K Number
K192026

Validate with FDA (Live)

Date Cleared
2019-11-26

(120 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. Peridot Intervertebral body fusion system is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

Device Description

The Peridot Intervertebral Body Fusion System is intervertebral body fixation devices intended for use as an aid in spinal fixation. The Peridot Intervertebral fusion system is made from PEEK as per ASTM F2026 and Tantalum marker as per ASTM F560. And some cage holders are made of titanium alloy as per ASTM F136.

X-ray markers system on the cages permits the identification of cage position and allows post-operative assessment.

The device is available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates.

The device is supplied non-sterile.

The device consists of cages differentiated by their approach, with varying dimensions and ancillary products for placement of the cages.

The device is supplied with their specific instrument. The specific instrument is supplied dedicated tray and non-sterile. The device and instrument are supplied separately.

The device must be used in combination with the dedicated instrument supplied.

The hyperlordotic lumbar cages (>20 degree) the form of supplemental fixation should be an anterior plate system.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the "Peridot Intervertebral body fusion system." It is a submission to the FDA to demonstrate substantial equivalence to legally marketed predicate devices, not a study presenting the performance of an AI/ML powered device against acceptance criteria.

The document does not contain any information about acceptance criteria or a study proving that an AI/ML powered device meets those criteria.

Instead, it describes the device, its intended use, and compares it to existing predicate devices to show substantial equivalence for regulatory clearance. It focuses on mechanical testing of the physical implant according to ASTM standards, not on the performance of a software algorithm.

Therefore, I cannot provide the requested information about acceptance criteria and a study proving an AI/ML device meets them based on the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" stacked vertically.

November 26, 2019

GBS Commonwealth Co., Ltd. Jimmy Kim RA C-309, Woolim Lion's Valley, 168 Gasan Digital 1-ro Geumcheon-Gu, Seoul 08507 Korea

Re: K192026

Trade/Device Name: Peridot Intervertebral body fusion system Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: October 22, 2019 Received: October 28, 2019

Dear Jimmy Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Brent Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192026

Device Name

Peridot Intervertebral body fusion system

Indications for Use (Describe)

Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. Peridot Intervertebral body fusion system is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters, the word "COMMONWEALTH" is written in a smaller, blue font. The background of the logo is a light gray globe.

510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

1. Device Identification

Submitter:GBS Commonwealth Co., Ltd.
#C-309, 168 Gasan Digital 1-ro, Geumcheon-Gu Seoul,
South Korea
Phone. 82-2-6925-4469
e-mail: Jimmy.kim@gbscommonwealth.com
Contact Person:Jimmy Kim
Date preparedJuly, 25, 2019
Trade NamePeridot Intervertebral body fusion system
Classification21 CFR 888.3080 Intervertebral body fusion device,Class IIProduct Code: MAX

Purpose of 510(k) 2.

The GBS Commonwealth Co. Ltd., here by submits this traditional 510(k): for Initial product Introduction of Peridot Intervertebral fusion system.

3. Predicate or legally marketed devices which are substantially equivalent

    1. Primary Predicate Device: K151140 LnK Lumbar Interbody Fusion Cage System
    1. Additional Predicate device: K092193 Shurfit Interbody device system

K153783 SpaceVision OLIF K110927 MectaLIF Oblique K161379 ELSA Spacers

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters is the word "COMMONWEALTH" in a smaller, bold, blue font. The logo is surrounded by a white background.

K151214, K161230 NuVasive Lumbar Interbody Implants

Description of the Device 4.

The Peridot Intervertebral Body Fusion System is intervertebral body fixation devices intended for use as an aid in spinal fixation. The Peridot Intervertebral fusion system is made from PEEK as per ASTM F2026 and Tantalum marker as per ASTM F560. And some cage holders are made of titanium alloy as per ASTM F136.

X-ray markers system on the cages permits the identification of cage position and allows post-operative assessment.

The device is available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates.

The device is supplied non-sterile.

The device consists of cages differentiated by their approach, with varying dimensions and ancillary products for placement of the cages.

The device is supplied with their specific instrument. The specific instrument is supplied dedicated tray and non-sterile. The device and instrument are supplied separately.

The device must be used in combination with the dedicated instrument supplied.

The hyperlordotic lumbar cages (>20 degree) the form of supplemental fixation should be an anterior plate system.

5. Indication for Use

Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters, the word "COMMONWEALTH" is written in a smaller, blue font. The background of the logo is a light gray.

bone graft. Peridot Intervertebral body fusion system is to be used supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

Comparison of the technological characteristics of the subject and predicate 6. devices

The Peridot Intervertebral body fusion system is considered substantially equivalent to the primary predicate LnK Lumbar Interbody Fusion Cage System (K151140) and additional predicate Shurfit interbody device system (K092193). They are similar in design, material, scientific technologies and indications for use. And the Peridot Intervertebral body fusion system is considered substantially equivalent to the additional predicates SpaceVision OLIF (K153783) and MectaLIF Oblique (K110927) and ELSA Spacers (K161379) and Nuvasive Lumbar Interbody Implants (K151214, K161230). They are similar in design, material, scientific technologies and indications for use.

7. Performance Testing

The Peridot Intervertebral body fusion system conform to class II special controls quidance document: Intervertevbral body fusion device – Document issued on : June 12, 2007.

Mechanical testing includes static compression, static compression - shear, fatigue compression, fatigue compression – shear performed according to ASTM F2077-14, subsidence testing performed according to ASTM F2267-04 and , expulsion testing performed according to Draft ASTM F-04.25.02.02. Results demonstrate comparable mechanical properties to the predicate device. No clinical date has been presented.

8. Conclusion

The Peridot Intervertebral body fusion system is substantially equivalent to legally marketed predicates.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.