) Number: K153061 Product Name: Juno VPAP ST-A

    1. Non-Clinical Test Conclusion
      Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • A AAMI/ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012 Medical Electrical Equipment-Part 1: General Requirements for Basic Safety and Essential Performance;

  • A IEC 60601-1-8:2012 Medical Electrical Equipment-Part 1-8: General requirements for basic safety and essential performance-Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

  • IEC 60601-1-11:2015 Medical Electrical Equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

  • A IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

  • ISO 80601-2-70:2015 Medical Electrical Equipment-Part 2-70: Particular Requirements for Basic Safety and Essential Performance of Sleep Apnea Breathing Therapy Equipment;

  • A ISO 80601-2-74:2017 Medical Electrical Equipment-Part 2-74: Particular Requirements for Basic

6

Safety and Essential Performance of Respiratory Humidifying Equipment;

  • ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity.

  • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization;

  • ISO 10993-12: 2012 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials

  • ISO 18562-2: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter

  • ISO 18562-3: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds

  • ISO 18562-4: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 4: Tests for leachables in condensate

7

Substantially Equivalent (SE) Comparison 8.

ItemProposed DevicePredicate DeviceReference Device
ClassificationIIIIII
Product CodeBZDBZDMNS
Regulation Number21 CFR 868.590521 CFR 868.590521 CFR 868.5895
Intended UseThe device provides positive
pressure therapy for the
treatment of adult obstructive
sleep apnea syndrome in self
breathing patients weighing
over 30kg (66lbs). This product
can be used in the home as well
as in clinical/hospital
environments.The S9 WANDA VPAP ST is
indicated for the treatment of
Obstructive Sleep Apnea (OSA)
in patients weighing more than
66lb (30kg). It is intended for
use in the hospital and home.The Juno VPAP ST-A is
indicated to provide
noninvasive ventilation for
patients weighing more than
30lbs (13 kg) with
respiratory insufficiency or
obstructive sleep apnoea
(OSA).
The ¡VAPS mode is
indicated for patients
weighing more than 66lbs
(30 kg). The Juno VPAP
ST-A is intended for home
and hospital use.
The humidifier is intended
for single patient use in the
home environment and
re-use in a
hospital/institutional
environment.
Environment of UseHospital/HomeHospital/HomeHospital/Home
Modes AvailableCPAP;
S;
T
S/T;CPAP
S
T
S/T
VAutoCPAP
S
T
S/T
PAC
¡VAPS
Pressure RangeCPAP: 4-20cmH2O
S, S/T, T: 4-30cmH2OCPAP: 4-20cmH2O
S, S/T, T: 2-25cmH2OCPAP: 4-20cmH2O
S, S/T, T: 3-30cmH2O
SterileNon-sterileNon-sterileNon-sterile
Single UseReuseReuseReuse
Prescription statusPrescriptionPrescriptionPrescription
Supplemental
oxygenNoLabeled for use with
Supplemental OxygenLabeled for use with
Supplemental Oxygen
AlarmYesNoYes
WirelessYesYesYes
HumidifierYesYesYes
RampUser selected as “Off” to 45
minutes in 5 minute incrementsUser selected as “Off” to 45
minutes in 5 minute incrementsUser selected as “Off” to 45
minutes in 5 minute
increments.
Transmit datatherapy data, therapy settings,
upgrade device softwaretherapy data, therapy settings,
upgrade device softwaretherapy data, therapy
settings, upgrade device
software
Data
transfer
mediumSD card, WirelessSD card, WirelessSD card, Wireless
Biocompatibility
CytotoxicityComply with ISO 10993-5
Skin IrritationComply with ISO 10993-10
SensitizationComply with ISO 10993-10
Particulate matterComply with ISO 18562-2Comply with ISO 10993-1, ISO
10993-5, ISO 10993-10 and ISO
10993-12 standardsComply with ISO 10993-1,
ISO 10993-5, ISO 10993-10
and ISO 10993-12 standards
Volatile
organic
compounds
(VOCs)Comply with ISO 18562-3Comply with ISO 10993-1, ISO
10993-5, ISO 10993-10 and ISO
10993-12 standardsComply with ISO 10993-1,
ISO 10993-5, ISO 10993-10
and ISO 10993-12 standards
Leachables in
condensateComply with ISO 18562-4
Comply with
IEC 60601-1-8: 2012;
AAMI/ANSI
ES60601-1:2005/(R) 2012 And
A1:2012, C1:2009/(R) 2012
And A2:2010/(R) 2012;
IEC 60601-1-2: 2014
IEC 60601-1-11: 2011
ISO 80601-2-70:
ISO 80601-2-74Comply with
IEC 60601-1:2005+AMD1:
2012
IEC 60601-1-2:2014
IEC 60601-1-8: 2006
IEC 60601-1-11:2010Comply with
IEC 60601-1:2005+AMD1:
2012
IEC 60601-1-2:2014
IEC 60601-1-8: 2006
IEC 60601-1-11:2010
Electrical Safety
and EMC

Table 1 Comparison of Technology Characteristics

8

    1. Substantially Equivalent (SE) Conclusion
      Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.