(87 days)
The VITEK®DENSICHEK® instrument is an accessory intended for use with the VITEK® 2 Systems to measure the optical density of a microorganism suspension. The instrument provides values in McFarland units, proportional to the microorganism concentration. VITEK® is intended for use with polystyrene tubes, and the reading range is 0.00 to 4.00 McFarland. The VITEK® DENSICHEK® has applications as an in vitro diagnostic medical device, or in an industry setting.
The VITEK® DENSICHEK® consists of the following accessories and components:
- . VITEK® DENSICHEK® Pod
- VITEK® DENSICHEK® Display Base ●
- VITEK® DENSICHEK® Connectivity Base ●
- . VITEK® DENSICHEK® McFarland Reference Kit
- USB Power Adaptor .
The VITEK® DENSICHEK® is an optical reading device intended for professional use by laboratory health professionals in clinical or industry laboratory settings. The device determines the McFarland value by evaluating the absorption of light in the test tube containing the suspension. The more turbid the organism suspension is, the higher the McFarland value displayed by the device. An algorithm assesses results obtained from the turbidimetric reading and translates them into a McFarland value.
The VITEK® DENSICHEK® instrument contains a Base unit with a detachable optical interface, the Pod. The Pod calculates and automatically transmits the optical McFarland reading to the Base unit for display. The user can confirm the McFarland value by pressing a button on the Pod, optionally saving the value to the VITEK® FLEXPREP™ Software on a laboratory PC, where it is recorded with specimen setup information for traceability.
The Pod McFarland readings are communicated to the Base through a Bluetooth Low Energy (BTLE) wireless interface. When docked on a Base, the Pod is automatically paired and recharged through a wireless contact connection. The Display Base displays McFarland readings from the Pod on an LCD resistive touch screen, and can be integrated with VITEK® 2 software to record readings for traceability. The Connectivity Base is available for customers who only want to use the VITEK® DENSICHEK® with the VITEK® 2 software to display and record McFarland readings. McFarland Reference Standards featuring radio-frequency identification (RFID) tags are used to verify the measurement accuracy performance of the VITEK® DENSICHEK®.
The provided text from the FDA 510(k) summary for VITEK® DENSICHEK® does not contain information typically found in an AI/ML device submission, such as a ground truth established by experts, MRMC studies, or training/test set details in the context of deep learning models. This document describes a medical device used to measure optical density for microbial suspensions, which is a physical measurement device, not an AI/ML algorithm. Therefore, many of the requested fields are not applicable in this context.
Based on the provided text, here's an analysis of the acceptance criteria and study data:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for reproducibility studies typically involve a certain percentage of agreement or reproducibility rate. While explicit numerical acceptance criteria are not tabulated, the text indicates the performance was "acceptable" based on established norms for this type of device.
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Study 1: Quality Control & Reproducibility | Acceptable reproducibility (>95%) | VITEK® DENSICHEK® demonstrated acceptable quality control performance of >95% and reproducibility for isolates. No exceptions were noted. |
| Study 2: Equivalency (PB2 vs. PB1) | High level of agreement, small difference in reproducibility rates | Exact Agreement (2-fold dilution): Nearly 90.0% Essential Agreement (2-fold dilution): 99.5% Combined Reproducibility Rate Difference: No more than 0.4% (both best and worst case) |
| Study 3: Abbreviated Reproducibility (PB2 vs. DensiCHEK® Plus) | Performance criteria of >95% across all sites reproducibility | VITEK® DENSICHEK® Product Build 2 met the performance criteria of >95% across all sites reproducibility with no observed exceptions. Agreement Rate within ± 1 doubling dilution (PB2 vs. Plus): Nearly 99% |
2. Sample Size Used for the Test Set and Data Provenance
The text refers to "isolates tested" and "strain / antimicrobial combination" but does not explicitly state the total number of unique samples (e.g., number of microbial suspensions tested) for each study.
- Study 1: "isolates tested with Gram-negative, Gram-positive, Streptococus, and Yeast AST card types." No specific sample size provided. Provenance likely clinical setting given the description.
- Study 2:
- Distribution of Dilution Differences:
- Auto: 190 strain/antimicrobial combinations
- Manual: 190 strain/antimicrobial combinations
- Combined: 380 strain/antimicrobial combinations
- Reproducibility Rates: The numbers derive from the same underlying data as the dilution differences.
- Data Provenance: Not explicitly stated, but implied to be from internal testing ("comparison of modes between the two product builds").
- Distribution of Dilution Differences:
- Study 3:
- Distribution of Dilution Differences & Reproducibility Rates:
- Auto: 190 strain/antimicrobial combinations
- Manual: 190 strain/antimicrobial combinations
- Combined: 380 strain/antimicrobial combinations
- Data Provenance: "A subset of the reproducibility data collected internally using VITEK® Product Build 2 was combined with data collected at two external sites." This indicates a combination of internal and external prospective clinical data from multiple sites. Country of origin not specified.
- Distribution of Dilution Differences & Reproducibility Rates:
3. Number of Experts Used to Establish the Ground Truth and Qualifications
Not applicable. This device measures optical density. The "ground truth" for optical density measurements would be established by physical standards and calibrations, not by human expert interpretation in the way it is for image-based diagnostic AI.
4. Adjudication method for the test set
Not applicable. This is not an interpretive device requiring human adjudication of results.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI/ML algorithm that assists human readers. It is a measurement instrument.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the core function of the device, which is an automated measurement of optical density. Its performance is evaluated as a standalone instrument. The data provided in the tables (e.g., dilution differences, reproducibility rates) represent the performance of the device itself (VITEK® DENSICHEK®) and its comparison to a predicate device (DensiCHEK® Plus) or different product builds, not a human-in-the-loop scenario.
7. The type of ground truth used
The "ground truth" in this context is the actual (or reference) McFarland values and the expected behavior/reproducibility of the optical density measurements. This would be established through:
- Reference Standards: The document mentions "McFarland Reference Standards" used to verify measurement accuracy. These are likely chemically or physically calibrated standards.
- Comparisons to Predicate Devices: The DensiCHEK® Plus serves as a comparative "ground truth" or reference for equivalency, implying its established performance is the benchmark.
- Reproducibility: The consistency of the device's own measurements across repeated tests provides the "ground truth" for its reproducibility.
It is not based on expert consensus, pathology, or outcomes data in the traditional sense of diagnostic imaging or clinical trials for disease states.
8. The sample size for the training set
Not applicable in the context of AI/ML training data. This device is a physical instrument, not an AI/ML model trained on a dataset. Its "training" would involve engineering design, calibration, and validation, not AI model training.
9. How the ground truth for the training set was established
Not applicable, as this is not an AI/ML device with a "training set" in that sense.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 31, 2019
bioMerieux, Inc. Nathan Hardesty Sr. Manager, Regulatory Affairs 595 Anglum Rd. Hazelwood, Missouri 63042
Re: K192110
Trade/Device Name: VITEK DENSICHEK Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON Dated: August 2, 2019 Received: August 5, 2019
Dear Nathan Hardesty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ribhi Shawar -S
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K192110
Device Name VITEK DENSICHEK
Indications for Use (Describe)
The VITEK DENSICHEK instrument is an accessory intended for use with the VITEK 2 Systems to measure the optical density of a microorganism suspension. The instrument provides values in McFarland units, proportional to the microorganism concentrations. VITEK DENSICHEK is intended for use with polystyrene tubes, and the reading range is 0.00 to 4.00 McFarland. The VITEK DENSICHEK has applications as an in vitro diagnostic medical device, or in an industry setting.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with the top half being blue and the bottom half being yellow-green. The company name, "BIOMÉRIEUX", is written in white, sans-serif font across the center of the circle, positioned on the blue portion.
510(k) SUMMARY
VITEK® DENSICHEK®
A. 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|---|---|
| Address: | 595 Anglum RoadHazelwood, MO 63042 |
| Contact Person: | Jennifer Jines |
| Phone Number: | 314-731-8352 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | August 2, 2019 |
| B. Device Name: | |
| Formal/Trade Name: | VITEK® DENSICHEK® |
| Regulation Section: | 21 CFR 866.1645 |
| Classification Name: | Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility System |
LON
Microbiology
VITEK® DENSICHEK®
Common Name:
C. Predicate Device:
Panel:
Product Code:
| Predicate Device Name: | DensiCHEK® Plus |
|---|---|
| ------------------------ | ----------------- |
Predicate Device Number: K083536
D. 510(k) Summary:
E. Intended Use:
The VITEK®DENSICHEK® instrument is an accessory intended for use with the VITEK® 2 Systems to measure the optical density of a microorganism suspension. The instrument provides values in McFarland units, proportional to the microorganism concentration. VITEK® is intended for use with polystyrene tubes, and the reading range is 0.00 to 4.00 McFarland. The VITEK® DENSICHEK® has applications as an in vitro diagnostic medical device, or in an industry setting.
F. Test Principle:
The VITEK® DENSICHEK® instrument is designed for use with the VITEK® 2 Systems to measure the optical density of a microorganism suspension. The instrument provides values in McFarland units,
{4}------------------------------------------------
proportional to the microorganism concentrations. It is intended for use with the VITEK® 2 Systems during the inoculum preparation phase of the Identification (ID) and Antimicrobial Susceptibility Test (AST) card testing process.
The VITEK®DENSICHEK® generates a McFarland value using basic colorimetry, which is a method of measurement that relates the amount of color in a transparent medium (liquid) to the amount of a particular substance in the liquid. In general the concentration of the substance being measured is proportional to the intensity of the solution. The darker the color becomes, the higher the concentration.
The VITEK® DENSICHEK® measures the turbidity of the saline and microorganism suspension using a single wavelength, 635nm, light emitting diode (LED). The absorption of light determines the McFarland value measured by the instrument. The more turbid the suspension, the higher the McFarland measurement reported by the device. The light source in the VITEK® emits a narrow range of wavelengths, which pass through the test tube containing the measured suspension and into a photodiode detector. The instrument is first zeroed using a test tube filled with saline (blank). The technician then places the well-mixed organism suspension into the instrument and slowly rotates the test tube. The instrument will display a series of dashes followed by a McFarland reading, based on the amount of energy that was detected at the photodiode.
G. Device Description:
The VITEK® DENSICHEK® consists of the following accessories and components:
- . VITEK® DENSICHEK® Pod
- VITEK® DENSICHEK® Display Base ●
- VITEK® DENSICHEK® Connectivity Base ●
- . VITEK® DENSICHEK® McFarland Reference Kit
- USB Power Adaptor .
The VITEK® DENSICHEK® is an optical reading device intended for professional use by laboratory health professionals in clinical or industry laboratory settings. The device determines the McFarland value by evaluating the absorption of light in the test tube containing the suspension. The more turbid the organism suspension is, the higher the McFarland value displayed by the device. An algorithm assesses results obtained from the turbidimetric reading and translates them into a McFarland value.
The VITEK® DENSICHEK® instrument contains a Base unit with a detachable optical interface, the Pod. The Pod calculates and automatically transmits the optical McFarland reading to the Base unit for display. The user can confirm the McFarland value by pressing a button on the Pod, optionally saving the value to the VITEK® FLEXPREP™ Software on a laboratory PC, where it is recorded with specimen setup information for traceability.
The Pod McFarland readings are communicated to the Base through a Bluetooth Low Energy (BTLE) wireless interface. When docked on a Base, the Pod is automatically paired and recharged through a wireless contact connection. The Display Base displays McFarland readings from the Pod on an LCD resistive touch screen, and can be integrated with VITEK® 2 software to record readings for traceability. The Connectivity Base is available for customers who only want to use the VITEK® DENSICHEK® with the VITEK® 2 software to display and record McFarland readings. McFarland Reference Standards featuring radio-frequency identification (RFID) tags are used to verify the measurement accuracy performance of the VITEK® DENSICHEK®.
{5}------------------------------------------------
H. FDA Recognized Consensus Standards Referenced:
- M100S, "Performance Standards for Antimicrobial Susceptibility Testing, 28th Edition", . 28th Edition (January 2018)
- M07-A10. " Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard - Tenth Edition," Vol. 35 No.2 (January 2015)
- . UL 61010-1: 2012 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
- EN 62471: 2008 Photobiological safety of lamps and lamp systems ●
l. Performance Characteristics:
Clinical Studies
The Premarket Notification (510[k]) presents data in support of VITEK® DENSICHEK®. The purpose of this study/trial is to evaluate the reproducibility performance of the VITEK® device as used with the VITEK®2 and VITEK®2 Compact Systems in a clinical setting.
Study 1
The purpose of Study 1 was to evaluate the reproducibility and quality control performance of the VITEK® DENSICHEK® device as used with the VITEK® 2 and VITEK® 2 Compact Systems in a clinical setting. Data was collected using VITEK® DENSICHEK® and the DensiCHEK® Plus. The VITEK® DENSICHEK® demonstrated acceptable quality control performance of >95% and reproducibility for isolates tested with Gram-negative, Gram-positive, Streptococus, and Yeast AST card types. There were no exceptions for this performance data. As such, the overall quality control and reproducibility performance of VITEK® DENSICHEK® was acceptable.
Study 2
The purpose of Study 2 was an Equivalency Study to validate the equivalence of VITEK® DENSICHEK® Product Build 1, and the VITEK® DENSICHEK® Product Build 2. Reproducibility data was collected using both product builds and the performance of the VITEK® DENSICHEK® Product Build 1 was compared to the performance of the VITEK® DENSICHEK® Product Build 2.
The table below shows the distribution of two-fold dilution differences between modes by dilution method and combined. For this overall comparison, any strain / antimicrobial combination having a mode that was offscale was ignored. The rationale for this is that variation between modes can be measured more accurately using only on-scale organisms.
| Dilution Method | Dilution Difference | Essential Agreement | ||||
|---|---|---|---|---|---|---|
| ≤-2 | -1 | 0 | +1 | ≥ +2 | ||
| Auto | 00.0% | 126.3% | 17290.5% | 63.2% | 00.0% | 190 / 190100.0% |
| Manual | 00.0% | 115.8% | 16888.4% | 94.7% | 21.1% | 188 / 19098.9% |
| Combined | 00.0% | 236.1% | 34089.5% | 153.9% | 20.5% | 378 / 38099.5% |
Distribution of Dilution Differences Between Product Build Modes ( PB2 - PB1)
{6}------------------------------------------------
The table below provides a summary of the overall reproducibility rates by dilution method and combined for each product build.
| Dilution | Best Case | Worst Case | ||||
|---|---|---|---|---|---|---|
| Method | ProductBuild 1 | ProductBuild 2 | Difference | ProductBuild 1 | ProductBuild 2 | Difference |
| Auto | 99.8% | 99.4% | 0.4% | 98.9% | 98.6% | 0.3% |
| Manual | 99.0% | 98.5% | 0.5% | 98.0% | 97.6% | 0.4% |
| Combined | 99.4% | 99.0% | 0.4% | 98.4% | 98.0% | 0.4% |
Overall Reproducibility Rate Comparison
The overall comparison of modes between the two product builds reveals that the exact agreement rate was nearly 90.0%, and the essential agreement rate was 99.5%. Furthermore, the combined reproducibility rates for both product builds differ by no more than 0.4%, overall, for both best and worst case. The observed level of agreement between modes indicates that both product builds are associated with results that occur at a similar location on the MIC scale. The small difference in reproducibility rates provides evidence that AST result variability is nearly identical for both PB1 and PB2.
The best-case reproducibility was acceptable for all drugs evaluated for a representative class of Gramnegative, Gram-positive, Streptococcus, and Yeast drugs.
Study 3
The purpose of Study 3 was an Abbreviated Reproducibility Study was to assess the performance of the VITEK® DENSICHEK® Product Build 2 when end-user variability was introduced. A subset of the reproducibility data collected internally using VITEK® Product Build 2 was combined with data collected at two external sites.
VITEK® DENSICHEK® Product Build 2 met the performance criteria of >95% across all sites reproducibility with no observed exceptions.
Data analysis was then completed to compare reproducibility directly between Product Build 2 (PB2) and DENSICHEK Plus (Plus). Table 1 provides a summary of the dilution differences between modes from the two devices by auto dilution, manual dilution, and combined. Table 2 shows a comparison of reproducibility rates from the two devices. As with the original analysis comparing Product Build 1 and Product Build 2, organism/antimicrobial combinations having a mode that was off-scale were removed from the analysis. Again, the rationale being reproducibility or variation between results can be measured more accurately using on-scale organisms.
{7}------------------------------------------------
| DilutionMethod | Dilution Difference | EssentialAgreement | ||||
|---|---|---|---|---|---|---|
| ≤ -2 | -1 | 0 | +1 | ≥ +2 | ||
| Auto | 21.1% | 2312.1% | 16084.2% | 52.6% | 00.0% | 188/19098.9% |
| Manual | 21.1% | 2010.5% | 16586.8% | 31.6% | 00.0% | 188/19098.9% |
| Combined | 41.1% | 4311.3% | 32585.5% | 82.1% | 00.0% | 376/38098.9% |
Table 1: Distribution of Dilution Differences Between Modes from Reproducibility Testing (PB2 – Plus)
| Table 2: Overall Reproducibility Rate Comparison | |||||
|---|---|---|---|---|---|
| -------------------------------------------------- | -- | -- | -- | -- | -- |
| DilutionMethod | Best Case | Worst Case | ||||
|---|---|---|---|---|---|---|
| Plus | PB2 | Difference | Plus | PB2 | Difference | |
| Auto | 99.4% | 99.7% | -0.3% | 98.6% | 98.8% | -0.2% |
| Manual | 99.2% | 99.1% | 0.1% | 98.2% | 98.2% | 0.0% |
| Combined | 99.3% | 99.4% | -0.1% | 98.4% | 98.5% | -0.1% |
From the comparison of modes, the agreement rate within ± 1 doubling dilution is nearly 99% indicating a high level of correlation between results from the two devices. For the comparison of reproducibility rates, the differences are small for both dilution methods.
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”