K Number
K192156
Manufacturer
Date Cleared
2019-09-04

(26 days)

Product Code
Regulation Number
888.4505
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Orthopedic surgical instrumentation supplied by Cartiva, Inc. is indicated for use in the press-fit fixation of the implantable medical device products manufactured by Cartiva, Inc.

Device Description

Orthopedic surgical instruments designed for osteochondral implants with press-fit fixation are devices intended to manipulate bone and cartilage tissue or the implant for the positioning, alignment, defect creation, and placement of press-fit osteochondral implants that utilize no additional means of fixation (e.g., suture fixation, adhesives). This type of device includes instruments specific to the geometry of the implant.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Cartiva® SCI Reusable Instrumentation Set. This submission is for orthopedic surgical instrumentation, specifically an instrumentation set, not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device studies (such as sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set information) are not applicable to this document.

The document focuses on demonstrating substantial equivalence to a predicate device for this instrumentation set, specifically the addition of two new sizes (6mm and 12mm) to an existing product line.

Here's an analysis of the provided information, tailored to the context of a medical instrumentation submission:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Type of Testing)Reported Device Performance (Summary of Results)
Biocompatibility per ISO 10993-1Cytotoxicity (ISO 10993-5): Data demonstrated safety.
Sensitization (ISO 10993-10): Data demonstrated safety.
Irritation (ISO 10993-10): Data demonstrated safety.
Reprocessing
- Cleaning validationCleaning validation per AAMI TIR30 and AAMI TIR12: Data demonstrated safety and performance was not affected.
- Sterilization validationSterilization validation per AAMI TIR39, ISO 17665-1 and ANSI/AAMI ST79: Data demonstrated safety and performance was not affected.
Performance/Functional Testing
- Ship studiesShip studies per ASTM D4169: Data demonstrated functionality.
- Simulated Use TestingSimulated Use Testing in beyond use conditions to demonstrate functionality: Data demonstrated functionality.
Overall ConclusionAll data demonstrated that the safety and performance of the Cartiva SCI Reusable Instrumentation is not affected by the proposed modification to add two additional sizes, 6mm and 12mm, to the Cartiva existing reusable instrumentation product line. The new Cartiva SCI Reusable Instrumentation sizes 6mm and 12mm have demonstrated to provide the same level of performance as the predicate device, Cartiva SCI Instrumentation. The modification does not raise any new questions of safety or effectiveness.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

This information is not provided in the document. The filing is for a physical medical device (instrumentation set), and the "testing" performed relates to material properties, cleaning, sterilization, and basic functionality, not clinical performance with a test set of data in the way an AI/ML device would be evaluated. The tests performed are likely in vitro and bench-top studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not applicable as the device is an instrumentation set and does not involve AI/ML or diagnostic assessment that would require clinician-established ground truth. Expert involvement would be in the design, testing protocols, and assessment of physical device performance, but not in establishing ground truth for diagnostic accuracy.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not applicable for the same reasons as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is an instrumentation set, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable as the device is an instrumentation set, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not applicable for the clinical context as the device is an instrumentation set. For the engineering performance testing, the "ground truth" would be established by validated test methods (e.g., ISO, AAMI, ASTM standards) against predetermined specifications for material properties, sterility assurance levels, cleaning efficacy, and functional integrity.

8. The sample size for the training set

This information is not applicable as the device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cartiva, Inc. Shriya Kafle Regulatory Specialist 6120 Windward Parkway, Suite 220 Alpharetta, Georgia 30005

Re: K192156

Trade/Device Name: Cartiva® SCI Reusable Instrumentation Set Regulation Number: 21 CFR 888.4505 Regulation Name: Orthopedic surgical instrumentation designed for osteochondral implants with pressfit fixation Regulatory Class: Class II Product Code: QBO Dated: August 8, 2019 Received: August 9, 2019

Dear Shriya Kafle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192156

Device Name Cartiva® SCI Reusable Instrumentation Set

Indications for Use (Describe)

Orthopedic surgical instrumentation supplied by Cartiva, Inc. is indicated for use in the press-fit fixation of the implantable medical device products manufactured by Cartiva, Inc.

Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "CARTIVA" in large, green letters. Below that, in smaller, blue letters, are the words "Synthetic Cartilage Implant". The background is a dark blue color.

WRIGHT.

K192156 Page 1 of 2

510(k) Number: K192156 Date Submitted: August 9, 2019

510(k) Summary

SUBMITTER'S INFORMATION

Owner:Cartiva, Inc.
Address:6120 Windward Parkway, Suite 220, Alpharetta, GA 30005
Phone:770-754-3817
Fax Numbers:770-754-3808
Contact Person:Shriya Kafle, Regulatory Specialist

DEVICE INFORMATION

Name of Device:Cartiva SCI Reusable Instrumentation Set
Common/Usual Name:Orthopedic Surgical Instrumentation
Classification Name:Orthopedic Surgical Instrumentation designed for osteochondralimplants with press-fit fixation (Product Code QBO)
Predicate Device(s):Cartiva SCI Instrumentation (Product Code QBO, RegulationNumber 21 CFR 888.4505)
Device Description:Orthopedic surgical instruments designed for osteochondralimplants with press-fit fixation are devices intended tomanipulate bone and cartilage tissue or the implant for thepositioning, alignment, defect creation, and placement of press-fit osteochondral implants that utilize no additional means offixation (e.g., suture fixation, adhesives). This type of deviceincludes instruments specific to the geometry of the implant.
Indication for Use:Orthopedic surgical instrumentation supplied by Cartiva, Inc. isindicated for use in the press-fit fixation of the implantablemedical device products manufactured by Cartiva, Inc
TechnologicalCharacteristics:The line extension of Cartiva SCI Reusable Instrumentation toadd two additional sizes (6mm and 12mm) to the existingreusable instrumentation product line do not raise new questionsof safety or effectiveness. All technological aspects of press-fitfixation are preserved.
Comparison to PredicateDevice:, Suite 220The Cartiva SCI Reusable Instrumentation set use andperformance characteristics are not altered by this modificationto add two additional sizes, (6mm and 12mm) to the existingCartiva reusable instrumentation product line.
Performance Data:Testing of the Cartiva Reusable Instrumentation set was carried out, including:
Biocompatibility per ISO 10993-1Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-10) Irritation (ISO 10993-10)
Reprocessing:Cleaning validation:Cleaning validation per AAMI TIR30 and AAMI TIR12 Sterilization ValidationSterilization validation per AAMI TIR39, ISO 17665-1 and ANSI/AAMI ST79
Performance/Functional Testing Ship studies per ASTM D4169 Simulated Use Testing in beyond use conditions to demonstrate functionality
All data demonstrated that the safety and performance of the Cartiva SCI Reusable Instrumentation is not affected by the proposed modification to add two additional sizes, 6mm and 12mm, to the Cartiva existing reusable instrumentation product line.
A Clinical Evaluation was determined not to be required as the device design, intended use and indication for use are all preserved.
Rationale for Substantial Equivalence:This modification falls within the FDA regulations for 510(k) review. The indication for use, intended use, principles of operation, and performance have not been altered. The minor change to add the new Cartiva SCI Reusable Instrumentation sizes 6mm and 12mm to the existing Cartiva Reusable Instrumentation portfolio (8mm and 10mm) do not raise any new questions of safety or effectiveness. The new Cartiva SCI Reusable Instrumentation sizes 6mm and 12mm have demonstrated to provide the same level of performance as the predicate device, Cartiva SCI Instrumentation, (Product Code QBO, Regulation Number 21 CFR 888.4505). Therefore, the new Cartiva SCI Reusable Instrumentation Set (6mm and 12mm) is substantially equivalent to the predicate Cartiva SCI Instrumentation, (Product Code QBO, Regulation Number 21 CFR 888.4505).
Conclusion:The Cartiva SCI Reusable Instrumentation, as modified by this 510(k), do not raise any new issues regarding safety or effectiveness, and therefore is suitable for commercial sale.

Wright Medical Group N 6120 Windward Parkway Alpharetta, GA 30005 877.336.4616

cartiva.net | wright.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "Cartiva Synthetic Cartilage Implant". The word "Cartiva" is in a large, light green font. Below that, the words "Synthetic Cartilage Implant" are in a smaller, white font.

TT WRIGHT.

K192156 Page 2 of 2

Wright Medical Group N.V.
6120 Windward Parkway, Suite 220
Alpharetta, GA 30005 877.336.4616

cartiva.net | wright.com

§ 888.4505 Orthopedic surgical instrumentation designed for osteochondral implants with press-fit fixation.

(a)
Identification. Orthopedic surgical instruments designed for osteochondral implants with press-fit fixation are hand-held devices intended to manipulate bone and cartilage tissue or the implant for the positioning, alignment, defect creation, and placement of press-fit osteochondral implants that utilize no additional means of fixation (e.g., suture fixation, adhesives). This type of device includes instruments specific to the geometry of the implant.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9. The special controls for this device are:(1) Technical specifications regarding geometry of the instruments must be specified and validated to demonstrate that the instruments can safely position and place the implant.
(2) The patient contacting components of the device must be demonstrated to be biocompatible.
(3) Labeling must include:
(i) Identification of implant(s) and instruments which have been validated for use together; and
(ii) Validated methods and instructions for reprocessing any reusable parts.