K Number
K190464
Manufacturer
Date Cleared
2019-04-26

(59 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Penumbra Reperfusion Catheters and Separators: As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device: As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing: As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The Penumbra System 3D Revascularization Device is designed to revascularize patients experiencing acute ischemic stroke as a component of the Penumbra System. The Penumbra System's fundamental mechanism of action is aspiration removes clot by drawing it into the Penumbra Reperfusion Catheter. The 3D Revascularization Device is designed to assist the Reperfusion Catheter with thrombus removal if needed. Intended users for this device are physicians who have received appropriate training in interventional neuro-endovascular techniques and treatment of acute ischemic stroke. The device is provided sterile, non-pyrogenic, and intended for single use only.

AI/ML Overview

The provided text describes the Penumbra System (3D Revascularization Device) and its substantial equivalence to a predicate device (Penumbra 3D Revascularization Device, K162901). The submission relies on non-clinical data for verification and validation, explicitly stating that "No clinical study was conducted."

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

AttributeAcceptance CriteriaReported Device Performance
Pouch Seal StrengthMinimum value per specificationPass
Dimensional / Visual InspectionUnits meet all product specifications.Pass
Simulated Use (Intracranial Access, Vessel Access Entry Performance & Clot Removal)Devices effectively remove clots in an anatomical model simulating neurovasculature tortuosity.Pass
Physician Evaluation (Deliverability & Clot Removal)Multiple Physician performance evaluation of the subject Penumbra System 3D Revascularization Device in a simulated neurovascular tortuosity model with the Penumbra System 3D Revascularization Device used as a baseline.Pass
Particulate Testing ($\ge$ 10 μm)$\le$ 6000 particlesPass
Particulate Testing ($\ge$ 25 μm)$\le$ 600 particlesPass
Particulate Testing ($\ge$ 75 μm)Measured for informational purposes only (FIFO)FIFO
Particulate Testing ($\ge$ 125 μm)Measured for informational purposes only (FIFO)FIFO
Radial PressureValue within the specification rangePass
TorsionMinimum value per specificationPass
Kink ResistanceNo kinking when formed in a defined radiusPass
Device AFMaximum value per specificationPass
CorrosionSpecimens will be free from signs of corrosionPass
Tensile Strength - MarkerbandMinimum value per specificationPass
Tensile Strength - Wire AttachmentMinimum value per specificationPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a distinct "test set" sample size in the clinical sense, as no clinical study was conducted for this specific submission. The evaluations were primarily bench-top testing and previously completed animal studies. The document does not specify sample sizes for individual bench tests beyond stating "units used in this Design Verification testing." The provenance of the data for bench testing is not explicitly stated beyond being conducted by Penumbra, Inc.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

For the "Physician Evaluation" in bench-top testing, the document mentions "Multiple Physician performance evaluation." However, it does not specify the exact number of physicians or their specific qualifications (e.g., years of experience, specialty).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe an adjudication method for the physician evaluations; it merely states that the evaluations were part of the testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was performed, as stated: "No clinical study was conducted." The device does not appear to be an AI-assisted diagnostic tool, but rather a mechanical revascularization device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable to the device described. The Penumbra System (3D Revascularization Device) is a physical medical device for revascularization, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the bench-top testing, the "ground truth" was established by predetermined acceptance criteria and product specifications for physical and mechanical properties. For the simulated use and physician evaluations, the ground truth was based on the effectiveness of the devices to remove clots in anatomical models and physician perception of deliverability and clot removal compared to a baseline.

For the animal study portion, the ground truth for vessel response was derived from existing GLP test results from rabbit vessel response studies on the predicate device.

8. The sample size for the training set

No "training set" in the context of an algorithm or AI is mentioned in the document, as no clinical study was conducted for this submission, and the device is a mechanical one. The previous animal studies used for the predicate device were not described as a "training set" in this context.

9. How the ground truth for the training set was established

Not applicable, as no training set (in the AI/algorithm sense) was used or described. The document relies on equivalence to a predicate device and extensive non-clinical bench testing and prior animal studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.

April 30, 2019

Penumbra, Inc. Michaela Mahl Senior Manager of Regulatory Affairs One Penumbra Place Alameda, California 94502

Re: K190464

Trade/Device Name: Penumbra System (3D Revascularization Device) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: February 25, 2019 Received: February 26, 2019

Dear Michaela Mahl:

This letter corrects our substantially equivalent letter of April 26, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190464

Device Name

Penumbra System (3D Revascularization Device)

Indications for Use (Describe)

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra System® 3D Revascularization Device.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Michaela Mahl Senior Manager, Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: mmahl@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

February 22, 2019

1.4 Device Trade or Proprietary Name

Penumbra System® (3D Revascularization Device)

1.5 Device Classification

Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR §870.1250 Product Code: NRY (Catheter, Thrombus Retriever)

1.6 Predicate and Reference Devices

510(k) Number/Clearance DateName of DeviceName of Manufacturer
Predicate Device
K162901 cleared on April 20, 2017Penumbra 3DRevascularization DevicePenumbra, Inc.One Penumbra PlaceAlameda, CA 94502 USA

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle.

Predicate Comparison 1.7

System NamePenumbra System®
Predicate DeviceSubject Device
Device NamePenumbra 3D Revascularization DevicePenumbra 3DRevascularizationDevice (Modified)
510(k) No.K162901K190464
ClassificationClass II, NRYSAME
Indication for UsePenumbra Reperfusion Catheters and SeparatorsAs part of the Penumbra System, the Reperfusion Catheters and Separators areindicated for use in the revascularization of patients with acute ischemic strokesecondary to intracranial large vessel occlusive disease (within the internalcarotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries)within 8 hours of symptom onset. Patients who are ineligible for intravenoustissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy arecandidates for treatment.Penumbra 3D Revascularization DeviceAs part of the Penumbra System, the Penumbra 3D Revascularization Deviceis indicated for use in the revascularization of patients with acute ischemicstroke secondary to intracranial large vessel occlusive disease (within theinternal carotid, middle cerebral – M1 and M2 segments) within 8 hours ofsymptom onset. Patients who are ineligible for intravenous tissue plasminogenactivator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.Penumbra Aspiration TubingAs part of the Penumbra System, the Penumbra Sterile Aspiration Tubing isindicated to connect the Penumbra Reperfusion Catheters to the PenumbraAspiration Pump.Penumbra Aspiration PumpThe Penumbra Aspiration Pump is indicated as a vacuum source forPenumbra Aspiration Systems.SAME
Reperfusion Catheter and SeparatorsSAME
System Components3D Revascularization DeviceSAME
Aspiration TubingSAME
Aspiration Pump & Pump CanisterSAME
MaterialsPenumbra 3D Revascularization DevicePenumbra 3DRevascularizationDevice (Modified)
Delivery-Wire
Core wireStainless SteelSAME
CoatingPTFE Green PC 8-403GSAME
Outer coilStainless SteelSAME
Proximal SolderSilver SolderSAME
System NamePenumbra System®
Distal Inner CoilN/APlatinum Alloy
Distal SolderSilver SolderSAME
Distal Tip
Outer coilStainless SteelSAME
Radiopaque MarkersPlatinum AlloySAME
Attachment TipStainless SteelSAME
Solder jointGold Solder / Silver SolderSilver Solder
3D Nitinol Device
3D latticeNitinolSAME
Introducer sheathFEPSAME
Dimensions
- Device NominalTotal Length26 mmSAME
- Device Nominal OD4.5 mmSAME
- Device NominalUsable Length20 mmSAME
- Delivery Wire DistalOD0.014"SAME
- Delivery WireLength200 cmSAME
SterilizationEOSAME
Shelf-Life36 MonthsSAME

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle.

Device Description 1.8

The Penumbra System 3D Revascularization Device is designed to revascularize patients experiencing acute ischemic stroke as a component of the Penumbra System. The Penumbra System's fundamental mechanism of action is aspiration removes clot by drawing it into the Penumbra Reperfusion Catheter. The 3D Revascularization Device is designed to assist the Reperfusion Catheter with thrombus removal if needed. Intended users for this device are physicians who have received appropriate training in interventional neuro-endovascular techniques and treatment of acute ischemic stroke. The device is provided sterile, non-pyrogenic, and intended for single use only.

1.9 Indications for Use

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending to the edge of the circle.

vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.10 Summary of Non-Clinical Data

As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.

Included in this section are descriptions of the testing, which substantiates the safe and effective performance of the subject Penumbra System 3D Revascularization Device as well as its substantial equivalence to the predicate device:

  • Biocompatibility
  • Design Verification (Bench-Top Testing) ●
  • Design Validation (Animal Study) ●
  • . Performance Data

The subject, Penumbra System 3D Revascularization Device met all established requirements.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a bold, red font. To the right of the name is a circular logo, also in red, with a stylized white "P" inside.

1.10.1 Previously Completed Biocompatibility Testing

The materials used in the subject Penumbra System 3D Revascularization Device [modified] are identical to those in the predicate device, Penumbra System 3D Revascularization Device [K162901]. The subject Penumbra System 3D Revascularization Device [modified] uses identical materials, equivalent processing, and identical sterilization methods as products that Penumbra has already successfully conducted biocompatibility testing per BS EN ISO 10993. The studies were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for a limited exposure (< 24 hours), external communicating device with circulating blood contact. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP).

The classification for the predicate, Penumbra System 3D Revascularization Device [K162901] and subject device, Penumbra System 3D Revascularization Device [modified] are identical. All biocompatibility risks have been mitigated through successful completion of prior biocompatibility testing per ISO 10993-1:2009/AC: 2010 which will be used to support the biocompatibility for the subject Penumbra System 3D Revascularization Device [modified]. Therefore, no further biocompatibility testing was conducted on the subject device.

1.10.2 Design Verification - Bench-top Testing

The physical and mechanical properties of the 3D Revascularization Device was assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:

AttributeSpecificationResult
Pouch Seal StrengthMinimum value per specificationPass
Dimensional / VisualInspectionThese evaluations confirm that the units used in this Design Verificationtesting meet all product specifications.Pass
Simulated Use[Intracranial Access,Vessel Access EntryPerformance & ClotRemoval]Simulated use testing of the Reperfusion Catheter and 3D RevascularizationDevice was performed with accessory devices in an anatomical modelwhich simulated the tortuosity of the neurovasculature. Devices weredelivered through the tortuous anatomical model to evaluate theeffectiveness of the devices to remove clots.Pass
Physician Evaluation[Deliverability & ClotRemoval]Multiple Physician performance evaluation of the subject Penumbra System3D Revascularization Device in a simulated neurovascular tortuosity modelwith the Penumbra System 3D Revascularization Device used as a baseline.Pass
Particulate TestingThe maximum number of particles:Pass

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the edge of the circle. The logo is simple and modern.

AttributeSpecificationResult
$\ge$ 10 μm will be $\le$ 6000 particles
$\ge$ 25 μm will be $\le$ 600 particlesPass
$\ge$ 75 μm will be measured for informational purposes only(FIFO)FIFO
$\ge$ 125 μm will be measured for informational purposes only(FIFO)FIFO
Radial PressureValue within the specification rangePass
TorsionMinimum value per specificationPass
Kink ResistanceNo kinking when formed in a defined radius
Device AfMaximum value per specificationPass
CorrosionSpecimens will be free from signs of corrosion
Tensile Strength -MarkerbandMinimum value per specificationPass
Tensile Strength -WireAttachmentMinimum value per specificationPass

1.10.3 Design Validation - Previously Completed Animal Study

The subject, 3D Revascularization Device [modified] uses the same materials, has the same indication for use, and has the same intended use as the predicate, 3D Revascularization Device. In addition, the subject, 3D Revascularization Device [modified] will be held to the same maximum radial pressure specification as the predicate, 3D Revascularization Device [K162901]. Therefore, the subject, 3D Revascularization Device [modified] presents an equivalent challenge for vessel response compared to the predicate device. Rabbit vessel response studies have been performed for the predicate, 3D Revascularization Device [K162901] with successful results. Based on the existing GLP test results, the interaction of the predicate, 3D Revascularization Device [K162901] with the rabbit vasculature is well understood. Therefore, the existing rabbit vessel response animal study results mitigate all vessel response risks for the subject, 3D Revascularization Device [modified].

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Penumbra, a medical device company. The word "Penumbra" is written in a bold, red font. To the right of the word is a red circle with a white "P" inside of it.

1.10.4 Performance Data - Clinical

No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the predicate device and other devices with similar dimensions used for clot removal. The literature review was used to support the determination of substantial equivalence by using clinical outcomes from devices that are considered technologically equivalent.

1.11 Summary of Substantial Equivalence

The subject, Penumbra System 3D Revascularization Device is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).