The EUROPA™ Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.
The EUROPA™ Pedicle Screw System is intended for posterior, noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The EUROPA™ Pedicle Screw System is comprised of both Open and Minimally Invasive Surgery (MIS) polyaxial pedicle screw and rod components that are available in different sizes to accommodate various patient anatomical and physiological requirements. The pedicle screws are manufactured from Ti-6Al-4V ELI alloy (ASTM F136). The rods are manufactured from Molybdenum-47.5Rhenium alloy (ASTM F3273) or cobalt chromium alloy (ASTM F1537).
This document is a 510(k) summary for the EUROPA™ Pedicle Screw System, seeking market clearance for sterile-packaged rod components. It is a Class II medical device, and the submission is a "Special 510(k)" because it is nearly identical to previously cleared devices (K180337, K182970), with the only change being the terminal sterilization method for the rod components.
Therefore, the document does not provide information about acceptance criteria or a study proving the device meets those criteria for the following reasons:
- No new performance data is presented or required. The submission explicitly states under "IX. PERFORMANCE DATA": "Performance data is not provided in this submission." This is because the device is largely identical to its predicate devices, and the change (sterilization method) is addressed through an "Adequate description of the sterilization method" rather than a clinical or performance study.
- This is a 510(k) summary. A 510(k) submission generally demonstrates substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive new performance studies as would be required for a PMA (Premarket Approval Application). A "Special 510(k)" limits the required information even further when the modifications do not affect the device's fundamental function, safety, or effectiveness.
Based on the provided text, I cannot extract the requested information regarding acceptance criteria and performance study details because the document explicitly states that no performance data is provided for this submission.
The document's purpose is to get clearance for a minor change (sterilization of a component) to an already cleared device, not to present novel performance data.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 19, 2019
MiRus, LLC Mr. Jordan Bauman Director of Regulatory Affairs and Quality 2150 Newmarket Parkway SE Marietta, Georgia 30067
Re: K191757
Trade/Device Name: EUROPA™ Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: June 28, 2019 Received: July 1, 2019
Dear Mr. Bauman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K191757
Device Name EUROPA™ Pedicle Screw System
Indications for Use (Describe)
The EUROPA™ Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.
The EUROPA™ Pedicle Screw System is intended for posterior, noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(c).
| I. SUBMITTER | MiRus™, LLC2150 Newmarket Parkway SESuite 108Marietta, Georgia 30067Tel: (678)-324-6272Fax: (678) 401-5607 |
|---|---|
| II. OFFICIAL CORRESPONDENT | Jordan BaumanDirector of Regulatory Affairs and QualityMiRus™, LLC2150 Newmarket Parkway SESuite 108Marietta, Georgia 30067Tel: (678)-324-6272Fax: (678) 401-5607 |
| III. DATE PREPARED | June 28, 2019 |
| IV. DEVICEName of DeviceCommon NameClassification NameRegulatory ClassProduct CodesSubmission Type | EUROPA™ Pedicle Screw SystemThoracolumbosacral pedicle screw system21 CFR §888.3070Class IINKBSpecial 510(k) |
| V. PREDICATE DEVICE | EUROPA™ Pedicle Screw System (K180337,K182970) (Primary Predicate) |
VI. DEVICE DESCRIPTION
The purpose of this 510(k) submission is to obtain market clearance of sterile packaged EUROPA™ Pedicle Screw System rod components. These devices are to be provided sterile via ethylene oxide.
The EUROPA™ Pedicle Screw System is comprised of both Open and Minimally Invasive Surgery (MIS) polyaxial pedicle screw and rod components that are available in different sizes to accommodate various patient anatomical and physiological requirements. The pedicle screws are manufactured from Ti-6Al-4V ELI alloy (ASTM F136). The rods are manufactured from Molybdenum-47.5Rhenium alloy (ASTM F3273) or cobalt chromium alloy (ASTM F1537).
VII. INDICATIONS FOR USE
The EUROPA™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.
{4}------------------------------------------------
The EUROPA™ Pedicle Screw System is intended for posterior, noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
VIII. PREDICATE DEVICE COMPARISON
The subject devices are identical to the predicate devices, except that the subject device rod components will be terminally sterilized via ethylene oxide. The design, materials, indications, and technology remain identical to the predicate devices.
IX. PERFORMANCE DATA
Performance data is not provided in this submission. Adequate description of the sterilization method has been provided.
X. CONCLUSONS
The EUROPA™ Pedicle Screw System with sterile packaged rod components is substantially equivalent to the predicate devices (K180337, K182970) in intended use, indications for use, technological characteristics and labeling.
N/A