K Number
K190649
Date Cleared
2019-05-07

(55 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EC-760S-V/L is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

Device Description

FUJIFILM Endoscope Model EC-760S-V/L is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the anqulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion contains glass fiber bundles, several channels and a complementary metal-oxide-semiconductor (CMOS) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

AI/ML Overview

The provided text is a 510(k) Summary for the FUJIFILM Endoscope Model EC-760S-V/L. It describes the device, its intended use, and its comparison to a predicate device to establish substantial equivalence.

Based on the information provided, here's an analysis of the acceptance criteria and study data:

1. A table of acceptance criteria and the reported device performance

The document explicitly states: "Fujifilm conducted the following performance testing on the proposed device EC-760S-V/L to ensure that the modified device performs equivalently to the predicate device:

  • Field of view
  • Bending capability
  • Rate of air supply
  • Rate of water supply
  • Suction rate
  • Working length
  • Forceps channel diameter
  • Viewing direction
  • Resolution
  • LG output

In all cases, the device met the pre-defined acceptance criteria for the test."

However, the document does not provide the specific acceptance criteria values for each of these performance metrics, nor does it provide the measured performance values for the proposed device or the predicate device. It only states that the device "met the pre-defined acceptance criteria."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Bench testing was conducted on the subject device to validate the Advanced Force Transmission, Adaptive Bending features" and lists other performance tests. These are bench tests and thus do not involve human subjects or clinical data in the traditional sense. The sample size for these engineering/performance tests (e.g., number of endoscopes tested) is not specified. Data provenance related to country of origin or retrospective/prospective is not applicable as it's not a clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a technical device performance study, not a perception or diagnostic accuracy study requiring expert human readers to establish ground truth. Therefore, this information is not applicable. The ground truth is based on engineering specifications and measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable as this is not a study involving human adjudication of results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a 510(k) for an endoscope itself, comparing its performance characteristics to a predicate endoscope. It does not involve AI or human reader performance comparisons.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this device is an endoscope, not an AI algorithm. Its performance is measured directly through physical and optical characteristics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for these performance tests (e.g., field of view, bending capability, resolution) are established based on engineering specifications, industry standards, and the performance characteristics of the predicate device. For instance, the "acceptance criteria" would be a defined range or value for a characteristic that the device must meet to be considered functional and equivalent to the predicate.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 7, 2019

FUJIFILM Corporation % Liza Dejesa-Jamanila Director, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

Re: K190649

Trade/Device Name: FUJIFILM Endoscope Model EC-760S-V/L Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: March 12, 2019 Received: March 13, 2019

Dear Liza Dejesa-Jamanila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190649

Device Name

FUJIFILM Endoscope Model EC-760S-V/L

Indications for Use (Describe)

EC-760S-V/L is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

FUJIFILM Endoscope Model EC-760S-V/L

Date: March 12, 2019

Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

Contact Person:

Liza Dejesa-Jamanila Director, Regulatory Affairs Telephone: (520) 414-6641 E-Mail: liza.dejesa-jamanila@fujifilm.com

Identification of the Proposed Device:

Proprietary/Trade Name:FUJIFILM Endoscope Model EC-760S-V/L
Common Name:Colonoscope
Device Class:Class II
Review Panel:Gastroenterology/Urology
Classification:Endoscope and accessories, 21 C.F.R. § 876.1500
Product Code:FDF

Predicate Device:

FUJIFILM Endoscope Model EC-760R-V/L (K172916)

Intended Use / Indications for Use:

EC-760S-V/L is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

Device Description:

FUJIFILM Endoscope Model EC-760S-V/L is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the anqulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion contains glass fiber bundles, several channels and a complementary metal-oxide-semiconductor (CMOS) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as

{4}------------------------------------------------

endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

Technological Characteristics:

The proposed device EC-760S-V/L differs from the predicate device in the following minor modifications:

  • . Dimensional and material changes to the insertion portion
  • . Introduction of the "Advanced Force Transmission" and "Adaptive Bending" features

Performance Data:

Electrical safety of the proposed device was evaluated using the following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2014, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

Biocompatibility of the proposed device was evaluated using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

The proposed device EC-760S-V/L was adopted into the reprocessing validation of the predicate device EC-760R-V/L using the following consensus standards: AAMI TIR12:2010, AAMI TIR30:2011. Validation of the cleaning, disinfection, and sterilization was performed in accordance with FDA's guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015.

The proposed device EC-760S-V/L was adopted into the photobiological safety testing of the predicate device EC-760R-V/L in accordance with IEC 62471:2006.

Bench testing was conducted on the subject device to validate the Advanced Force Transmission, Adaptive Bending features.

Fujifilm conducted the following performance testing on the proposed device EC-760S-V/L to ensure that the modified device performs equivalently to the predicate device:

  • . Field of view
  • Bending capability
  • . Rate of air supply
  • Rate of water supply
  • Suction rate
  • Working length

{5}------------------------------------------------

  • Forceps channel diameter
  • . Viewing direction
  • Resolution
  • . LG output

In all cases, the device met the pre-defined acceptance criteria for the test.

Substantial Equivalence:

The company's EC-760S-V/L has the same intended use and indications for use as the previously cleared predicate EC-760R-V/L (K172916). In addition, the proposed device has similar technological characteristics and principles of operation as the predicate. The minor differences between the proposed and predicate devices do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed device EC-760S-V/L is substantially equivalent to the predicate device.

Conclusions:

The modified EC-760S-V/L is substantially equivalent to the predicate EC-760R-V/L and conforms to applicable medical device safety and performance standards.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.