K Number
K191581
Manufacturer
Date Cleared
2020-01-08

(208 days)

Product Code
Regulation Number
888.3080
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.

The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment prior to treatment with the ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device. Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. The Interbody Device is indicated to be used with autograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft bone marrow aspirate, or a combination thereof. The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Devicemust be used with an integrated Lateral Plate and Bone Screw and additionally must be used with posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.

Device Description

The ENDOSKELETON® TL Interbody Fusion Device implants are available in a variety of lengths, widths, and heights for treatment in Lateral Lumbar Interbody Fusion and are designed with a large hollow region in the center to house autograft bone or allograft bone comprised of cancellous and/ or corticocancellous bone, demineralized allograft bone marrow aspirate, or a combination thereof. The new bone formation through the implant is intended to provide long-term structural support and fusion at the implanted disc space. The design incorporates "windows" through the implant to permit visualization of device placement and over time formation of new bone. The superior and inferior surfaces include either the Chemtex® surface treatment or nanoLOCK® surface treatment (MMN™) designed to improve fixation to the adjacent bone. The nanoLOCK® surface technology (MMN™) provides a microscopic roughened surface with nano-scale features. The implants are composed of ASTM F136 Ti 6Al-4V ELI titanium alloy and provided either sterile. An implant holding feature has been incorporated into the trailing surface of the implant to mate with the implant inserter and to facilitate placement of the implant into the interbody space.

The subject submission introduces the ENDOSKELETON® TL Hyperlordotic Interbody Fusion Devices are available in a variety of lengths, widths, and heights for treatment in Lateral Lumbar Interbody Fusion and are designed with a large hollow region in the center to house autograft bone comprised of cancellous and/or corticocancellous bone. The design incorporates "windows" through the Interbody Device to permit visualization of the Interbody Device placement and over time formation of new bone. The new bone formation through the Interbody Device is intended to provide long-term structural support and fusion at the implanted disc space. The Interbody Device is treated with nanoLOCK® Surface Technology (MMN™) designed to improve fixation to the adjacent bone. The nanoLOCK® Surface Technology (MMN™) provides a microscopic roughened surface with nano-scale features.

The subject submission also includes Lateral Plate and Bone Screws which are available in a variety of sizes for stabilizing the Interbody Device. The Lateral Plate has one or two hole(s) for receiving integrated Bone Screws and a feature for attachment to the Interbody Device. The Lateral Plate incorporates a lock cover to resist the integrated Bone Screws from backing out. Holding features have been incorporated into the Interbody Device and associated device components to facilitate placement of the system into the interbody space. The Interbody Device and associated device components are composed of ASTM F136 or ASTM F3001 Ti6Al4V ELI titanium alloy and provided sterile.

AI/ML Overview

The provided text describes the regulatory filing for the Endoskeleton® TL Interbody Fusion Device and Endoskeleton® TL Hyperlordotic Interbody Fusion Device. This document is a 510(k) summary, which is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device.

Crucially, this document highlights a mechanical device (intervertebral body fusion device) for spinal fusion, not an AI or software-as-a-medical-device (SaMD) product. Therefore, the acceptance criteria and study designs typically associated with AI/SaMD (e.g., performance metrics like sensitivity/specificity, MRMC studies, ground truth establishment by experts, training/test sets) are not applicable here.

The "performance testing" section refers to biomechanical and material testing (e.g., static and dynamic testing per ASTM standards, particulate/wear analysis, bacterial endotoxin testing) to ensure the device's physical properties and biocompatibility are acceptable, and that it can withstand the forces it will encounter in the body.

Here's how to address your request given the provided context:

The acceptance criteria and study detailed in this document are for a physical medical device (spinal implant), not a software/AI medical device. Therefore, the questions related to AI/SaMD performance metrics, ground truth, expert review, and training/test sets are not applicable.

Here's what can be extracted and inferred from the document regarding the acceptance criteria and the study that proves the device meets them, adapted for a mechanical device:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria CategorySpecific Criteria / Test StandardReported Device Performance / Outcome
Biomechanical PerformanceStatic Testing (accordance with ASTM F2077-18)Performed on worst-case construct; demonstrated compliance with standard.
Dynamic Testing (accordance with ASTM F2077-18)Performed on worst-case construct; demonstrated compliance with standard.
Bone Screw Testing (accordance with ASTM F543-17)Worst-case bone screw configurations tested; demonstrated compliance with standard.
Material PerformanceParticulate and Wear Analysis (on dynamically loaded specimens, per ASTM F1877-16)Conducted; results met criteria for wear debris and particulate generation. (Implied: acceptable levels)
BiocompatibilityBacterial Endotoxin Testing (LAL test)Conducted; results met criteria for endotoxin levels. (Implied: acceptable levels)
Substantial EquivalenceSimilar indications for use, geometry, material, manufacturing processes, construction to predicate devices (K163269, K140055, K170399, K080615, K192018, K073109)Concluded that the subject device does not raise new questions about safety and effectiveness compared to predicates.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not explicitly stated as a "test set" in the context of clinical data for AI. For biomechanical testing, the "worst-case construct" and "worst-case bone screw configurations" were tested. This implies a representative selection of device variations. The exact number of samples tested per ASTM standard is not provided in this summary but would be detailed in the full testing report submitted to the FDA.
  • Data Provenance: Not applicable in the sense of patient data. The "data" comes from laboratory and physical testing of the device prototypes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This is a physical device; ground truth is established through adherence to engineering standards (ASTM) and physical measurements/tests, not through expert clinical consensus of images or outcomes as would be done for an AI device.

4. Adjudication method for the test set:

  • Not Applicable. No human adjudication process is described for this type of device testing. Compliance with ASTM standards is determined by the results of the physical tests.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

  • No. An MRMC study is relevant for evaluating the impact of AI on human reader performance, typically in diagnostic imaging. This device is a surgical implant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical implant, not an algorithm.

7. The type of ground truth used:

  • Mechanical and Biological Standards: Ground truth is established by adherence to recognized industry standards (ASTM), material specifications (ASTM F136 Ti 6Al-4V ELI, ASTM F3001 Ti6Al4V ELI), and biological compatibility testing requirements (LAL for endotoxins). The "ground truth" is that the device physically performs according to these established engineering and safety benchmarks.

8. The sample size for the training set:

  • Not Applicable. There is no "training set" in the context of machine learning for a mechanical device. The design and manufacturing process are refined through engineering principles and iterative testing, not by "training" an algorithm on a dataset.

9. How the ground truth for the training set was established:

  • Not Applicable. As there's no training set, there's no ground truth to establish in this context. The design process for a physical medical device relies on established engineering principles, material science, and prior knowledge from predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 8, 2020

Titan Spine, Inc. % Christine Scifert Exec VP MRC/X, LLC 6075 Poplar Avenue Memphis. Tennessee 38119

Re: K191581

Trade/Device Name: Endoskeleton® TL Interbody Fusion Device, Endoskeleton® TL Hyperlordotic Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX Dated: December 6, 2019 Received: December 11, 2019

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent L. Showalter, PhD Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191581

Device Name

Endoskeleton® TL Interbody Fusion Device Endoskeleton® TL Hyperlordotic Interbody Fusion Device

Indications for Use (Describe)

The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone marrow aspirate.

The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device (≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment prior to treatment with the ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device. Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. The Interbody Device is indicated to be used with autograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft bone marrow aspirate, or a combination thereof. The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Devicemust be used with an integrated Lateral Plate and Bone Screw and additionally must be used with posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Endoskeleton® TL Interbody Fusion Device Endoskeleton® TL Hyperlordotic Interbody Fusion Device January 7, 2020

Company:Titan Spine, Inc.6140 West Executive Drive, Suite AMequon, WI 53092, USA
Establishment Registration:3006340236
Primary Contact:Christine ScifertPhone: 901-831-8053
Company Contact:Kelly McDonnellPhone: 866-822-7800Fax: 262-242-7802
Trade Name:Endoskeleton® TL Interbody Fusion DeviceEndoskeleton® TL Hyperlordotic Interbody Fusion Device
Common Name:Intervertebral Body Fusion Device with Bone Graft, LumbarIntervertebral Fusion Device With Integrated Fixation, Lumbar
Classification:Class II
Regulation Number:21 CFR 888.3080 (Intervertebral body fusion device)
Panel:87- Orthopedic
Product Code:MAX, OVD
Predicate Devices:Primary Predicate:Titan Spine Endoskeleton® TAS / TASH – K163269Secondary Predicates:Endoskeleton® TL Interbody Fusion Device – K140055Endoskeleton® TO Interbody Fusion Device – K170399Endoskeleton® TA Interbody Fusion Device – K080615Surgicraft STAFLIF TT – K073109Endoskeleton® Interbody Fusion Devices - K192018Reference Predicates:Titan Spine Allograft Indications update - K183557Titan Spine Sterile TAS / TCS screws – K173535

{4}------------------------------------------------

Device Description:

The ENDOSKELETON® TL Interbody Fusion Device implants are available in a variety of lengths, widths, and heights for treatment in Lateral Lumbar Interbody Fusion and are designed with a large hollow region in the center to house autograft bone or allograft bone comprised of cancellous and/ or corticocancellous bone, demineralized allograft bone marrow aspirate, or a combination thereof. The new bone formation through the implant is intended to provide long-term structural support and fusion at the implanted disc space. The design incorporates "windows" through the implant to permit visualization of device placement and over time formation of new bone. The superior and inferior surfaces include either the Chemtex® surface treatment or nanoLOCK® surface treatment (MMN™) designed to improve fixation to the adjacent bone. The nanoLOCK® surface technology (MMN™) provides a microscopic roughened surface with nano-scale features. The implants are composed of ASTM F136 Ti 6Al-4V ELI titanium alloy and provided either sterile. An implant holding feature has been incorporated into the trailing surface of the implant to mate with the implant inserter and to facilitate placement of the implant into the interbody space.

The subject submission introduces the ENDOSKELETON® TL Hyperlordotic Interbody Fusion Devices are available in a variety of lengths, widths, and heights for treatment in Lateral Lumbar Interbody Fusion and are designed with a large hollow region in the center to house autograft bone comprised of cancellous and/or corticocancellous bone. The design incorporates "windows" through the Interbody Device to permit visualization of the Interbody Device placement and over time formation of new bone. The new bone formation through the Interbody Device is intended to provide long-term structural support and fusion at the implanted disc space. The Interbody Device is treated with nanoLOCK® Surface Technology (MMN™) designed to improve fixation to the adjacent bone. The nanoLOCK® Surface Technology (MMN™) provides a microscopic roughened surface with nano-scale features.

The subject submission also includes Lateral Plate and Bone Screws which are available in a variety of sizes for stabilizing the Interbody Device. The Lateral Plate has one or two hole(s) for receiving integrated Bone Screws and a feature for attachment to the Interbody Device. The Lateral Plate incorporates a lock cover to resist the integrated Bone Screws from backing out. Holding features have been incorporated into the Interbody Device and associated device components to facilitate placement of the system into the interbody space. The Interbody Device and associated device components are composed of ASTM F136 or ASTM F3001 Ti6Al4V ELI titanium alloy and provided sterile.

Indication for Use:

The ENDOSKELETON® TL Interbody Fusion Device is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation that has been cleared by the FDA for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft

{5}------------------------------------------------

bone and/or allograft bone comprised of cancellous and/or corticocancellous bone and/or demineralized allograft bone with bone marrow aspirate.

The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device (≥ 16") is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment prior to treatment with the ENDOSKELETON® TL Hyperlordotic Interbody Fusion Device. Patients with previous non-fusion spinal surgery at the involved level(s) may be treated with the device. The Interbody Device is indicated to be used with autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft bone marrow aspirate, or a combination thereof. The ENDOSKELETON® TL Hyperlordotic Interbody Fusion Devicemust be used with an integrated Lateral Plate and Bone Screw and additionally must be used with posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.

Substantial Equivalence:

Primary predicate Endoskeleton TAS / TAS Hyperlordotic (K163269), Secondary predicates including Endoskeleton TL (K140055), Endoskeleton TO (K170399), Endoskeleton TA (K080615), Endoskeleton® Interbody Fusion Devices (K192018) and Centinel Spine / Surgicraft STAFLIF TT (K073109) and Reference predicates include Titan Spine Allograft indications update (K183557), Titan Spine sterile TAS/ TCS screws (K173535), and Titan Spine sterile only (K142589).

The subject devices share similar indications for use, geometry, material, manufacturing processes and construction with the predicate devices. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.

Performance Testing:

The subject Endoskeleton® TL and TL Hyperlordotic device constructs underwent static and dynamic testing in accordance with ASTM F2077-18 and ASTM F2267-04(R18), where testing was performed on the worst-case construct. The worst-case bone screw configurations were additionally tested per ASTM F543-17. Particulate and wear analysis on dynamically loaded specimens was conducted per ASTM F1877-16. Bacterial endotoxin testing (LAL) was conducted for these devices.

Conclusion:

Based upon the information contained in this submission and the similarities of the subject and predicate devices, the subject Endoskeleton® TL Interbody Fusion Device and Endoskeleton® TL Hyperlordotic Fusion Devices are substantially equivalent to the predicate devices.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.