(59 days)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
2. UV Irradiation:
- During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
-
- 0.1 micron filtration:
- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical heat-sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The device was originally cleared under K090036. Seven Special 510(k)s were cleared that subsequently made minor modifications to hardware, software, specifications, labeling, sterilant and accessories. A Traditional 510(k) was cleared in 2017 adding a new Ultrasound Processing Tray C3000XL and modifying the Indications for Use. The current submission is provided to describe the addition of a second supplier for a proprietary component of S40 Sterilant Concentrate.
The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical heat-sensitive medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe the addition of a second supplier for of a proprietary component of S40 Sterilant Concentrate.
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. After LCS processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels.
The provided document is a 510(k) Pre-Market Notification for the SYSTEM 1E Liquid Chemical Sterilant Processing System (K190104) and the SYSTEM 1 endo Liquid Chemical Sterilant Processing System (K190104). This submission describes modifications to the S40 Sterilant Concentrate used with these systems, specifically the addition of a second supplier for a proprietary component of the concentrate.
The document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices (K180342 for SYSTEM 1E and K173256 for SYSTEM 1 endo). As such, the studies described are verification tests for the modified component of the sterilant rather than a full stand-alone performance study of the entire device.
Here's the breakdown of the acceptance criteria and the studies that prove the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For S40 Sterilant Concentrate with alternate source of a proprietary component:
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Biocompatibility of S40 Sterilant Concentrate with alternate source of a proprietary component | Medical devices shall be non-cytotoxic after exposure to S40 Sterilant Concentrate containing the alternate source of a proprietary component | Pass |
| Stability of S40 Sterilant Concentrate with alternate source of a proprietary component | S40 Sterilant Concentrate shall meet acceptance criteria after 6-months of storage. | Pass |
Study Details:
The document clearly states that "New testing was performed to evaluate the modified device," specifically referring to the change in the supplier for a component of the S40 Sterilant Concentrate. The provided tables are comparison tables with predicate devices, indicating that the overall system performance (effectiveness as a sterilant) for the proposed device is considered "identical" to the predicate devices which had already demonstrated efficacy through extensive testing. The new testing is therefore focused on verifying that the change in sterilant component supplier does not negatively impact the established performance.
For the Biocompatibility and Stability Tests (Table 3 on pages 16 and 28):
-
Sample size used for the test set and the data provenance:
- The document does not specify the exact sample size for the biocompatibility or stability tests regarding the alternate component.
- Data Provenance: The studies were "new testing" performed to evaluate the modified device, implying prospective testing for this specific change. The country of origin for the data is not specified, but the submission is to the US FDA, so it's likely part of US-based regulatory activities.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as these are laboratory-based physical/chemical/biological tests rather than expert interpretation of medical images or clinical data.
-
Adjudication method for the test set:
- Not applicable; these are objective laboratory tests with pre-defined pass/fail criteria.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a liquid chemical sterilizer, not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a medical device for sterilization, not an algorithm. The "standalone" performance is implied by the "Pass" results of the biocompatibility and stability tests of the sterilant itself. The overall sterilization efficacy of the SYSTEM 1E and SYSTEM 1 endo Processors using S40 Sterilant is established through prior predicate device clearances and the underlying efficacy claims of the sterilant (e.g., sporicidal, fungicidal, bactericidal, virucidal, tuberculocidal activity, and simulated-use tests, summarised in the comparison tables as "Meets efficacy requirements").
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the new tests: The ground truth is laboratory-determined objective measurements against pre-defined regulatory and performance standards for biocompatibility (non-cytotoxicity) and chemical stability.
- For the overall device's sterilization efficacy (as implied by the identical comparison with predicate devices): This would have been established through a combination of in vitro microbiological assays (e.g., AOAC methods for sporicidal, fungicidal, bactericidal activity), in vitro virucidal testing (EPA method), simulated-use tests, and clinical in-use studies. The references to specific methods (e.g., AOAC Official Method 966.04) indicate standardized laboratory protocols are the ground truth.
-
The sample size for the training set:
- Not applicable in the context of this 510(k) submission, as it's not an AI/machine learning device.
-
How the ground truth for the training set was established:
- Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
March 22, 2019
Steris Corporations Jennifer Nalepka Senior Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K190104
Trade/Device Name: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 Endo Liquid Chemical Sterilant Processing System Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants/High Level Disinfectants Regulatory Class: Class II Product Code: MED Dated: March 5, 2019 Received: March 6, 2019
Dear Jennifer Nalepka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190104
Device Name
SYSTEM 1E Liquid Chemical Sterilant Processing System
Indications for Use (Describe)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
2. UV Irradiation:
- During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
-
- 0.1 micron filtration:
- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
IX Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K190104
Device Name
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
Indications for Use (Describe)
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical heat-sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is written in a serif font in black. Below the word is a graphic of several horizontal wavy lines in blue. The registered trademark symbol is in the upper right corner of the logo.
510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060
Contact: Jennifer Nalepka Senior Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
Summary Date: March 5, 2019
Premarket Notification Number: K190104
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
1. Device Name
| Trade Name: | SYSTEM 1E Liquid Chemical SterilantProcessing System |
|---|---|
| Device Classification: | Class II |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid ChemicalSterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
Predicate Device 2.
SYSTEM 1E Liquid Chemical Sterilant Processing System, K180342
3. Description of Device
The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The device was originally cleared under K090036. Seven Special 510(k)s were cleared that subsequently made minor modifications to hardware, software, specifications, labeling, sterilant and accessories. A Traditional 510(k) was cleared in 2017 adding a new Ultrasound Processing Tray C3000XL and modifying the Indications for Use. The current submission is provided to describe the addition of a second supplier for a proprietary component of S40 Sterilant Concentrate.
The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
{6}------------------------------------------------
S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.
Indications for Use 4.
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semicritical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- 米 The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- Pre-filter A is a gross depth filter that removes approximately 2.5 . micron or larger particles/contaminants.
- Pre-filter B is a surface filter that removes particles/contaminants > . 0.1 micron.
-
- UV Irradiation:
- During transit through the UV water treatment chamber, a UV dose . sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
-
- 0.1 micron filtration:
- The water prepared by pre-filtration and UV irradiation is filtered . through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
-
{7}------------------------------------------------
5. Technological Characteristic Comparison Table
The SYSTEM 1E Liquid Chemical Sterilant Processing System is the same as the predicate device; the specific modification described in this submission is for an alternate supplier for a component of S40 Sterilant Concentrate. The differences between the proposed and predicate devices are limited to the addition of an alternate supplier for a proprietary component of S40 Sterilant Concentrate.
There were no changes to the SYSTEM 1E Processor.
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Indicationsfor Use | The SYSTEM 1E LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable critical and semi-critical heat-sensitive medicaldevices in healthcarefacilities.The SYSTEM 1E Processordilutes the S40 SterilantConcentrate to its use dilution(>1820 mg/L peracetic acid),liquid chemically sterilizesthe load during a controlled6-minute exposure at 45.5 to60°C, and rinses the load withextensively treated* potablewater. After completion of acycle, critical devices shouldbe used immediately; semi-critical devices should beused immediately or may behandled and stored in amanner similar to that of highlevel disinfected endoscopes.Critical devices not usedimmediately should beprocessed again before use.The SYSTEM 1E Processoruses only S40 Sterilant | The SYSTEM 1E LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable critical and semi-critical heat-sensitive medicaldevices in healthcarefacilities.The SYSTEM 1E Processordilutes the S40 SterilantConcentrate to its use dilution(>1820 mg/L peracetic acid),liquid chemically sterilizesthe load during a controlled6-minute exposure at 45.5 to60°C, and rinses the load withextensively treated* potablewater. After completion of acycle, critical devices shouldbe used immediately; semi-critical devices should beused immediately or may behandled and stored in amanner similar to that of highlevel disinfected endoscopes.Critical devices not usedimmediately should beprocessed again before use.The SYSTEM 1E Processoruses only S40 Sterilant | Identical | |||||||
| Concentrate to liquidchemically sterilize medical | Concentrate to liquidchemically sterilize medical | |||||||||
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
| devices. | devices. | |||||||||
| * The extensive treatment ofEPA potable water consistsof:1. Pre-filtration throughtwo pre-filters:• Pre-filter A is a grossdepth filter thatremovesapproximately 2.5micron or largerparticles/contaminants.• Pre-filter B is a surfacefilter that removesparticles/contaminants> 0.1 micron.2. UV Irradiation:• During transit throughthe UV watertreatment chamber, aUV dose sufficient toachieve a ≥ 6-logreduction of MS2 virus | * The extensive treatment ofEPA potable water consistsof:1. Pre-filtration throughtwo pre-filters:• Pre-filter A is a grossdepth filter thatremovesapproximately 2.5micron or largerparticles/contaminants.• Pre-filter B is a surfacefilter that removesparticles/contaminants> 0.1 micron.2. UV Irradiation:• During transit throughthe UV watertreatment chamber, aUV dose sufficient toachieve a ≥ 6-logreduction of MS2 virus | |||||||||
| is delivered to the | is delivered to the | |||||||||
| water. | water. | |||||||||
| 3.0.1-micron | 3.0.1-micron | |||||||||
| filtration: | filtration: | |||||||||
| • The water prepared by | • The water prepared by | |||||||||
| pre-filtration and UVirradiation is filtered | pre-filtration and UVirradiation is filtered | |||||||||
| through redundant, | through redundant, | |||||||||
| 0.1-micron (absolute | 0.1-micron (absolute | |||||||||
| rated) membranes to | rated) membranes to | |||||||||
| remove bacteria, fungi | remove bacteria, fungi | |||||||||
| and protozoa > 0.1 | and protozoa > 0.1 | |||||||||
| micron. | micron. | |||||||||
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
| OperatingPrinciples/Technology | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processingchamber in which the load isplaced. Devices with internallumens are interfaced with theprocessor using connectors.Sterilant Concentrate isplaced in a specializedcompartment and when theprocessor fills with water, itcreates the sterilant usedilution from the single usesterilant cup. The processormonitors and controls the usedilution temperature andcontact time. The processorautomatically rinses the loadwith extensively treated waterto remove sterilant residuals. | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processingchamber in which the load isplaced. Devices with internallumens are interfaced with theprocessor using connectors.Sterilant Concentrate isplaced in a specializedcompartment and when theprocessor fills with water, itcreates the sterilant usedilution from the single usesterilant cup. The processormonitors and controls the usedilution temperature andcontact time. The processorautomatically rinses the loadwith extensively treated waterto remove sterilant residuals. | Identical | |||||||
| ProcessParameters | Standardized cycleparameters cannot be alteredby operator. The criticalprocess parameters are:• Contact Time• Use Dilution• Temperature• Peracetic acidconcentration• Bacterial retentive waterfilter integrity• UV irradiation | Standardized cycleparameters cannot be alteredby operator. The criticalprocess parameters are:• Contact Time• Use Dilution• Temperature• Peracetic acidconcentration• Bacterial retentive waterfilter integrity• UV irradiation | Identical | |||||||
| ProcessMonitors | • Cycle Printoutdocuments successfulcycle completion oridentifies fault if cycleaborts• Alarms if thermocouplesindicate temperature outof specification• Alarms if pressureswitch indicates thathigh pressure pump isnot operating | • Cycle Printoutdocuments successfulcycle completion oridentifies fault if cycleaborts• Alarms if thermocouplesindicate temperature outof specification• Alarms if pressureswitch indicates thathigh pressure pump isnot operating | Identical | |||||||
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
| Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | |||||||||
| DesignFeatures | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered, UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter. | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered. UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter. | Identical | |||||||
| Processing Cycle | ||||||||||
| Incomingwatertemperature | $\geq 43°C$ | $\geq 43°C$ | Identical | |||||||
| Temperatureto startexposure | $\geq 46°C$ | $\geq 46°C$ | Identical | |||||||
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
| phase | ||||||||||
| Temperaturealarm pointduring theexposurephase | <45.5 or >60°C | <45.5 or >60°C | Identical | |||||||
| Temperaturerange during atypical LiquidChemicalSterilantProcessingCycle | 46 - 55°C | 46 - 55°C | Identical | |||||||
| ExposureTime | 6 minutes | 6 minutes | Identical | |||||||
| Rinse waterpreparation | Hot potable tap water is:pre-filteredflowed through a UVLight treatment chamberto achieve ≥ a 6-logreduction of virusFiltered throughredundant 0.1-micronfilter membranes | Hot potable tap water is:pre-filteredflowed through a UVLight treatment chamberto achieve ≥ a 6-logreduction of virusFiltered throughredundant 0.1-micronfilter membranes | Identical | |||||||
| Number ofrinses | 2 | 2 | Identical | |||||||
| Air Purge | Aids in removing excesswater from instrument lumensafter rinsing | Aids in removing excesswater from instrument lumensafter rinsing | Identical | |||||||
| Water FilterIntegrity Test | Conducted at the end of everyliquid chemical sterilantprocessing cycle and duringthe Diagnostic cycle | Conducted at the end of everyliquid chemical sterilantprocessing cycle and duringthe Diagnostic cycle | Identical | |||||||
| ApproximateCycle time | 25 minutes | 25 minutes | Identical | |||||||
| Feature | Proposed DeviceSYSTEM 1E Processor | Predicate DeviceSYSTEM 1E Processor(K180432) | Comparison | |||||||
| DiagnosticCycle | Performs 15 tests onprocessor's systemsconfirming proper function(same tests as predicatedevice except for an addedUV monitor test).Recommended to performevery 24 hours. After a failedDiagnostic cycle, a liquidchemical sterilant processingcycle cannot be performeduntil the problem is rectifiedand a successful Diagnosticcycle has been completed. | Performs 15 tests onprocessor's systemsconfirming proper function(same tests as predicatedevice except for an addedUV monitor test).Recommended to performevery 24 hours. After a failedDiagnostic cycle, a liquidchemical sterilant processingcycle cannot be performeduntil the problem is rectifiedand a successful Diagnosticcycle has been completed. | Identical | |||||||
| Interchangeable Components | ||||||||||
| ProcessingTray /Containers | Uses interchangeableprocessing trays/containers• Universal FlexibleProcessing Tray (C1160E)• General ProcessingContainer/Tray (C1200)• Directed Flow ProcessingContainer/Tray (C1220)• Flexible EndoscopeProcessing Container / Tray(C1140)• Ultrasound Processing Tray(C3000XL) | Uses interchangeableprocessing trays/containers• Universal FlexibleProcessing Tray (C1160E)• General ProcessingContainer/Tray (C1200)• Directed Flow ProcessingContainer/Tray (C1220)• Flexible EndoscopeProcessing Container / Tray(C1140)• Ultrasound Processing Tray(C3000XL) | Identical | |||||||
| Accessories | ||||||||||
| SterilantConcentrate | Uses S40 SterilantConcentrate | Uses S40 SterilantConcentrate | Identical | |||||||
| QuickConnects | Uses Quick Connects to adaptinstrument lumens to theTray/Container ports | Uses Quick Connects to adaptinstrument lumens to theTray/Container ports | Identical | |||||||
| ChemicalIndicator | VERIFY Chemical Indicatorfor the SYSTEM 1EProcessor (K102217) | VERIFY Chemical Indicatorfor the SYSTEM 1EProcessor (K102217) | Identical | |||||||
| Spore TestStrip | VERIFY Spore Test Strip forS40 Sterilant Concentrate(K100049) | VERIFY Spore Test Strip forS40 Sterilant Concentrate(K100049) | Identical | |||||||
| OperatorMaintenanceRequirements | Periodic replacement ofprinter tape, water filters andair filter | Periodic replacement ofprinter tape, water filters andair filter | Identical | |||||||
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K180432) | Comparison | |||||||
| Indications forUse | S40 Sterilant Concentrate isfor use only with theSYSTEM 1E LiquidChemical SterilantProcessing System | S40 Sterilant Concentrate isfor use only with theSYSTEM 1E LiquidChemical SterilantProcessing System | Identical | |||||||
| Germicidalclaim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical | |||||||
| GermicideExposureTime (min) forintended use | 6 | 6 | Identical | |||||||
| UseTemperature | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at<43°C | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at<43°C | Identical | |||||||
| Reuse | Single use | Single use | Identical | |||||||
| HumanFactors | Dispensed ready to use.Container is opened anddiluted by the processor,thus limiting user exposureto the active ingredient | Dispensed ready to useContainer is opened anddiluted by the processor,thus limiting user exposureto the active ingredient | Identical | |||||||
| ActiveIngredient | 35% peroxyacetic (peracetic)acid automatically dilutedfor use in the SYSTEM 1EProcessor. | 35% peroxyacetic (peracetic)acid automatically dilutedfor use in the SYSTEM 1EProcessor. | Identical | |||||||
| Mode ofAction | It is believed that peraceticacid exerts its germicidaleffect by severalmechanisms:-oxidizing sulfhydryl andsulfur bonds in proteins andenzymes, particularly in thecell walls 1-hydroxyl radicals producedfrom PAA are bactericidal 2-PAA damages the viralcapsid and viral nucleicacid3,4. | It is believed that peraceticacid exerts its germicidaleffect by severalmechanisms:-oxidizing sulfhydryl andsulfur bonds in proteins andenzymes, particularly in thecell walls1-hydroxyl radicals producedfrom PAA are bactericidal2-PAA damages the viralcapsid and viral nucleicacid3,4 | Identical | |||||||
| Rinses | Automatic, UV-irradiated,dual 0.1 micron filtered | Automatic, UV-irradiated,dual 0.1 micron filtered | Identical | |||||||
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K180432) | Comparison | |||||||
| potable hot water. | ||||||||||
| Microbial EfficacySporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacyrequirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacyrequirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical | |||||||
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacyrequirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacyrequirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical | |||||||
| FungicidalActivity ofDisinfectantsAOACOfficialMethod 955.17 | Solution is fungicidal.TrichophytonmentagrophytesTesting conducted in vitro | Solution is fungicidal.TrichophytonmentagrophytesTesting conducted in vitro | Identical | |||||||
| Use DilutionMethodAOAC,OfficialMethods955.14, 955.15,964.02 | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Identical | |||||||
| EPA ViricidalTesting(DIS/TSS-7,Nov. 1981) | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Identical | |||||||
| TuberculocidalActivityAscenziQuantitativeSuspensionTest | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Identical | |||||||
| Simulated-UseTest | Meets efficacy requirement.> 6 log reductionGeobacillusstearothermophilus spores ina manual application | Meets efficacy requirement.> 6 log reductionGeobacillusstearothermophilus spores ina manual application | Identical | |||||||
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K180432) | Comparison | |||||||
| Clinical In-Use | No survivingmicroorganisms onrepresentative medicaldevices tested | No survivingmicroorganisms onrepresentative medicaldevices tested | Identical | |||||||
| Biocompatibility | ||||||||||
| CytotoxicityDevice Extracts | Two rinses with UV treated,dual 0.1-micron membranefiltered water effectivelyreduce sterilant residues tosafe levels. | Two rinses with UV treated,dual 0.1-micron membranefiltered water effectivelyreduce sterilant residues tosafe levels. | Identical | |||||||
| ResidueReduction | Automatic within theSYSTEM 1E Processor:Two rinses with UV treated,dual 0.1-micron membranefiltered water effectivelyreduce sterilant residues tosafe levels. | Automatic within theSYSTEM 1E Processor:Two rinses with UV treated,dual 0.1-micron membranefiltered water effectivelyreduce sterilant residues tosafe levels. | Identical | |||||||
| DeviceMaterialCompatibility | Compatible with medicaldevices as established bytesting finished flexibleendoscopes through 300cycles and rigid devicesthrough 150 cycles.No functional changes haveoccurred to flexible devices.Some materials showcosmetic changes such asfading of black anodizedaluminum without harm tothe base material. | Compatible with medicaldevices as established bytesting finished flexibleendoscopes through 300cycles and rigid devicesthrough 150 cycles.No functional changes haveoccurred to flexible devices.Some materials showcosmetic changes such asfading of black anodizedaluminum without harm tothe base material. | Identical | |||||||
| ChemicalMonitoring | VERIFY Chemical Indicatorfor S40 Sterilant (K173428) | VERIFY Chemical Indicatorfor the SYSTEM 1EProcessor (K102217) | K173428included useof the CI(with namechange) in theSYSTEM 1endoLCSPS.Notechnologicalcharacteristicschanged. |
Table 1. Processor Device Comparison Table
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
Table 2. S40 Sterilant Concentrate Device Comparison Table
1 Block, S. ed., Disinfection, Sterilization, and Preservation. 5th edition, 2001.
2 Clapp et al., Free Rad. Res., (1994) 21:147-167.
3 Maillard et. al., J. Med. Microbiol (1995) 42:415-420.
4 Maillard et. al., J. Appl Bacteriol (1996) 80:540-554.
{14}------------------------------------------------
5 McDonnell et al., J. AOAC International (2000) 83:269-276.
6 McDonnell et al., J. AOAC International (2000) 83:269-276.
{15}------------------------------------------------
{16}------------------------------------------------
The proposed device and its predicate have identical intended use and technological characteristics. New testing was performed to evaluate the modified device and the results are summarized in Table 3.
| Test | Acceptance Criteria | Result |
|---|---|---|
| Biocompatibility of S40Sterilant Concentratewith alternate source of aproprietary component | Medical devices shall be non-cytotoxicafter exposure to S40 SterilantConcentrate containing the alternatesource of a proprietary component | Pass |
| Stability of S40 SterilantConcentrate withalternate source of aproprietary component | S40 Sterilant Concentrate shall meetacceptance criteria after 6-months ofstorage. | Pass |
| Table 3. Summary of verification activities. | ||||
|---|---|---|---|---|
| -- | -- | -- | ---------------------------------------------- | -- |
6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K180342).
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is written in a serif font in all caps. Below the word is an image of several horizontal wavy lines stacked on top of each other. The lines are blue.
510(k) Summary For SYSTEM 1 endo Liquid Chemical Sterilant Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060
Contact: Jennifer Nalepka Senior Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
March 5, 2019 Summary Date:
Premarket Notification Number: K190104
{18}------------------------------------------------
1. Device Name
| Trade Name: | SYSTEM 1 endo Liquid Chemical SterilanProcessing System |
|---|---|
| Device Class: | Class II |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid ChemicalSterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
2. Predicate Device
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, K173256. Reference Device: SYSTEM 1 endo Liquid Chemical Sterilant Processing System (Model 6900), K182827
3. Description of Device
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical heat-sensitive medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe the addition of a second supplier for of a proprietary component of S40 Sterilant Concentrate.
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. After LCS processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if
{19}------------------------------------------------
required, facilitate delivery of the sterilant use-solution and rinse water to internal channels. Table 1 compares the proposed and predicate devices.
Indications for Use 4.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical heatsensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
ട. Technological Characteristic Comparison Table
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is the same as the predicate device; the specific modification described in this submission is for an alternate supplier for a component of S40 Sterilant Concentrate. The differences between the proposed and predicate devices are limited to the addition of an alternate supplier for of a proprietary component of S40 Sterilant Concentrate.
The similarities and differences between the proposed and predicate devices are identified in Table 1 and Table 2 below.
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
|---|---|---|---|
| IntendedUseIndicationsfor Use | The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable semi-critical heat-sensitive medical devices andtheir accessories in healthcarefacilities. | The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable semi-critical heat-sensitive medical devices andtheir accessories in healthcarefacilities. | Identical |
| The SYSTEM 1 endoProcessor automatically dilutesthe S40 Sterilant Concentrateto its use dilution (> 1820 | The SYSTEM 1 endoProcessor automaticallydilutes the S40 SterilantConcentrate to its use dilution | ||
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
| mg/L peracetic acid), liquidchemically sterilizes the loadduring a controlled 6-minuteexposure at 45.5 to 60°C, andrinses the load with 0.2 micronfiltered water.The SYSTEM 1 endo | (> 1820 mg/L peracetic acid),liquid chemically sterilizesthe load during a controlled6-minute exposure at 45.5 to60°C, and rinses the load with0.2 micron filtered water.The SYSTEM 1 endo | ||
| Processor uses only S40Sterilant Concentrate to liquidchemically sterilize medicaldevices. | Processor uses only S40Sterilant Concentrate to liquidchemically sterilize medicaldevices. | ||
| OperatingPrinciples/Technology | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processing chamberin which the load is placed.Devices with internal lumensare interfaced with theprocessor using connectors,i.e. Quick Connects.S40 Sterilant is placed in aspecialized compartment andwhen the processor fills withwater, it creates the sterilantuse dilution from the singleuse sterilant cup.The processor monitors andcontrols the use dilutiontemperature and contact time.The processor automaticallyrinses the load with 0.2 micronfiltered water to removesterilant residuals. | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processingchamber in which the load isplaced.Devices with internal lumensare interfaced with theprocessor using connectors,i.e. Quick Connects.S40 Sterilant is placed in aspecialized compartment andwhen the processor fills withwater, it creates the sterilantuse dilution from the singleuse sterilant cup.The processor monitors andcontrols the use dilutiontemperature and contact time.The processor automaticallyrinses the load withextensively treated filteredwater to remove sterilantresiduals. | Identical |
| ProcessParameters | Standardized cycle parameterscannot be altered by operator.The critical process parametersare:• Use dilution contact time• Use dilution temperature• Peracetic acidconcentration | Standardized cycleparameters cannot be alteredby operator. The criticalprocess parameters are:• Use dilution contact time• Use dilution temperature• Peracetic acidconcentration | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
| • Integrity of the internalwater filter (tested by thesystem) | • Integrity of the internalwater filter (tested by thesystem) | ||
| ProcessMonitors: | • Cycle Printout documentssuccessful cyclecompletion or identifiesfault if cycle aborts• Alarms if thermocouplesindicate temperature out ofspecification• Alarms if pressure switchindicates that high pressurepump is not operating• Alarms if conductivityprobe indicatedconductivity specificationnot met• Alarms if pressuretransducer indicatescirculation pressure is outof specification inDiagnostic cycle• Alarms if pressuretransducer indicatesinternal water filter failedintegrity test | • Cycle Printout documentssuccessful cyclecompletion or identifiesfault if cycle aborts• Alarms if thermocouplesindicate temperature outof specification• Alarms if pressure switchindicates that highpressure pump is notoperating• Alarms if conductivityprobe indicatedconductivity specificationnot met• Alarms if pressuretransducer indicatescirculation pressure is outof specification inDiagnostic cycle• Alarms if pressuretransducer indicatesinternal water filter failedintegrity test | IdenticalIn theReferenceDevice,K182827,cycle recordsare electronic |
| DesignFeatures | • Microprocessor controlledunalterable andstandardized liquidchemical sterilization andDiagnostic cycles• Intended for use only withS40 Sterilant Concentrate• Automated dilution anddelivery of S40 Sterilant• Processor provides 0.2micron filtered water forliquid chemicalsterilization and rinsing | • Microprocessor controlled unalterable and standardized liquid chemical sterilization and Diagnostic cycles • Intended for use only with S40 Sterilant Concentrate • Automated dilution and delivery of S40 Sterilant • Processor provides 0.2 micron filtered water for liquid chemical sterilization and rinsing | IdenticalAdditionaldesignfeatures arepart of theReferenceDevice inK182827:Thereferencedevice usesan updatedcontroller |
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
| Make-up air for processor during drain sequences is filtered through a 0.2 micron membrane air filter Has no barcode scanner, employs a membrane touch panel interface and small display, has no USB drive. Includes an onboard printer for cycle information | Make-up air for processor during drain sequences is filtered through a 0.2 micron membrane air filter Has no barcode scanner, employs a membrane touch panel interface and small display, has no USB drive. Includes an onboard printer for cycle information | and its respective software for identical processing parameters.The proposed device uses an updated touchscreen display, has a barcode reader and USB drive for electronic cycle records, and facilitates use of a web-based data mgmt system.Use of printer is optional. | |
| Cycle Parameters | Comparison | ||
| Incomingwater temp. | $\geq 43°C$ | $\geq 43°C$ | Identical |
| Temperatureto startsterilantexposure | $\geq 46°C$ | $\geq 46°C$ | Identical |
| Temperaturealarm pointduring LCSexposure | < 45.5 or > 60°C | < 45.5 or > 60°C | Identical |
| Temperaturerange oftypical LCScycle | 46 - 55°C | 46 - 55°C | Identical |
| ExposureTime - S40use dilution | 6 minutes | 6 minutes | Identical |
| Rinse water | Hot potable tap water | Hot potable tap water | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
| preparation | is pre-filtered is filtered through 0.2 micron bacterial retentive membrane filter | is pre-filtered is filtered through 0.2 micron bacterial retentive membrane filter | |
| Number ofrinses | 2 | 2 | Identical |
| Air Purge | Aids in removing excess waterfrom instrument lumens afterrinsing | Aids in removing excesswater from instrument lumensafter rinsing | Identical |
| InternalWater FilterIntegrityTest | Conducted during theDiagnostic cycle | Conducted during theDiagnostic cycle | Identical |
| ApproximateCycle Time | 18 - 20 minutes | 18 - 20 minutes | Identical |
| DiagnosticCycle | Performs 14 tests onprocessor's systemsconfirming proper function.Recommended to performeach day of use. After a failedDiagnostic cycle, a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cyclehas been completed. | Performs 14 tests onprocessor's systemsconfirming proper function.Recommended to performeach day of use. After a failedDiagnostic cycle, a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cyclehas been completed. | Identical |
| Accessories | |||
| Sterilant | Uses S40 Sterilant Concentrate- see Table 2 | Uses S40 SterilantConcentrate - see Table 2 | Identical |
| ProcessingTrays andContainers | Uses interchangeableprocessing trays/containers Universal Flex Processing Tray General Processing Container & Tray Directed Flow Processing Container & Tray Flexible Endoscope Processing Container & Tray Ultrasound Processing Tray | Uses interchangeableprocessing trays/containers Universal Flex Processing Tray General Processing Container & Tray Directed Flow Processing Container & Tray Flexible Endoscope Processing Container & Tray Ultrasound Processing Tray | Identical |
| QuickConnects | Uses Quick Connects to attachinstrument lumens to the | Uses Quick Connects toattach instrument lumens to | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor | Predicate DeviceSYSTEM 1 endo Processor(K173256) | Comparison |
| Tray/Container ports | the Tray/Container ports | ||
| ChemicalIndicator | VERIFY Chemical Indicatorfor S40 Sterilant is availablefor use in SYSTEM 1 endoLCSPS | VERIFY Chemical Indicatorfor S40 Sterilant is availablefor use in SYSTEM 1 endoLCSPS | Identical |
| Spore TestStrip | VERIFY Spore Test Strip forS40 Sterilant for use inSYSTEM 1 endo LCSPS | VERIFY Spore Test Strip forS40 Sterilant for use inSYSTEM 1 endo LCSPS | Identical |
| OperatorMaintenance | Periodic replacement of waterfilters and air filter.Periodic replacement ofprinter tape. | Periodic replacement of waterfilters and air filter.Periodic replacement ofprinter tape. | IdenticalIn ReferenceDeviceK182827, theprinter is anexternaloption. |
Table 1. Processor Comparison Table
{20}------------------------------------------------
{21}------------------------------------------------
{22}------------------------------------------------
{23}------------------------------------------------
{24}------------------------------------------------
{25}------------------------------------------------
| Feature | S40 Sterilant Concentratein S1 endo LCSPS | S40 Sterilant Concentratein S1 endo LCSPS(K173256) | Comparison |
|---|---|---|---|
| Intended Use | For use in S1E or S1 endoLCSPS | For use in S1E or S1 endoLCSPS | Identical |
| Germicidalclaim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical |
| ExposureTime | 6 minutes | 6 minutes | Identical |
| UseTemperature | 45.5 - 60°C - allowable rangePotency and simulated useevaluations conducted at ≤43°C | 45.5 - 60°C - allowable rangePotency and simulated useevaluations conducted at ≤43°C | Identical |
| Reuse | Single use | Single use | Identical |
| HumanFactors | Ready to use. Container isopened and diluted by theprocessor, limiting userexposure to the sterilantconcentrate | Ready to use. Container isopened and diluted by theprocessor, limiting userexposure to the sterilantconcentrate | Identical |
| ActiveIngredient | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the Processor | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the Processor | Identical |
| Mode ofAction | It is believed that peraceticacid exerts its germicidaleffect by severalmechanisms:-oxidizing sulfhydral andsulfur bonds in proteins andenzymes, particularly in thecell walls 1-hydroxyl radicals producedfrom PAA are bactericidal 2-PAA damages the viralcapsid and viral nucleicacid3,4 | It is believed that peraceticacid exerts its germicidaleffect by severalmechanisms:-oxidizing sulfhydral andsulfur bonds in proteins andenzymes, particularly in thecell walls1-hydroxyl radicals producedfrom PAA are bactericidal2-PAA damages the viralcapsid and viral nucleicacid3.4 | Identical |
| Rinses | 2 equivalent automated rinseswith pre-filtered, dual 0.2micron membrane filtered,potable hot water | 2 equivalent automated rinseswith pre-filtered, dual 0.2micron membrane filtered,potable hot water | |
| Feature | S40 Sterilant Concentratein S1 endo LCSPS | S40 Sterilant Concentratein S1 endo LCSPS(K173256) | Comparison |
| SporicidalActivity ofDisinfectantsAOACOfficialMethod966.04 | Meets efficacy requirements5Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements5Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod966.04 | Meets efficacy requirements5Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements5Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| FungicidalActivity ofDisinfectantsAOACOfficialMethod955.17 | Solution is fungicidalTrichophytonmentagrophytesTesting conducted in vitro | Solution is fungicidalTrichophyton mentagrophytesTesting conducted in vitro | Identical |
| Use DilutionMethodAOAC,OfficialMethods955.14,955.15, 964.02 | Solution is bactericidalSalmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Solution is bactericidalSalmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Identical |
| EPA VirucidalTesting(DIS/TSS-7,Nov. 1981) | Solution is virucidalHerpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Solution is virucidalHerpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Identical |
| TuberculocidalActivity ofDisinfectantsAOACOfficialMethod965.12 | Not performed; a quantitativesuspension Tuberculocidaltest was conducted | Not performed; a quantitativesuspension Tuberculocidaltest was conducted | Identical |
| Feature | S40 Sterilant Concentratein S1 endo LCSPS | S40 Sterilant Concentratein S1 endo LCSPS(K173256) | Comparison |
| TuberculocidalActivityAscenziQuantitativeSuspensionTest | Solution is tuberculocidalMycobacterium terraetesting conducted in vitro | Solution is tuberculocidalMycobacterium terraetesting conducted in vitro | Identical |
| Simulated Use | Meets efficacy requirement.>= 6 log10 reductionGeobacillusstearothermophilus spores ina manual application. | Meets efficacy requirement.> 6 log10 reductionGeobacillusstearothermophilus spores ina manual application. | Identical |
| Clinical In Use | No surviving microorganismson representative medicaldevices tested in S1 endoLCSPS | No surviving microorganismson representative medicaldevices tested in S1 endoLCSPS | Identical |
| CytotoxicityDeviceExtracts | Two Processor controlledrinses with pre-filtered, dual0.2 micron filtered potablewater effectively reducesterilant residues to non-cytotoxic levels. | Two Processor controlledrinses with pre-filtered, dual0.2 micron filtered potablewater effectively reducesterilant residues to non-cytotoxic levels. | Identical |
| ResidueReduction | Two Processor controlledrinses with pre-filtered, dual0.2 micron filtered potablewater effectively reducesterilant residues to safelevels. | Two Processor controlledrinses with pre-filtered, dual0.2 micron filtered potablewater effectively reducesterilant residues to safelevels. | Identical |
| MaterialCompatibility | Compatible with medicaldevices as established bytesting finished medicaldevices through 300 cycles.No functional changesoccurred to devices. Somematerials show cosmeticchanges such as fading ofexternal markings (yet allremained legible) andbleaching of black anodizedaluminum without harm tothe base material. | Compatible with medicaldevices as established bytesting finished medicaldevices through 300 cycles.No functional changesoccurred to devices. Somematerials show cosmeticchanges such as fading ofexternal markings (yet allremained legible) andbleaching of black anodizedaluminum without harm tothe base material. | Identical |
Table 2. S40 Sterilant Concentrate Comparison Table
1 Block, S. ed., Disinfection, Sterilization, and Preservation. 5th edition, 2001
2 Clapp et al., Free Rad. Res., (1994) 21:147-167
3 Maillard et. al., J. Med. Microbiol (1995) 42:415-420
4 Maillard et. al., J. Appl Bacteriol (1996) 80:540-554
{26}------------------------------------------------
5 McDonnell et al., J. AOAC International (2000) 83:269-275
{27}------------------------------------------------
The proposed device and its predicate have identical intended use and technological characteristics. New testing was performed to evaluate the modified device and the results are summarized in Table 3.
{28}------------------------------------------------
| Test | Acceptance Criteria | Result |
|---|---|---|
| Biocompatibility of S40Sterilant Concentrate withalternate source of aproprietary component | Medical devices shall be non-cytotoxicafter exposure to S40 SterilantConcentrate containing the alternatesource of a proprietary component | Pass |
| Stability of S40 SterilantConcentrate withalternate source of aproprietary component | S40 Sterilant Concentrate shall meetacceptance criteria after 6-months ofstorage. | Pass |
Table 3. Summary of Verification Testing
6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective, and performs as well or better than the legally marketed predicate device (K173256).
§ 880.6885 Liquid chemical sterilants/high level disinfectants.
(a)
Identification. A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.(b)
Classification. Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.