K Number
K191833
Manufacturer
Date Cleared
2019-10-07

(90 days)

Product Code
Regulation Number
862.1355
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dexcom G6 Pro Continuous Glucose Monitoring System) is a real time continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. The Dexcom G6 Pro System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the real-time Dexcom G6 Pro System results should be based on the glucose trends and several sequential readings over time.

The Dexcom G6 Pro System may also be used as a retrospective glucose recording device indicated for assessing glycemic variability in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. Retrospective interpretation of data recorded by the Dexcom G6 Pro System should be conducted solely by a healthcare professional.

The Dexcom G6 Pro System aids in detecting glucose excursions facilitating care plan adjustments. The Dexcom G6 Pro System is also intended to interface with digitally connected devices. The Dexcom G6 Pro System can be used alone or in conjunction with these digitally connected medical devices or assessing glycemic variability.

Device Description

The Dexcom G6 Pro Continuous Glucose Monitoring (G6 Pro) System is a continuous glucose monitor that offers an introduction to CGM for users who would benefit from the supervision of their qualified Healthcare Professional (HCP) during early or initial use of CGM. The predicate of the Dexcom G6 Pro CGM System is the Dexcom G6 CGM System. G6 Pro consists of three main components: a sensor/applicator delivery system, a transmitter, and a mobile application (app). The sensor/applicator is identical to the sensor/applicator used in the G6 CGM System (K191450). The sensor is a small and flexible wire inserted into subcutaneous tissue where it converts glucose into electrical current.

The G6 Pro Transmitter is connected to the sensor and is worn on the body. It samples the electrical current produced by the sensor and converts these measurements into glucose readings using an onboard algorithm. The G6 Pro Transmitter's firmware includes an auto-start feature which enables the transmitter to start a session immediately upon attachment of the transmitter to the on-body wearable. The G6 Pro Transmitter can be used as a retrospective CGM data logger and it can also send realtime estimated glucose values to the G6 Mobile Application. The HCP selects which type of CGM session the patient receives (retrospective vs. real-time). The G6 Pro Transmitter Firmware supports a single-use 10-day sensor session per transmitter.

The G6 App displays the current glucose reading (updated every 5 minutes) and glucose trends (up to 24 hours) from the transmitter.

AI/ML Overview

The provided document is a 510(k) summary for the Dexcom G6 Pro Continuous Glucose Monitoring System. It describes the device, its intended use, and its equivalence to a predicate device. However, the document does not contain the specific details of performance testing, acceptance criteria, or study results in the format requested.

The "Summary of Performance Testing" section (5.8) states: "The proposed Dexcom G6 Pro Continuous Glucose Monitoring System was verified and validated according to Dexcom's internal design control process and in accordance with special controls for integrated continuous glucose monitors. This testing demonstrated that the system performed according to its specifications, and that the technological and performance criteria are comparable to the predicate device."

This statement confirms that performance testing was conducted, and the device met its specifications, indicating that there are acceptance criteria and study results. However, the details of these criteria and the reported performance are not included in this summary document.

Therefore, I cannot extract the specific information requested in your prompt based solely on the provided text. To answer your questions, I would need access to the full submission or specific performance study reports that detail:

  1. A table of acceptance criteria and the reported device performance.
  2. Sample size used for the test set and the data provenance.
  3. Number of experts used to establish the ground truth and their qualifications.
  4. Adjudication method.
  5. Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done and any effect size.
  6. Whether a standalone performance study was done.
  7. The type of ground truth used.
  8. The sample size for the training set.
  9. How the ground truth for the training set was established.

Without this detailed performance data, I can only state that the document confirms such testing was performed and the device met its specifications.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" written in blue.

October 7, 2019

Dexcom, Inc. Jacob Nardone Staff Regulatory Affairs Specialist 6340 Sequence Dr. San Diego, CA 92121

Re: K191833

Trade/Device Name: Dexcom G6 Pro Continuous Glucose Monitoring System Regulation Number: 21 CFR 862.1355 Regulation Name: Integrated Continuous Glucose Monitoring System Regulatory Class: Class II Product Code: QII Dated: July 8, 2019 Received: July 9, 2019

Dear Jacob Nardone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191833

Device Name

Dexcom G6 Pro Continuous Glucose Monitoring System

Indications for Use (Describe)

The Dexcom G6 Pro Continuous Glucose Monitoring System) is a real time continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. The Dexcom G6 Pro System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the real-time Dexcom G6 Pro System results should be based on the glucose trends and several sequential readings over time.

The Dexcom G6 Pro System may also be used as a retrospective glucose recording device indicated for assessing glycemic variability in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. Retrospective interpretation of data recorded by the Dexcom G6 Pro System should be conducted solely by a healthcare professional.

The Dexcom G6 Pro System aids in detecting glucose excursions facilitating care plan adjustments. The Dexcom G6 Pro System is also intended to interface with digitally connected devices. The Dexcom G6 Pro System can be used alone or in conjunction with these digitally connected medical devices or assessing glycemic variability.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is: K191833

5.1 Submitter:

Dexcom, Inc. 6340 Sequence Dr. San Diego, CA 92121

Contact: Jacob Nardone Staff Regulatory Affairs Specialist Phone: 858.203.6337 Fax: 858.332.0204 jacob.nardone@dexcom.com Email:

Secondary Contact: Holly Drake Senior Manager, Regulatory Affairs Phone: 858.875.5326 858.332.0204 Fax: Email: hdrake@dexcom.com

Date Prepared:

5.2 Device Names and Classification:

Proprietary NameDexcom G6 Pro Continuous Glucose Monitoring System
Common NameIntegrated Continuous Glucose Monitoring System, FactoryCalibrated
ClassII
Classification Regulation21 CFR 862.1355
Product CodeQII
Review PanelClinical Chemistry

{4}------------------------------------------------

5.3 Predicate Device:

Dexcom G6 Continuous Glucose Monitoring System (K191450)

5.4 Device Description:

5.4.1 Dexcom G6 Pro Continuous Glucose Monitoring System

The Dexcom G6 Pro Continuous Glucose Monitoring (G6 Pro) System is a continuous glucose monitor that offers an introduction to CGM for users who would benefit from the supervision of their qualified Healthcare Professional (HCP) during early or initial use of CGM. The predicate of the Dexcom G6 Pro CGM System is the Dexcom G6 CGM System. G6 Pro consists of three main components: a sensor/applicator delivery system, a transmitter, and a mobile application (app). The sensor/applicator is identical to the sensor/applicator used in the G6 CGM System (K191450). The sensor is a small and flexible wire inserted into subcutaneous tissue where it converts glucose into electrical current.

The G6 Pro Transmitter is connected to the sensor and is worn on the body. It samples the electrical current produced by the sensor and converts these measurements into glucose readings using an onboard algorithm. The G6 Pro Transmitter's firmware includes an auto-start feature which enables the transmitter to start a session immediately upon attachment of the transmitter to the on-body wearable. The G6 Pro Transmitter can be used as a retrospective CGM data logger and it can also send realtime estimated glucose values to the G6 Mobile Application. The HCP selects which type of CGM session the patient receives (retrospective vs. real-time). The G6 Pro Transmitter Firmware supports a single-use 10-day sensor session per transmitter.

The G6 App displays the current glucose reading (updated every 5 minutes) and glucose trends (up to 24 hours) from the transmitter.

{5}------------------------------------------------

5.5 Indications for Use:

Dexcom G6 Pro Continuous Glucose Monitoring System 5.5.1

The Dexcom G6 Pro Continuous Glucose Monitoring System (Dexcom G6 Pro System) is a real time continuous glucose monitoring device indicated for the management of diabetes in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. The Dexcom G6 Pro System is intended to replace fingerstick blood glucose testing for diabetes treatment decisions. Interpretation of the real-time Dexcom G6 Pro System results should be based on the glucose trends and several sequential readings over time.

The Dexcom G6 Pro System may also be used as a retrospective glucose recording device indicated for assessing glycemic variability in persons age 2 years and older in a home environment while under the supervision of a healthcare professional. Retrospective interpretation of data recorded by the Dexcom G6 Pro System should be conducted solely by a healthcare professional.

The Dexcom G6 Pro System aids in detecting glucose excursions facilitating care plan adjustments. The Dexcom G6 Pro System is also intended to interface with digitally connected devices. The Dexcom G6 Pro System can be used alone or in conjunction with these digitally connected medical devices for managing diabetes or assessing glycemic variability.

{6}------------------------------------------------

5.6 Comparison with the Predicate Device:

5.6.1 Dexcom G6 Pro Continuous Glucose Monitoring System (subject device) to Dexcom G6 Continuous Glucose Monitoring System (K191450)

DeviceDexcom G6 Continuous Glucose Monitoring System (K191450)Dexcom G6 Pro Continuous Glucose Monitoring System(subject device)
Trade NameDexcom G6 Continuous Glucose Monitoring SystemDexcom G6 Pro Continuous Glucose Monitoring System
ManufacturerDexcom, Inc.Same
Intended UseThe System is an integrated continuous glucose monitoringsystem (iCGM) intended to automatically measure glucose inbodily fluids continuously or frequently for a specified period oftime. The System is designed to reliably and securely transmitglucose measurement data to digitally connected devices,including automated insulin dosing systems, and is intended to beused alone or in conjunction with these digitally connectedmedical devices for the purpose of managing a disease orcondition related to glycemic control.Same
Indications for UseThe Dexcom G6 Continuous Glucose Monitoring System (DexcomG6 CGM System) is a real time, continuous glucose monitoringdevice indicated for the management of diabetes in persons age2 years and older.The Dexcom G6 System is intended to replace fingerstick bloodglucose testing for diabetes treatment decisions. Interpretationof the Dexcom G6 System results should be based on the glucosetrends and several sequential readings over time. The Dexcom G6System also aids in the detection of episodes of hyperglycemiaand hypoglycemia, facilitating both acute and long-term therapyadjustments.The Dexcom G6 Pro Continuous Glucose Monitoring System(Dexcom G6 Pro System) is a real time continuous glucosemonitoring device indicated for the management of diabetesin persons age 2 years and older in a home environment whileunder the supervision of a healthcare professional. TheDexcom G6 Pro System is intended to replace fingerstickblood glucose testing for diabetes treatment decisions.Interpretation of the real-time Dexcom G6 Pro System resultsshould be based on the glucose trends and several sequentialreadings over time.The Dexcom G6 Pro System may also be used as aretrospective glucose recording device indicated for assessing
DeviceDexcom G6 Continuous Glucose Monitoring System (K191450)Dexcom G6 Pro Continuous Glucose Monitoring System(subject device)
The Dexcom G6 System is also intended to autonomouslycommunicate with digitally connected devices, includingautomated insulin dosing (AID) systems. The Dexcom G6 Systemcan be used alone or in conjunction with these digitallyconnected medical devices for the purpose of managing diabetes.glycemic variability in persons age 2 years and older in a homeenvironment while under the supervision of a healthcareprofessional. Retrospective interpretation of data recorded bythe Dexcom G6 Pro System should be conducted solely by ahealthcare professional.
The Dexcom G6 Pro System aids in detecting glucoseexcursions facilitating care plan adjustments. The Dexcom G6Pro System is also intended to interface with digitallyconnected devices. The Dexcom G6 Pro System can be usedalone or in conjunction with these digitally connected medicaldevices for managing diabetes or assessing glycemicvariability.
Clinical applicationManagement of diabetes mellitusManagement of diabetes mellitus or assessing glycemicvariability
Clinicalsetting/sites ofuseHome useHome use (sensor insertion, transmitter attachment andretrospective glucose data download occurs in a clinic withthe supervision of a healthcare professional)
Principle ofOperationAmperometric measurement of current proportional to glucoseconcentration in interstitial fluid via glucose oxidase chemicalreactionSame
Data PresentedEstimated Glucose Value (EGV): The EGV is the nominal glucosevalue presented to the user.Glucose Trend: Based off the glucose rate of change, users areshown their glucose trend with a corresponding arrow.Historical Glucose Data: Users can view their previous three, six,twelve, or twenty-four hours of glucose data.Same
FeaturesConnect to Dexcom Share: Users can share their glucose datawith followers.Same
DeviceDexcom G6 Continuous Glucose Monitoring System (K191450)Dexcom G6 Pro Continuous Glucose Monitoring System(subject device)
Human FactorsEasy to understand UI/UX.Commonly understood navigation tools and features.Color-coded graphics.Same
Transmitter UseLifeReusable (90 days)Single-Use (10-day)
Supported CGMData MethodsReal-TimeReal-Time and Retrospective
FactoryCalibrationYesYes
OptionalCalibrationYesNo
Compatibility withintendedenvironmentsAndroid OS version 7.0.0 - 9.0.0iOS 11.0.0 - 12.1.4Same

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

5.7 Technological Characteristics

The proposed Dexcom G6 Pro Continuous Glucose Monitoring System shares the same technological characteristics as the predicate Dexcom G6 Continuous Glucose Monitoring System (K191450). Both Systems are used to measure glucose values via an amperometric measurement of current proportional to glucose concentration in interstitial fluid via a glucose oxidase chemical reaction.

5.8 Summary of Performance Testing

The proposed Dexcom G6 Pro Continuous Glucose Monitoring System was verified and validated according to Dexcom's internal design control process and in accordance with special controls for integrated continuous glucose monitors. This testing demonstrated that the system performed according to its specifications, and that the technological and performance criteria are comparable to the predicate device.

5.9 Conclusions

The Dexcom G6 Pro Continuous Glucose Monitoring System is a professional product designed to assist patients in a home environment better manage diabetes mellitus while under HCP supervision. The Dexcom G6 Pro Continuous Glucose Monitoring System may also be used as a retrospective glucose recording device that allows HCPs to assess glycemic variability of patients in a home environment. The Dexcom G6 Pro Continuous Glucose Monitoring System is substantially equivalent to the Dexcom G6 Continuous Glucose Monitoring System as they are identical with regard to intended use and there are no differences in indications, technological characteristics or performance that raise new questions of safety and effectiveness.

§ 862.1355 Integrated continuous glucose monitoring system.

(a)
Identification. An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
e.g., an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
e.g., abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.