K Number
K192062
Device Name
Nucleoss T6 Dental Implant System
Date Cleared
2020-08-27

(392 days)

Product Code
Regulation Number
872.3640
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Nucleoss T6 Dental Implants are intended to be surgically placed in the bone of the maxillary and/or mandibular arches to provide support for prosthetic restorations (crowns, bridges or overdenture) in edentulous or partially edentulous patients to restore a patients' chewing function. Nucleoss T6 Dental Implants are intended for delayed loading after 12 weeks. Nucleoss Abutments and Prosthetic parts are intended for use with Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures for edentulous or partially edentulous patients.
Device Description
The Nucleoss T6 Dental Implant is a bone level implant constructed of unalloyed titanium (ISO 5832-2). The surface of the T6 implant is a sand-blast and acid-etch (SLA) surface treatment. T6 Implants: Nucleoss T6 implants have a cylindrical form design, with a double lead thread form and two helical anti-rotation grooves. The thread structure is a reverse buttress. The internal structure is designed as a conical internal hex connection with 140 degrees. The Nucleoss Dental Abutments are intended for use with the Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients. Abutments: The Nucleoss Dental Abutments consist of the following designs: Standard Straight, Esthetic Straight, Esthetic Angled, Ball and Equator, Cover, Gingiva Former and Screw. Nucleoss Dental Abutments are cement-retained and screw-retained restorations and intended for placement on Nucleoss T6 implants with diameters of 3.5, 4.1 and 4.8 mm. The connection to the dental implant is achieved by an internal hexagon. These devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3. Covers and Gingiva Formers: The Nucleoss Cover and Gingiva Former are used during the healing and prosthesis period following surgical placement of the dental implant. All of these devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3. Screws: The Nucleoss Dental Implant and Abutments devices also include screws for securing abutments to the implant to provide a secure coupling for the denture prosthetic attachment to the abutment.
More Information

No
The summary describes a physical dental implant system and does not mention any software, algorithms, or data processing that would indicate the use of AI or ML.

No.
The device is intended to restore chewing function through prosthetic restorations, which is a restorative rather than a therapeutic function.

No

The device is a dental implant system intended for surgical placement to support prosthetic restorations, restoring chewing function. It is a therapy device, not a diagnostic one.

No

The device description clearly outlines physical components made of titanium and titanium alloy, such as implants, abutments, covers, gingiva formers, and screws. There is no mention of software as the primary or sole component.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the Nucleoss T6 Dental Implants and associated parts are surgically placed in the bone to support prosthetic restorations and restore chewing function. This is a direct therapeutic intervention on the patient's body.
  • Device Description: The description details the physical components of the implants, abutments, covers, gingiva formers, and screws, all of which are physical devices implanted or used in the mouth.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue, etc.) to provide information for diagnosis, monitoring, or treatment.

IVD devices are used outside the body to analyze samples and provide diagnostic information. This device is a surgical implant used inside the body for structural support and functional restoration.

N/A

Intended Use / Indications for Use

Nucleoss T6 Dental Implants are intended to be surgically placed in the bone of the maxillary and/or mandibular arches to provide support for prosthetic restorations (crowns, bridges or overdenture) in edentulous or partially edentulous patients to restore a patients' chewing function. Nucleoss T6 Dental Implants are intended for delayed loading after 12 weeks.

Nucleoss Abutments and Prosthetic parts are intended for use with Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures for edentulous or partially edentulous patients.

Product codes (comma separated list FDA assigned to the subject device)

DZE, NHA

Device Description

The Nucleoss T6 Dental Implant is a bone level implant constructed of unalloyed titanium (ISO 5832-2). The surface of the T6 implant is a sand-blast and acid-etch (SLA) surface treatment.

Nucleoss T6 implants have a cylindrical form design, with a double lead thread form and two helical anti-rotation grooves. The thread structure is a reverse buttress. The internal structure is designed as a conical internal hex connection with 140 degrees.

The Nucleoss Dental Abutments are intended for use with the Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.

The Nucleoss Dental Abutments consist of the following designs: Standard Straight, Esthetic Straight, Esthetic Angled, Ball and Equator, Cover, Gingiva Former and Screw. Nucleoss Dental Abutments are cement-retained and screw-retained restorations and intended for placement on Nucleoss T6 implants with diameters of 3.5, 4.1 and 4.8 mm. The connection to the dental implant is achieved by an internal hexagon. These devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3.

The Standard Straight Abutment is intended for cement-retained single and multiple restorations. There is no angulation. The Standard Straight abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Standard Straight abutment is secured on the implant with a screw.

The Esthetic Straight Abutment is intended for cement-retained single and multiple restorations. There is no angulation. The Esthetic Straight abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Esthetic Straight abutment is secured on the implant with a screw. The front and back gingival heights of these abutments are different from each other. The front heights are shorter and the back heights are greater. These height differences offer the dentist an esthetically beneficial option as required.

The Esthetic Angled Abutment is intended for cement-retained single and multiple restorations. The Esthetic Angled Abutment is provided in angles of 15 and 25 degrees for cases where angle correction is required. The Esthetic Angled abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Esthetic Angled abutment is secured on the implant with a screw. The front and back gingival heights of these abutments are different from each other. The front heights are shorter and the back heights are greater. These height differences offer the dentist an esthetically beneficial option as required.

The Angled Abutment is intended for cement-retained single and multiple restorations. The Angled Abutment is provided in angles of 15 and 25 degrees for cases where an angle correction is required. The Angled Abutment can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Angled Abutment is secured on the implant with a screw.

The Ball abutment is intended for overdenture restorations. The top part of the abutment is designed as a ball with a dimeter of 2.25 mm. There is no angulation. The Ball Abutment can be used in anterior and posterior maxillary and/or mandibular arches. The Ball abutment is inserted and combined with the fixture in order to support the overdenture.

The Equator abutment is intended for overdenture restorations. There is no angulation. The Equator Abutment can be used in anterior and posterior maxillary and/or mandibular arches. The Equator abutment is inserted and combined with the fixture in order to support the overdenture.

The Abutments are single-use devices. The Abutments are provided non-sterile for end-user steam sterilization.

The Nucleoss Cover and Gingiva Former are used during the healing and prosthesis period following surgical placement of the dental implant. All of these devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3. Covers and Gingiva Formers are single-use devices. The Cover is provided sterile in the same packaging tube as the Nucleoss T6 Dental Implant. The Gingiva Formers are provided nonsterile for end-user steam sterilization.

The Nucleoss Dental Implant and Abutments devices also include screws for securing abutments to the implant to provide a secure coupling for the denture prosthetic attachment to the abutment. They are provided non-sterile and are packaged together with the abutments.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

maxillary and/or mandibular arches

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical testing of the Nucleoss T6 Implant System was performed following the FDA guidance: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments, and applicable ISO and ASTM standards.

Bench Testing: The Pull-out test was performed to compare the Nucleoss T6 Implant and reference device in accordance with ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws with acceptable test results. A comparative surface area analysis was conducted on the Nucleoss T6 Implant and reference device with the original length and 3mm bone loss.

Fatigue Testing: Fatigue testing conducted in accordance with ISO 14801:2016 Dentistry- Implants-Dynamic Fatigue Test for Endosseous Dental Implants found the durability of the Nucleoss T6 Implant and Abutment combinations was acceptable. Testing was conducted on the worst-case implant abutment combinations.

Sterilization Validation: Nucleoss T6 Implants are sterilized using a gamma ray sterilization process that has been validated in accordance with ISO 11137-2 Sterilization of healthcare products-Radiation-Part 2 establishing the sterilization dose to ensure a SAL of 10-6. Nucleoss Dental Abutments are provided non-sterile for end-user steam sterilization using a traditional cycle process. Since the raw material, production process, parameters and process materials of both Tpure and the predicate device are exactly same, the same steam sterilization parameters apply to the T6 Dental abutments as was used in the predicate device.

Accelerated Aging Test and Packaging Validation Test: Nucleoss T6 implants will be packaged and labeled based on new testing that was conducted to support a shelf life of 5 years on the predicate device. Package Validation and Simulated transportation testing on the predicate device that was accelerated aged to an equivalent 5 years found that the sterility of the dental implants was maintained. Accelerated aging and packaging validation testing was not repeated for the T6 Dental Implants since the raw material, production process, parameters, process materials and all packaging materials/type of the T6 Dental Implants are exactly same as those of the predicate device.

Biocompatibility: All the materials used the T6 Dental Implant and the manufacturing processes were identical to those used in the predicate device. Biocompatibility testing was not repeated for the T6 Dental Implants since the raw material, production process, parameters and process materials of the T6 Dental Implants and abutments and the predicate device are exactly the same.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K160850

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K171784, K093615

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.

August 27, 2020

Sanlilar Tibbi Cihazlar Medikal Kimya Sanayi Ticaret Ltd. Sti % Semih Oktay President CardioMed Device Consultants 3168 Braverton St., Suite 200 Edgewater, Maryland 21037

Re: K192062

Trade/Device Name: Nucleoss T6 Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 24, 2020 Received: July 27, 2020

Dear Semih Oktay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known): K192062

Device Name: Nucleoss T6 Dental Implant System

Indications for Use:

Nucleoss T6 Dental Implants are intended to be surgically placed in the bone of the maxillary and/or mandibular arches to provide support for prosthetic restorations (crowns, bridges or overdenture) in edentulous or partially edentulous patients to restore a patients' chewing function. Nucleoss T6 Dental Implants are intended for delayed loading after 12 weeks.

Nucleoss Abutments and Prosthetic parts are intended for use with Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures for edentulous or partially edentulous patients.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

3

510(k) Summary [as required by 21 CFR 807.92(c)]

Nucleoss T6 Dental Implant 510(k) K192062

DATE PREPAREDAugust 27, 2020
APPLICANT INFORMATIONEzgi Ozbudak
10018 Sok. No: 7 ITOB Organize Sanayi Bölgesi Tekeli
Menderes/IZMIR TURKEY
Phone: +90 232 799 03 04
Email: kalite@nucleoss.com
CONTACT INFORMATIONSemih Oktay, Ph.D,
President and CEO, CardioMed Device Consultants
Phone:(410) 674-2060
Email: soktay@cardiomedllc.com
TRADE NAMENucleoss T6 Dental Implant System
DEVICE CLASSIFICATIONClass II per 21 CFR §872.3640
CLASSIFICATION NAMEEndosseous dental implant
PRODUCT CODEDZE
NHA
PREDICATE DEVICETPure Dental Implant (K160850)
REFERENCE DEVICESStraumann Dental Implant System (K171784)
Thommen Medical (K093615)

INDICATIONS FOR USE

Nucleoss T6 Dental Implants are intended to be surgically placed in the bone of the maxillary and/or mandibular arches to provide support for prosthetic restorations (crowns, bridges or overdenture) in edentulous or partially edentulous patients to restore a patients' chewing function. Nucleoss T6 Dental Implants are intended for delayed loading after 12 weeks.

Nucleoss Abutments and Prosthetic parts are intended for use with Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures for edentulous or partially edentulous patients.

4

DEVICE DESCRIPTION

The Nucleoss T6 Dental Implant is a bone level implant constructed of unalloyed titanium (ISO 5832-2). The surface of the T6 implant is a sand-blast and acid-etch (SLA) surface treatment.

T6 Implants

Nucleoss T6 implants have a cylindrical form design, with a double lead thread form and two helical anti-rotation grooves. The thread structure is a reverse buttress. The internal structure is designed as a conical internal hex connection with 140 degrees.

The Nucleoss Dental Abutments are intended for use with the Nucleoss T6 Dental Implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.

Abutments

The Nucleoss Dental Abutments consist of the following designs: Standard Straight, Esthetic Straight, Esthetic Angled, Ball and Equator, Cover, Gingiva Former and Screw. Nucleoss Dental Abutments are cement-retained and screw-retained restorations and intended for placement on Nucleoss T6 implants with diameters of 3.5, 4.1 and 4.8 mm. The connection to the dental implant is achieved by an internal hexagon.

These devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3.

The Standard Straight Abutment is intended for cement-retained single and multiple restorations. There is no angulation. The Standard Straight abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Standard Straight abutment is secured on the implant with a screw.

The Esthetic Straight Abutment is intended for cement-retained single and multiple restorations. There is no angulation. The Esthetic Straight abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Esthetic Straight abutment is secured on the implant with a screw. The front and back gingival heights of these abutments are different from each other. The front heights are shorter and the back heights are greater. These height differences offer the dentist an esthetically beneficial option as required.

The Esthetic Angled Abutment is intended for cement-retained single and multiple restorations. The Esthetic Angled Abutment is provided in angles of 15 and 25 degrees for cases where angle correction is required. The Esthetic Angled abutments can be used in anterior and posterior maxillary and/or mandibular arches. Prior to cementing in place, the Esthetic Angled abutment is secured on the implant with a screw. The front and back gingival heights of these abutments are different from each other. The front heights are shorter and the back heights are greater. These height differences offer the dentist an esthetically beneficial option as required.

The Angled Abutment is intended for cement-retained single and multiple restorations. The Angled Abutment is provided in angles of 15 and 25 degrees for cases where an angle correction is required. The Angled Abutment can be used in anterior and posterior maxillary and/or

5

mandibular arches. Prior to cementing in place, the Angled Abutment is secured on the implant with a screw.

The Ball abutment is intended for overdenture restorations. The top part of the abutment is designed as a ball with a dimeter of 2.25 mm. There is no angulation. The Ball Abutment can be used in anterior and posterior maxillary and/or mandibular arches. The Ball abutment is inserted and combined with the fixture in order to support the overdenture.

The Equator abutment is intended for overdenture restorations. There is no angulation. The Equator Abutment can be used in anterior and posterior maxillary and/or mandibular arches. The Equator abutment is inserted and combined with the fixture in order to support the overdenture.

The Abutments are single-use devices. The Abutments are provided non-sterile for end-user steam sterilization.

Covers and Gingiva Formers

The Nucleoss Cover and Gingiva Former are used during the healing and prosthesis period following surgical placement of the dental implant. All of these devices are constructed of titanium alloy (Ti6Al4V ELI) per ISO 5832-3.

Covers and Gingiva Formers are single-use devices. The Cover is provided sterile in the same packaging tube as the Nucleoss T6 Dental Implant. The Gingiva Formers are provided nonsterile for end-user steam sterilization.

Screws

The Nucleoss Dental Implant and Abutments devices also include screws for securing abutments to the implant to provide a secure coupling for the denture prosthetic attachment to the abutment. They are provided non-sterile and are packaged together with the abutments.

COMPARISON WITH PREDICATE DEVICES

A comparison of the Subject Device and its primary predicate has been conducted. The indications for use of the T6 is identical to that of the predicate device.

A comparison of the T6 and predicate device shows that the technological characteristics such as the components, design, materials, sterilization method, and operating principle of the T6 are similar to the currently marketed Tpure predicate device.

6

| | Subject Device | Primary
Predicate Device | Reference Device | Reference Device |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Nucleoss T6 Dental
Implant System | Nucleoss Tpure
Implant System
(K160850) | Straumann Dental
Implant System
(K171784) | Thommen Medical
SPI Dental Implant,
ELEMENT
(K093615) |
| Indications
for Use
Statement | Nucleoss T6
Dental Implants
are intended to be
surgically placed
in the bone of the
maxillary and/or
mandibular arches
to provide support
for prosthetic
restorations
(crowns, bridges
or overdenture) in
edentulous or
partially
edentulous patients
to restore a
patients' chewing
function. Nucleoss
T6 Dental
Implants are
intended for
delayed loading
after 12 weeks. | Nucleoss Tpure
Implant system are
medical devices
intended to be
surgically placed
in the bone of the
maxillary and/or
mandibular arches
to provide support
for prosthetic
restorations
(crowns, bridges
or overdenture) in
edentulous or
partially
edentulous patients
to restore a
patients' chewing
function.
Nucleoss Tpure
Dental Implants
are intended for
delayed loading
after 12 weeks. | Indicated for oral
endosteal implantation
in the upper and lower
jaw arches and for the
functional and esthetic
oral rehabilitation of
edentulous and partially
dentate patients.
Straumann dental
implants are also
indicated for immediate
or early implantation
following extraction or
loss of natural teeth.
Implants can be placed
with immediate
function on single-tooth
and/or multiple tooth
applications when good
primary stability is
achieved and with
appropriate occlusal
loading to restore
chewing function. The
prosthetic restorations
used are single crowns,
bridges and partial or
full dentures, which are
connected to the
implants through the
corresponding
components
(abutments) | SPI® Dental Implant,
ELEMENT is for one-
stage or two-stage
surgical procedures.
SPI
Dental Implant,
ELEMENT is intended
for immediate
placement and function
on singletooth
and/or multiple tooth
applications when good
primary stability is
achieved and with
appropriate occlusal
loading, to restore
chewing function.
Multiple tooth
applications
may be rigidly splinted.
In the case of
edentulous patients,
four or more implants
must be
used in immediately
loaded cases. |
| Technological | SUBJECT
DEVICE | PRIMARY
PREDICATE
DEVICE | REFERENCE
DEVICE | REFERENCE
DEVICE |
| Characteristics | Nucleoss T6 Dental
Implant System | Nucleoss Tpure
Implant System
(K160850) | Straumann Dental
Implant System
(K1171784) | Thommen Medical
SPI Dental Implant,
ELEMENT
(K093615) |
| Material | Grade 4
commercially pure
titanium conforming
with ISO 5832-2 | Grade 4
commercially pure
titanium conforming
with ISO 5832-2 | Ti-13Zr alloy (trade
named Roxolid) or Ti
Grade 4 | CP Titanium
(ISO 5832-2) |
| Surface | Sand blasted Large
grit Acid etched
(SLA) | Sand blasted Large
grit Acid etched
(SLA) | Sand blasted Large
grit Acid etched
(SLActive) | Sand blasted, thermal
acid etched |
| Implant Design | Two component /
Bone level/
Cylindrical form | Two component /
Bone level/Conical
form | Bone level / Tissue
level
Internal connection | Bone level
Internal connection / Root
form |
| Implant body
diameter (mm) | 3.5, 4.1, 4.8 | 3.4, 3.8, 4.2, 5.0 | 3.3, 4.1, 4.8 | 3.5, 4.0, 4.5, 5.0, 6.0 |
| Implant length
(mm) | 6.5, 8, 10, 12, 14, 17 | 8, 10, 12, 14 | 8, 10, 12, 14, 16, 18 | 6.5, 8, 9.5, 11, 12.5 |
| Diameter x
Length (mm) | 3.5x8, 3.5x10,
3.5x12, 3.5x14,
3.5x17,
4.1x6.5, 4.1x8,
4.1x10, 4.1x12,
4.1x14, 4.1x17,
4.8x6.5, 4.8x8,
4.8x10, 4.8x12,
4.8x14, 4.8x17 | 3.4x10, 3.4x12,
3.4x14,
3.8x8, 3.8x10,
3.8x12, 3.8x14,
4.2x8, 4.2x10,
4.2x12, 4.2x14,
5.0x8, 5.0x10,
5.0x12, 5.0x14 | 3.3x8, 3.3x10,
3.3x12, 3.3x14,
3.3x16, 3.3x18
4.1x6, 4.1x8, 4.1x10,
4.1x12, 4.1x14,
4.1x16, 4.1x18
4.8x6, 4.8x8, 4.8x10,
4.8x12, 4.8x14,
4.8x16, 4.8x18 | 3.5x8, 3.5x9.5, 3.5x11,
3.5x12.5,
4.0x6.5, 4.0x8.0, 4.0x9.5,
4.0x11, 4.0x12.5
4.5x6.5, 4.5x8.0, 4.5x9.5,
4.5x11, 4.5x12.5,
5.0x6.5, 5.0x8.0, 5.0x9.5,
5.0x11,5.0x12.5
6.0x6.5, 6.0x8.0, 6.0x9.5,
6.0x11, 6.0x12.5 |
| Restoration | For single or
multiple restorations | For single or
multiple restorations | For single or multiple
restorations | For single or multiple
restorations |
| Implant
connection | Internal | Internal | Internal | Internal |
| Surgical
technique | Two stage surgical
technique | Two stage surgical
technique | One-stage or two-
stage surgical
technique | One-stage or two-stage
surgical technique |
| Sterilization | Sterile (Gamma
irradiation) | Sterile (Gamma
irradiation) | Sterile (Gamma
irradiation) | Sterile (Gamma
irradiation) |

Comparison of Indications for Use Statements to Predicate and Reference Devices

7

Comparison of Technological Characteristics to Predicate and Reference Devices

8

| Tecnological
Characteristics | SUBJECT DEVICE | PRIMARY PREDICATE
DEVICE |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Nucleoss T6 Dental Implant System | Nucleoss Tpure Implant System
(K160850) |
| Device | Nucleoss T6 Standard Straight,
Esthetic Straight, Esthetic Angled,
Angled, Ball and Equator abutments,
Gingiva Former and Cover | Nucleoss Tpure Straight, Angled, Ball,
and Equator abutments,
Gingiva Former and Cover |
| Material | Titanium alloy (Ti6Al4V ELI) per ISO
5832-3 | Titanium alloy (Ti6Al4V ELI) per ISO
5832-3 |
| Abutment Diameter (mm) | Standard Straight, Esthetic Straight,
Esthetic Angled, Angled abutment:
4.3, 5.0, 6.0 mm

Ball and Equator Abutment: 4.5 mm

Gingiva Former: 4.3, 5.0, 6.0

Cover: 4.3, 5.0, 6.0 | Straight and Angled abutment:
4.2, 5.0, 6.0 mm

Ball and Equator Abutment Abutment:
4.5 mm

Gingiva Former: 4.2, 5.0, 6.0

Cover: 4.2, 5.0, 6.0 |
| Gingival Height (mm) | Standard straight abutment:
1.0, 1.5, 2.0, 3.0, 4.0, 4.5, 5.0 mm

Esthetic straight abutment:
1.0-2.0, 2.0-3.0,
1.5-2.5, 2.5-3.5 mm

Angled abutment:
1.0, 1.5, 2.0, 3.0, 4.0, 5.0 mm | Straight abutment:
1.0, 2.0, 3.0, 5.0 mm

Angled abutment:
1.0, 2.0, 3.0, 4.0 mm

Ball and equator abutment:
1.5, 3.0, 5.0 mm |
| | Esthetic angled abutment:
1.0-2.0, 2.0-3.0,
3.0-4.0 mm | Gingiva Former: 2.0, 4.0, 6.0 |
| | Ball and equator abutment:
1.5, 3.0, 5.0 mm

Gingiva Former: 2.0, 4.0, 6.0

Cover: 0 | Cover: 0 |
| Technological
Characteristics | SUBJECT DEVICE | PRIMARY PREDICATE
DEVICE |
| | Nucleoss T6 Dental Implant System | Nucleoss Tpure Implant System
(K160850) |
| Abutment Design | Straight or Angled | Straight or Angled |
| Angulation | Standard straight abutment: 0° | Straight abutment: 0° |
| | Esthetic straight abutment: 0° | Angled abutment: 10°-20°-30° |
| | Angled abutment: 15° and 25° | Ball and equator abutment: 0° |
| | Esthetic angled abutment: 15° and 25° | Gingiva Former: 0° |
| | Ball and equator abutment: 0° | Cover: 0° |
| | Gingiva Former: 0° | |
| | Cover: 0° | |
| Restoration | The Standard straight abutment, Esthetic
straight abutment, Angled abutment,
Esthetic angled abutment are for single or
multi unit restorations. | The Straight abutment, Angled
abutment are for single or multi unit
restorations. |
| | The Ball and Equator abutments are only
for multi unit restorations. | The Ball and Equator abutments are
only for multi unit restorations. |
| | There is no restoration type for cover and
gingiva formers. | There is no restoration type for cover
and gingiva formers. |
| Implant-Abutment
connection | Internal | Internal |
| Sterilization | Nucleoss T6 Standard Straight, Esthetic
Straight, Esthetic Angled, Angled, Ball
Equator abutments and gingiva formers
are provided non-sterile.
Cover is packaged with the dental implant
and sterilized via gamma irradiation. | The Nucleoss Tpure Straight
abutment, Angled abutments, and
gingiva formers are provided non-
sterile.
Cover is packaged with the dental
implant and sterilized via gamma
irradiation. |

Comparison of Technological Characteristics of Abutments

9

Comparison of Technological Characteristics of Abutments

10

NON-CLINICAL TESTING / PERFORMANCE DATA

Non-clinical testing of the Nucleoss T6 Implant System was performed following the FDA guidance: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments, and applicable ISO and ASTM standards.

Bench Testing

The Pull-out test was performed to compare the Nucleoss T6 Implant and reference device in accordance with ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws with acceptable test results. A comparative surface area analysis was conducted on the Nucleoss T6 Implant and reference device with the original length and 3mm bone loss.

Fatigue Testing

Fatigue testing conducted in accordance with ISO 14801:2016 Dentistry- Implants-Dynamic Fatigue Test for Endosseous Dental Implants found the durability of the Nucleoss T6 Implant and Abutment combinations was acceptable. Testing was conducted on the worst-case implant abutment combinations.

Sterilization Validation

Nucleoss T6 Implants are sterilized using a gamma ray sterilization process that has been validated in accordance with ISO 11137-2 Sterilization of healthcare products-Radiation-Part 2 establishing the sterilization dose to ensure a SAL of 10-6.

Nucleoss Dental Abutments are provided non-sterile for end-user steam sterilization using a traditional cycle process. Since the raw material, production process, parameters and process materials of both Tpure and the predicate device are exactly same, the same steam sterilization parameters apply to the T6 Dental abutments as was used in the predicate device.

Accelerated Aging Test and Packaging Validation Test

Nucleoss T6 implants will be packaged and labeled based on new testing that was conducted to support a shelf life of 5 years on the predicate device. Package Validation and Simulated transportation testing on the predicate device that was accelerated aged to an equivalent 5 years found that the sterility of the dental implants was maintained.

Accelerated aging and packaging validation testing was not repeated for the T6 Dental Implants since the raw material, production process, parameters, process materials and all packaging materials/type of the T6 Dental Implants are exactly same as those of the predicate device.

Biocompatibility

All the materials used the T6 Dental Implant and the manufacturing processes were identical to those used in the predicate device.

Biocompatibility testing was not repeated for the T6 Dental Implants since the raw material, production process, parameters and process materials of the T6 Dental Implants and abutments and the predicate device are exactly the same.

11

CONCLUSION

Data presented in this 510(k) Submission support the substantial equivalence of the T6 Dental Implant to the predicate devices. The Nucleoss T6 Dental Implant System have the similar indication for use, material composition, design, and surface treatment to the primary predicate device and reference predicate devices. Based on the comparative assessment including minor differences with the predicate devices, and based on performance test data, the Nucleoss T6 Dental Implant System is determined to be substantially equivalent to the predicate devices.