(170 days)
DemeDIOX Barbed polydioxanone suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
DemeDIOX Barbed Absorbable Surgical Suture is an absorbable, sterile surgical monofilament suture composed of polyester, poly (p-dioxanone). The DemeTECH Suture meets all requirements in the latest edition of the USP monograph for absorbable surgical sutures with the exception of diameter. DemeDIOX Barbed Suture consists of an absorbable monofilament strand thread with spiral unidirectional barbs and is available with or without needles in a suture size 5-0 to 2. The material is dyed with D&C Violet No.2. and contains no additives. The barbing allows for tissue approximation without the use of surgical knots.
The provided text describes the submission of a 510(k) premarket notification for a medical device: DemeDIOX Barbed Absorbable Surgical Suture. This type of submission relies on demonstrating substantial equivalence to a previously legally marketed device (predicate device) rather than proving de novo safety and effectiveness through extensive clinical trials.
Therefore, the study design described is focused on non-clinical testing to demonstrate conformance to established standards and comparison to predicate devices, rather than a clinical study measuring human-in-the-loop performance or outcomes in a clinical setting.
Based on the provided information, here's an analysis of the acceptance criteria and the "study" (non-clinical testing) that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are largely derived from established standards for absorbable surgical sutures, primarily the United States Pharmacopeia (USP) and ISO 10993. The reported device performance demonstrates adherence to these standards and equivalence to predicate devices.
| Acceptance Criteria (Derived from USP 42, ISO 10993, and predicate equivalence) | Reported Device Performance (as stated in the 510(k) Summary) |
|---|---|
| Material Composition | Composed of polyester, poly (p-dioxanone), same as predicate devices. |
| Sterility | Sterile. Achieved via Ethylene Oxide (EO) sterilization. |
| Biocompatibility | Conforms to ISO 10993-1. Material (polydioxanone) has |
| "known biocompatibility" and "established history of use" in surgical sutures. | |
| In-vitro Resorption | Testing conducted to further demonstrate substantial equivalence to predicate devices. (No specific numerical results provided, but stated to support equivalence). |
| USP 42 Requirements for "Absorbable Surgical Suture" | Suture Material meets or exceeds requirements, with the exception of diameter (due to barbs). This includes: |
| a. Suture Diameter | Meets USP 42 requirements except for diameter. This deviation is explicitly acknowledged and attributed to the presence of barbs on the suture. |
| b. Suture Length | Meets or exceeds the performance requirements defined in USP for "Suture Length Requirement" (95% of stated length). |
| c. Tensile Strength (<881> USP) | Meets or exceeds the performance requirements defined in USP 42 for "Tensile Strength." |
| d. Needle Attachment (<871> USP) | For sutures supplied with needles, meets or exceeds the performance requirements defined in USP 42 for "Needle Attachment." |
| Extractable Color | Assured conformance to methods outlined in USP 42. (The device is dyed with D&C Violet No.2). |
| Physical Properties & Functionality | Assured conformance to suture diameter and length. (Reiterates USP conformance with the noted diameter exception due to barbs). The barbing allows for tissue approximation without knots. |
| Labeling | Sterile, Single Use. Packaging and labeling conform to 21 CFR standards. |
| Substantial Equivalence | Demonstrated to be substantially equivalent to the predicate devices (Ethicon PDS Barbed Suture K113004 and DemeTECH Polydioxanone Synthetic Monofilament K082097) in terms of intended use, technical characteristics, material, sterilization, and performance (with the acknowledged exception for diameter due to barbs). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes (e.g., number of sutures tested) for each non-clinical test (tensile strength, diameter, length, etc.). It generally states that "Non-clinical testing was conducted... to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and in-vitro resorption." and "Physical properties and functionality testing assured that the device conformed with suture diameter and suture length, extractable color and sterility to methods outlined in USP 42." These tests typically involve a statistically relevant number of samples to ensure robust results, as outlined in the respective standards (USP, ISO).
- Data Provenance: The data is generated from non-clinical laboratory testing performed specifically for this 510(k) submission. No information on the country of origin of the data or whether it was retrospective or prospective is given, but non-clinical tests would inherently be prospective for the purpose of demonstrating conformance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable to this type of submission. Ground truth, in the context of diagnostic AI or image analysis, typically involves expert human annotation or clinical outcomes. For a medical device like a surgical suture, the "ground truth" and acceptance criteria are defined by established performance standards (USP, ISO) and comparison to predicate device characteristics as detailed in their respective 510(k) clearances. The performance is measured against these objective standards through physical and chemical testing, not through expert consensus on a test set.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used in studies where human experts are making subjective assessments, often in diagnostic accuracy studies. For testing a physical device like a suture, the "adjudication" is inherent in the standardized testing process where results are objectively measured against defined criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a MRMC comparative effectiveness study was not done.
- This type of study is relevant for diagnostic devices, particularly those involving human interpretation (e.g., AI-assisted radiology).
- The device in question, DemeDIOX Barbed Absorbable Surgical Suture, is a physical surgical suture, not a diagnostic or AI-powered device that assists human readers.
- The submission explicitly states: "No clinical trials were conducted."
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No, a standalone study (in the context of an algorithm or AI) was not done.
- This is not an AI/algorithm-based device.
- The performance assessment for this device is based on its physical properties, material characteristics, and biological compatibility as measured directly through laboratory tests and comparisons to predicate devices, not through an algorithm.
7. Type of Ground Truth Used
The "ground truth" used for this device's acceptance is based on:
- Established Industry Standards: Primarily the United States Pharmacopeia (USP 42) monographs for absorbable surgical sutures, which define physical and chemical performance requirements (tensile strength, diameter, length, etc.).
- International Standards: ISO 10993-1 for biocompatibility.
- Predicate Device Characteristics: Demonstrating that the new device is "substantially equivalent" to legally marketed predicate devices (Ethicon PDS Barbed Suture and DemeTECH Polydioxanone Synthetic Monofilament Suture) in terms of intended use, technology, and safety/performance profile.
8. Sample Size for the Training Set
This information is not applicable. This submission is for a physical medical device (surgical suture), not a machine learning model or an AI algorithm. Therefore, there is no "training set" in the sense of data used to train an algorithm. The development of the suture material and manufacturing process would involve internal R&D and quality control, but not a formally defined "training set" for an AI.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as there is no "training set" for an AI algorithm in this context. The "ground truth" for the development and testing of the suture is established by the well-defined and often long-standing scientific and engineering principles of material science, biomaterials, and surgical product design, all guided by regulatory standards like USP and ISO.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 7, 2019
DemeTECH Corporation Tracy Chadwrick Director of Quality 14175 NW 60th Ave Miami Lakes, Florida 33014
Re: K191361
Trade/Device Name: DemeDIOX Barbed Absorbable Surgical Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: May 10, 2019 Received: May 21, 2019
Dear Tracy Chadwrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191361
Device Name
DemeDIOX Barbed Absorbable Surgical Suture
Indications for Use (Describe)
DemeDIOX Barbed polydioxanone suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| ☒ Registration Use (Part 31 CFR 901.3, Subpart B) | ☐ One-Time Contact Use (31 CFR 901.3, Subpart C) |
|---|---|
| --------------------------------------------------------------------------------------------------- | -------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for "DemeTECH". To the left of the company name is a blue circular logo with a white triangle inside. The company name is written in a bold, black font.
510(K) Summary
| Submission Date: | 11/07/2019 |
|---|---|
| K191361 | |
| SUBMITTER INFORMATION: | |
| Company Name: | DemeTECH Corporation |
| Company Address: | 14175 NW 60th Avenue, Miami Lakes, FL 33014 |
| Contact Person: | Tracy ChadwrickPhone: 305-824-1048 Ext 131Tracy.Chadwrick@demetech.us |
| Device Trade Name: | DemeDIOX Barbed Absorbable Surgical Suture |
| Device Common Name: | POLYDIOXANONE monofilament synthetic absorbable sutures |
| Class: | Class II |
| Classification: | 21 CFR 878.4840Absorbable Barbed Polydioxanone Surgical Suture |
| Product Code: | NEW |
Predicate Devices:
The predicate is Ethicon PDS Barbed Suture, 510(k) - K113004 (The primary predicate), and the original DemeTECH Polydioxanone Synthetic Monofilament (PDO) Absorbable Suture, 510(k) -K082097 (Additional predicate).
Device Description:
DemeDIOX Barbed Absorbable Surgical Suture is an absorbable, sterile surgical monofilament suture composed of polyester, poly (p-dioxanone). The DemeTECH Suture meets all requirements in the latest edition of the USP monograph for absorbable surgical sutures with the exception of diameter. DemeDIOX Barbed Suture consists of an absorbable monofilament strand thread with spiral unidirectional barbs and is available with or without needles in a suture size 5-0 to 2. The material is dyed with D&C Violet No.2. and contains no additives. The barbing allows for tissue approximation without the use of surgical knots.
Intended Use:
DemeDIOX Barbed polydioxanone suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "DemeTECH". To the left of the company name is a light blue triangular shape. The company name is in bold black font.
Non-Clinical Tests Performed:
Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and in-vitro resorption to further demonstrate substantial equivalence to the predicate devices. Physical properties and functionality testing assured that the device conformed with suture diameter and suture length, extractable color and sterility to methods outlined in USP 42.
Substantial Equivalence:
| Comparison Items | DemeTECHDemeDIOX BarbedAbsorbable Suture | DemeTECHDemeDIOX AbsorbableSuture | EthiconPDS Barbed Suture(Primary Predicate> |
|---|---|---|---|
| FDA ID# | K191361 | K082097 | K113004 |
| Product Code | NEW | Same | Same |
| Common Name | Polydioxanone BarbedAbsorbable Suture | Same | Same |
| Suture Characteristic | Barbed AbsorbablePolydioxanone surgicalsuture | AbsorbablePolydioxanone surgicalsuture | Same |
| Labeling | Sterile, Single Use | Same | Same |
| Intended Use | Soft tissue approximation | Same | Same |
| Technical Characteristics | Monofilament, syntheticspiral unidirectionalBarbed absorbablesuture is prepared frompolyester, poly-(p-dioxanone) | Monofilament, syntheticnon-Barbed absorbablesuture is prepared frompolyester, poly-(p-dioxanone) | Same |
| Material | Prepared from polyester,poly-(p-dioxanone) | Same | Same |
| Sizes | 5-0, 4-0, 3-0, 2-0, 1-0, 0, 1, 2 | 7-0, 6-0, 5-0, 4-0, 3-0, 2-0, 1-0, 0, 1, 2 | 3-0, 2-0, 1-0, 0, 1 |
| Sterilization | Ethylene Oxide (EO) | Same | Same |
| USP Requirements | Suture Material meets orexceeds requirements for"Absorbable SurgicalSuture" in USP 42 exceptdiameter. | Suture Material meets orexceeds requirements for"Absorbable SurgicalSuture" in USP 42 | Suture Material meets orexceeds requirements for"Absorbable SurgicalSuture" in USP 42 exceptdiameter |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "K191361 Page 3 of 3". This text appears to be a document identifier and page number. The document identifier is K191361, and the page number is 3 of 3, indicating that this is the last page of the document.
Image /page/5/Picture/1 description: The image shows the DemeTECH logo. The logo consists of a light blue circular icon on the left and the company name "DemeTECH" in bold, black letters on the right. A registered trademark symbol is located to the right of the company name.
| Comparison Items | DemeTECHDemeDIOX BarbedAbsorbable Suture | DemeTECHDemeDIOX AbsorbableSuture | EthiconPDS Barbed Suture(Primary Predicate> |
|---|---|---|---|
| Tensile StrengthRequirements | Suture Material meets orexceeds the performancerequirements defined inthe USP 42 for "TensileStrength" <881> | Same | Same |
| Needle AttachmentRequirements | Suture Material meets orexceeds the performancerequirements defined inthe USP 42 for "NeedleAttachment" <871> | Same | Same |
| Suture LengthRequirements | Suture Material meets orexceeds the performancerequirements defined inUSP for "Suture LengthRequirement" (95% of | Same | Same |
| Suture Packaging | Suture Material ispackaged in a same orequivalent manner withsterile single or doublepackage having labelingconforming to 21 CFR | Same | Same |
Clinical Tests Performed:
No clinical trials were conducted
Conclusion:
DemeDIOX Barbed Absorbable Surgical Suture is composed of the same material as the predicated devices DemeTECH DemeDIOX Polydioxanone Absorbable Suture and the Ethicon PDS Barbed Suture. It also has the same design being a sterile, flexible, monofilament absorbable thread meeting all the requirements of the United States Pharmacopeia with the exception diameter due to the small barbs created on the suture. DemeTECH's material used was selected based on known biocompatibility (per ISO 10993) and established history of use in the surgical suture industry.
The biocompatibility data and the results of performance testing presented, demonstrate the substantial equivalence of DemeTECH DemeDIOX Barbed Absorbable Surgical Suture to that of the predicate devices. It further demonstrates conformance with the USP, ISO 10993 and FDA Guidance for Surgical Suture 510(k).
§ 878.4840 Absorbable polydioxanone surgical suture.
(a)
Identification. An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.