K Number
K191763

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-23

(266 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.

MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

Device Description

MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens is available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.

The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.

When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

AI/ML Overview

This document is an FDA 510(k) clearance letter for a contact lens, not a medical device that relies on AI or complex algorithms requiring extensive performance studies with human readers or AI-specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design for AI models (e.g., sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies) is not applicable to this submission.

However, I can extract the relevant performance data and the conclusion regarding substantial equivalence for this specific device.

Device Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens (with e-PL in the packaging solution)

Predicated Device: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens (K131378)

Acceptance Criteria and Reported Device Performance (Non-Clinical/Bench Testing):

The submission aimed to demonstrate substantial equivalence to a predicate device, with the primary difference being the addition of 100 ppm Epsilon-Polylysine (e-PL) to the blister packing solution as a preservative. The acceptance criteria are implicitly met by demonstrating that the physical, mechanical, optical, and toxicological properties of the new device are equivalent to the predicate device and comply with relevant standards.

Here's a table summarizing the types of performance testing done and the overall conclusion:

Acceptance Criteria Category (Implicitly Met by Testing)Reported Device Performance (Conclusion)
Physicochemical Properties:
- Water content (per ISO18369-2:2012, ISO 18369-4:2006)Equivalent to predicate device
- Refractive index (per ISO 18369-2:2012, ISO 16369-4:2006)Equivalent to predicate device
- Light transmittance (per ISO 18369-2:2012, ISO 18369-3:2006)Equivalent to predicate device
- Non-polymeric residuals in lens and packaging solution (per ISO 18369-4:2006)Equivalent to predicate device
- Total extractable (per ANSI Z80.20-2010, ISO 18369-4:2006)Equivalent to predicate device
- Contact angle (per ANSI Z80.20-2010)Equivalent to predicate device
- Mechanical properties (per ANSI Z80.20-2010, ASTM D1708-02a)Equivalent to predicate device
Preservative Efficacy:
- Preservative efficacy testing (PET) (per ISO 14730:2014)Demonstrated function of e-PL as a preservative
- Analytical testing (uptake and release) (per ISO 11986:2010)Demonstrated function of e-PL as a preservative
- In-vitro microbiological testing (lens handling)Demonstrated function of e-PL as a preservative
Biocompatibility Testing:
- Cytotoxicity (per ISO 10993-5:2009)All biocompatibility tests passed, demonstrating equivalence.
- Systemic Toxicity (per ISO 10993-11:2006)
- Sensitization (per ISO 10993-10:2010)
- Irritation (per ISO 10993-10:2010)
- Irritation (22 Day) (per ISO 9394:2012)

Regarding the other requested information (which is not directly applicable to this non-AI device submission):

  1. Sample size used for the test set and the data provenance: Not explicitly stated as a distinct "test set" in the context of an AI model. The bench testing data would involve samples of the manufactured contact lenses. Data provenance is implied to be from laboratory testing conducted by the manufacturer.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a contact lens's physical properties is established by objective measurements based on specified standards, not expert consensus on interpretations.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Evaluation is done by standardized laboratory testing.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI or imaging device.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI or algorithmic device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth was established through adherence to recognized international and industry standards (ISO, ANSI, ASTM) for measuring physicochemical properties and conducting biocompatibility tests (e.g., specific gravity, refractive index, water content, cytotoxicity).
  7. The sample size for the training set: Not applicable. This device does not involve a training set as it's not an AI/machine learning product.
  8. How the ground truth for the training set was established: Not applicable.

Overall Conclusion from the Submission:

"Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physical/mechanical/optical properties of the lens. Based on the performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). The performance testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

March 23, 2020

CooperVision, Inc. Ms. Marie Dutton Senior Regulatory Affairs Specialist 5870 Stoneridge Drive, Suite 1 Pleasanton, CA 94588

Re: K191763

Trade/Device Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens

Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 24, 2020 Received: February 25, 2020

Dear Ms. Marie Dutton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidanceregulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-devicesafety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-andconsumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191763

Device Name

MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens

Indications for Use (Describe)

MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.

MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER:

CooperVision, Inc. 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588

Contact Person:

Marie Dutton Senior Regulatory Affairs Specialist CooperVision, Inc. 5870 Stoneridge Drive. Suite 1 Pleasanton, CA 94588 Phone: Phone: (925) 251-6645 Fax: (925) 251-6643 MDutton@coopervision.com E-mail:

Date Prepared:

June 28, 2019

II. DEVICE:

Trade Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens Common Name: Soft (hydrophilic) Contact Lens Classification Name: Lens, Contact, (Disposable) [21 CFR 886.5925 (b) (1)] Regulatory Class: II LPL, MVN Product Code: Classification Panel: Ophthalmic

III. PREDICATE DEVICE:

The predicate device is CooperVision's MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens, submitted as SUS (stenfilcon A) Soft (hydrophilic) Contact Lenses for Single Use Daily Wear under 510(k) K131378 and received FDA clearance on August 30, 2013.

IV. DEVICE DESCRIPTION:

The device description for the subject device and the predicate device is identical.

MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens is available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.

The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.

{4}------------------------------------------------

When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

MyDay (stenfilcon A) contact lens parameters are:

• Chord Diameter:13.0 mm to 15.5mm
• Base Curve:8.4 ± 0.5 mm and 8.7 ± 0.5 mm
• Center Thickness:0.08 mm to 0.218 mm (varies with power)
• Powers:-20.00D to +20.00D
• Cylinder Powers:-0.25D to -10.00D
• Axis:0° to 180° in 10° increments
• Add Power Range:+0.50 to +4.00

The physical/optical properties of the lens are:

Specific Gravity:1.033
Refractive Index:1.401
Light Transmittance:96%
Surface Character:Hydrophilic
Water Content:54%
Oxygen Permeability:80x10-11 [(cm²/sec)x(ml O2)/(ml x mm Hg)]

V. INDICATIONS FOR USE:

There is no difference between the subject device and the predicate device with respect to indications and intended use.

MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

{5}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The technological principle for both the subject and the predicate device is identical. The subject and predicate device are based on the same technological elements. The only technological difference that exists between the subject and predicate device is the addition of 100 ppm Epsilon-Polylysine (also known as epolylysine or e-PL) to the blister packing solution formulation.

The e-PL is being added to the packaging solution as a preservative. Changes to the labeling are being made to reflect the addition of e-PL to the packaging solution.

Technology/Material Comparison Predicate Device Subject Device Product Name MyDay (stenfilcon A) Soft MyDay (stenfilcon A) Soft (Hydrophilic) Daily (Hydrophilic) Daily Disposable Contact Lens Disposable Contact Lens (with e-PL in the packaging solution) Material USAN Name stenfilcon A Same 510(k) Number K131378 TBD - Current Submission FDA Category (Group) Silicone Hydrogel Same Manufacturing Method Molded Same Sterilization Moist Heat Same Packaging Materials Injection molded Same polypropylene blisters covered by aluminum foil laminate; blister strips are packed into printed cartons Packaging Solution Phosphate Buffered Saline Phosphate Buffered Saline Solution with Tween Solution with Tween and 100 ppm e-PL Visibility Tint Reactive Blue #246 (RB246) Same UV Blocker Norbloc Same

The technological characteristics of the subject device and the predicate device are compared in the table below.

{6}------------------------------------------------

VII. PERFORMANCE DATA:

Results from non-clinical studies were provided in support of the substantial equivalence determination.

Performance testing - bench:

In accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, issued on May 12, 1994 and amended June 28, 1994, the following battery of physicochemical testing was performed. All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols. Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated:

  • Water content per ISO18369-2:2012 and ISO 18369-4:2006 ●
  • Refractive index per ISO 18369-2:2012 and ISO 16369-4:2006 ●
  • Light transmittance per ISO 18369-2:2012 and ISO 18369-3:2006
  • Non-polymeric residuals in lens and packaging solution per ISO 18369o 4:2006
  • Total extractable per ANSI Z80.20-2010 and ISO 18369-4:2006 ●
  • o Contact angle per ANSI Z80.20-2010
  • Mechanical properties per ANSI Z80.20-2010 and ASTM D1708-02a .

Additionally, preservative efficacy testing (PET) per ISO 14730:2014, analytical testing (uptake and release) per ISO 11986:2010, and in-vitro microbiological testing (lens handling) were performed to support the function of e-PL as that of a preservative.

Biocompatibility testing:

In accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, issued on May 12, 1994 and amended June 28, 1994 and with the GLP regulation (21 CFR Part 58), the following battery of biocompatibility testing was performed:

  • Cytotoxicity per ISO 10993-5:2009 ●
  • Systemic Toxicity per ISO 10993-11:2006 ●
  • Sensitization per ISO 10993-10:2010 ●
  • Irritation per ISO 10993-10:2010 ●
  • o Irritation (22 Day) per ISO 9394:2012

{7}------------------------------------------------

Clinical testing:

Data provided in this 510(k) is sufficient to adequately characterize the subject device in terms of its physical/mechanical/optical and toxicological performance characteristics when compared to the predicate device. The results are equivalent, determining that additional clinical performance data was not required to complete the substantial equivalence determination. Additionally, the technical characteristics and manufacturing and sterilization processes of the subject lens are equivalent to MyDay (stenfilcon A) contact lens currently marketed by CooperVision; therefore, it was confirmed that no clinical data is required.

VIII. CONCLUSIONS:

Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physical/mechanical/optical properties of the lens. Based on the performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). The performance testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.