(89 days)
Not Found
No
The document describes a standard blood glucose monitoring system and does not mention any AI or ML capabilities in its intended use, device description, or performance studies.
No.
The device is a diagnostic tool used to measure glucose levels, aiding in monitoring the effectiveness of diabetes control. It does not provide treatment or directly manage the disease.
Yes
The device is intended for in-vitro diagnostic use to quantitatively measure glucose in blood samples, which is a diagnostic purpose.
No
The device description explicitly lists hardware components such as a Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device, and Sterile lancets. This indicates it is a system with physical components, not solely software.
Yes, this device is an IVD (In Vitro Diagnostic).
The Intended Use statement explicitly states: "The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use)..." and "The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use)...".
This clearly indicates that the device is designed to be used outside of the body to diagnose or monitor a condition, which is the definition of an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.
The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
Product codes (comma separated list FDA assigned to the subject device)
NBW
Device Description
Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB, consist of the following devices:
Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device and Sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.
Rightest Meter GM700S and Rightest Meter GM700SB (with Bluetooth), when used with the Rightest Test Strips GM700S, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB (with Bluetooth) are verified by the Rightest Control Solution GC700.
The Rightest Blood Glucose Monitoring System GM700S, GMS700SB are electronic device that uses electrochemical methodologies. The system quantitatively measures blood glucose levels using an amperometric method. The system employs flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
fingertips, forearm or palm
Indicated Patient Age Range
Not Found
Intended User / Care Setting
individuals with diabetes at home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
User Performance Study: A User performance study was performed to demonstrate that lay users could obtain accurate results using the subject device. The study was performed using capillary whole blood from fingertip, palm and forearm sample sites. A total of 354 patients participated. The study result shows substantial equivalence to predicate device used in finger, palm and forearm position.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
Image /page/0/Picture/1 description: The image shows the words "Convert to PDF" in a simple, sans-serif font. The text is horizontally aligned and appears to be a title or heading. The background is a plain, light color, providing contrast for the dark text.
Image /page/0/Picture/2 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 2, 2019
Bionime Corporation c/o Feng-Yu Lee Dynamic Biotech Inc. dba IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675
Re: K190564
Trade/Device Name: Rightest Blood Glucose Monitoring System GM700S Rightest Blood Glucose Monitoring System GM700SB Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: February 22, 2019 Received: March 5, 2019
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K190564
Device Name
Rightest Blood Glucose Monitoring System GM700S
Indications for Use (Describe)
The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700.
The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.
The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Indications for Use
510(k) Number (if known) K190564
Device Name
Rightest Blood Glucose Monitoring System GM700SB
Indications for Use (Describe)
The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter GM700SB and the Rightest Blood Glucose Test Strips GS700.
The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.
The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Bionime Corporation
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K190564
-
- Submitter's Identification:
BIONIME CORPORRATION NO 100, Sec. 2, Daging St., South Dist., 40242 Taichung City, Taiwan Phone Number: 886-4-23692388 FAX Number: 886-4-22617568
- Submitter's Identification:
c/o IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-949-218-0928
Date Summary Prepared: May 2nd, 2019
-
- Name of the Device: Rightest Blood Glucose Monitoring System GM700S Rightest Blood Glucose Monitoring System GM700SB
-
- Common or Usual Name: Blood Glucose Monitoring System
Product Code | Classification | Regulation Section | Panel |
---|---|---|---|
NBW; System, Test, Blood | |||
Glucose, Over-the-Counter | Class II | 21 CFR 862.1345 | Clinical Chemistry |
75 |
4. Device Description:
Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB, consist of the following devices:
Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device and Sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.
Rightest Meter GM700S and Rightest Meter GM700SB (with Bluetooth), when used with the Rightest Test Strips GM700S, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System
5
GM700SB (with Bluetooth) are verified by the Rightest Control Solution GC700.
Intended Use: న.
Rightest Blood Glucose Monitoring System 700S (I)
The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.
The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
(II) Rightest Blood Glucose Monitoring System 700SB
The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter 700SB and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
6. Predicate Device Information:
The subject devices are substantially equivalent to the brand of Rightest Blood Glucose Monitoring System noted below.
Name: | Rightest Blood Glucose Monitoring System GM720 |
---|---|
Device Company: | Bionime Corporation |
510(K) Number: | K140210 |
-
- Comparison to Predicate Devices:
6
Bionime Corporation
Subject Devices | Predicate Device | ||
---|---|---|---|
Specification | Rightest GM700S | Rightest GM700SB | |
(GM700S w/ Bluetooth) | Rightest GM720 | ||
Indications for Use | Same | Quantitative measurement of glucose | |
in fresh capillary whole blood samples | |||
drawn from the fingertip, palm and | |||
forearm as an aid to monitor the | |||
effectiveness of diabetes control | |||
Minimum Sample | |||
Volume | 0.75 microliter | ||
Test Time | 5 seconds | ||
Hematocrit | 20 - 60 % | 20 - 65 % | |
Memory Capacity | 500 blood glucose test results with date and time | ||
Power Saving | Turn off automatically 2 minutes after last user action / Press the main button for 3 seconds. | ||
Coding | Auto coding | ||
Measurement | |||
Technology | Dehydrogenase Electrochemical Sensor | ||
Sample | Fresh capillary whole blood from fingertips, palm, or forearm. | ||
Measuring Range | 20 - 600 mg/dL | ||
Operating | |||
Temperature | 10 ~ 40°C (50 ~104 °F) | ||
Operating Relative | |||
Humidity | 10~90% | ||
Monitor | LCD display | ||
Meter Storage | |||
Conditions | -10 ~ 60°C (14~140 °F) | ||
Test Strip Storage | |||
Conditions | 4 ~ 30℃ (39~ 86 °F), 10 - 90% relative humidity | ||
Unit of Measurement | |||
Data | Fixed on mg/dL | ||
Backlight | No | Yes | |
Strip Shelf Life - | |||
Open Vial | 4 months | ||
Control Solution | Rightest Control Solution GC700 | ||
Interference | Ascorbic acid ≥ 3 mg/dL | ||
Uric acid ≥ 10 mg/dL | |||
Xylose ≥ 20 mg/dL | Ascorbic acid (≥ 3 mg/dL), | ||
Dopamine (≥1.25 mg/dL), | |||
L-Dopa (≥2 mg/dL), | |||
Xylose (≥10 mg/dL) | |||
Uric Acid (≥16 mg/dL) | |||
Reagent | 1. FAD-Glucose dehydrogenase 12.4 % |
- Potassium Ferricyanide 49.6 %
- Non-reactive Ingredients 38.0 % | 1. FAD-Glucose dehydrogenase 12.1%
- Potassium ferricyanide 48.5%
- Non-reactive ingredients 39.4% | |
| Wireless module | No | No | |
| Power Supply | One CR2032 battery | Two CR2032 batteries | |
| Battery Life | 1000 tests | About 600 tests | |
| Meter Dimension | 82mm45mm15.5mm | 71.5 mm *39.5 mm 14.0 mm | |
| LCD display area | 34 mm27.5mm | 25.02 mm *32.7 mm | |
| Meter Weight | $43.0\pm 5$ g with batteries | $50 \pm 5$ g with batteries | |
| Color | White/Gray | Black | |
7
Bionime Corporation
8. Technology Characteristics:
The Rightest Blood Glucose Monitoring System GM700S, GMS700SB are electronic device that uses electrochemical methodologies. The system quantitatively measures blood glucose levels using an amperometric method. The system employs flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.
9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
Verification and validation of test results were evaluated to establish the performance, functionality and reliability of the subject devices. The evaluation included precision, linearity, interference, sample volume, hematocrit, and operated environment.
10. Discussion of Clinical Tests Performed:
User Performance Study:
A User performance study was performed to demonstrate that lay users could obtain accurate results using the subject device. The study was performed using capillary whole blood from fingertip, palm and forearm sample sites. A total of 354 patients participated. The study result shows substantial equivalence to predicate device used in finger, palm and forearm position.
-
- Conclusions:
Results of performance evaluation of the Rightest Blood Glucose Monitoring System subject devices demonstrate that the candidate devises are substantial equivalence to the predicate device, Rightest Blood Glucose Monitoring System GM720.
- Conclusions: