(288 days)
Not Found
No
The device description and performance studies focus on the mechanical function and material properties of single-use endoscope valves, with no mention of AI or ML.
No.
Explanation: The device, Single Use Endoscope Valves Set, is designed to enable an endoscope operator to control the functions of endoscope working channels/ports and prevent retrograde flow. It facilitates diagnostic (e.g., biopsy) and procedural maneuvers during GI endoscopy, but it does not directly treat a disease or condition itself.
No
The device description indicates that the Single Use Endoscope Valves Set is used to control the function of endoscope working channels/ports (air/water, suction, biopsy) and prevent retrograde flow. Its purpose is to facilitate endoscopic procedures, not to diagnose conditions.
No
The device description clearly states that the device is a "Single Use Endoscope Valves Set" with three main components: Air/Water Valve, Suction Valve, and Biopsy Valve. These are physical components designed to fit onto endoscopes. The performance studies also describe testing of physical properties like tensile strength, flow rates, and pressure maintenance. There is no mention of software as a component or function of this device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the valves are used to control the function of endoscope working channels/ports during a GI endoscopic procedure. This is a direct interaction with the patient's body (in vivo) and not the examination of specimens derived from the human body (in vitro).
- Device Description: The description reinforces the function of controlling air/water, suction, and accommodating instruments during an endoscopic procedure.
- Lack of IVD Characteristics: There is no mention of analyzing samples, detecting substances, or providing diagnostic information based on in vitro testing.
The device is a medical device used during an in vivo procedure, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials.
The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.
Product codes
ODC
Device Description
The Single Use Endoscope Valves Set collects three types of valve products into one package unit, and which are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials.
This is single-use device and supplied sterile.
There are three main components included, which are Air/Water Valve, Suction Valve and Biopsy Valve.
The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.
There are three models for fitting to varied endoscopes as AMH-EV-01; AMH-EV-02; AMH-EV-03. The main differences between models are physical size of components for varied endoscopes interface, but they share same indication for use, materials and components form.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
Biological Testing was conducted as following standards, the results shown that the proposed devices has same biocompatibility with predicate devices.
ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity.
ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity;
Sterile Testing was conducted as following standards, the results shown that the proposed devices has same sterilization performance with predicate devices.
ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
USP40 NF35 Bacterial Endotoxins Test.
Package Testing was conducted as following standards, the results shown that the proposed devices has same package performance with predicate devices.
ASTM D 3078-02(2013) Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission
ASTM F 1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
DIN 58953-6:2016 Sterilization-Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilizaed.
ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
The shelf Life validation was conducted as following standard, the results shown that the proposed devices meet the requirements of claimed shelf life.
ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
The performance testing was conducted for the performance of tensile strength, water and gas flow rates, pressure maintenance, water and air flow leakage, backflow, insertion and removal, endoscope compatibility and insertion force etc., this testing was a test that compares the proposed device with the predicate devices. The results shown that the proposed device has same performance with predicate device.
No clinical study is included in this submission.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
0
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue and white.
May 14, 2020
Anrei Medical (Hangzhou) Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. R912, B#15, XiYueHui, No.5, YiHe North Road FangShan District Beijing, 102401 CHINA
Re: K192048
Trade/Device Name: Single Use Endoscope Valves Set Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC Dated: April 13, 2020 Received: April 15, 2020
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
K192048
Device Name
Single Use Endoscope Valves Set
Indications for Use (Describe)
The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials.
The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
3
Tab #3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K192048
-
- Date of Preparation: 04/13/2020
-
- Sponsor Identification
Anrei Medical (HangZhou) Co., Ltd. No.3 Ave 8, HEDA, HangZhou City, ZheJiang, China 310018
Contact Person: Yang HuiBing Position: Regulations & Quality System Director Tel: +86-571-8691333 ext. 8658 Fax: +86-571-87603502 Email: huibing.yang@anrei.com.cn
- Designated Submission Correspondent 3.
Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10- 56335780 Email: ray.wang@believe-med.com
4
4. Identification of Proposed Device
Trade Name: Single Use Endoscope Valves Set Common Name: Endoscope and Accessories Model(s): AMH-EV-01; AMH-EV-02; AMH-EV-03
Regulatory Information Classification Name: Endoscope and Accessories Classification: 2 Product Code: ODC Regulation Number: 21 CFR 876.1500 Review Panel: Gastroenterology/Urology;
Indications for Use Statement:
The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials.
The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.
Device Description
The Single Use Endoscope Valves Set collects three types of valve products into one package unit, and which are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials.
This is single-use device and supplied sterile.
There are three main components included, which are Air/Water Valve, Suction Valve and Biopsy Valve.
The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.
The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure.
The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.
There are three models for fitting to varied endoscopes as AMH-EV-01; AMH-EV-02; AMH-EV-03. The main differences between models are physical size of components for varied endoscopes interface, but they share same indication for use, materials and components form.
5
ર. Identification of Predicate Device(s)
510(k) Number: K102409 Product Name: DEFENDO Disposable Air/Water Valve for GI Endoscopes
510(k) Number: K102581 Product Name: DEFENDO Disposable Suction Valve for GI Endoscopes
510(k) Number: K090851 Product Name: DEFENDO Biopsy Valve (Model #100301 and I00302)
6. Non-Clinical Test Discussion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
Biological Testing was conducted as following standards, the results shown that the proposed devices has same biocompatibility with predicate devices.
ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity.
ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity;
Sterile Testing was conducted as following standards, the results shown that the proposed devices has same sterilization performance with predicate devices.
ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
USP40 NF35 Bacterial Endotoxins Test.
Package Testing was conducted as following standards, the results shown that the proposed devices has same package performance with predicate devices.
ASTM D 3078-02(2013) Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission
ASTM F 1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
DIN 58953-6:2016 Sterilization-Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilizaed.
ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
The shelf Life validation was conducted as following standard, the results shown that the proposed
6
devices meet the requirements of claimed shelf life.
ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
The performance testing was conducted for the performance of tensile strength, water and gas flow rates, pressure maintenance, water and air flow leakage, backflow, insertion and removal, endoscope compatibility and insertion force etc., this testing was a test that compares the proposed device with the predicate devices. The results shown that the proposed device has same performance with predicate device.
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
7
Substantially Equivalent (SE) Compariso ం
ble 1 Comparison of Technology Characteristic
Item | Device name | Classification Name | Regulation Number | Predicate Device (K102409) | Predicate Device(K102581) | Predicate Device (K090851) | Remark |
---|---|---|---|---|---|---|---|
Proposed Device(s) | |||||||
Single Use Endoscope Valves | |||||||
Set | Endoscope and accessories | 876.1500 | DEFENDO Disposable Air/Water | ||||
Valve for GI Endoscopes | DEFENDO Disposable Suction | ||||||
Valve for GI Endoscopes | DEFENDO Biopsy Valve | ||||||
(Model #100301 and 100302) | / | ||||||
Endoscope and accessories | 876.1500 | Endoscope and accessories | Endoscope and accessories | SAME | |||
876.1500 | SAME | ||||||
Indications for Use | The Single Use Endoscope | ||||||
Valves Set are used to fit to | |||||||
multiple endoscope working | |||||||
channels/ports to enable an | |||||||
endoscope operator control the | |||||||
function of the working | |||||||
channels/ports and prevent | |||||||
retrograde flow of the fluids, | |||||||
gases, and other materials. | The DEFENDO Disposable | ||||||
Air/Water Valve is intended to be | |||||||
used to control the air/water | |||||||
function on an endoscope during a | |||||||
GI endoscopic procedure. | The DEFENDO Disposable | ||||||
Suction Valve is intended to be | |||||||
used to control the suction | |||||||
function on an endoscope | |||||||
during a GI endoscopic | |||||||
procedure. | DEFENDO Disposable Biopsy | ||||||
Valve is indicated for covering | |||||||
the endoscope biopsy port | |||||||
during an endoscopy procedure. | |||||||
The valve provides access for | |||||||
endoscopic device passage and | |||||||
exchange, helps maintain | |||||||
sufflation, and minimizes | |||||||
leakage of biomaterial from the | |||||||
biopsy port throughout the | |||||||
endoscopic procedure. | SAME |
K192048
Page 5 of 9
8
| | endoscopic procedure.
The Biopsy Valve is intended
to accommodate various sizes
of endoscopic accessory
instruments while still
providing a seal around the
instrument channel inlet. | | | | |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------|
| Prescription/OTC | Prescription Use | Prescription Use | Prescription Use | Prescription Use | SAME |
| Components | Air/Water Valve, Suction
Valve, Biopsy Valve | Air/Water Valve | Suction Valve | Biopsy Valve | SAME |
| Supplied Sterile | Yes | Yes | Yes | Yes | SAME |
| Single use | Yes | Yes | Yes | Yes | SAME |
| Feature | Sterile and single-use;
Compatible with multiple
endoscope working
channels/ports;
Enable an endoscope operator
control the function of the
working channels/ports and
prevent retrograde flow of the
fluids, gases, and other
materials; | Sterile and single-use;
Compatible with multiple
endoscope working channels/ports;
Enable an endoscope operator
control the function of the working
channels/ports and prevent
retrograde flow of the fluids, gases,
and other materials; | Sterile and single-use;
Compatible with multiple
endoscope channels/ports;
Enable an endoscope operator
control the function of the
working channels/ports and
prevent retrograde flow of the
fluids, gases, and other
materials; | Compatible with multiple
endoscope biopsy port during an
endoscopy procedure | SAME |
| Materials | Air/Water Valve:
ABS+Silica+SUS304
Suction Valve:
ABS+Silica+SUS304 | Not Available | Not Available | Not Available | Difference |
9
K192048
Page 7 of 9
10
Summary | |
---|---|
( | |
0(k | |
l | |
S |
| | 140/160/180/240/260/290
series;
Fujinon 500 and 600 series;
Pentax i10 and 90 series. | series endoscopes | series endoscopes | Endoscopes | |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------|---------------|------------|
| | The air/water valves, suction
valve and biopsy valve are
housed in a single tray and
packaged in sealed initial
package. | Suction and air/water valves are
housed in a single tray and
packaged in a sealed Tyvek
pouch | | | |
| Packaging | Not Available | Not Available | | | SAME |
| Sterilization Method | EO | EO | | Not Available | SAME |
| Shelf Life | 3 year | 1 year | | Not Available | Difference |
| Biocompatibility | Cytotoxicity, Sensitization and
Irritation | Cytotoxicity, Sensitization and
Irritation | | Not Available | SAME |
| Difference Analysis | | | | | |
atibility The proposed device has different on the for this differs, the performance estable to teacher and closes devices and fect the safety and effectiveness of proposed device
this different d fe: The proposed device has different shelf life than predicate device, the shelf life validation of 3 years has been conduced, which included perfo after accelerated aging. The testing results shown that 3 years shelf life would not affect the performance of proposed device. So, et the safety and effectiveness of proposed device
Materials: Beause the detail materials are not wailable, or could be onsidered as different the tiffers, the Bioompitility testing has betactives and oroposed device
11
9. Substantially Equivalent (SE) Conclusion
The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device predicate