K Number
K191412
Manufacturer
Date Cleared
2019-08-23

(87 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VariAx 2 Mini Fragment System is indicated for fracture fixation, replantation, stabilization, reduction, fusions, osteotomies, mal-unions of small bones and small bone fragments including normal and osteopenic bones in adult and adolescent (12 - 21 years) patients. The system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • Intra- and extra-articular fractures
  • · Compression fracture
  • Displaced fracture
  • · Reconstruction
  • · Replantation
  • · Arthrodesis
  • Osteotomy
Device Description

The VariAx 2 Mini Fragment System is used for internal fixations and is composed of sterile and nonsterile plates. Additionally, the plates are provided in locking and non-locking options, as well as various lengths, thicknesses, and configurations.

The VariAx 2 System is used for internal fixation applications and is composed of sterile and nonsterile screws, washers, and instruments. In addition to independent use, screws of this system are also used with compatible, internal fixation systems. These devices are made of commercially pure titanium and titanium alloy with Type III anodization, and are available in a variety of sizes and diameters, as well as locking and non-locking types.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Stryker VariAx 2 System and VariAx 2 Mini Fragment System. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria as might be seen for novel devices.

Therefore, the document does not contain the detailed information requested regarding acceptance criteria and a study proving the device meets them in the way this request envisions for a software or AI-based medical device.

However, I can extract the information provided about the non-clinical testing performed to demonstrate substantial equivalence, which serves a similar purpose in the context of a 510(k) submission.

Here's the breakdown of what is available based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria. Instead, it states that "The aforementioned testing demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the predicate devices." This implies that the performance of the new devices in these tests was comparable to or better than the predicate devices, satisfying the FDA's requirement for substantial equivalence for mechanical properties.

Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence to predicate devices in:Torsion: Demonstrated substantial equivalence to predicate devices.
- TorsionInsertion Torque: Demonstrated substantial equivalence to predicate devices.
- Insertion TorquePull-out: Demonstrated substantial equivalence to predicate devices.
- Pull-outFour-point bending: Demonstrated substantial equivalence to predicate devices.
- Four-point bendingMRI Safety (Magnetically-induced displacement force, Magnetically-induced torque, RF-induced heating, and Image artifacts): Demonstrated to be MR conditional.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes (e.g., number of screws, plates tested) for the non-clinical mechanical tests. It also does not provide information on data provenance as these are physical device tests, not data-driven studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This was a non-clinical, mechanical testing study of physical medical devices, not a study involving expert interpretation of data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This was a non-clinical, mechanical testing study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This was a non-clinical, mechanical testing study, not a human reader study involving AI.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This was a non-clinical, mechanical testing study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical tests, the "ground truth" or reference for comparison was the performance of the identified predicate devices (Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (K132886) and Synthes (USA) Modular Mini Fragment LCP System (K063049) for the Mini Fragment System, and VariAx 2 System (K180500) for the VariAx 2 System). The tests measured physical properties and compared them against known standards or the predicate's performance. For MRI safety, the ground truth would be established by industry standards or FDA guidance for MR compatibility.

8. The sample size for the training set

Not applicable. This was a non-clinical, mechanical testing study, not an AI or machine learning study requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This was a non-clinical, mechanical testing study.


Summary of the Study:

The study referenced in the 510(k) submission is a non-clinical mechanical testing study. Its purpose was to demonstrate substantial equivalence of the VariAx 2 Mini Fragment System and the VariAx 2 System to their respective predicate devices.

Tests Performed:

  • Torsion
  • Insertion Torque
  • Pull-out
  • Four-point bending
  • MRI assessments (magnetically-induced displacement force, magnetically-induced torque, RF-induced heating, and image artifacts)

Conclusion of the Study:

The non-clinical testing "demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the predicate devices." Additionally, the devices were found to be MR conditional. Clinical testing was not required for this submission.

{0}------------------------------------------------

August 23, 2019

Stryker GmbH Sanja Jahr Senior Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K191412

Trade/Device Name: VariAx 2 System, VariAx 2 Mini Fragment System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HRS Dated: May 24, 2019 Received: May 28, 2019

Dear Sanja Jahr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali. MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

List of Cleared Devices in K191412

{2}------------------------------------------------

List of Cleared Devices in K191412

    1. VariAx 2 Mini Fragment System
    1. VariAx 2 System

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K191412

Device Name VariAx 2 Mini Fragment System

Indications for Use (Describe)

The VariAx 2 Mini Fragment System is indicated for fracture fixation, replantation, stabilization, reduction, fusions, osteotomies, mal-unions of small bones and small bone fragments including normal and osteopenic bones in adult and adolescent (12 - 21 years) patients. The system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191412

Device Name VariAx 2 System

Indications for Use (Describe)

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • · Intra- and extra-articular fractures
  • · Compression fracture
  • Displaced fracture
  • · Reconstruction
  • · Replantation
  • Arthrodesis
  • · Osteotomy
Type of Use (Select one or both, as applicable)

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Proprietary Name:VariAx 2 Mini Fragment System
VariAx 2 System
Common Name:Bone screw
Bone plate
Regulation Description:Single/multiple component metallic bone fixation appliances
and accessories.
Smooth or threaded metallic bone fixation fastener.
Regulation Number:21 CFR 888.3030
21 CFR 888.3040
Product Code:HRS
HWC
Device Class:Class II
Sponsor:Stryker GmbH
Bohnackerweg l
2545 Selzach / Switzerland
Contact Person:Sanja Jahr
Senior Regulatory Affairs Specialist
325 Corporate Drive
Mahwah, NJ 07430
Phone: 201-831-6797
Fax: 201-831-6020
Date Prepared:May 24, 2019
Predicate Device:VariAx 2 Mini Fragment System
Primary Predicate: Smith & Nephew Variable-Angle Locking
Mini-Fragment Plating System (K132886)
Reference Device: Synthes (USA) Modular Mini Fragment LCP
System (K063049)
VariAx 2 SystemPrimary Predicate: VariAx 2 System (180500)

510(k) Summary

{6}------------------------------------------------

Description

The VariAx 2 Mini Fragment System is used for internal fixations and is composed of sterile and nonsterile plates. Additionally, the plates are provided in locking and non-locking options, as well as various lengths, thicknesses, and configurations.

The VariAx 2 System is used for internal fixation applications and is composed of sterile and nonsterile screws, washers, and instruments. In addition to independent use, screws of this system are also used with compatible, internal fixation systems. These devices are made of commercially pure titanium and titanium alloy with Type III anodization, and are available in a variety of sizes and diameters, as well as locking and non-locking types.

Indications for Use

VariAx 2 Mini Fragment System

The VariAx 2 Mini Fragment System is indicated for fracture fixation, replantation, stabilization, reduction, fusions, osteotomies, mal-unions, and non-unions of small bones and small bone fragments including normal and osteopenic bones in adult and adolescent (12 – 21 years) patients. The system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.

VariAx 2 System

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • Intra- and extra-articular fractures
  • · Compression fracture
  • Displaced fracture
  • · Reconstruction
  • · Replantation
  • · Arthrodesis
  • Osteotomy

Summary of Technologies

A comparison of the systems demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the previously cleared Synthes (USA) Modular Mini Fragment LCP System (K063049), Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (K132886), and VariAx 2 System (K180500) in regard to intended use, material, design, and operational principles.

Non-Clinical Testing

Comparative mechanical testing to the predicate system demonstrated substantial equivalence. The following tests were performed:

  • Torsion

{7}------------------------------------------------

VariAx 2 Mini Fragment System & VariAx 2 System Stryker GmbH

  • . Insertion Torque
  • . Pull-out
  • . Four-point bending

The aforementioned testing demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the predicate devices. Additionally, MR assessments of magnetically-induced displacement force, magnetically-induced torque, RF-induced heating, and image artifacts demonstrate that the VariAx 2 Mini Fragment System and the VariAx 2 System are MR conditional.

Clinical Testing

This submission does not require clinical testing.

Conclusion

The VariAx 2 Mini Fragment System and the VariAx 2 System described in this submission have the same indications, intended use, target patient population, technological characteristics, and materials as the predicate devices. The mechanical testing demonstrates the performance of the proposed devices is equivalent to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.