(529 days)
The "ORION" Spinal System is a non-cervical, pedicle screw system intended to provide posterior, non-cervical immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine : degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; scoliosis; kyphosis; spinal stenosis; tumor; pseudarthrosis and/or failed previous fusion.
In addition, the "ORION" Spinal System is also indicated for the treatment of the skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, having fusion by autogenous bone graft, with implant fixed or attached to the lumbar and sacral spine (L3 to Sacrum), and for whom the intended to be removed after solid fusion is achieved.
The "ORION" Spinal System is a system that is intended to be used for posterior thoracolumbar fusion procedures. The system is manufactured from Ti-6Al-4V which complies with ASTM F136 and PEEK which complies with ASTM F2026. The components, which are included as part of the system, include screws, rods, links, and accessory connection components.
This is a 510(k) premarket notification for the "ORION" Spinal System. This document does not describe the acceptance criteria and study for an AI/ML device. It's for a traditional medical device (a pedicle screw system). Therefore, the requested information regarding AI/ML device testing criteria, such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, is not applicable and not present in this document.
The document discusses the substantial equivalence of the "ORION" Spinal System to a predicate device based on its indications for use, technological characteristics, and performance data from engineering tests.
Here's what information is provided:
1. A table of acceptance criteria and the reported device performance:
This document does not provide a table comparing acceptance criteria and reported device performance in the way typically expected for AI/ML performance metrics (e.g., sensitivity, specificity, AUC). Instead, it lists the types of engineering tests performed on the "ORION" Spinal System:
| Test Mode | Standard |
|---|---|
| Axial Grip | ASTM F1798 |
| Torsional Grip | ASTM F1798 |
| Static Compression Bending | ASTM F1717 |
| Static Torsion | ASTM F1717 |
| Dynamic Compression Bending | ASTM F1717 |
The document states that the "subect device is substantially equivalent to the predicate device," implying that the performance in these tests met the established criteria for substantial equivalence to the predicate device and relevant ASTM standards. However, the specific numerical results or acceptance thresholds from these tests are not detailed in this summary.
2. Sample sized used for the test set and the data provenance: Not applicable. This is an engineering test of a physical device, not an AI/ML algorithm evaluated on a data set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is an engineering test of a physical device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This is an engineering test of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an engineering test of a physical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is an engineering test of a physical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to an AI context. For this device, the "ground truth" implicitly refers to the device's ability to withstand mechanical forces and maintain structural integrity as per relevant ASTM standards, demonstrating substantial equivalence to the predicate device.
8. The sample size for the training set: Not applicable. This is an engineering test of a physical device, not an AI/ML algorithm.
9. How the ground truth for the training set was established: Not applicable. This is an engineering test of a physical device, not an AI/ML algorithm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 18, 2020
Orion Biotech Inc. Tun-Ling Li Chen Official Correspondent 5F-1, No.88, Sec.2, Chong-qing N. Rd., Datong Dist. Taipei, 103 Taiwan
Re: K190600
Trade/Device Name: "ORION" Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: July 20, 2020 Received: July 20, 2020
Dear Tun-Ling Li Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190600
Device Name "ORION" Spinal System
Indications for Use (Describe)
The "ORION" Spinal System is a non-cervical, pedicle screw system intended to provide posterior, non-cervical immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine : degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; scoliosis; kyphosis; spinal stenosis; tumor; pseudarthrosis and/or failed previous fusion.
In addition, the "ORION" Spinal System is also indicated for the treatment of the skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, having fusion by autogenous bone graft, with implant fixed or attached to the lumbar and sacral spine (L3 to Sacrum), and for whom the intended to be removed after solid fusion is achieved.
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
age icy ay information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 er
{3}------------------------------------------------
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R. §807.92.
| Preparation Date: | July 20, 2020 |
|---|---|
| Applicant/Sponsor: | Orion Biotech Inc.5F-1, No.88, Sec.2, Chong-qing N. Rd., Datong DistTaipei 103, Taiwan |
| Contact Person: | Tung-Ling Li ChenTEL: +886 2 2550 3366 ext. 101Fax : +886 2 2559 3366Email: tunglingli@orionspine.com |
| Proprietary Name: | "ORION" Spinal System |
| Common Name: | Pedicle Screw System |
| Classification Name: | 21 CFR 888.3070Thoracolumbosacral Pedicle Screw System |
Classification Identification: Class II
NKB Product Codes:
Predicate Device: Facilis Spinal System K161231
Device Description:
The "ORION" Spinal System is a system that is intended to be used for posterior thoracolumbar fusion procedures. The system is manufactured from Ti-6Al-4V which complies with ASTM F136 and PEEK which complies with ASTM F2026. The components, which are included as part of the system, include screws, rods, links, and accessory connection components.
{4}------------------------------------------------
Indications for Use:
- · The "ORION" Spinal System is a non-cervical, pedicle screw system intended to provide posterior, non-cervical immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine : degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; dislocation; scoliosis; spinal stenosis; tumor; pseudarthrosis and/or failed previous fusion.
- · In addition, the "ORION" Spinal System is also indicated for the treatment of the skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, having fusion by autogenous bone graft, with implant fixed or attached to the lumbar and sacral spine (L3 to Sacrum), and for whom the implants is intended to be removed after solid fusion is achieved.
Technological Characteristics
The subject and predicate devices have similar technological characteristics and the minor differences do not raise any different issues of safety and effectiveness. The following characteristics are the same between the subject and predicates: Indications for use, Principles of operation, Materials of manufacture, Implant sizes, and Surgical technique.
Performance Data
The "ORION" Spinal System has been testing in the following test modes:
- Axial Grip per ASTM F1798
- Torsional Grip per ASTM F1798
- Static Compression Bending per ASTM F1717
- Static Torsion per ASTM F1717
- Dynamic Compression Bending per ASTM F1717
Conclusion
The information provided in this submission demonstrate that the subject device is substantially equivalent to the predicate device.
N/A