K Number
K191311

Validate with FDA (Live)

Date Cleared
2019-08-12

(89 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ATEC Lateral Interbody System is indicated for spinal fusion procedures in skeletally mature patients at one or two contiguous levels in the thoracolumbar spine.

Thoracic: T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic degenerative disc disease (DDD), including thoracic disc hemiation (myelopathy with or without axial pain). The lateral approach is limited to levels T5-6 to T11-T12.

Lumbar: L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The ATEC Lateral Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.

LIF Anti-Migration Plate (AMP) may be used with ATEC Lateral Interbody spacers to provide integrated fixation. Spacers with >20° lordosis must be used with LIF AMP integrated fixation to supplemental fixation.

Device Description

The ATEC Lateral Interbody System includes IdentiTi LIF Porous Ti Interbody, Battalion LLIF Spacer System and LIF Interbody System. Any of the three subsystems may be used with LIF AMP integrated fixation. The ATEC Lateral Interbody System is an intervertebral body fusion system implanted from an anterolateral or lateral approach. The interbody implants consist of various lengths, widths, heights and degrees of lordosis to accommodate individual patient anatomy. These implants are manufactured PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, titanium coating per ASTM F1580, tantalum per ASTM F560, titanium alloy (Ti- 6Al-4V ELI) per ASTM F136, and commercially pure titanium (CPTi Grade 2) per ASTM F67. The device includes rows of teeth on the surface of each end of the device which serve to grip the adjacent vertebrae to resist migration and expulsion of the device. Additionally, the commercially pure titanium implants are offered with a microstructure due to the layering of material that forms the porous geometry. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. All interbody implants feature an internal graft aperture for placement of bone graft to promote fusion through the cage.

The ATEC Lateral Interbody System includes LIF Anti-Migration Plate (AMP) integrated fixation that may be used with all ATEC Lateral Interbody offerings. The LIF AMP integrated fixation includes fixation plates, center locking screws and bone screws manufactured from titanium alloy per ASTM F136. This 510(k) submission seeks clearance to add size options to the LIF AMP portfolio.

AI/ML Overview

This document is a 510(k) premarket notification from the FDA for a medical device called the "ATEC Lateral Interbody System." It is not a document about an AI/ML powered device, therefore, the requested information about acceptance criteria and study proving device meets the criteria (particularly regarding elements like "human readers improve with AI vs without AI assistance," "standalone algorithm performance," and "ground truth establishment for training set") are not applicable to this submission.

This 510(k) is for a physical medical device (implants for spinal fusion), and its acceptance criteria and performance data are based on nonclinical testing and comparison to legally marketed predicate devices, not on AI/ML performance.

Therefore, most of the specific questions about AI/ML device testing cannot be answered from the provided text.

Here is what can be extracted from the document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states that the device was tested to support substantial equivalence. The "acceptance criteria" are implied to be meeting the performance standards and being substantially equivalent to predicate devices for the tests conducted.

Acceptance Criteria (Implied)Reported Device Performance
Meet dynamic axial compression standards per ASTM F2077Passed
Meet dynamic compression-shear standards per ASTM F2077Passed
Satisfactory subsidence analysisPassed
Satisfactory push-out analysisPassed
Meet particulate and gravimetric analysis standards per ASTM F1877 and ASTM F1714Passed
Meet bacterial endotoxin testing standards per ANSI/AAMI ST72:2011/(R)2016Passed
Substantially equivalent to predicate devices regarding indications for use, intended use, design, technology, and performanceAchieved

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not specified in terms of number of devices tested for each nonclinical test. The tests are typically conducted on a representative sample of the devices.
  • Data Provenance: Not applicable in the context of clinical data. The tests are performed in a laboratory setting for the physical device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

  • Not applicable. Ground truth in the context of AI/ML models (e.g., image annotation by radiologists) is not relevant for this physical device. The "ground truth" here is the pass/fail result of standardized mechanical and biological tests.

4. Adjudication Method for the Test Set:

  • Not applicable. This relates to human interpretation of data, which is not the primary assessment method for device performance in this context. The tests have predefined pass/fail criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No. This type of study is for evaluating human performance (e.g., diagnostic accuracy) with or without AI assistance. It is not applicable to a physical spinal implant. The document explicitly states: "Clinical Information: Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data."

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/ML algorithm.

7. The Type of Ground Truth Used:

  • Not applicable in the AI/ML sense. For this physical device, the "ground truth" for proving performance are the established and recognized ASTM and ANSI/AAMI standards for mechanical and biological properties of intervertebral body fusion devices.

8. The Sample Size for the Training Set:

  • Not applicable. There is no "training set" as this is not an AI/ML device.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. There is no "training set" or "ground truth" in the AI/ML sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 12, 2019

Alphatec Spine, Inc. Cynthia Adams Project Manager, Regulatory Affairs 5818 El Camino Real Carlsbad, California 92008

Re: K191311

Trade/Device Name: ATEC Lateral Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, PHM, OVD Dated: July 16, 2019 Received: July 17, 2019

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191311

Device Name ATEC Lateral Interbody System

Indications for Use (Describe)

The ATEC Lateral Interbody System is indicated for spinal fusion procedures in skeletally mature patients at one or two contiguous levels in the thoracolumbar spine.

Thoracic: T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic degenerative disc disease (DDD), including thoracic disc hemiation (myelopathy with or without axial pain). The lateral approach is limited to levels T5-6 to T11-T12.

Lumbar: L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The ATEC Lateral Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.

LIF Anti-Migration Plate (AMP) may be used with ATEC Lateral Interbody spacers to provide integrated fixation. Spacers with >20° lordosis must be used with LIF AMP integrated fixation to supplemental fixation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for ATEC. The logo consists of a green square with a white "a" inside of it, followed by the word "tec" in white. The background of the logo is a dark blue color.

I. SUBMITTER:Alphatec Spine, Inc.5818 El Camino RealCarlsbad, CA 92008Phone: (760) 431-9286Fax: (760) 431-0289
Contact Person:Cynthia AdamsProject Manager, Regulatory AffairsContact Phone: (760) 494-6740
Date Summary Prepared:May 14, 2019
II. DEVICE
Name of Device:ATEC Lateral Interbody System
Common or Usual Name:Intervertebral fusion device with bone graft, lumbarIntervertebral fusion device with integrated fixation lumbar
Classification Name:Intervertebral body fusion device(21 CFR 888.3080)
Regulatory Class:Class II
Product Code:MAX, PHM, OVD

LEGALLY MARKETED PREDICATE DEVICES III.

510(k)Product CodeTrade NameManufacturer
Primary Predicate Device
K183705PHM, MAX, OVD, ODPIdentiTi Porous Ti Interbody SystemAlphatec Spine
Additional Predicate Devices
K182746PHM, MAX, OVDATEC ALIF and LLIF Spacer SystemAlphatec Spine
K173892MAX, OVDXLX Interbody SystemNuVasive, Inc

IV. DEVICE DESCRIPTION

The ATEC Lateral Interbody System includes IdentiTi LIF Porous Ti Interbody, Battalion LLIF Spacer System and LIF Interbody System. Any of the three subsystems may be used with LIF AMP integrated fixation. The ATEC Lateral Interbody System is an intervertebral body fusion system implanted from an anterolateral or lateral approach. The interbody

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Atec. The logo consists of a green square with a white curved line inside, followed by the word "atec" in white. The letters are lowercase, and there is a trademark symbol after the "c". The background is a dark blue color.

implants consist of various lengths, widths, heights and degrees of lordosis to accommodate individual patient anatomy. These implants are manufactured PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, titanium coating per ASTM F1580, tantalum per ASTM F560, titanium alloy (Ti- 6Al-4V ELI) per ASTM F136, and commercially pure titanium (CPTi Grade 2) per ASTM F67. The device includes rows of teeth on the surface of each end of the device which serve to grip the adjacent vertebrae to resist migration and expulsion of the device. Additionally, the commercially pure titanium implants are offered with a microstructure due to the layering of material that forms the porous geometry. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. All interbody implants feature an internal graft aperture for placement of bone graft to promote fusion through the cage.

The ATEC Lateral Interbody System includes LIF Anti-Migration Plate (AMP) integrated fixation that may be used with all ATEC Lateral Interbody offerings. The LIF AMP integrated fixation includes fixation plates, center locking screws and bone screws manufactured from titanium alloy per ASTM F136. This 510(k) submission seeks clearance to add size options to the LIF AMP portfolio.

INDICATIONS FOR USE V.

The ATEC Lateral Interbody System is indicated for spinal fusion procedures in skeletally mature patients at one or two contiguous levels in the thoracolumbar spine.

Thoracic: T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic degenerative disc disease (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain). The lateral approach is limited to levels T5-6 to T11-T12.

Lumbar: L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The ATEC Lateral Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.

LIF Anti-Migration Plate (AMP) may be used with ATEC Lateral Interbody spacers to provide integrated fixation. Spacers with >20° lordosis must be used with LIF AMP integrated fixation in addition to supplemental fixation.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Atec. The logo consists of a green square with a white stylized "a" inside of it. To the right of the square is the word "tec" in white font. There is a trademark symbol to the upper right of the "c".

VI. TECHNOLOGICAL COMPARISON TO PREDICATES

The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent.

PERFORMANCE DATA VII.

Nonclinical testing performed on the ATEC Lateral Interbody System supports substantial equivalence to other predicate devices. The following testing was performed:

  • Dynamic axial compression testing per ASTM F2077
  • Dynamic compression-shear testing per ASTM F2077 ●
  • . Subsidence analysis
  • Push-out analysis
  • . Particulate and gravimetric analysis per ASTM F1877 and ASTM F1714
  • . Bacterial endotoxin testing (BET) per ANSI/AAMI ST72:2011/(R)2016

The results demonstrate that the subject ATEC Lateral Interbody System is substantially equivalent to other predicate devices for nonclinical testing.

Clinical Information

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

VIII. CONCLUSION

Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regards to indications for use, intended use, design, technology, and performance.

N/A